18. Diamond Microelectrodes H. Olivia, B. V. Sarada, T. N. Rao and A. Fujishima 18.1. Introduction Microelectrodes have attracted much attention recently in electrochemistry due to their superior properties, which enable them to outperform conventional macroelectrodes and extend the experimental range to several new fields, such as fast-scan measurements and analysis in poorly conducting media [1-4]. The history of microelectrodes actually started more than 60 years ago, when 1942, Davies and Brink [5] reported the use of platinum microdisk electrodes for the measurement of oxygen in muscle tissues. In their work, microelectrodes were used to minimize the damage to the muscle, and to limit the current flowing through the electrode. Since then, several reviews [6-8] and books [9,10] about microelectrodes have been published. Denuault, in his review [8], defined the term ”microelectrode” as an electrode with at least one dimension in the range of 0.1 to 50 m. The small size of microelectrodes makes them possible to be used for in vivo detection, which is usually performed with very small volumes of samples, such as those for neurotransmitter monitoring in the brain. Moreover, due to its small size, at relatively long experimental timescales, the thickness of the diffusion layer is comparable to the dimensions of the microelectrode, and a spherical (or hemi-spherical) diffusion field controls the fast mass transport of reactants and products to and from the electrode surface. Accordingly, a steady-state response (or pseudo-steady state response) can be observed with cyclic voltammetry at low sweep rates. Another interesting feature of microelectrodes is their small interfacial capacitance. Capacitance decreases with electrode area, and therefore, due to its small area, microelectrodes have a reduced capacitance and hence small charging current, allowing fast and sensitive response. Furthermore, voltammetry using microelectrodes often completely eliminates iR drop, which enhances the use of media such as organic solvents [11], nonelectrolyte solutions [12-14], and even gases and solids [15,16], which are generally excluded from any measurements using macroelectrodes. The various geometries of microelectrodes include microdisks, microfibers, microarrays, microbands, and microrings. Among these, the microdisk is the most popular geometry, because of its simple fabrication and the possibility of treatment by polishing. However, the current response at microdisk electrodes is often small enough that it limits the range of measurements, augmenting the need for techniques to fabricate microfiber, microband and microarray electrodes, which provide larger signals. The most commonly used electrode materials for microelectrodes include platinum, gold and carbon. Carbon fiber microelectrodes are widely used for electroanalysis in aqueous media, as they exhibit a relatively wide potential window. However, similar to metal electrodes, carbon has several serious limitations, including high background current and deactivation via fouling, especially during the detection of compounds in complex biological fluids, as reported by Baur et al.[17]. It is an inherent property of carbon to undergo deactivation upon exposure to the laboratory environment or working solution, which is due to factors such as surface oxidation and adsorption of contaminants and reaction products. Diamond is one of the more recent of the carbon allotropes that has been examined as an electrode material. It exhibits several superior properties, including low background current, wide potential window, long-term stability, relative insensitivity towards the presence of dissolved oxygen in the solution, and biocompatibility [18-20]. Thus, diamond is becoming an interesting material to consider for electroanalysis. Cooper et al. (1998) reported for the first time the fabrication and the use of boron-doped diamond (BDD) microelectrodes in nonaqueous electrolytes [21]. Considering the advantages of BDD mentioned above, the Fujishima group undertook the application of BDD microelectrodes, especially BDD microdisk [22], microfiber [23] and microdisk array [24] electrodes in aqueous solutions. 18.2. Preparation of Diamond Microelectrodes 18.2.1. Fabrication of diamond microdisk and microfiber electrodes Diamond microfibers were prepared by depositing boron-doped diamond on electrochemically polished tungsten fibers. Diamond deposition was carried out using a microwave plasma chemical vapor deposition (MPCVD) system at a hydrogen pressure of 50-80 Torr and microwave power of 1500-3000 W for 3-8 h on tungsten fibers. Different powers and deposition times resulted in the variation of the crystal size and the film thickness, respectively. The crystal size varies from 5 to 40 m, while the film thickness varies from 5 to 20 m. Prior to deposition, the tips of the tungsten wires (= 30 m) were etched in 2 M NaOH at 3 V for 45 s in order to reduce the diameter of the fiber to ~10 m, and these tips were nucleated by ultrasonicating in solution containing a suspension of 100-nm diamond particles for 60 min. The diamond-deposited tungsten wire was then inserted into a pre-pulled glass capillary (= 50-100 m) and was sealed using epoxy. The ohmic contact to the diamond fiber was made using a copper wire with either mercury or silver paste. In the case of the microdisk electrode [22], the diamond fiber was preliminarily fully sealed by the use of epoxy, and the tip was then polished until the diamond was just exposed, while for the microfiber electrode [23], a ~300-m length of fiber was left exposed. 18.2.2. Characterization Successfully fabricated diamond fibers were characterized by use of scanning electron microscopy (SEM) and Raman spectroscopy, while the roughness factor of the diamond fiber was calculated based on double-layer capacitance measurements. SEM images of diamond fibers are shown in Fig 18.1. Figure 18.1(a) shows a suitable diamond fiber for microdisk electrode fabrication, while for the microfiber electrode, full coverage of diamond polycrystallites on the tungsten fiber was necessary (Fig 18.1(b)). Raman spectra (not shown) indicated the high quality and purity of the diamond. Fig 18.1. Suitable diamond for (a) microdisc electrode an d (b) microfiber electrode The double layer capacitance of a diamond microelectrode is calculated based on the equation Ic = C d where Ic is the charging current, is the potential sweep rate, and Cd is the double layer capacitance. By plotting Ic as a function of , the double-layer capacitance Cd can be obtained from the slope. The Cd value obtained for a diamond microfiber electrode was 8 nF, and the capacitance density was calculated to be 7.02 F cm-2. Considering the capacitance density of a smooth (100) homoepitaxial diamond electrode (ca. 3 F cm-2), the roughness factor of the diamond fiber was estimated to be 2.34. 18.3. Electrochemical Behavior 18.3.1. Electrochemical behavior of diamond microdisk electrodes The simplest way to investigate the electrochemical behavior of an electrode is by studying its cyclic voltammetric curves. Figure 18.2 shows cyclic voltammograms for the oxidation of ferrocyanide at BDD microdisk electrodes with two different radii in aqueous electrolyte. Fig 18.2. Cyclic voltammograms at diamond micro-electrodes for the oxidation of 1 mM K4Fe(CN)6 in 0.1 M KCl (potential sweep rate, 10 mV s-1); electrode radii: (a) 20 and (b) 6 μm. The sigmoidal shapes of the curves and lack of hysteresis, i.e., steady state-type behavior, is characteristic of voltammetry at low potential sweep rates for microelectrodes [1,3]. The half-wave potential was +0.210 vs. SCE. This value agrees well with that reported at conventional macro-type diamond electrodes by Jolley et al. (+0.230 V) [25]. The radius of each microelectrode was calculated from the equation ilim = 4nFDCr 18.1 where ilim is the limiting current, C is the concentration, D is the diffusion coefficient, r is the radius of the electrode, F is the Faraday constant, and n is the number of electrons, in this case, one. The radii of the microelectrodes were calculated to be 20 and 6 m using a value of 6.5 10-6 cm2s-1 for the diffusion coefficient for ferrocyanide [26]. Similar steady-state type voltammograms were also obtained for the oxidation of Ru(NH3)63+, for which a diffusion coefficient of 6.0 10-6 cm2s-1 was used [27]. Owing to the steady-state nature of the spherical diffusion at the microelectrode, the limiting current should be independent of potential sweep rates at lower sweep rates. As the sweep rate increases, the contribution of planar diffusion increases. The value of sweep rate at which planar diffusion begins to significantly interfere depends on the size of microelectrode. One of the most promising features expected for BDD microelectrodes is very low background current, due to a combination of the effect of the microelectrode size [1] plus the intrinsic properties of diamond [28]. One way this effect can be tested is by examining the detection limit for a relatively simple redox couple at slow sweep rates. Figure 18.3(a) shows a voltammogram for a BDD microelectrode (r=20 m) in a 200 nM ferrocyanide (0.1 M KCl) solution, compared with the background current. The voltammogram is very well defined, even at this low concentration, indicating its potential use for electrochemical sensor applications. Limiting currents increased linearly with increasing ferrocyanide concentration up to 1.2 M (Fig 18.3(b)). Fig 18.3(a) Cyclic voltammogram for a diamond microelectrode of radius 20μm for the oxidation of 200 nM K4Fe(CN )6 in 0.1 M KCl (sweep rate, 2 mVs-1). (b) Calibration curve for K4Fe(CN)6 oxidation in 0. 1M KCl In contrast, the high background at glassy carbon microelectrodes did not allow well-defined voltammograms to be observed at low analyte concentrations. For example, for a ferrocyanide concentration of 200 nM, the increment in the current due to the analyte was only ~25% of the background current, whereas for the BDD microelectrode of similar radius, the corresponding value was ~200%. 18.3.2. Electrochemical microfiber electrodes behavior of diamond A BDD microfiber (BDDMF) electrode was characterized by performing voltammetric experiments using an outer-sphere redox couple. Figure 18.4 shows the cyclic voltammogram for 1 mM ruthenium hexaamine trichloride at a BDDMF electrode in 0.1 M phosphate buffer (pH 7.1) at a sweep rate of 10 mV s-1. The voltammogram shows the pseudo-steady state response, a characteristic of microfiber electrodes. For a sweep rate of 100 mV s-1, a peak-shaped voltammogram was observed, indicating that planar diffusion is dominating the mass transport in the vicinity of the electrode at relatively high scan rates. Potential (vs SCE) / -400 -300 -200 mV -100 0 100 200 300 400 Current /nA (b) 90 nA (a) Fig 18. 4. Cyclic voltammogr am for (A) 1 mM ruthenium hexaamine trichloride in 0.1 M phosphate buffer and (B) 0.1 M phosphate buffer at a diamond microfiber elec trodes. Sweep rate 10 mV s-1. The current density of the fiber electrode was estimated from the following equation, given for linear sweep voltammetry at cylindrical microelectrodes[29,30]: I = (n2F2Ca/RT)(0.446p-1 + 0.335p-1.85) 18.2 where I is the diffusion current density, a is the microelectrode radius, is the potential sweep rate, and p=(nFa2/RTD)1/2 is a dimensionless parameter that characterizes the type of diffusion. In the theoretical calculation, the value of 6.0 × 10-6 cm2 s-1 was used [27] for the diffusion coefficient of ruthenium hexaamine trichloride and 25 μm for the fiber radius, giving a current density of 1380 nA mm-2. The experimental current density, calculated by considering the fiber length of 0.8 mm, is 3916.34 nA mm-2. The difference between measured and calculated current density (ca. 2.8) can be mainly attributed to the roughness factor. The roughness factor of the diamond fiber calculated from the doublelayer capacitance measurement was 2.34. The other possibility is that the rough surface of the electrode does not conform to the microfiber model, and therefore, the formula above is not strictly valid for diamond microfiber electrodes. 18.4. Electroanalytical Applications of Diamond Microelectrodes 18.4.1. Detection of H2O2 at metal-modified diamond microelectrodes Despite its several superior properties, as mentioned above, diamond has several limitations compared to metal electrodes, such as slow kinetics for reactions involving adsorption and multielectron transfer processes, including hydrogen and oxygen evolution reactions. However, since the low rates of the hydrogen and oxygen evolution reactions result in the wide potential window [31,32], this can be considered to be an advantage of using diamond, especially in aqueous media. Another important multielectron transfer reaction is the oxidation and reduction reaction of H2O2, which is generally enzymatically generated from the oxidation reactions of biological materials, such as glucose, lactate, pyruvate, and cholesterol. Therefore, the detection of H2O2 is important for a wide range of applications in the electroanalytical field. Since diamond is inactive for the oxidation and reduction reactions of H2O2, modification of the electrode is required to make diamond suitable for the enzyme-based biosensor application. Tatsuma, et al. [33] reported the use of heme peptide and horseradish peroxidase, types of redox enzymes, based on the direct electron transfer between the diamond electrode and the redox enzyme. Another promising approach is the deposition of metal nanoparticles that have catalytic activity for the H2O2 oxidation- reduction reaction. The modification of a BDDMF electrode with platinum nanoparticles and its use for H2O2 detection are discussed in the present chapter, based on the following reactions: H2O2 H2O2+2H++2ePlatinum deposition Pt Pt on O2+2H++2e2H2O diamond microelectrodes was performed electrochemically in 0.1 M H2SO4 containing 100 µM K2PtCl6 by cycling between the potentials of –0.2 V and 1.2 V at 50 mVs-1. The electrode was then dipped into 0.1 M H2SO4, and the same cycling potentials were applied until a stable cyclic voltammogram was achieved, indicating the complete cleaning of the Pt active area. The Pt active area was calculated from the charge density for the hydrogen desorption reaction [34] between 0 and –0.2 V, using a standard value of 210 C cm-2 for polycrystalline Pt [35]. We found that the Pt active area increased linearly with deposition time (Fig 18.5), and the signal-tobackground ratio (s/b) for 1 mM H2O2 achieved its maximum value for a 20-min Pt deposition (Fig. 18.6). 120 100 s/b 80 60 40 20 0 10 20 30 40 deposition time/min 50 60 Fig 18. 5. Plots of s/b value at 0.6 V as a function of Pt deposition tim e, calculated from the cyclic voltammogram for 1 mM H 2O2 at Pt-BDDMF electrode. 1.2 y = 0.0236x - 0.3995 2 R = 0.9946 Pt active area /cm 2 1 0.8 0.6 0.4 0.2 0 -0.2 10 20 30 40 50 60 deposition tim e /m in Fig 18.6. Plots of Pt active area as a function of Pt deposition time. Pt active area was calculated from the charge density of hydrogen desorption reaction in 0.1 M H 2SO4. This can be explained from the SEM images below. Figures 18.7(a) and (b) show SEM images of Pt-modified diamond microfiber electrodes after 1 h and 20 min deposition, respectively. It can be seen from both images that Pt nanoparticles were distributed uniformly on the diamond surface. From Fig. 18.7(a), after a 1-h deposition time, the amount of Pt loading was large, with an average diameter (D1h) of 500 nm, and the number of Pt particles deposited per unit real area (N1h) of ca. 8.0 107 particles cm-2 [taking into account the roughness factor of the diamond microfiber electrode [23] (ca. 2.34)]. The number of exposed surface Pt atoms was estimated to be 5.82 × 1014 Pt atoms cm-2 from the background CV. As the average distance between each particle (estimated to be 1.12 m from N1h) is becoming small, the diffusion layers of reactant were assumed to overlap each other [Fig 18.8(a)], resulting in decreased electrochemical activity. a b Fig 18.7. SEM images of Pt-BDDMF electrodes for (a) 1 hour and (b) 20 min Pt deposition time. (a) (b) Fig 18.8. (a) Overlapping diffusion layers and (b) Ideal spherical diffusion layers In contrast, after a 20-min deposition [Fig. 18.7(b)], the amount of Pt loaded on BDDMF was less (3.22 × 1013 Pt atoms cm-2) with D20min = 400 nm and = 6.0 106 particles cm-2. The average distance between each particle for this electrode (4.09 m from ) is 3.7 times greater than that for a 1-h deposition, and this distance is considered to be close to the optimum value [36], allowing an ideal spherical diffusion field to occur in the vicinity of each particle. In this case, the electrode can be considered as a micro-array of platinum particles that are uniformly distributed on the diamond microfiber electrode (Fig. 18.8(b)). Comparison experiments were performed using a bare platinum microelectrode, a Pt-deposited diamond macroelectrode, and a bare Pt macroelectrode. The s/b values for each electrode taken for 1 mM H2O2 in 0.1 M PBS are summarized in Fig 18.9. Fig 18.9. Comparison of S/B values at 0.6V calculated from cyclic voltammogram for 1 mM H 2 O2 at Pt-diamond microfiber, Pt microfiber electrodes, Pt-diam ond macroelectrodes, and Pt macroelectrodes. The Pt-modified and bare Pt microelectrodes show higher s/b values than the Pt-modified and bare Pt macroelectrodes, respectively. This can be attributed to the properties of microelectrodes: non-planar diffusion profile, low background current, and high sensitivity. Depositing the proper amount of metal catalyst on the microelectrode further effectively enhances the sensitivity of the microelectrode. Moreover, with the use of diamond as a stable supporting material for metal deposition, the resulting microelectrode overwhelmingly outperforms the platinum microelectrodes. 18.5. Summary Microelectrodes, due to their unique properties, such as small size, non-planar diffusion, small capacitance, and small current, have provided a major breakthrough in electrochemistry. Taking advantage of the superior properties of diamond, the diamond microelectrode is a promising new device for several applications in electrochemistry. Furthermore, modification of the diamond microelectrode with appropriate metal nanoparticles increases the quality of the measurements, in the terms of sensitivity, stability, and selectivity. References 1. R.J. Forster, Chem. Soc. Rev., 23 (1994) 289. 2. D.J. Wiedemann, K.T. Kawagoe, R.T. Kennedy, E.L. Ciolkowski, R.M. Wightman, Anal. Chem., 63 (1991) 2965. 3. A.M. Bond, M. Fleischmann, J. and Robinson, J.Electroanal. Chem., 168 (1984) 299. 4. K. Pihel, Q.D. Walkel, R.M. Wightman, Anal. Chem., 68 (1996) 2084. 5. P.W. Davies, F. Brink, Rev. Sci. Instum., 13 (1942) 524. 6. R.M. Wightman, Anal. Chem., 53 (1981) 1125A. 7. A.M. Bond, Analyst (Cambridge, U.K.), 119 (1994) R1. 8. G. Denuault, Chem. Ind., (1996, 16 September) 678. 9. Ultramicroelectrodes, ed. M. Fleischmann, S. Pons, D.R. Rolison & P.O. Schmidt, Morganton, NC: Datatech Systems, 1987. 10. Microelectrodes: Theory and Applications, ed. M.I. Montenegro, M.A. Queiros & J.L. Daschbach, Dortrecht: Kluwer Academic Press, NATO ASI series E, Volume 197. 11. J.O. Howell, R.M. Wightman, J. Phys. Chem., 88 (1984) 3915. 12. D.L. Luscombe, A.M. Bond, Talanta, 38 (1991) 65. 13. J.B. Cooper, A.M. Bond, Anal. Chem., 65 (1993) 2724. 14. A.A. Bumber, Y.N. Nechitailov, I.A. Profatilova, E.G. Ignatenko, J. Anal. Chem., 55 (2000) 666. 15. R. Brina, S. Pons, M. Fleischmann, J. Electroanal. Chem., 244 (1988) 81. 16. T.T. Wooster, M.L. Longmire, H. Zhang, M. Watanabe, R.W. Murray, Anal. Chem., 64 (1992) 1132. 17. J.E. Baur, E.W. Kristensen, L.J. May, D.J. Wiedemann, R.M. Wightman, Anal. Chem., 60 (1988) 1268. 18. M. C. Granger, J. Xu, J.W. Strojek, G.M. Swain, Anal. Chim. Acta, 397(1999) 145. 19. C. Prado, G. Flechsig, P. Grundler, J. S. Ford, F. Marken, and R. G. Compton, Analyst (Cambridge, U.K.), 127 (2002) 329. 20. T. N. Rao, I. Yagi, T. Miwa, D. A. Tryk, and A. Fujishima, Anal. Chem., 71 (1999) 2506. 21. J.B. Cooper, S. Pang, S. Albin, J. Zheng, R.M. Johnson, R.M., Anal. Chem., 70 (1998) 464. 22. B.V. Sarada, B.V., T.N. Rao, D.A. Tryk, A. Fujishima, J. Electrochem. Soc., 146 (1999) 1469. 23. H. Olivia, B.V. Sarada, D. Shin, T.N. Rao, A. Fujishima, Analyst (Cambridge, U.K.), 127 (2002) 1572. 24. K. Tsunozaki, Y. Einaga, T.N. Rao, A. Fujishima, Chem. Lett., (2002) 502. 25. S. Jolley, M. Koppang, J. Jackson, G.M. Swain, Anal. Chem., 69 (1997) 4099. 26. J. Koryta, J. Dvorak, L. Kavan, Principles of Electrochemistry, p. 116, John Wiley & Sons, Chichester, 1993. 27. K.R. Wehmeyer, R.M. Wightman, Anal. Chem., 57 (1985) 1989. 28. G.M. Swain, R. Ramesham, R., Anal. Chem., 65 (1993) 345. 29. L. Agui, L., J.E.L. Guzman, A. Gonzales, A., P. Yanez, J.M. Pingarron, Anal. Chim. Acta, 385 (1999) 241. 30. K. Aoki, K. Honda, K. Tokuda, H. Matsuda, J. Electroanal. Chem., 182 (1985) 267. 31. T. Yano, E. Popa, D.A. Tryk, K. Hashimoto, A. Fujishima, J. Electrochem. Soc., 146 (1999) 1081. 32. C.H. Goeting, J.S. Foord, F. Marken, R.G. Compton, Diamond Relat. Mater., 8 (1999) 824. 33. T. Tatsuma, H. Mori, A. Fujishima, Anal. Chem., 72 (2000) 2919. 34. K. Honda, M. Yoshimura, T.N. Rao, D.A. Tryk, A. Fujishima, K. Yasui, Y. Sakamoto, K. Nishio, H. Masuda, J. Electroanal. Chem., 514 (2001) 35. 35. T. Biegler, D.A.J. Rand, R. Woods, R., J. Electroanal. Chem., 21 (1971) 269. 36. H.J. Lee, C. Beriet, R. Ferrigno, H.H. Girault, J. Electroanal. Chem., 502 (2001) 138.
© Copyright 2026 Paperzz