GS-1160 Handheld Spectroradiometer for Display Color, Intensity

About Gamma
Scientific:
Since 1961 Gamma
Scientific has produced
LED, display and light
measurement test
solutions for production
and R&D environments.
With over 50 years of
experience in
developing commercial
light measurement
instruments, Gamma
Scientific is trusted by
the world’s leading
organizations to provide
accurate and
repeatable test
systems. Gamma
Scientific instruments
are trusted by leading
global organizations
that require high speed,
precision
measurements and
custom configurations
for the most challenging
environments. Gamma
Scientific also operates
a NVLAP accredited
laboratory that performs
LM-79/LM-80 testing
and is ISO 17025
compliant for calibration
(NVLAP Code
200823-0).
GS-1160 Handheld Spectroradiometer for Display
Color, Intensity, and Flicker Measurement
Compact, high speed, and lightweight
spectrometer/flicker meter for accurate and
repeatable Display Measurements
• Suitable for color and intensity measurement of all display types including LCD,
LED,OLED, Quantum Dot.
• Spectroradiometer with superior accuracy and repeatability relative to Filter
Typecolor analyzer. Avoids measurement errors of Filter type analyzers which
can beinaccurate if exacting coating tolerances are not met.
• Portable and battery-powered. Can be used as a portable display testHU or
calibrationsystem or as a desktop system operated with Gamma Scientific's
LightTouch usSpectrum and LightTouch Flicker software.
• Easy to use with integrated touchscreen display operation. Simply position the
optical sensor near the display and a press of a button displays the
measurementson a 3.5” color display.
• Backed by ISO 17025 Certified and NIST- traceable calibration.
To view the complete
line of test and
measurement solutions
from Gamma Scientific,
please visit our website
at www.gammasci.com
GS-1160 Spectral Sensing Technology
• Delivers key display measurement capabilities for Color Chromaticity,
Gamma, White Balance Adjustment, Contrast, Flicker (VESA & JEITA), and
uniformity.
• Luminance measurement range: 0.05 to 5,000 cd/m2.
Color measurement in Chromaticity: x/y/u’/v’, XYZ, x10, y10, u10, v10.
• Wide range of Flicker Measurement: JEITA, VESA, CONTRAST (Max/Min),
RMS, Time Domain, FFT.
• Full spectral data and spectrum graphics.
• Fast measurement: 100us to 5000ms for productivity demands.
• Automatic dark calibration.
9925 Carroll Canyon Rd. San Diego, CA 92131 USA | Phone +1 800-637-2758 | Fax: +1 858-576-9286
www.gamma-sci.com
GS-1160
Application: Measure wavelength shift on various production displays
Flexible software for color & flicker analysis
Specifications
Spectrum
Sensor
CMOS Linear Image Sensor
Wavelength Range
380 to 780 nm
Wavelength Data Increment
1 nm
Spectral Bandwidth
Approximately 12 nm (Half Bandwidth)
Receptor Size
Ø 10 mm
Acceptance angle
±1°
Wavelength Reproducibility
± 1 nm
Display Range
0.001 to 5000 cd/m2
Measurement range
2
0.05 to 5000 cd/m
±2% 100 to 5000 cd/m2
±3% 0.2 to 100 cd/m2
Accuracy
±4% 0.05 to 0.2 cd/m2
±0.2% 100 to 5000 cd/m2
Luminance
±0.5% 0.2 to 100 cd/m2
Repeatability (2σ)
±0.8% 0.05 to 0.2 cd/m2
Measurement range
0.05~5000 cd/m2
±0.002 in CIE1931 x, y for white 100 to 5000 cd/m2
Accuracy
±0.003 in CIE1931 x, y for white 0.2 to 100 cd/m2
±0.005 in CIE1931 x, y for white 0.05 to 0.2 cd/m2
Color
±0.0005 in CIE1931 x, y for white 100 to 5000 cd/m2
Repeatability (2σ)
2
±0.001 in CIE1931 x, y for white 0.2 to 100 cd/m
±0.002 in CIE1931 x, y for white 0.05 to 0.2 cd/m2
System Dimensions
Height
220 mm
Width
81 mm
Depth
33 mm
Weight (with battery)
330 g
9925 Carroll Canyon Rd. San Diego, CA 92131 USA | Phone +1 800-637-2758 | Fax: +1 858-576-9286