ILab 600 - 650 instructions rev 425.0.2 - 2010-09-01 Chema Diagnostica Via Campania 2/4 I-60030 Monsano | An Italy T +39 0731 605064 | F +39 0731 605672 www.chema.com | [email protected] Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 ILab 600 - 650 instructions rev 425.0.2 - 2010-09-01 Warning! These programs should be used only as guideline. To check the reagent correct setting on the instrument, use only good quality control sera and work accordingly to with good laboratory practice. Instrumental factor must be checked. Chema Diagnostica will not be responsible for instrumental bad programming. IUS.425.0.2 Directions for use Edition 2010-09-01 page 3 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Ilab 600 - 650 ACID PHOSPHATASE - Codes AC 0120 TC POWDER SINGLE REAGENT. FOLLOW PREPARATION PROGRAM INDICATED IN INSERT SHEET. INSTALL AS R1 REFRIGERATE AT 2-8°C Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** ACP-T Serum U/L, 1 Standard OFF Rate, 19/33 1 Wavelenght 405 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 30 0/0 15 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Serum Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 0.0 - 4.7 ** ** 0 - 75 *** Positive Above, 3000 N/A 40 1/0 OFF Calib. Conditions (option 1: K-factor) Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 300/0, yes *** 10 N/A N/A N/A N/A Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Note: * ** *** N/A IUS.425.0.2 Factor: 743, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Calib. Conditions (option 2: calibrator) Reagent Volumes R1 R2 Ranges and Eval. Criteria Directions for use 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 4 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Ilab 600 - 650 Application: Preparation: Storage: Stability: ALBUMIN - Code BC 0100 - 0500 - 1000 - 1500 CH R1 - LIQUID READY TO USE SINGLE REAGENT ROOM TEMPERATURE OR REFRIGERATE (2-30°C) AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** ALB Serum g/dl, 1 Standard OFF End Point, 8 2 Wavelenght 600/700 Ranges and Eval. Criteria Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report Serum 3.8 - 5.1 3.8 - 5.1 ** 0.5 - 8.0 *** Positive Above, 4000 N/A N/A 1/0 OFF Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 3 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 300/0, yes *** 60 N/A N/A N/A N/A Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 5 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 ALKALINE PHOSPHATASE FL DGKC - Codes AL F080 / F245 / F400 / F600 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** ALP Serum U/L, O Standard OFF Rate, 21/27 1 Wavelenght 405 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 5 0/0 3 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Serum Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 64 - 306 80 - 306 ** 5 - 2000 *** Positive Above, 4000 N/A 40 1/0 OFF Calib. Conditions (option 1: K-factor) Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 250/0, yes *** 14 ** 62/0, yes *** 14 Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Note: * ** *** N/A IUS.425.0.2 Factor: 3.433, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Calib. Conditions (option 2: calibrator) Reagent Volumes R1 R2 Ranges and Eval. Criteria Directions for use 1 Point, Linear, 3 Reps 14 Autocal H, * *** *** *** *** N/A ON see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 6 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 ALKALINE PHOSPHATASE FL IFCC - Codes AF F080 / F245 / F400 / F600 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** ALP Serum U/L, O Standard OFF Rate, 21/27 1 Wavelenght 405 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 5 0/0 3 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Serum Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 35 - 104 40 - 129 ** 5 - 3000 *** Positive Above, 4000 N/A 40 1/0 OFF Calib. Conditions (option 1: K-factor) Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 250/0, yes *** 14 ** 62/0, yes *** 14 Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Note: * ** *** N/A IUS.425.0.2 Factor: 3.433, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Calib. Conditions (option 2: calibrator) Reagent Volumes R1 R2 Ranges and Eval. Criteria Directions for use 1 Point, Linear, 3 Reps 14 Autocal H, * *** *** *** *** N/A ON see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 7 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Ilab 600 - 650 Application: Preparation: Storage: Stability: AMYLASE FL - Codes AM F060 / F120 / F245 CH R1 - LIQUID READY TO USE SINGLE REAGENT REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points ** AMY Serum U/L, 0 ** AMY UR Urine U/L, 0 Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. Standard OFF Rate 6/17 1 Wavelenght 405 Standard OFF Rate 6/17 1 Wavelenght 405 8 0/0 8 10/90 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 8 20/80 8 10/90 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 0 - 96 0 - 96 ** 5 - 3000 *** Positive Above, 4000 N/A 40 1/0 OFF 0 - 480 0 - 480 ** 5 - 15000 *** Positive Above, 4000 N/A 40 5/0 OFF Factor: 3.178, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Factor: 3.178, 3 Reps N/A N/A *** N/A N/A N/A N/A ON 1 Point, Linear, 3 Reps 30 Autocal H, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 5, 3 Reps N/A ** N/A N/A N/A N/A N/A ON Calib. Conditions (option 1: K-factor) Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Calib. Conditions (option 2: calibrator) Reagent Volumes R1 R2 Ranges and Eval. Criteria Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 320/0, yes *** 30 N/A N/A N/A N/A ** 320/0, yes *** 30 N/A N/A N/A N/A Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle NOTA: il metodo CNP-G3 non é sempre disponibile tra i valori del calibratore. In mancanza, utilizzare il valore EPS, programmando i parametri come di seguito: Normal range-Male Normal range-Female Valid Range Note: * ** *** N/A IUS.425.0.2 Directions for use 28 - 100 28 - 100 2 - 1500 see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 8 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 AMYLASE EPS FL - Codes EA F080 / F245 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points ** AMY Serum U/L, 0 ** AMY UR Urine U/L, 0 Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. Standard OFF Rate 21/27 1 Wavelenght 405 Standard OFF Rate 21/27 1 Wavelenght 405 10 0/0 10 10/90 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 10 20/80 10 10/90 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 28 - 100 28 - 100 ** 5 - 1500 *** Positive Above, 4000 N/A 40 1/0 OFF 0 - 460 0 - 160 ** 5 - 7500 *** Positive Above, 4000 N/A 40 5/0 OFF Factor: 3480, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Factor: 3480, 3 Reps N/A N/A *** N/A N/A N/A N/A ON 1 Point, Linear, 3 Reps 30 Autocal H, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 5, 3 Reps N/A ** N/A N/A N/A N/A N/A ON Calib. Conditions (option 1: K-factor) Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Calib. Conditions (option 2: calibrator) Reagent Volumes R1 R2 Ranges and Eval. Criteria Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 240/0, yes *** 30 ** 60/0, yes *** 30 ** 240/0, yes *** 30 ** 60/0, yes *** 30 Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 9 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 PANCREATIC ISOAMYLASE FL - Code PA F080 / F245 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points ** PANC Serum U/L, 0 ** PANC UR Urine U/L, 0 Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. Standard OFF Rate 21/27 1 Wavelenght 405 Standard OFF Rate 21/27 1 Wavelenght 405 6 0/0 6 10/90 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 6 20/80 6 10/90 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 13 - 53 13 - 53 ** 5 - 2500 *** Positive Above, 4000 N/A 40 1/0 OFF 0 - 350 0 - 350 ** 5 - 12500 *** Positive Above, 4000 N/A 40 5/0 OFF Factor: 6280, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Factor: 6280, 3 Reps N/A N/A *** N/A N/A N/A N/A ON 1 Point, Linear, 3 Reps 30 Autocal H, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 5, 3 Reps N/A ** N/A N/A N/A N/A N/A ON Calib. Conditions (option 1: K-factor) Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Calib. Conditions (option 2: calibrator) Reagent Volumes R1 R2 Ranges and Eval. Criteria Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 240/0, yes *** 30 ** 60/0, yes *** 30 ** 240/0, yes *** 30 ** 60/0, yes *** 30 Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 10 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Ilab 600 - 650 Application: Preparation: Storage: Stability: BICARBONATE FL - Codes BR F060 / F245 / F400 CH R1 - LIQUID READY TO USE SINGLE REAGENT ROOM TEMPERATURE OR REFRIGERATE (2-30°C) AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** BR Serum mmol/l, 1 Standard OFF End Point, 32 2 Wavelenght 415/505 Ranges and Eval. Criteria Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report Serum 22 - 29 22 - 29 ** 1 - 50 *** Positive Above, 4000 N/A N/A 1/0 OFF Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 3 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle 1 Point, Linear, 3 Reps 10 Standard, * *** *** *** *** N/A ON Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 300/0, yes *** 60 N/A N/A N/A N/A Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 11 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 BILIRUBIN TOTAL FL - Codes DT F125 / F500 CH R1 - USE LIQUID READY TO USE REAGENT A R2 - USE LIQUID READY TO USE REAGENT A REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** BILT FL Serum mg/dl, 2 Standard OFF End Point, 16/32 1 Wavelenght 510 Ranges and Eval. Criteria Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report Serum 0.10 - 1.20 0.10 - 1.20 ** 0.10 - 20.00 *** Positive Above, 4000 N/A N/A 1/0 OFF Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 15 0/0 10 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle 1 Point, Linear, 3 Reps 7 Autocal H, * *** *** *** *** N/A ON Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 300/0, yes *** 10 ** 75/0, yes *** 10 Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 12 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 BILIRUBIN DIRECT FL - Codes DD F125 / F500 CH R1 - USE LIQUID READY TO USE REAGENT A R2 - USE LIQUID READY TO USE REAGENT A REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** BILD FL Serum mg/dl, 2 Standard OFF End Point, 16/32 1 Wavelenght 546 Ranges and Eval. Criteria Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report Serum 0.00 - 0.2 0.00 - 0.2 ** 0.00 - 13.00 *** Positive Above, 4000 N/A N/A 1/0 OFF Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 15 0/0 10 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle 1 Point, Linear, 3 Reps 7 Autocal H, * *** *** *** *** N/A ON Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 300/0, yes *** 10 ** 75/0, yes *** 10 Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 13 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Ilab 600 - 650 Application: Preparation: Storage: Stability: CALCIUM ASX - Code CA 0100 - 0500 CH R1 - LIQUID READY TO USE SINGLE REAGENT ROOM TEMPERATURE OR REFRIGERATE (2-30°C) AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points ** CA Serum mg/dl, 2 ** CA UR Urine mg/dl, 0 Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. Standard OFF End Point, 28 2 Wavelenght 660/700 Standard OFF End Point, 28 2 Wavelenght 660/700 3 0/0 3 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 3 20/80 3 10/90 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 ** 300/0, yes *** 30 N/A N/A N/A N/A ** 300/0, yes *** 30 N/A N/A N/A N/A Ranges and Eval. Criteria Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 8.80 - 10.80 8.80 - 10.80 ** 1.0 - 20.0 *** Positive Above, 4000 N/A N/A 1/0 OFF ** ** ** 2 - 120 *** Positive Above, 4000 N/A N/A 1/0 OFF 1 Point, Linear, 3 Reps 3 Autocal H, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 5, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 14 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 CALCIUM - Code CA 0305 - 0505 CH R1 - INSTALL LIQUID READY TO USE REAGENT B R2 - INSTALL LIQUID READY TO USE REAGENT A REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points ** CA Serum mg/dl, 2 ** CA UR Urine mg/dl, 0 Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. Standard OFF End Point, 28 2 Wavelenght 570/660 Standard OFF End Point, 28 2 Wavelenght 570/660 6 0/0 3 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 6 20/80 6 10/90 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 ** 180/0, yes *** 30 ** 180/0, yes *** 30 ** 180/0, yes *** 30 ** 180/0, yes *** 30 Ranges and Eval. Criteria Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 8.80 - 10.80 8.80 - 10.80 ** 1.0 - 20.0 *** Positive Above, 4000 N/A N/A 1/0 OFF ** ** ** 2 - 120 *** Positive Above, 4000 N/A N/A 1/0 OFF 1 Point, Linear, 3 Reps 3 Autocal H, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 5, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 15 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Ilab 600 - 650 Application: Preparation: Storage: Stability: CHOLESTEROL FL - Codes CT F100 / F400 / 100F / 150F CH R1 - LIQUID READY TO USE SINGLE REAGENT REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** CHOL Serum mg/dl, 0 Standard OFF End Point, 33 2 Wavelenght 510/660 Ranges and Eval. Criteria Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report Serum 140 - 200 140 - 200 ** 10 - 800 *** Positive Above, 4000 N/A N/A 1/0 OFF Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 3 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 300/0, yes *** 60 N/A N/A N/A N/A Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 16 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 HDL DIRECT FL - Code HD F080 / F245 / F400 CH R1 - USE LIQUID RFADY TO USE REAGENT A R2 - USE LIQUID RFADY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** HDL-C Serum mg/dl, 1 Standard OFF End Point, 33 2 Wavelenght 600/700 Ranges and Eval. Criteria Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report Serum 35 - 79 42 - 88 ** 5 - 220 *** Positive Above, 3500 N/A N/A 1/0 OFF Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 3 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 270/0, yes *** 60 ** 90/0, yes *** 60 Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 17 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 LDL DIRECT FL - Code DL F080 CH R1 - USE LIQUID RFADY TO USE REAGENT A R2 - USE LIQUID RFADY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** LDL-C Serum mg/dl, 1 Standard OFF End Point, 33 2 Wavelenght 600/700 Ranges and Eval. Criteria Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report Serum 76 - 218 76 - 218 ** 5 - 400 *** Positive Above, 3500 N/A N/A 1/0 OFF Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 3 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 270/0, yes *** 60 ** 90/0, yes *** 60 Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 18 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 CHOLINESTERASE FL (DGKC) - Codes CH F096 - F245 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** CHE Serum U/L, O Standard OFF Rate, 22/27 1 Wavelenght 405 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 5 0/0 3 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Serum Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 5600 - 11200 4200 - 10800 ** 50 - 25000 *** Negative Below, 2500 N/A 40 1/0 OFF Calib. Conditions (option 1: K-factor) Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 250/0, yes *** 14 ** 50/0, yes *** 14 Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Note: * ** *** N/A IUS.425.0.2 Factor: -65.800, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Calib. Conditions (option 2: calibrator) Reagent Volumes R1 R2 Ranges and Eval. Criteria Directions for use 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 19 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 CK-NAC FL - Codes CK F060 / F120 / F245 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** CKNAC Serum U/L, O Standard OFF Rate, 21/29 1 Wavelenght 340 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 12 0/0 6 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Serum Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 0 - 190 0 - 165 ** 5 - 1600 *** Positive Above, 2500 N/A 40 1/0 OFF Calib. Conditions (option 1: K-factor) Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 240/0, yes *** 14 ** 60/0, yes *** 14 Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Note: * ** *** N/A IUS.425.0.2 Factor: 4.127, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Calib. Conditions (option 2: calibrator) Reagent Volumes R1 R2 Ranges and Eval. Criteria Directions for use 1 Point, Linear, 3 Reps 30 Autocal H, * *** *** *** *** N/A ON see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 20 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 CK-MB FL - Code MB F060 / F120 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** CKMB Serum U/L, O Standard OFF Rate, 21/29 1 Wavelenght 340 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 12 0/0 6 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Serum Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 0 - 24 0 - 24 ** 2 - 1600 *** Positive Above, 3000 N/A 40 1/0 OFF Calib. Conditions (option 1: K-factor) Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 240/0, yes *** 14 ** 60/0, yes *** 14 Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Note: * ** *** N/A IUS.425.0.2 Factor: 8.254, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Calib. Conditions (option 2: calibrator) Reagent Volumes R1 R2 Ranges and Eval. Criteria Directions for use 1 Point, Linear, 3 Reps 30 MB Cal, * *** *** *** *** N/A ON see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 21 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 COPPER - Code CU 0100 CH R1 - INSTALL LIQUID READY TO USE REAGENT B R2 - INSTALL LIQUID READY TO USE REAGENT A ROOM TEMPERATURE (Warning! The reagent A could precipitate if refrigerate) AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points ** CU Serum microg/dl, 0 ** CU UR Urine microg/dl, 0 Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. Standard OFF End Point, 28 2 Wavelenght 570/660 Standard OFF End Point, 28 2 Wavelenght 570/660 25 0/0 10 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 25 0/0 10 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 ** 188/0, yes *** 30 ** 188/0, yes *** 30 ** 188/0, yes *** 30 ** 188/0, yes *** 30 Ranges and Eval. Criteria Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 70 - 140 80 - 155 ** 1 - 500 *** Positive Above, 4000 N/A N/A 1/0 OFF ** ** ** 1 - 500 *** Positive Above, 4000 N/A N/A 1/0 OFF 1 Point, Linear, 3 Reps 3 XXX * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 5, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 22 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 CREATININE - Code CR 0500 / 1000 / 1500 CH R1 - INSTALL LIQUID READY TO USE REAGENT B R2 - INSTALL LIQUID READY TO USE REAGENT A ROOM TEMPERATURE OR REFRIGERATE (2-30°C) AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points ** CREA Serum mg/dl, 2 ** CREA UR Urine mg/dl, 1 Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. Standard OFF End Point, 19/23 2 Wavelenght 510/570 Standard OFF End Point, 19/23 2 Wavelenght 510/570 20 0/0 10 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 20 8/395 20 4/395 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 ** 150/0, yes *** 10 ** 150/0, yes *** 10 ** 150/0, yes *** 10 ** 150/0, yes *** 10 Ranges and Eval. Criteria Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 0.60 - 1.20 0.50 - 1.00 ** 0.10 - 20.0 *** Positive Above, 3500 N/A N/A 1/0 OFF ** ** ** 2 - 800 *** Positive Above, 3500 N/A N/A 1/0 OFF 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 50, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 23 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 GAMMA-GT FL - Codes GT F080 / F245 / F400 / F600 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** GGT Serum U/L, O Standard OFF Rate, 21/27 1 Wavelenght 405 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 30 0/0 15 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Serum Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 10 - 47 7 - 30 ** 2 - 700 *** Positive Above, 4000 N/A 40 1/0 OFF Calib. Conditions (option 1: K-factor) Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 240/0, yes *** 14 ** 60/0, yes *** 14 Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Note: * ** *** N/A IUS.425.0.2 Factor: 1.158, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Calib. Conditions (option 2: calibrator) Reagent Volumes R1 R2 Ranges and Eval. Criteria Directions for use 1 Point, Linear, 3 Reps 14 Autocal H, * *** *** *** *** N/A ON see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 24 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Ilab 600 - 650 Application: Preparation: Storage: Stability: GLUCOSE FL - Code GL F400 / 100F / 150F CH R1 - LIQUID READY TO USE SINGLE REAGENT REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points ** GLUC Serum mg/dl, 0 ** GLUC UR Urine mg/dl, 0 Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. Standard OFF End Point, 33 2 Wavelenght 510/660 Standard OFF End Point, 33 2 Wavelenght 510/660 3 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 3 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 ** 300/0, yes *** 14 N/A N/A N/A N/A ** 300/0, yes *** 14 N/A N/A N/A N/A Ranges and Eval. Criteria Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 60 - 110 60 - 110 ** 10 - 500 *** Positive Above, 3500 N/A N/A 1/0 OFF ** ** ** 10 - 500 *** Positive Above, 3500 N/A N/A 1/0 OFF 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 1, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 25 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 GLUCOSE UV FL - Code GL F601 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points ** GLUC Serum mg/dl, 0 ** GLUC UR Urine mg/dl, 0 Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. Standard OFF End Point, 33 2 Wavelenght 340/660 Standard OFF End Point, 33 2 Wavelenght 340/660 3 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 3 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 ** 240/0, yes *** 14 ** 60/0, yes *** 14 ** 240/0, yes *** 14 ** 60/0, yes *** 14 Ranges and Eval. Criteria Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 60 - 110 60 - 110 ** 10 - 500 *** Positive Above, 3500 N/A N/A 1/0 OFF ** ** ** 10 - 700 *** Positive Above, 3500 N/A N/A 1/0 OFF 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 1, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 26 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 GOT/AST FL - Codes GO F080 / F245 / F400 / F600 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** GOT Serum U/L, O Standard OFF Rate, 21/27 1 Wavelenght 340 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 30 0/0 15 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Serum Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 0 - 40 0 - 31 ** 2 - 400 *** Negative Below, 2000 N/A 40 1/0 OFF Calib. Conditions (option 1: K-factor) Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 240/0, yes *** 14 ** 60/0, yes *** 14 Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Note: * ** *** N/A IUS.425.0.2 Factor: -1.746, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Calib. Conditions (option 2: calibrator) Reagent Volumes R1 R2 Ranges and Eval. Criteria Directions for use 1 Point, Linear, 3 Reps 14 Autocal H, * *** *** *** *** N/A ON see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 27 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 GPT/ALT FL - Codes GP F080 / F245 / F400 / F600 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** GPT Serum U/L, O Standard OFF Rate, 21/27 1 Wavelenght 340 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 30 0/0 15 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Serum Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 0 - 42 0 - 32 ** 2 - 400 *** Negative Below, 2000 N/A 40 1/0 OFF Calib. Conditions (option 1: K-factor) Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 240/0, yes *** 14 ** 60/0, yes *** 14 Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Note: * ** *** N/A IUS.425.0.2 Factor: -1.746, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Calib. Conditions (option 2: calibrator) Reagent Volumes R1 R2 Ranges and Eval. Criteria Directions for use 1 Point, Linear, 3 Reps 14 Autocal H, * *** *** *** *** N/A ON see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 28 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Ilab 600 - 650 Application: Preparation: Storage: Stability: IRON FZ - Code FE F245 / F400 CH R1 - USE LIQUID RFADY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B (preparation: read insert sheet) REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** IRON Serum microg/dl, 1 Standard OFF End Point, 16/33 2 Wavelenght 600/700 Ranges and Eval. Criteria Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report Serum 59 - 158 37 - 145 ** 5 - 1000 *** Positive Above, 3500 N/A N/A 1/0 OFF Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 40 0/0 20 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 240/0, yes *** 15 ** 60/0, yes *** 15 Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 29 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 LDH FL - Codes LD F060 / F120 / F245 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** LDH Serum U/L, O Standard OFF Rate, 21/27 1 Wavelenght 340 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 3 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Serum Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 225 - 450 225 - 450 ** 10 - 4000 *** Negative Below, 300 N/A 40 1/0 OFF Calib. Conditions (option 1: K-factor) Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 240/0, yes *** 14 ** 60/0, yes *** 14 Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Note: * ** *** N/A IUS.425.0.2 Factor: -16.030, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Calib. Conditions (option 2: calibrator) Reagent Volumes R1 R2 Ranges and Eval. Criteria Directions for use 1 Point, Linear, 3 Reps 14 Autocal H, * *** *** *** *** N/A ON see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 30 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 LIPASE FL - Code LP F060 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** LIP Serum U/L, O Standard OFF Rate, 21/31 1 Wavelenght 570 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 3 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Ranges and Eval. Criteria Serum Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 0 - 63 0 - 63 ** 1 - 250 *** Positive Above, 4000 N/A 40 1/0 OFF Calib. Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle 1 Point, Linear, 3 Reps 14 LIP Calib, * *** *** *** *** N/A ON Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 240/0, yes *** 14 ** 60/0, yes *** 14 Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 31 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 MAGNESIUM - Code MG 0200 / 0500 CH R1 - INSTALL LIQUID READY TO USE REAGENT B R2 - INSTALL LIQUID READY TO USE REAGENT A ROOM TEMPERATURE OR REFRIGERATE (2-30°C) AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points ** MG Serum mEq/l, 2 ** MG UR Urine mEq/l, 1 Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. Standard OFF End Point, 28 2 Wavelenght 510/660 Standard OFF End Point, 28 2 Wavelenght 510/660 4 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 4 50/50 4 10/90 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 ** 200/0, yes *** 10 ** 200/0, yes *** 10 ** 200/0, yes *** 10 ** 200/0, yes *** 10 Ranges and Eval. Criteria Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 1.30 - 2.10 1.30 - 2.10 ** 0.10 - 8.00 *** Positive Above, 4000 N/A N/A 1/0 OFF ** ** ** 0.1 - 15.0 *** Positive Above, 4000 N/A N/A 1/0 OFF 1 Point, Linear, 3 Reps 3 Autocal H, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 2, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 32 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Ilab 600 - 650 Application: Preparation: Storage: Stability: MAGNESIUM XL - Codes MX 0300 / 0500 CH R1 - LIQUID READY TO USE SINGLE REAGENT REFRIGERATE (2-8°C) AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** MX Serum mEq/l, 2 ** MX UR Urine mEq/l, 1 Standard OFF End Point, 30 2 Wavelenght 546/660 Standard OFF End Point, 30 2 Wavelenght 546/660 3 0/0 3 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 3 20/80 3 10/90 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 ** 300/0, yes *** 30 N/A N/A N/A N/A ** 300/0, yes *** 30 N/A N/A N/A N/A Ranges and Eval. Criteria Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 1.30 - 2.10 1.30 - 2.10 ** 0.30 - 6.00 *** Positive Above, 4000 N/A N/A 1/0 OFF ** ** ** 0.3 - 15.0 *** Positive Above, 4000 N/A N/A 1/0 OFF 1 Point, Linear, 3 Reps 3 Autocal H, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 2, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 33 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Ilab 600 - 650 Application: Preparation: Storage: Stability: PHOSPHORUS - Code PH 0100 / 0500 CH R1 - LIQUID READY TO USE SINGLE REAGENT REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points ** PHOS Serum mg/dl, 1 ** PHOS UR Urine mg/dl, 0 Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. Standard OFF End Point, 33 2 Wavelenght 340/375 Standard OFF End Point, 33 2 Wavelenght 340/375 3 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 3 10/90 3 5/95 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 ** 300/0, yes *** 30 N/A N/A N/A N/A ** 300/0, yes *** 30 N/A N/A N/A N/A Ranges and Eval. Criteria Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 3.0 - 4.5 3.0 - 4.5 ** 1.0 - 15.0 *** Positive Above, 4000 N/A N/A 1/0 OFF ** ** ** 2 - 120 *** Positive Above, 4000 N/A N/A 1/0 OFF 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 10, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 34 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Ilab 600 - 650 Application: Preparation: Storage: Stability: PROTEINS HS - Code HS 0100 / 0500 CH R1 - LIQUID READY TO USE SINGLE REAGENT ROOM TEMPERATURE OR REFRIGERATE (2-30°C) AS INDICATED IN THE LABEL Photometric Test Parameters Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** HS Urine mg/dl, 0 Standard OFF End Point, 33 2 Wavelenght 600/700 Ranges and Eval. Criteria Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report Urine 28 - 141 28 - 141 ** 10 - 500 *** Positive Above, 4000 N/A N/A 1/0 OFF Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 2 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle 1 Point, Linear, 3 Reps 10 Standard, * *** *** *** *** N/A ON Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 400/0, yes *** 30 N/A N/A N/A N/A Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 35 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Ilab 600 - 650 Application: Preparation: Storage: Stability: TOTAL PROTEINS - Codes TP 0100 / 0500 / 1000 / 1500 CH R1 - LIQUID READY TO USE SINGLE REAGENT ROOM TEMPERATURE OR REFRIGERATE (2-30°C) AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** TP Serum g/dl, 1 Standard OFF End Point, 33 2 Wavelenght 546/700 Ranges and Eval. Criteria Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report Serum 6.3 - 8.4 6.3 - 8.4 ** 0.5 - 12.0 *** Positive Above, 4000 N/A N/A 1/0 OFF Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 3 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 300/0, yes *** 30 N/A N/A N/A N/A Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 36 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Ilab 600 - 650 Application: Preparation: Storage: Stability: TRIGLYCERIDES FL - Code TR F100 / F400 / 100F / 150F CH R1 - LIQUID READY TO USE SINGLE REAGENT REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** TRIG Serum mg/dl, 0 Standard OFF End Point, 33 2 Wavelenght 510/660 Ranges and Eval. Criteria Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report Serum 10 - 190 10 - 190 ** 2 - 1000 *** Positive Above, 4000 N/A N/A 1/0 OFF Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS 3 0/0 2 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle 1 Point, Linear, 3 Reps 20 Autocal H, * *** *** *** *** N/A ON Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) ** 300/0, yes *** 60 N/A N/A N/A N/A Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 37 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 UREA UV FL - Codes AZ F080 / F245 / F400 / F600 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points ** UREA Serum mg/dl, 0 ** UREA UR Urine mg/dl, 0 Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. Standard OFF End Point, 19/22 1 Wavelenght 340 Standard OFF End Point, 19/22 1 Wavelenght 340 3 0/0 10 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 3 8/395 2 4/395 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 ** 240/0, yes *** 14 ** 60/0, yes *** 14 ** 240/0, yes *** 14 ** 60/0, yes *** 14 Ranges and Eval. Criteria Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 10 - 50 10 - 50 ** 2 - 300 *** Negative Below, 250 N/A N/A 1/0 OFF ** ** ** 2 - 10000 *** Negative Below, 250 N/A N/A 1/0 OFF 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 50, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 38 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 URIC ACID AOX FL - Codes AX F100 / F250 / F600 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. ** AOX Serum mg/dl, 1 ** AOX UR Urine mg/dl, 0 Standard OFF End Point, 28 2 Wavelenght 546/660 Standard OFF End Point, 28 2 Wavelenght 546/660 12 0/0 6 10/90 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 12 20/80 12 10/90 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 ** 240/0, yes *** 30 ** 60/0, yes *** 30 ** 240/0, yes *** 30 ** 60/0, yes *** 30 Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Ranges and Eval. Criteria Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 3.5 - 7.2 2.6 - 6.0 ** 1.0 - 35.0 *** Positive Above, 4000 N/A N/A 1/0 OFF ** ** ** 2 - 120 *** Positive Above, 4000 N/A N/A 1/0 OFF 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 5, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 39 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Ilab 600 - 650 Application: Preparation: Storage: Stability: URIC ACID T FL - Code AU F402 CH R1 - MIX REAGENT A + REAGENT B AS INDICATED IN INSERT SHEET REFRIGERATE AT 2-8°C 90 DAYS Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points ** URIC Serum mg/dl, 1 ** URIC UR Urine mg/dl, 0 Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. Standard OFF End Point, 28 2 Wavelenght 546/660 Standard OFF End Point, 28 2 Wavelenght 546/660 12 0/0 6 10/90 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 12 20/80 12 10/90 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 ** 300/0, yes *** 30 N/A N/A N/A N/A ** 300/0, yes *** 30 N/A N/A N/A N/A Ranges and Eval. Criteria Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 3.5 - 7.0 2.4 - 5.7 ** 1.0 - 25.0 *** Positive Above, 4000 N/A N/A 1/0 OFF ** ** ** 2 - 120 *** Positive Above, 4000 N/A N/A 1/0 OFF 1 Point, Linear, 3 Reps 10 Autocal H, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 5, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 40 of 41 Chema Diagnostica - Guidelines for automatic analyzers - Ilab 600 - 650 Analyzer: Application: Preparation: Storage: Stability: Ilab 600 - 650 ZINC - Code ZN 0125 CH R1 - INSTALL LIQUID READY TO USE REAGENT A R2 - INSTALL LIQUID READY TO USE REAGENT B REFRIGERATE AT 2-8°C AS INDICATED IN THE LABEL Photometric Test Parameters Serum Urine Test no. Test Name, Test Code Sample Type Reporting Units, Dec. Points ** ZN Serum microg/dl, 0 ** ZN UR Urine microg/dl, 0 Reaction Cycle Twin Analysis Method. Type, Meas. Point Photometric Methodology Primary/Secondary Wav. Standard OFF End Point, 28 2 Wavelenght 570/660 Standard OFF End Point, 28 2 Wavelenght 570/660 15 0/0 5 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 15 0/0 5 0/0 *** Water N/A Sampling 1 *** *** Sampling 2 *** N/A Sampling 2 ** 240/0, yes *** 30 ** 240/0, yes *** 30 ** 240/0, yes *** 30 ** 240/0, yes *** 30 Ranges and Eval. Criteria Serum Urine Normal range-Male Normal range-Female Normal range-Other Valid Range Hem./Icterus/Lip. Lmt. Reaction Slope Absorbance Limit, Prozone Limit Non Linear Limit Slope/Intercept Corr. Qualitative Report 70 - 150 70 - 150 ** 1 - 1000 *** Positive Above, 4000 N/A N/A 1/0 OFF ** ** ** 1 - 1000 *** Positive Above, 4000 N/A N/A 1/0 OFF 1 Point, Linear, 3 Reps 3 XXX, * *** *** *** *** N/A ON K-Accept, Partner**, Coeff: 5, 3 Reps N/A N/A *** N/A N/A N/A N/A ON Sampling Conditions Sampling 1 Smp. Vol. Smp./Dil. Vol. Sampling 2 Smp.Vol. Smp./Dil.Vol. Sampling 3 and Sampling 4 Diluent Code Diluent Warning Limit First Run Below/Above Normal Range Panic L Panic H Noise Prozone HIGH!, ABS Calibration Conditions Calibration Stability (days) Calibrator, Conc. R-Blank Limit (mAbs) Cal. Reps Range (%) Min. Cal. Resp. (mAbs) Cal. Factor Change (%) M-Point Curve Fit (%) Auto R-Blank by Bottle Reagent Volumes R1 R2 Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Code Rgt/Dil. Vol., Stir. Low Vol. W/ Limit Stability (days) Note: * ** *** N/A IUS.425.0.2 Directions for use see the lot-assigned value user defined optional not applicable Edition 2010-09-01 page 41 of 41
© Copyright 2025 Paperzz