From Helium to Hydrogen: Case Study in the GC-MS Analysis of VOCs and SVOCs Sergio Guazzotti, Alexander N. Semyonov, Jessie Butler, Thermo Fisher Scientific, Austin, Texas, USA Overview Results Purpose: EPA Methods 8270, 524, and 525 were optimized with hydrogen carrier gas Stabilization Bake out for H Methods: Hydrogen carrier gas was configured by using a smaller ID column and adding a hydrogen kit to a mass spectrometer When switching from helium no gas filters are necessary. of hydrogen is used, grade U spectrometer. The source te is turned on for one hour dur source temperature was set contamination off the entire f present in the gas lines, a m ionization and protonation ar observed in the calibration g Results: Peak shape, resolution, and run time improved Introduction The global helium shortage and price increase of this non-renewable noble gas cause more and more laboratories to migrate to the use of hydrogen as a carrier gas. This transition is easy for the GC methods utilizing FID, TCD, ECD, and other non-mass-selective detectors. However, for the GC-MS methods, especially complex and regulated ones, migration to hydrogen carrier gas presents significant challenges. In addition to the changes in chromatographic conditions of the run due to the physical property differences of hydrogen, its reactivity brings about chemical reactions in the mass spectrometer’s ion source that do not occur with the use of helium. The analyses of semi-volatile organic compounds (SVOCs) in wastewater by EPA Method 82701 and volatile organic compounds (VOCs) in drinking water by EPA Methods 5242 and 5253 involve identification and quantitation of well over a hundred analytes of varying chemical structure, polarity, and volatility. The diversity of the analytes in this method presents particular challenges when migrating from helium to hydrogen carrier. GC-MS analysis of SVOCs was performed on a Thermo Scientific™ TRACE™ 1310 GC coupled to a Thermo Scientific™ ISQ™ Series single quadrupole mass spectrometer utilizing helium and hydrogen carrier gases. Key modifications to both GC and MS hardware and methods are necessary for successful migration to hydrogen carrier gas. The final, optimized EPA methods were migrated to hydrogen carrier gas with improved peak shape, resolution, and run time. FIGURE 2. Air/water spectr m/z 18 Methods m/z 19 Sample Preparation For EPA Method 8270, standards were prepared in methylene chloride. A performance mix containing pentachlorophenol, DFTPP, benzidine, and p,p’-DDT was made at a concentration of 50 ppm. A calibration curve was prepared from 1 to 200 ppm with internal standards and surrogates at 40 ppm. For EPA Method 524, a 5 mL volume of water was used in a purge and trap concentrator (Eclipse 4660 Purge-and-Trap Sample Concentrator, OI Analytical),and a calibration curve from 0.4 to 200 ppb was run. For EPA Method 525, a 1 L volume of water was extracted with ethyl acetate. A calibration curve was run from 0.1 to 10 ppm. m/z 28 Gas Chromatography FIGURE 4. Cal Gas (FC-43) For EPA Methods 525 and 8270, an Instant Connect split/splitless injector module was used in the split mode. A one microliter injection was made. The standard Thermo Scientific™ TraceGOLD TG-5MS capillary column (30 m × 0.25 mm × 0.5 µm film) was replaced with a TraceGOLD TG-5MS column of smaller ID (20 m × 0.18 mm × 0.36 µm film). For EPA Method 524, a purge and trap (P&T) adapter was used to make the injection of gases purged from water sample of 5 mL volume onto a TraceGOLD TG-VMS column of smaller ID (20 m × 0.18 mm × 1.0 µm film). Mass Spectrometry The ISQ Series mass spectrometer was equipped with a hydrogen kit. The instrument was preconditioned by baking it out with an elevated hydrogen carrier flow rate with the EI ion source filament turned on for one hour. Data Analysis Data was acquired using Thermo Scientific™ Xcalibur™ 2.0 software and processed using either Thermo Scientific™ Target 4.14 or Thermo Scientific™ TraceFinder™ 3.0 software. The Xcalibur software was configured for direct target processing. A typical chromatogram of a 40 ppm standard is shown in Figure 1. FIGURE 1. TIC of 40 ppm EPA Method 8270 standard with hydrogen carrier gas. RT: 0.00 - 13.27 Selecting the Appropriate NL: 5.35E8 TIC MS level8 6.83 100 95 Due to the lower viscosity of min of column flow rate. By the reduction in the vapor vo 90 85 3.85 80 8.98 9.19 7.90 70 4.56 65 Relative Abundance FIGURE 6. Comparison of 10.24 75 6.26 5.18 60 55 5.65 6.08 6.45 3.88 45 8.41 11.16 4.61 3.34 40 4.67 5.02 11.44 9.76 7.01 4.26 35 11.19 8.06 5.34 50 12.35 10.27 7.64 6.97 9.45 7.37 11.72 10.79 8.84 12.55 5.95 30 9.97 25 3.31 20 FIGURE 7. Methylene Chlo 15 2.25 1.80 2.98 RT: 0.000 - 13.008 2.51 100 5 8.48 10.54 12.17 0 0 1 2 3 4 5 6 7 Time (min) 8 9 10 11 12 90 12.70 80 13 Relative Abundance 10 4.039 70 60 50 3.016 40 30 1.368 20 10 2.002 3.281 0 100 90 80 70 3.730 60 50 2.719 40 30 2 From Helium to Hydrogen: Case Study in the GC-MS Analysis of VOCs and SVOCs 20 10 1.079 1.526 0 0 1 2.681 2 3.000 3 4.3 4 Peak Shape and Resolution Results The advantage of hydrogen carrier g spending less time in the stationary narrower peak width, and the critical Stabilization Bake out for Hydrogen Carrier Gas and adding a hydrogen kit to a mass When switching from helium to hydrogen carrier gas, a brief stabilization period is required. If using a hydrogen generator, no gas filters are necessary. Gas lines should be plumbed with new, pre-cleaned 1/8 in. stainless steel tubing. If a cylinder of hydrogen is used, grade UHP 5.0 or better is required with in-line gas purifiers. The hydrogen kit is installed in the mass spectrometer. The source temperature is set to 350 °C for bake out. The hydrogen carrier is set to 4 mL/min. The filament is turned on for one hour during the bake out period. After bake out, the hydrogen flow is reduced to 1 mL/min, and the source temperature was set to 325 °C. Figures 2–5 show the effect of bake out. The hydrogen gas desorbs the contamination off the entire flow path, and the reactive hydrogen species formed in the ion source clean it out. If water is present in the gas lines, a m/z of 19 is seen. This is due to protonation of water (M+1). Other ions typical of chemical ionization and protonation are also seen, m/z 29 and m/z 41 (Figure 2). Elevated levels of low mass hydrocarbons are observed in the calibration gas spectrum (Figure 4). These all are reduced after the one-hour bake out. as cause more and more laboratories to GC methods utilizing FID, TCD, ECD, especially complex and regulated ones, to the changes in chromatographic reactivity brings about chemical e of helium. FIGURE 8. Improvement of Peak S critical pair of benzo[g&h]fluorant RT: 11.87 - 13.34 12.29 100 12.50 90 H 80 Relative Abundance arrier gas 70 60 50 40 30 20 10 0 100 11.88 11.97 12.02 12.11 12.14 12.53 12.56 12.63 12.65 12.34 12.37 12.26 12.72 12.76 12.81 90 H 80 70 60 50 EPA Method 82701 and volatile organic identification and quantitation of well The diversity of the analytes in this en carrier. GC-MS analysis of SVOCs ermo Scientific™ ISQ™ Series single Key modifications to both GC and MS arrier gas. The final, optimized EPA esolution, and run time. 40 30 20 10 0 11.90 11.94 12.03 12.05 12.09 12.16 12.21 12.25 12.27 11.9 FIGURE 2. Air/water spectrum before bake out. 12.0 12.1 12.2 12.72 12.37 12.42 12.46 12.52 12.54 12.3 12.4 12.5 12.6 Time (min) 12.7 12.75 12.8 12.8 FIGURE 3. Air/water spectrum after bake out. Linearity m/z 18 m/z 28 m/z 19 m/z 29 formance mix containing ation of 50 ppm. A calibration curve was m. For EPA Method 524, a 5 mL volume Trap Sample Concentrator, OI hod 525, a 1 L volume of water was m/z 19 m/z 28 FIGURE 4. Cal Gas (FC-43) spectrum before bake out. dule was used in the split mode. A one TG-5MS capillary column (30 m × 0.25 er ID (20 m × 0.18 mm × 0.36 µm film). njection of gases purged from water m × 0.18 mm × 1.0 µm film). A linearity study was performed in t results are shown in Figures 9 and than hydrogen but still met the crite m/z 18 m/z 29 FIGURE 9. Comparison of linea R2 fit with hydrogen). FIGURE 5. Cal Gas (FC-43) spectrum after bake out. strument was preconditioned by baking nt turned on for one hour. FIGURE 10. Typical linearity plo cessed using either Thermo Scientific™ software was configured for direct target 1. ier gas. Selecting the Appropriate Column Dimensions and Inlet Pressure NL: 5.35E8 TIC MS level8 Due to the lower viscosity of hydrogen carrier gas compared to helium, a lower head pressure is required to obtain 1 mL/ min of column flow rate. By reducing the ID of the column, the pressure may be set to a usable level (Figure 6 ). Also note the reduction in the vapor volume of methylene chloride at higher pressures (Figure 7). FIGURE 6. Comparison of inlet pressure for helium and hydrogen for a 0.25 mm and 0.18 mm ID column. 10.24 A linearity study was performed in results for the PAHs is shown in F range of concentration from 0.4 to 8.98 9.19 12.35 10.27 11.19 FIGURE 11. Typical linearity of .41 11.16 Compound 11.44 9.76 9.45 11.72 10.79 8.84 12.55 9.97 FIGURE 7. Methylene Chloride solvent tailing at low vs. high inlet pressures. RT: 0.000 - 13.008 9 10 12.17 11 12 8.596 90 12.70 80 13 Relative Abundance 10.54 NL: 2.12E8 TIC MS level1 7.031 100 8.48 5.795 4.039 8 psi 9.808 70 60 50 3.016 40 30 1.368 20 10 2.002 3.281 6.358 4.698 5.057 7.075 7.874 0 100 8.953 9.423 10.099 11.114 11.520 12.801 NL: 1.55E8 TIC MS 1ng 8.309 6.802 90 80 9.338 70 13 psi 5.430 3.730 60 50 2.719 40 30 6.1 2-methylnaphthalene 8.7 1-methylnaphthalene 7.7 acenaphthylene 8.5 acenaphthene 8.5 flourene 8.5 phenanthene 3.8 anthracene 7.0 pyrene 3.6 acenaphthene-d10 4.6 chrysene-d12 9.7 Sensitivity The sensitivity of the EPA Method The average Instrument Detection Figure 12 shows most of the IDLs Thermo Scientific Poster Note • EAS2013_PN10360_E 11/13S 3 20 10 1.079 1.526 0 0 1 2.681 2 3.000 3 4.367 4.721 4 5 6.035 6.849 7.633 6 7 Time (min) 9.036 8 9 10.411 10.742 11.648 12.000 9.577 10 11 12 13 %RSD naphthalene Compound 9.76 9.45 11.72 10.79 8.84 naphthalene 12.55 2-methylnaphthalene 9.97 1-methylnaphthalene FIGURE 7. Methylene Chloride solvent tailing at low vs. high inlet pressures. RT: 0.000 - 13.008 s and SVOCs 8.48 9 10 12.17 11 8.596 90 12.70 12 80 13 acenaphthene NL: 2.12E8 TIC MS level1 7.031 100 flourene 5.795 4.039 phenanthene 8 psi 9.808 70 Relative Abundance 8 10.54 acenaphthylene 60 anthracene 50 3.016 pyrene 40 30 1.368 acenaphthene-d10 20 10 Texas, USA 2.002 3.281 6.358 4.698 5.057 7.075 7.874 0 100 8.953 9.423 11.114 11.520 10.099 NL: 1.55E8 TIC MS 1ng 8.309 6.802 90 chrysene-d12 12.801 80 Sensitivity 9.338 70 13 psi 5.430 3.730 60 The sensitivity of the EPA M The average Instrument De Figure 12 shows most of th 50 2.719 40 30 20 10 1.079 1.526 0 0 1 2.681 2 3.000 3 6.035 4.367 4.721 4 5 6.849 7.633 6 9.036 7 8 10.411 10.742 11.648 12.000 9.577 9 10 11 12 13 Time (min) FIGURE 12. Peak Shape and Resolution FIGURE 8. Improvement of Peak Shape for Indeno[1,2,3-cd]pyrene & benzo[g,h,i]perylene and separation of the critical pair of benzo[g&h]fluoranthenes. RT: 11.87 - 13.34 NL: 5.23E6 m/z= 275.50276.50 MS level5 12.29 100 12.50 90 Hydrogen 80 70 Helium 60 50 Hydrogen 40 30 20 RT: 10.26 - 10.43 10 0 100 100 11.88 11.97 12.02 12.11 12.14 12.53 12.56 12.63 12.65 12.34 12.37 12.26 12.72 12.76 12.81 90 12.87 12.93 13.01 13.06 13.12 13.17 Helium 80 NL: 7.26E6 m/z= 275.50276.50 MS level5 90 60 50 95 90 85 80 75 75 70 70 50 45 55 50 45 40 30 25 25 20 60 35 35 30 30 11.11 65 60 55 40 40 NL: 6.19E6 m/z= 251.50252.50 MS level5 11.14 100 85 80 65 Relative Abundance 70 NL: 1.19E7 m/z= 251.50252.50 MS level5 10.36 10.33 95 13.26 13.23 RT: 11.06 - 11.20 Relative Abundance Relative Abundance s required. If using a hydrogen generator, d 1/8 in. stainless steel tubing. If a cylinder s. The hydrogen kit is installed in the mass gen carrier is set to 4 mL/min. The filament en flow is reduced to 1 mL/min, and the The hydrogen gas desorbs the in the ion source clean it out. If water is (M+1). Other ions typical of chemical d levels of low mass hydrocarbons are the one-hour bake out. The advantage of hydrogen carrier gas is that higher than helium linear velocities can be used without sacrificing resolution. By spending less time in the stationary phase, the late eluting polynuclear aromatic hydrocarbons (PAHs) come out with a narrower peak width, and the critical pair of benzo[g&h]flouranthenes was easily separated (Figure 8). 20 20 15 15 10 0 10 10 11.90 11.94 12.03 12.05 12.09 12.16 12.21 12.25 12.27 11.9 12.0 12.1 12.2 12.3 12.72 12.37 12.42 12.46 12.52 12.54 12.4 12.5 12.6 Time (min) 12.7 12.75 12.84 12.87 12.94 12.98 13.03 12.8 12.9 13.0 5 13.13 13.1 5 13.2 13.3 0 10.42 10.26 10.26 10.27 10.28 10.29 10.30 10.31 10.32 10.33 10.34 10.35 Time (min) 10.36 10.37 10.38 10.39 10.40 10.41 0 11.06 11.07 11.07 11.16 11.08 11.08 11.09 11.10 11.11 10.42 11.12 11.13 Time (min) 11.14 11.15 11.16 11.17 11.17 11.19 11.18 11.20 11.19 Instrument de IDL was 0.07 level as 200 /water spectrum after bake out. Linearity A linearity study was performed in the range of concentration from 1 to 200 ppm in CH2Cl2 for EPA Method 8270. The results are shown in Figures 9 and 10 below. Helium carrier gas gave lower relative percent standard deviations (%RSD) than hydrogen but still met the criteria for the method. A few compounds required the fit of least squares (R2 = > 0.99). m/z 18 m/z 28 FIGURE 13. RT: 0.80 - 1.81 SM: 7B 100 50 Compounds: ≤ 0.99 • nitrophenols • nitroanilines • dinitrophenols • 4,6-dinityrophenol-2-methylphenol • t richlorophenols • dinitrotoluenes • hexachlorocyclopentadiene • t ribromophenol • pentachlorophenol • butylbenzylphthalate • di-n-octylphthalate • bis(2 ethylhexyl)phthalate 0 100 R2 al Gas (FC-43) spectrum after bake out. 50 0 Relative Abundance m/z 19 m/z 29 FIGURE 9. Comparison of linearity with helium and hydrogen carrier (the compounds on the insert required R2 fit with hydrogen). 100 50 0 100 50 0 100 50 0 100 50 0 0.80 FIGURE 10. Typical linearity plots with hydrogen carrier gas. 0.85 Robustness The stability robustness s samples also the internal s All check sta FIGURE 14. head pressure is required to obtain 1 mL/ set to a usable level (Figure 6 ). Also note ure 7). 5 mm and 0.18 mm ID column. A linearity study was performed in the range of concentration from 0.05 to 10 ppm for EPA Method 525. An excerpt of in theforGC-MS Analysis of VOCs and11. SVOCs 4 From Helium to Hydrogen: Case Study results the PAHs is shown in Figure Similarly, for EPA Method 524, a linearity study was performed in the range of concentration from 0.4 to 200 ppb. (Figure 6 ). Also note ID column. A linearity study was performed in the range of concentration from 0.05 to 10 ppm for EPA Method 525. An excerpt of results for the PAHs is shown in Figure 11. Similarly, for EPA Method 524, a linearity study was performed in the range of concentration from 0.4 to 200 ppb. FIGURE 11. Typical linearity of PAHs in EPA Method 525 with hydrogen carrier gas. Compound %RSD Compound %RSD naphthalene 6.1 flouranthene 6.7 2-methylnaphthalene 8.7 benzo(a)anthracene 8.2 1-methylnaphthalene 7.7 chrysene 5.9 acenaphthylene 8.5 benzo(b)fluoranthene 8.6 100 acenaphthene 8.5 benzo(k)fluoranthene 9.2 90 80 FIGURE 15. TIC 5 % diesel sam RT: 0.00 - 13.26 8.5 benzo(a)pyrene 7.8 phenanthene 3.8 indeno(1,2,3,c,d)pyrene 5.1 anthracene 7.0 dibenz(a,h)anthracene 5.4 pyrene 3.6 benzo(g,h,i)perylene 6.0 acenaphthene-d10 4.6 phenanthene-d10 4.5 chrysene-d12 9.7 perylene-d12 7.4 4.55 85 3.43 5.0 4.01 75 70 65 Relative Abundance flourene 95 60 55 50 45 40 35 4.92 30 25 3.75 20 4.35 2.80 3.21 15 10 5 Sensitivity 3.19 2.63 1.70 1.96 2.15 2.57 0 0 1 2 3 4 5 The sensitivity of the EPA Method 8270 using hydrogen carrier was within a factor of two to that achieved with helium. The average Instrument Detection Limit (IDL) with helium was 0.08 ppm and 0.15 ppm with hydrogen. The chart in Figure 12 shows most of the IDLs are around 0.1 ppm. Replicate injections of the standard were made at 1 ppm. FIGURE 12. EPA Method 8270 Instrument Detection Limits (ppm) Average = 0.15 ppm. FIGURE 16. Stability of internal standard are ithout sacrificing resolution. By PAHs) come out with a ure 8). and separation of the Butylbenzyl phthalate Di-n-octylphthalate Hydrogen Bis-2-chloroisopropylether NL: 6.19E6 m/z= 251.50252.50 MS level5 11.14 11.11 11.16 11.10 11.11 11.12 11.13 Time (min) 11.14 11.15 11.16 11.17 11.17 11.19 11.18 11.20 11.19 Instrument detection limits for EPA Method 524 were determined by running replicate injections at 0.4 ppb. The average IDL was 0.074 ppb (74 ppt) vs. 0.048 ppb (48 ppt) with helium. Excellent sensitivity for the first six gases, even at as low level as 200 ppt, was obtained (Figure 13). EPA Method 8270. The andard deviations (%RSD) t squares (R2 = > 0.99). RT: 0.80 - 1.81 SM: 7B dichlorodifluoromethane RT: 0.93 AA: 9279 BP: 85 0 pentadiene ol late te phthalate RT: 1.04 AA: 47618 BP: 50 100 ≤ 0.99 0 chloromethane RT: 1.09 AA: 19284 BP: 62 100 vinylchloride RT: 1.29 AA: 21231 BP: 94 100 bromomethane RT: 1.38 AA: 14294 BP: 64 RT: 1.08 AA: 5325 BP: 64 100 50 0 RT: 1.47 AA: 16876 BP: 101 100 50 0 0.80 0.85 0.90 0.95 1.00 1.05 1.10 1.15 1.20 1.25 1.30 Time (min) 1.35 1.40 90 1.45 1.50 1.55 1.60 1.65 1.70 1.75 3.85 80 4.56 70 6.26 5.18 60 3.34 40 30 3.88 4.26 7 6.97 5.65 50 4.61 7.01 3.31 20 1.80 10 2.25 2.98 2.51 0 100 NL: 1.11E4 m/z= 101-102 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC trichlorofuoromethane 6.83 100 NL: 9.91E3 m/z= 64-65 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC chloroethane 12/5/2012 2:04:03 PM RT: 0.00 - 20.01 NL: 1.30E4 m/z= 94-95 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC 50 0 C:\chem\...\level8.d\level8 NL: 1.51E4 m/z= 62-63 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC 50 0 FIGURE 17. Comparison of run time with hyd NL: 2.67E4 m/z= 50-51 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC 50 Relative Abundance ol-2-methylphenol NL: 7.88E3 m/z= 85-86 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC 50 n the insert required nds: The run time for EPA Method 8270 was reduce x 0.36 µm) and hydrogen carrier gas. The more mm x 30 m x 0.5µm. (Figure 16). FIGURE 13. EPA Method 524 first six gases at 200 ppt with hydrogen carrier gas. 100 R2 Run Time Relative Abundance 11.09 7 6.63 7.46 90 6.14 80 70 1.80 60 5.34 Robustness and Ion Ratio Stability 7.22 5.36 6.04 50 6.76 40 The stability of ion ratios for DFTPP and BFB was studied. The values were very stable (Figure 14). Next, a robustness study was run for EPA Method 8270 by injecting 30 of 5% diesel samples in methylene chloride. The samples also contained 40 ppm of the internal standards. The diesel sample TIC is shown in Figure 15. A chart with the internal standards plotted is shown in Figure 16. A check standard was run after each set of 10 diesel samples. All check standards met the QC criteria for the method. FIGURE 14. Ion ratio stability for DFTPP (left) and BFB (right). m/z Criteria/Batch 01 02 30 3.84 20 2.82 10 2.65 0 0 1 2 3.24 3 4.17 4 5 Thermo Scientific Poster Note • EAS2013_PN10360_E 11/13S 5 03 04 05 06 07 08 09 10 11 6 7 di-n-octylphthalate bis(2 ethylhexyl)phthalate 0 RT: 1.47 AA: 16876 BP: 101 100 50 0 0.80 0.85 0.90 0.95 1.00 1.05 1.10 1.15 1.20 1.25 1.30 Time (min) 1.35 1.40 1.45 1.50 1.55 1.60 1.65 1.70 1.75 100 NL: 1.11E4 m/z= 101-102 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC trichlorofuoromethane 90 80 70 1.80 60 50 Robustness and Ion Ratio Stability 40 The stability of ion ratios for DFTPP and BFB was studied. The values were very stable (Figure 14). Next, a robustness study was run for EPA Method 8270 by injecting 30 of 5% diesel samples in methylene chloride. The samples also contained 40 ppm of the internal standards. The diesel sample TIC is shown in Figure 15. A chart with the internal standards plotted is shown in Figure 16. A check standard was run after each set of 10 diesel samples. All check standards met the QC criteria for the method. 30 3.84 20 2.82 10 2.65 0 0 1 2 3.24 3 4.17 4 FIGURE 14. Ion ratio stability for DFTPP (left) and BFB (right). m/z Criteria/Batch 01 02 03 04 05 06 07 08 09 10 11 50 15-40% 28 26 26 24 27 26 25 26 28 25 25 75 >30% 52 51 48 48 51 52 49 48 48 50 53 95 100% 96 ppm for EPA Method 525. An excerpt of nearity study was performed in the 5-9% 8.0 7.3 7.9 7.7 8.2 7.2 7.7 7.4 7.8 7.5 7.7 173 < 2% of 174 0.4 0.6 0.7 1.5 0.4 0.4 0.8 0.7 0.5 1.1 0.3 174 >50% 76 75 69 69 75 70 72 72 71 74 72 175 5-9% of 174 7.3 7.4 7.4 5.9 7.6 8.5 6.4 7.7 6.8 6.4 6.7 carrier gas. 95-101% of 96 95 100 96 93 96 98 95 96 101 96 174 176 177 5-9% of 176 5.1 6.3 6.6 5.2 6.4 6 5.1 5.5 6.6 6.5 6.0 FIGURE 15. TIC 5 % diesel sample. RT: 0.00 - 13.26 NL: 6.22E8 TIC MS diesel5 7.76 100 6.21 6.77 8.18 95 7.30 90 5.65 3.43 7.31 5.09 4.01 75 70 8.92 Relative Abundance 65 55 7.79 50 9.25 7.01 5.97 45 5.50 5.29 35 9.56 4.92 7.47 25 3.75 20 9.85 4.35 2.63 5 1.70 1.96 2.15 3.19 0 1 2 10.68 3 4 5 6 = 0.15 ppm. 7 8 9 10 12 3. EPA Method 525.2, D capillary column gas c © 2013 Thermo Fisher Scientific This information is not intended t Run Time The run time for EPA Method 8270 was reduced by using a smaller ID column (TraceGOLD TG-5MS 0.18 mm x 20 m x 0.36 µm) and hydrogen carrier gas. The more commonly used column with helium is the TraceGOLD TG-5MS 0.25 mm x 30 m x 0.5µm. (Figure 16). ier gas. NL: 7.88E3 m/z= 85-86 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC FIGURE 17. Comparison of run time with hydrogen and helium at 1 mL/min. C:\chem\...\level8.d\level8 NL: 2.67E4 m/z= 50-51 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC 12/5/2012 2:04:03 PM RT: 0.00 - 20.01 Hydrogen 1 mL/min (TraceGOLD TG-5MS 0.18 mm x 20 m, 0.36 µm) 6.83 100 90 3.85 Relative Abundance 80 6.26 5.18 60 3.34 40 30 3.88 4.26 9.19 11.19 8.06 8.41 4.61 9.76 7.01 1.80 11.44 12.55 2.25 2.98 2.51 12.70 0 100 7.58 6.63 7.46 90 8.06 6.14 80 1.80 50 11.16 10.79 8.84 3.31 20 60 12.35 10.27 7.64 6.97 5.65 50 10 NL: 5.35E8 TIC MS level8 10.24 7.90 8.98 4.56 70 8.80 9.44 10.65 10.71 8.59 9.01 in the GC-MS Analysis of VOCs and SVOCs 6 From Helium to Hydrogen: Case Study 70 1.75 EPA Method 524.2, M chromatography/mass 13 plicate injections at 0.4 ppb. The average itivity for the first six gases, even at as low 1.70 2. 11.62 12.01 12.80 13.23 FIGURE 16. Stability of internal standard areas during injection of 5% diesel. 1.65 EPA Method 8270D, S MS) REV. 4, 2007, En 11.47 10.95 11 e 1.60 1. Time (min) actor of two to that achieved with helium. 0.15 ppm with hydrogen. The chart in f the standard were made at 1 ppm. 1.55 Shorter run time 10.24 2.57 0 trichlorofuoromethane Good ion ratio stability 10.13 10 NL: 1.11E4 m/z= 101-102 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC Sensitivity was within 2.80 3.21 15 chloroethane Linearity met the crite 6.55 30 NL: 9.91E3 m/z= 64-65 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC Good resolution and i 60 40 ethane References 8.57 80 NL: 1.30E4 m/z= 94-95 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC GC-MS analysis of SVOCs wa an ISQ single quadrupole mas MS hardware and methods ne optimized EPA Method 8270 w time. The ISQ Series MS was were: 4.55 85 NL: 1.51E4 m/z= 62-63 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC Conclusion 5.34 9.56 6.04 NL: 1.12E8 TIC MS 03091102_ 110309112 925 13.21 13.84 13.15 11.27 14.62 13.96 15.37 18.08 12.57 10.02 6.76 12.43 9.61 7.22 5.36 10.89 12.13 15.33 15.85 g replicate injections at 0.4 ppb. The average ensitivity for the first six gases, even at as low Run Time The run time for EPA Method 8270 was reduced by using a smaller ID column (TraceGOLD TG-5MS 0.18 mm x 20 m x 0.36 µm) and hydrogen carrier gas. The more commonly used column with helium is the TraceGOLD TG-5MS 0.25 mm x 30 m x 0.5µm. (Figure 16). arrier gas. NL: 7.88E3 m/z= 85-86 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC FIGURE 17. Comparison of run time with hydrogen and helium at 1 mL/min. C:\chem\...\level8.d\level8 NL: 2.67E4 m/z= 50-51 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC 90 47 6876 01 1.50 1.55 Relative Abundance NL: 9.91E3 m/z= 64-65 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC 1.60 1.65 1.70 1.75 6.26 5.18 60 3.88 4.26 3.34 40 30 9.19 12.35 10.27 7.64 6.97 5.65 50 11.19 8.06 8.41 4.61 9.76 7.01 11.16 11.44 10.79 8.84 12.55 3.31 1.80 10 2.25 2.98 2.51 12.70 0 100 8.80 7.58 6.63 7.46 90 9.44 10.65 10.71 8.06 6.14 80 8.59 9.01 9.56 10.89 12.13 12.43 60 5.34 14.62 13.15 13.96 11.27 15.37 6.04 50 18.08 7.22 5.36 NL: 1.12E8 TIC MS 03091102_ 110309112 925 13.21 13.84 9.61 70 1.80 12.57 15.33 10.02 15.85 6.76 18.71 40 e very stable (Figure 14). Next, a l samples in methylene chloride. The le TIC is shown in Figure 15. A chart with run after each set of 10 diesel samples. NL: 5.35E8 TIC MS level8 10.24 7.90 8.98 4.56 70 20 NL: 1.11E4 m/z= 101-102 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC trichlorofuoromethane 3.85 80 NL: 1.30E4 m/z= 94-95 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC chloroethane Hydrogen 1 mL/min (TraceGOLD TG-5MS 0.18 mm x 20 m, 0.36 µm) 6.83 100 NL: 1.51E4 m/z= 62-63 F: {0,0} + c EI Full ms [47.00-300.00] MS ICIS 200pptC omethane 12/5/2012 2:04:03 PM RT: 0.00 - 20.01 30 3.84 20 15.92 2.82 10 2.65 0 0 1 2 3.24 3 11.32 4.17 4 5 6 7 8 9 10 Time (min) 11 12 18.91 16.42 17.11 13 14 15 16 17 18 19 20 Helium 1 mL/min (TraceGOLD TG-5MS 0.25 mm x 30 m, 0.5 µm) /Batch 01 02 03 04 05 06 07 08 09 10 11 0% 28 26 26 24 27 26 25 26 28 25 25 % 52 51 48 48 51 52 49 48 48 50 53 100% % 8.0 7.3 7.9 7.7 8.2 7.2 7.7 7.4 7.8 7.5 7.7 of 174 0.4 0.6 0.7 1.5 0.4 0.4 0.8 0.7 0.5 1.1 0.3 % 76 75 69 69 75 70 72 72 71 74 72 of 174 7.3 7.4 7.4 5.9 7.6 8.5 6.4 7.7 6.8 6.4 6.7 1% of 96 95 100 96 93 96 98 95 96 101 96 4 of 176 5.1 6.3 6.6 5.2 6.4 6 5.1 5.5 6.6 6.5 6.0 NL: 6.22E8 TIC MS diesel5 Conclusion GC-MS analysis of SVOCs was performed in accordance with EPA Method 8270 on a TRACE 1310 GC coupled with an ISQ single quadrupole mass spectrometer utilizing hydrogen carrier gases. Key modifications to both the GC and MS hardware and methods necessary for successful migration to hydrogen carrier were performed. The final, optimized EPA Method 8270 was fully migrated to hydrogen carrier gas with improved peak shape, resolution, and run time. The ISQ Series MS was stable and maintained good ion ratio stability after injecting 5% diesel samples. Results were: Good resolution and improved peak shape Linearity met the criteria of the method Sensitivity was within a factor of 2 of that obtained with helium carrier gas Good ion ratio stability and robustness Shorter run time References 1. EPA Method 8270D, Semi-volatile Organic Compounds by Gas Chromatography/Mass Spectrometry (GC/ MS) REV. 4, 2007, Environmental Protection Agency. 2. EPA Method 524.2, Measurement of purgeable organic compounds in water by capillary column gas chromatography/mass spectrometry. Rev. 4.1, 1995. 3. EPA Method 525.2, Determination of organic compounds in drinking water by liquid-solid extraction and capillary column gas chromatography/mass spectrometry. Rev.2, 1995. 13.23 13 © 2013 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others. www.thermoscientific.com ©2013 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific, Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa +43 1 333 50 34 0 Australia +61 3 9757 4300 Austria +43 810 282 206 Belgium +32 53 73 42 41 Brazil +55 11 3731 5140 Canada +1 800 530 8447 China 800 810 5118 (free call domestic) 400 650 5118 Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland +358 9 3291 0200 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9494 Italy +39 02 950 591 Japan +81 6 6885 1213 Korea +82 2 3420 8600 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00 Thermo Fisher Scientific, Sunnyvale, CA USA is ISO 9001:2008 Certified. Russia/CIS +43 1 333 50 34 0 Singapore +65 6289 1190 Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00 Taiwan +886 2 8751 6655 UK/Ireland +44 1442 233555 USA +1 800 532 4752 EAS2013_PN10360_E 11/13S
© Copyright 2025 Paperzz