Nuclear Instruments and Methods in Physics Research A 405 (1998) 269-273 Fast, parallax-free, one-coordinate X-ray detector OD3 V.M. Aulchenko, M.A. Bukin, Yu.S. Velikzhanin, Ya.V. Gaponenko, M.S. Dubrovin, V.M. Titov, A.I. Ancharov, Yu.A. Gaponov, O.V Evdokov, B.P. Tolochko, M.R. Sharafutdinov Budker Institute of Nuclear Physics, Novosibirsk 630090, Russia Institute of Solid State Chemistry, Novosibirsk 630090, Russia Abstract A second version of fast, parallax-free, one-coordinate X-ray detector OD3.2 destined for X-ray diffraction experiments is presented. The construction, basic principles of operation, results of the test with isotopic source 5SFe and X-ray beam test are being discussed. 1. Introduction Since 1994 in Novosibirsk, BINP R&D work under the project of new one-coordinate X-ray detector has been started. The detector is aimed for angular measurements in the diffraction experiments at the synchrotron X-ray beam. The distinctive features of the detector are as follows: - the registration rate up to 10^7 photon/s; - parallax-free in the angular range from 0° to 30°; - variable focal distance from 40 cm to the infinity; - photon coordinate resolution in transverse to the beam direction σ ~ 100 mkm (the angular resolution depends on the focal distance). The main idea of the detector with the above listed parameter was at first suggested in Ref. [1]. The test of the first proportional chamber OD3.1 with focal distance 45 cm shows its good reliability [2]. An electronics of the first detector was ready in the summer of 1995. After that, the tests with X-ray tube and at the synchrotron radiation beam of the channel 5b of VEPP-3 storage ring was started. The results of these tests were presented in the proceedings of the conference SRI95 [3]. The physical results, obtained in 1995-1996 with OD3.1 are presented in another report of the present conference. Since the autumn of 1995, the production of the second detector OD3.2 with a focal distance of 1.5 m has been started according to request of Japanese scientists for the Photon Factory of KEK. Proportional chamber and preamplifiers of OD3.2 have essentially improved the design compared with OD3.1. Some modification was also applied in electronics. At present, the detector is tested at the beam of VEPP-3. In the report we describe the construction details of OD3.2 and the last test results. 2. The basic principles of operation and construction 2.1. Proportional chamber Photons scattered by the scanning object came over the Be window with a thickness of 0.8 mm and dimensions 260x25 mm2 into the drift gap of the proportional chamber (see Fig. 1) and interact with atoms of the gas due to photoeffect. Appearing at photoabsorption secondary electrons drift in the electric field to the anode wires where the avalanche multiplication takes place. A coordinate of an avalanche is measured using charge distribution over the cathode strips. All electrodes of the proportional chamber are placed in a leak-proof box, holding the pressure from 0 to 0.2 MPa. The materials for the chamber promotion are chosen specially for long term operation without gas flow. The anode wires are joined into three groups which are provided by own anode pre-amplifier for signal control. It should be mentioned that a coordinate resolution and photon registration efficiency depends on photon energy, gas mixture and its pressure. Fig. 1. Schematic view of the OD3.2 proportional chamber and front-end electronics. 2.2. Electronics Charge sensitive pre-amplifiers of signals from the cathode strips are located on the distance of 5 cm from the strips outside the gas volume. They have a conversion coefficient k=10V/pC, and dynamic range of output signal from -0.15V to +6.5 V. Analogous signals from pre-amplifiers are transmitted via shielded twisted pairs with a length of 10 m into the crate with shapers and fast analog-to-digital converters. A process of the event selection and fast co-ordinate definition was described in detail in Refs. [2,3] and briefly as follows. The amplitudes of the signals from 54 strips are digitized simultaneously by 8-bit FADCs with time discrete 30ns and transmitted into the buffer for conveyer processing. An event selection scheme analyzes the buffer and searches for the signals from separate photons. The photons with close location in space and in time are rejected by the trigger logic. In spite of this, few photons separated by sufficient space (at least 3 strips) could be registered simultaneously. The number of strips with maximal amplitude N, two amplitudes from neighboring strips A(N-1), A(N+1) and time of the event with respect to the start of registration are fixed for the selected events. The values of N, A(N-1), A(N+1) are used in coordinate processor (see Fig. 2) for the photon coordinate (channel number) and energy definition. The tables, written in RAMs, are used for all main conversions. The amplitude on the neighboring strips directly depends on the avalanche coordinate, so, using these amplitudes as the addresses to read the data from RAM1, there is a possibility to define the coordinate inside the strip with maximal amplitude. This coordinate together with a strip number N are used as addresses for RAM2 to define the general coordinate in the device range. Fig. 2. A scheme of the coordinate reconstruction algorithm in OD3.2 A coordinate information is accumulated in the increment RAM with capacity of 64K 32-bit words. There arc several modes of data accumulation in 16, 32, 64, 128, 256 or 512 frames with size of 4K, 2K, IK, 0.5K, 0.25K or 0.125K channels correspondingly. The frame size is defined by monitoring program and it could contain the all device scale, any part of it or several parts. The frame duration could be chosen by program or according to the external timer signals. The minimal frame duration is 1 mks. An electronics of the detector is located in two CAMAC crates. One of them contains a set of special modules with shapers, FADC and event selection logic joined together in one supermodule by the fast "pin-channel" bus and via the interface connects to the second crate, which contains standard CAMAC modules and controller. We used two types of controllers for connection with PC-486 and SUN-station. 2.3. Software At the present time there are two main program packages for the detector control using PC486 and SUN-station. In both cases the window interface is exploited for the CAMAC modules management and presentation of an amplitude, coordinate and service information in the graphic and digital form [4]. 3. The results of the detector tests 3.1. Test of the proportional chamber The reliability of the proportional chamber was tested with gas mixture Ar-CO2 10% at pressure of 1 and 2 atm. Counting rate characteristics with and without isotopic source Fe55 for one of the wire groups are presented in Fig. 3 and show that PC has its own noise ~ 10 Hz and wide range ~ 500 V of working voltage. The dependences of an average amplitude, corresponding to the photon energy 5.9 keV, as a function of anode and drift voltage are shown in Fig. 4. A drift voltage variation above 2000V does not influence on the signal amplitude while the dependence of amplitude on the anode voltage has an exponential behavior. At gas pressure of 2 atm (0.2 MPa) the working voltage increased ~ 600V compared with 1 atm case. Fig. 3. Counting rate characteristics of the chamber with and without isotopic source Fe55. Fig. 4. Anode signal amplitude, corresponding to the 5.9 keV energy deposition, as a function of the anode and drift voltage. The amplitude conversion coefficients in the different shaper channels were adjusted uniform with precision better than 1%. After that the spectral measurements have been done. In Fig. 5a a spectrum of sum amplitude from three nearest to the avalanche strips is shown. It was obtained at irradiation of the chamber volume by photons from isotopic source Fe55 without any collimation. A resolution of 5.9 keV peak on FWHM is 31%. An amplitude of the peak is changing less than 5% at movement of the isotopic source in front of the window. All of these confirm good geometrical quality of the proportional chamber. The same condition was used to obtain the spectrum from the strip located at left of the strip with maximal amplitude (Fig. 5b). This figure shows, that amplitude of the signal on the side strip is variated ~ 2.5 times depending on the avalanche position under the central strip. The noise signals spectra (Fig. 5c) was accumulated during 5 minutes. It has an exponential decrease with amplitude, that confirms normal chamber mode of operation. Fig. 5. Cathode spectra: (a) sum of amplitude from the three strips nearest to the avalanche produced by photons from isotopic source S5Fc; (b) amplitude of the side strip with respect to the strip with maximal amplitude; (c) sum amplitude of the noise signals; (d) sum amplitude from the synchrotron radiated photons with energy of 8.3 keV. 3.2. Beam test of the detector OD3.2 In July 1996 detector OD3.2 was tested at synchrotron radiation channel 5b of the VEPP-3 storage ring. The detector was installed on the goniometer and horizontal (in transverse to the beam direction) translation stage, which together provide the rotation with respect to the focal point, located at 1.5m in front of the chamber. Photons with energy 8.3 keV were selected by the crystal monochromator. The beam divergence and its size were controlled by the set of horizontal and vertical slits before the detector. In Fig. 5d a spectrum of sum amplitude from three strips is shown. It was accumulated at the photon beam width of 0.2mm and observed counting rate of ~ 233 kHz/channel, that is higher than maximum project frequency 200kHz/ channel. Obtained resolution of the main peak FWHM is 28%. A prolongation of the spectral distribution to the large amplitude region could be explained by the photons with higher energy in the beam. It was proved by suppression of the photons of main harmonic using the absorber. Before the coordinate measurements, a calibration of the detector was done. This procedure was described in detail in Ref. [3]. At calibration, the horizontal stage moved the detector with constant speed in transverse to the beam direction. A distribution of values ln(A(N+1)/A(N-1)) is accumulated in memory. Using this information, the calibration arrays are calculated in suggestion of uniform events distribution on coordinate. Then these arrays are used to fill RAM1 and RAM2. Two histograms in the top of Fig. 6 show the coordinate distribution, obtained just after calibration at back movement of the translation stage. Differential non-linearity, obtained from this distribution is r.m.s. 2.5% (statistics contribution is 0.9%) over all scale. The histogram in the bottom part of Fig. 6, was obtained for the discrete horizontal shifts of the detector with steep of 0.5 mm. This figure shows the detector resolution of the close, narrow peaks. Fig. 6. Two top histograms show about 2/3 of the detector range with a coordinate distribution obtained at movement of the detector with constant speed in transverse to the beam direction. The bottom histogram, was obtained for discrete horizontal translation of the detector with steep of 0.5 mm. Fig. 7. A coordinate distribution is obtained for detector rotation around the focal point with steep of 0.1°. Two top histograms show about 2/3 of the detector range. Two bottom histograms are the extended view of the distribution edges. A coordinate distribution in Fig. 7 was obtained at detector rotation around the focal point with steep of 0.1°. It is clear, that the coordinate resolution does not depend on the photon angle (parallax-free) and equal to 5 channel (FWHM) or σz=150 mkm. A rate performance of the detector is also tested. The region, corresponding to the one cathode strip, was irradiated by the synchrotron photon beam. An intensity of the photons controlled by the number of absorber layers at fixed width of the slit gap and slowly decreasing current in the storage ring. The total rate of the signals from the anode wires f_mode was measured due to additional scalers; an average amplitude from 8.3 keV photons was fixed using spectrum; the rate of events reading by the cathode electronics f_read was also stored. The anode rate as a function of the absorber layer number is shown in Fig. 8a. At low photon flux (thick absorber) the anode rate should coincide with a rate of the photon conversion f_γ in the chamber drift gap. Extrapolating this dependence for the thin absorber and taking in to account the value of current in the storage ring, one could calculate the real photon conversion rate f_γ for any photon flux. In Fig. 8b-d the dependences of the peak amplitude, event reading rate, and mode signal rate as a function of the photon conversion rate are shown. It is clear, that even at four times frequency overload of the detector (800 KHz/strip instead of 200 KHz/strip) the amplitude drop is 20%; about 15% of events are loosed at reading while the anode rate is steel coincide with photon conversion rate. The similar graphics in Fig. 9 but for much more overload are shown. These histograms demonstrate that the saturation rate of the channel is about 2 MHz. However, the normal operation with correct amplitude measurements is possible at much lower frequency, due to the strong signals overlapping at high rate. Fig. 8. Rate performance of the OD3.2 for the 4-times overload of the channel: (a) anode signals rate as a function of absorber layer number; dependence of (b) pulse height, (c) rate of event reading (d) anode rate versus the rate of photon interaction in the chamber drift volume. Fig. 9. The same as that in Fig. 8, but for larger overload of channel. 4. Conclusion Described tests are the small part of all, that was doing with detector. They confirm that parameters of the detector OD3.2 are close to the expected one. The coordinate resolution σ = 150 mkm was obtained for the photon energy 8.3 keV and gas mixture Ar CO2 10% at normal condition. The detector is completely parallax-free for the focal distance 1.5 m. Narrow peaks in the coordinate distribution could be distinguished at the distance of 0.5mm. The r.m.s. differential non-linearity is 2.5% if the anode rate does not exceed 200 kHz/channel. References [1] [2] [3] [4] Proceedings of the European Workshop on X-Ray Detectors for Synchrotron Radiation Sources, Aussois, France, September 30-October 4, 1991. S.E. Baru. The One and Two Coordinate X-Rays Detectors. V.M. Aulchenko. S.E. Baru et al., Nucl. Instr. and Meth. in Phys. Res. A 367 (1995) 79-82. V.M. Aulchenko. Yu_S. Velikzhanin et al.. Position Sensitive X-ray Detector OD-3, in Proc. Synchrotron Radiation Instruments 1995 Conference, Argonne, Review of Scientific Instruments on CD-ROM. V.M. Aulchenko. M.A. Bukin et al.. The data acquisition and on-line visualisation system for the OD-3 fast one-coordinate X-ray detector, these Proceedings (11 th Nat. Synchrotron Radiation Conf.. Novosibirsk, Russia, 1996) Nucl. Instr. and Meth. A 405 (1998) 282.
© Copyright 2025 Paperzz