Webcast Notes • Type your questions in the “Submit Question” box, located below the slide window • You can enlarge the slide window at any time by clicking on the “Enlarge Slides” button, located below the presentation window – the slides will advance automatically throughout the event • If you are experiencing technical problems with viewing or hearing the event, please click on the “Help” button, located below the slide window Speakers John V Hinshaw GC Dept. Dean CHROMacademy Tony Taylor Technical Director CHROMacademy Moderator Dave Walsh Editor In Chief LCGC Magazine Understanding and Optimising Detectors for Capillary GC Aims & Objectives 1. Review of Common Detector Types for Capillary GC 2. Principles and Practice of: - Flame Ionization Detectors (FID) -Nitrogen Phosphorous Detectors (NPD) - Electron Capture Detectors (ECD) - Thermal Conductivity Detectors (TCD) 3. Learn about gases, flow rates and parameter settings 4. Maximising detector sensitivity 5. Troubleshooting detector problems 6. Relate detector issues to chromatographic problems The Flame Ionisation Detector (FID) 1. Carrier and detector ‘fuel’ gases mixed to support a FLAME at the anode tip 2. Potential difference applied between the Anode (Jet) and Collector (Cathode) which is electrically isolated 3. Analyte burns in the flame to produce cations which move toward the cathode producing a very small current (pA) 4. Signal is amplified and recorded as a chromatographic peak page_06.flv FID Gases 1. Requires three gases FUEL gas, typically Hydrogen OXIDISER gas, typically Air MAKEUP gas, typically Nitrogen or Helium 2. Gas flow rates are important (e.g.) Carrier Gas: 1ml/min. Makeup Gas: 30ml/min. Hydrogen: 30ml/min. Air: 300ml/min. 3. Make-up gas required to obtain optimum flow into detector from GC column of 20-40 ml/min page_08.flv FID Detector Sensitivity (I) 1. Critical Factors Include a. Combustion gas flow rates (Gas Stoichiometry) b. Makeup gas flow rate too low or too high c. Flame jet exit diameter (0.3mm v’s 0.5-0.7mm) d. Relative positions of jet and collector (fixed by manufacturer) e. Position of the column relative to the jet exit (avoid dead volume) FID Detector Sensitivity (II) f. Carrier Gas Flow Rate -Constant Pressure - Constant Flow - Ramped Make-up g. Detector temperature >150oC 20-30oC> highest oven temperature Constant Pressure Constant Flow page_11.flv FID Detector Troubleshooting (I) Baseline Wander Baseline Noise Consider Detector Gas Quality FID Detector Troubleshooting (II) Baseline Wander Baseline Noise Consider Detector Cleanliness 1. Breakdown of electrical isolation of collector (Cathode) 2. Dirt on the collector insulation 3. Dirt in the collector inside surface 4. Clean with solvent (methanol) 5. Clear plugged jet tip FID Detector Troubleshooting (III) Baseline Spikes Issues & Remedies 1. 2. 3. 4. Incorrect positioning of column in jet (protruding) Use correct jet tip diameter Dirt in the collector inside surface Clean with solvent (methanol) FID Detector Troubleshooting (IV) Peak Shape / Broadening Issues & Remedies 1. Column too low in jet - leading to large dead volume 2. Makeup gas flow rate too low 3. Shouldered peaks may indicate blocked jet tip FID Detector Performance Compounds giving little or no FID response Typical FID Performance 1. Minimum Detectability (LOD) – 10-11 g (~50ppb) 2. Linearity – 107 (very wide dynamic range) 3. Response – organic carbon containing compounds (exceptions shown above) The Nitrogen Phosphorous Detector (NPD) 1. Is an Ionising Detector like FID but uses different principle 2. Shows enhanced sensitivity to Nitrogen and Phosphorous containing compounds 3. Main difference is the rubidium silicate ‘BEAD’ which is resistively heated to emit thermionic electrons 4. Hydrogen flow much lower than in FID detectors – too low to sustain a flame at the jet tip 5. Plasma of Fuel, Oxidiser, Carrier and makeup is formed which partially combusts around the bead heating filament page_18.flv The NPD Operating Principle 5. Partially combusted Nitrogen and Phosphorous compounds adsorb on the bead surface 6. This reduces the beads ‘work function’ – increasing the electron density at a given potential / temperature 7. Increased electron density = increased current = detector response page_20.flv NPD Operation and Optimisation 1. Run at higher temperatures (260 – 350oC) to extend bead lifetime 2. Run at the lowest acceptable bead power (start at 2V and increase in 10mV steps if unsure) 3. Hydrogen flow MUST be low enough to ensure a continuous flame – otherwise Nitrogen response will be zero 4. Sensitive to variation in Hydrogen flow – ensure constant flow supply 5. Performance decreases with time – requiring an increase in bead voltage to obtain suitable sensitivity NPD Operation and Optimisation 6. Moisture formed during combustion hydrolyses the alkali silicate bead to the metal hydroxide and silica (whitening of the bead!!) 7. Rubidium hydroxide is volatile and constantly lost from the bead 8. Conserve bead life by interrupting hydrogen flow: • When solvent peak elutes • During Oven cool down 9. Detector is Mass / Flow sensitive – recommend constant flow operation to avoid drifting baselines NPD Detector Performance 1. Minimum Detectability (LOD) – 10-12 g/ml (P), 10-11 g/ml (N) 2. X500 increase in sensitivity v’s FID for N and P compounds 3. 500pg of pesticide easily achievable 4. Linearity 106 5. 400oC temperature limit 6. Requires constant carrier flow and stable hydrogen supply The Electron Capture Detector (ECD) 1. Measures electrical conductivity of the effluent after exposure to ionising radiation 2.Especially sensitive to ‘Electron Capturing’ species – Halogens 3.Radionuclide is 63Ni foil which emits low energy electrons (Beta particles) 4.These Beta particles collide with the carrier to produce higher energy electrons 5.This establishes a high standing current between detector body (foil) and a centrally located collector The Electron Capture Detector (ECD) 6. Applied potential difference ~ 20 – 100V (dc) 7. Halogenated analytes elute and ‘capture’ some of the electron density 8. Negative ions formed are low energy and are not collected by the cathode 9. Reduction in the standing current occurs page_26.flv ECD Operation and Optimisation 1. Detector gases should be clean (99.999%+) and dry as both oxygen and water are electronegative and contribute to noisy baselines 2.Nitrogen or Argon (5% Methane) are used as make up gases and the make up gas and flow can be critical to sensitivity 3.Better sensitivity and linearity is achieved using the detector in ‘Pulsed’ mode 4.Square wave pulse applied to maintain a constant current 5.As analyte elutes pulse frequency is increased to maintain current 6.Signal is proportional to frequency ECD Troubleshooting Baseline Noise 1. Use of impure gases (especially make up gas) 2.Incorrect make up gas flow rate Linearity / Peak Shape 1.ECD is non-destructive and foil / collector become contaminated 2.Background current reduces, but current reduction per analyte molecule remains constant 3. Contaminated detector may become MORE SENSITIVE!! 4. Contamination will case decrease in detector linear range 5. Contaminated detectors usually give rise to negative ‘dip’ after each peak ECD Detector Performance 1. Minimum Detectability (LOD) – 10-9 g/ml (10ppm) 2. Highly selective for Halogenated analytes 3. Linearity 103 - 104 4. 400oC temperature limit 5. Susceptible to low grade gases and leaks in the detector body 6. Extreme caution required when cleaning and reconditioning ECD at 250oC Make up – Nitrogen at 30ml/min. 20ppb Chloroacetic acid The Thermal Conductivity Detector (TCD) 1. Simple detector which will respond to almost any analyte 2.Popular for permanent gas and inorganic compound detection 3.Two filaments in separate blocks with Heated filaments connected to a tradition Wheatstone bridge 4.Pure carrier flows over one cell whilst the column effluent flows over the other and the Wheatsone bridge is ‘balanced’ to produce a baseline signal The Thermal Conductivity Detector (TCD) 5. Analyte eluting into the detector will change the thermal conductivity of the gas in the analytical cell 6. Rate of heat loss from the filament changes 7. Current applied to balance the Wheatstone bridge 8. Size of the cell is critical to detector sensitivity page_32.flv TCD Operation and Optimisation 1. Dual cell detectors not suitable for capillary GC due to large cell size (140ul) 2.Most manufactures offer a single cell version (5ul) with ‘switchable’ gas supply (5-10 Hz) 3.Reference gas flow is switched out when analyte elutes 4.Filament temperature should be > detector body temperature 5.Higher filament temperature means larger thermal conductivity changes and better sensitivity but shorter filament lifetime TCD Operation and Optimisation 6. Susceptible to variation in detector body temperature and carrier gas flow 7.Detector requires thermal cladding, constant temperature supply and carrier operating in constant flow mode 8. Sensitivity depends upon the magnitude of difference in thermal conductivity between carrier and analyte 9. Hydrogen and helium preferred as carrier 10. Area normalisation should not be used for quantitative measurement using TCD TCD Detector Performance 1. Minimum Detectability (LOD) – 10-9 g/ml (10ppm) 2. Responds to all compounds 3. Linearity 104 4. 400oC temperature limit 5. Susceptible to ambient temperature change and carrier gas flow variation CHROMacademy & CHROMmunity Today's Webcast was brought to you by CHROMacademy: http://www.CHROMacademy.com Join CHROMacademy for $250 per year for a full subscription and to see all of the archived webcast content All questions will be answered within CHROMmunity: http://chrommunity.chromacademy.com The webcast will be available for on-demand viewing for 2weeks from today Question and Answer Type your question in the “Submit Question” box below your slide window. Thank You for Attending We Need Your Feedback! Please complete the post-webcast evaluation form that will immediately appear in your window.
© Copyright 2026 Paperzz