LIMNOLOGY and OCEANOGRAPHY: METHODS Limnol. Oceanogr.: Methods 7, 2009, 730–739 © 2009, by the American Society of Limnology and Oceanography, Inc. Rapid and precise δ13C measurement of dissolved inorganic carbon in natural waters using liquid chromatography coupled to an isotope-ratio mass spectrometer Jay A. Brandes Skidaway Institute of Oceanography, Savannah, GA, USA Abstract The measurement of the carbon isotopic composition (δ13C) of dissolved inorganic carbon (DIC) in natural systems has provided a tool for examining a variety of processes, from net primary production to anthropogenic signatures of CO2 uptake in the oceans. Over the past decade, new δ13C-DIC methods have steadily decreased the sample analysis time and volumes for this measurement. The development of a new interface capable of inline acidification and extraction of CO2 from a liquid stream (LC-Isolink™; Thermo Electron) provides breakthroughs in sample size, ease of sample handling, and speed of analyses. Typical sample size injections of 25 µL provide precision and accuracy of 0.04‰, comparable to that obtained by off-line extractions/dual inlet analyses on much larger samples. Single-sample injection runs can be accomplished within 4 min, whereas sample preparation is limited to filtration to remove particulates. Thus combined runs of 200 samples over a 20-h period are possible. Highest-precision triplicate injection analyses of single samples require 11 min. Manually injected marine sample sizes of 2.5 µL provide nearly the same precision and accuracy as the larger loop sizes. For maximum precision, care must be taken to match sample and standard sizes to correct for linearity effects. Over wider concentration ranges typically found within estuarine systems, the overall precision of the method decreases to 0.06‰ with such correction. Introduction than direct pCO2 and DIC concentration measurements (Takahashi et al. 2006). Measurement of the isotopic composition of oceanic DIC requires meeting several challenges. Oligotrophic mixed layer δ13C-DIC values typically exhibit a 0.25‰ seasonal cycle (Gruber et al. 1998; Quay and Stutsman 2003). The uptake of anthropogenically released CO2 (the 13C Suess effect) into the oceans has generated slightly larger shifts (1.0‰) over the past 30 years toward lighter δ13C-DIC values (Gruber et al. 1999; Quay et al. 1992, 2007; Sonnerup et al. 2007). Thus investigations into both of these processes require measurement techniques of very high precision, generally better than 0.05‰. On the other hand, measurement of porewater δ13C-DIC values, useful for constraining the balance between organic C remineralization, methane oxidation, and carbonate dissolution (Gehlen et al. 1999; Hu and Burdige 2007; McCorkle and Klinkhammer 1991), have wide isotopic shifts but generally are limited to volumes less than 2–3 mL for fineresolution profiling. And in all these cases, sample throughput is a challenge, as detailed basin-wide or seasonal sampling can generate thousands of individual samples. Measurement of δ13C-DIC has evolved with the development of new isotope ratio mass spectrometry (IRMS) interfaces, much as the measurement of isotopes within other sample Oceanic dissolved inorganic carbon (DIC) represents the largest active pool of C on Earth’s surface (Hedges and Keil 1995). Thus a great deal of attention has gone into examining different aspects of DIC cycling, especially in the areas of the uptake of fossil fuel–derived C, the interaction between DIC and organic carbon pools, and sediment burial/diagenesis and dissolution processes. Radioisotopic and stable isotopic measurements have proven useful in constraining various aspects of all of these processes (Butler et al. 2009; Sonnerup et al. 1999), but their use is less widespread *Corresponding author: E-mail: [email protected] Acknowledgments This material is based on work supported by the National Science Foundation under grants NSF OCE-0722604 and GEO-0807387. Any opinions, findings, and conclusions or recommendations expressed in this material are those of the author and do not necessarily reflect the views of the National Science Foundation. The author thanks Dr. Paul Work and the students of the inaugural Introduction to Coastal Oceanography class at Georgia Tech for assistance in sampling and data reduction. 730 Brandes Rapid isotopic analysis of DIC normally used for persulfate oxidant, was not used. The Isolink reactor was set to 80°C. Samples were first collected into plastic syringes from 10-L Rosette Niskin bottles. Samples and standards were subsequently filtered through 0.2-µm pore-size nylon syringe filters into 2.0-mL glass crimp-seal vials. A needle was fitted onto the end of the syringe filter and samples filled from the bottom of vials slowly and with a minimum of turbulence. The first 1 mL was used as a filter/needle wash and discarded. Vials were overfilled, then sealed with red butyl rubber/PTFE septa/ aluminum ring tops (Discovery Sciences cat. no. 73070) taking care to not capture any bubbles within the vial. Vials so prepared were isotopically unchanged over a period of >1 month at room temperature. Analytical runs were performed with three injections per vial, spaced 225 s apart (Fig. 1). The three peaks were typically averaged, although if one peak was found to differ in peak area by more than 5% from the mean of the other two, it was discarded. Three CO2 gas injections by the Isolink interface were also used to provide internal references for the mass spectrometer’s Isodat software. Total run time was 11 min per sample. Overall precision of this method is ±0.04‰ (SD) on replicate samples (n = 10). An internal standard seawater solution, consisting of a filtered, autoclaved coastal seawater sample allowed to equilibrate with the atmosphere at room temperature for 1 week before encapsulation in vials, was used periodically to provide a reference. This internal standard was calibrated to NBS 19 CaCO3 standard defined as having an isotopic composition of 1.95‰ versus vPDB (Coplen 1996) using the method of Torres et al. (2005). types has changed over the years. The earliest efforts involved extraction of CO2 from large-volume (50-mL) samples using purge and trap systems after acidification (Kroopnick 1974; Quay et al. 1992). Purified CO2 from such samples was then measured via dual-inlet IRMS systems, resulting in high precision (±0.025‰ SD; Quay et al. 1992) but this is overall a lowthroughput and highly labor-intensive process. Smaller samples could be measured this way (Mccorkle and Klinkhammer 1991) but require still more care in extraction and analysis. The advent of automated headspace extraction interfaces in the late 1990s provided an alternative with much smaller sample size requirements. However, significant sample preparation is still required for these methods, including vial purging, addition of concentrated H3PO4, and equilibration times of many hours after acid addition (Torres et al. 2005). Here I present a simple, non–labor intensive method to rapidly (4–11 min/sample) and precisely (<0.05‰ SD) measure δ13C-DIC within aqueous samples on sizes as small as 2.5 µL. The technique takes advantage of a relatively new liquid chromatography gas extraction interface developed primarily for the analysis of nonvolatile organic solutes. The combination of the new interface with modern IRMS instrumentation provides significant improvements in sample throughput, minimizing sample size requirements for porewater and other volume-limited samples, and attaining the precision needed for surface ocean δ13C-DIC analyses. Materials and procedures The instrumentation used was a Thermo Electron SurveyorLite autosampler equipped with a 25-µL sample loop, a Surveyor MS HPLC pump, and a LC Isolink interface coupled to a Thermo Scientific Delta V plus stable isotope mass spectrometer. The Isolink is an integrated chemical reactor/gas extraction manifold and standard reference gas injection device. In normal operation, an incoming liquid stream from the HPLC is mixed with both phosphoric acid and persulfate oxidant (not used here), and the resulting fluid passed through a heated stainless-steel capillary column reactor to convert organic compounds to CO2. The mixture is cooled and subsequently passed through a gas–liquid separation manifold, allowing the CO2 to be extracted into a He stream. This He + CO2 stream passes through two Nafion driers and an open split before introduction to the IRMS. For further details on the Isolink’s design, see Krummen et al. (2004). The system was run in isocratic mode with no separation column. The mobile phase was degassed (90 min under vacuum in a 40°C ultrasonic bath) deionized water with 1 drop of 85% H3PO4 added (Fisher Scientific HPLC grade). The same solution composition was used as the autosampler wash solution. Mobilephase flow rate was 300 µL min–1. A phosphoric acid solution (10% vol/vol of 85% H3PO4 in distilled water, followed by degassing under vacuum as described for the mobile phase) was continuously mixed into the mobile phase within the Isolink at a rate of 50 µL min–1. The second Isolink LC pump, Assessment Although the Isolink LC interface was developed primarily as a tool for measuring the isotopic composition of organic compounds, its design also provides for the rapid and simple measurement of dissolved inorganic carbon. Indeed, without the blank associated with the oxidation of residual DOC in mobile phases, separation column bleed, and reagents, the system as described above routinely achieved baseline values of 20 mV on mass 44, less than 5% of the specified background level in organic analysis mode. This provides much higher signal-to-noise ratios for sample measurements. In addition, the lack of an oxidant means that O2 production and concomitant mass spectrometer sample filament longevity issues are insignificant (Hettmann et al. 2007). High-precision measurements place a special emphasis on sample storage and processing. Previous studies have indicated that standard polytetrafluoroethylene (PTFE)/silicone septa allow CO2 diffusion (Nelson 2000; Taipale and Sonninen 2009), and measurements of marine samples using such septa showed distinctive isotopic shifts after 3–4 h of storage at room temperature. However, use of butyl rubber/PTFE septa with crimp seal caps effectively provided isolation from atmospheric CO2 over periods of at least 1 month (Table 1). 731 Brandes Rapid isotopic analysis of DIC Fig. 1. Sample chromatogram. Upper traces are isotopic ratios 46:44 and 45:44; bottom trace is mass 44 signal versus time. Square peaks are reference CO2 gas injections. Blank concentrations after 2 weeks were <0.5% of a typical seawater sample, rising to about 4% after 35 days. The isotopic value of the blank vial DIC was –0.1 ± 0.5‰ at the end of this time period. Several other septa types were tested including black fluoroelastomer, butyl rubber without PTFE backing, PTFE/chlorobutyl, and Finneran Versa-Vials with PTFE/gray chlorobutyl plugs. None of these alternatives proved as resistant to leakage as the red butyl/PTFE seals. This is in keeping with the results for storage of air samples for CO2 analysis (Knohl et al. 2004). In addition, it is recommended that only standard crimp-top vials be used, as wide-mouth and snap ring vials exhibited higher leakage rates. Screw-top septa were also unacceptable, as the autosampler had a tendency to push the septa into the sample rather than puncturing it. Samples, after filtration, were isotopically unchanged after 35 days of storage at room temperature (Table 1). In contrast to blanks, no change in DIC concentrations was noticeable after this time. However, samples which are far from equilibrium or far from “normal” isotopic values may be more strongly affected by CO2 diffusion. Optimization of the DIC isotopic method required examination of flow rates, acid concentrations, and other parameters. The intrinsic precision (SD) of the Isolink for repeated analyses of a single DIC sample is 0.04‰ (Table 2) for samples with peak heights >4 V. This peak height translates into an injected amount of C of roughly 17 nmol C. However, precisions of better than 0.1‰ were achievable for samples as small as 5 nmol (Table 3). For seawater, it is possible to achieve this level of precision on injections of only 2.5 µL, and it may be possible to achieve this on 1-µL injections done manually. Indeed, the sample blank measured after 35 days was roughly 50 µM, corresponding to an injected amount of <1.1 nmol C, and exhibited a precision of ±0.5‰. For such small samples, of interest in studies of pore waters and other volume- or concentration-limited environments, the difficulty is in finding an effective storage container; they also possibly require the use of an injector designed specifically for small samples. Over longer terms, the accepted method of storage in glass-stoppered bottles with HgCl2– poison is recommended, and such treatment did not exhibit any noticeable effect on the isotopic values of samples analyzed by LC-IRMS. 732 Brandes Rapid isotopic analysis of DIC Table 1. Effect of storage on blank (degassed deionized water) and Skidaway River estuarine samples. Time, h 0 19 42 192 240 840 Area, V-s Blank Amount injected, nmol δ13C Area, V-s Dock standard δ13C 0.14 0.24 0.43 0.82 0.78 4.25 0.04 0.06 0.11 0.21 0.20 1.11 20 33 23 9.8 10.2 –0.1 134.9 134.5 135.2 138.8 137 134.2 –1.03 –1.03 –1.03 –1.02 –1.07 –1.07 Table 2. Number of injections per vial versus precision (SD). Injections/vial SD isotope, ‰ SD area, % Vials, n 0.020 0.036 0.026 0.025 0.79 0.39 0.31 0.13 4 4 4 4 1 2 3 4 Precisions are given as SD of measurement divided by sample peak area. Table 3. Comparison of loop volume on isotopic precision, manual injections. Loop volume, µL 2.5 5 10 20 25 Height, mV Area, V-s Amount C injected, nmol δ13C precision, ‰ Injections, n 1202 2149 4178 6267 7289 18.2 34.3 67.3 102.5 124.5 4.8 9.0 17.7 27.0 32.7 0.05 0.025 0.05 0.05 0.05 5 5 5 5 4 autosampler to analyze a series of standards with different injection volumes to bracket sample peak areas. Changing mobile-phase flow rates has an effect on isotopic composition, peak height, peak area, and linearity. Peak areas were inversely dependent on total flow (Table 4), suggesting incomplete extraction within the gas manifold from the sample stream. Peak widths were negatively correlated with total solution flow rate (mobile phase + acid) (Table 4). Note that the chief limiting factor in determining peak widths is the Isolink gas extraction manifold, as calculated peak widths assuming instantaneous extraction would be on the order of 15 s based on injector loop volume and reagent flow rates. Peak heights were influenced by both total flow and acid flow rates, with lower acid flows at a given total flow rate producing higher peaks. The acid flow rates chosen for the method provide a large excess (~150 times) compared to the DIC within injected samples, and no significant sample area or isotopic shift was noted with acid flow rates as low as 10 µL min–1. However, higher flow rates and lower acid flow rates, while producing better apparent chromatography, resulted in worse linearity (e.g., 400 µL min–1 data in Fig. 2a). In all cases, the total flow rate affects the isotopic composition of the For high-precision analyses, the primary concern therefore was the linearity of the system and in particular the Isolink interface. Tests over a wide range of DIC injection amounts, starting from seawater amended with NaHCO3 to triple present-day equilibrium levels, indicated that the Isolink interface did not deviate from the intrinsic linearity of the Delta plus IRMS from peaks of >35 V on mass 44 to about 3 V peaks (or about 100 V-s peak areas). From 3 to 1.5 V (areas of 100 to 50 V-s) the system exhibited a stronger fractionation, approximately a shift of –0.5‰ (Fig. 2a). This shift was repeatable and reproducible day to day and thus can be corrected for in samples by running standards that bracket the concentration range of samples analyzed. Although these tests were primarily conducted using the Surveyor autoinjector’s partial loop injection mode, this pattern was verified by manually injecting seawater solutions using different size loops (5, 10, 20, 25 µL; Fig. 2a). For routine measurement of low DIC concentration samples, it is suggested that a larger sample loop be used to maintain sample peak areas >100 V-s and that standards be run with concentration values bracketing the concentrations of samples. An alternative is to use the partial loop injection mode of the 733 Brandes Rapid isotopic analysis of DIC sample by –0.18‰ for every 100-mL increase in flow rate (data from Table 2, r 2 = 0.95 linear regression). Therefore the choice of 300 µL min–1 (mobile phase) and 50 µL min–1 (acid) was chosen as the best compromise between peak width, sensitivity, and linearity. Given the need to prevent sample degassing, vials must be filled with no headspace. This led to inconsistent peak areas in replicate samples, especially in refrigerated samples. An investigation of autosampler performance led to the conclusion that peak area consistency could be improved by overfilling the loop. Peak areas increased up to a doubling of loop volume (Fig. 2b). This led to a more effective washing of mobile phase from the loop and a decrease in the influence of fluctuations in sample peak area in pressurized vials, from an SD of 5% to an SD of 1%. Peak areas were more consistent in samples collected from warmer estuarine waters and in laboratory standards made from equilibrated seawater, down to an SD of 0.5%, which is the reported precision given by the manufacturer (Table 2). Sample area precision also was improved in samples warmed to room temperature before subsampling into sealed vials, but this increases the risk of bubble formation and subsequent isotopic fractionation in typical oceanic samples. Sample area precisions might be improved by the use of a refrigerated sample holder, but the system used lacked that capability. The system is quite linear throughout typical estuarine-marine DIC concentrations, with an r 2 = 0.9999 determined using Na2CO3 standards (Fig. 2c). This provides the possibility of obtaining both isotopic and concentration information from samples (e.g., Salata et al. 2000), although much more rapid shipboard methods exist for DIC concentration determinations (Feely et al. 2006). In the case of volume-limited samples, however, this capability will be useful. Again, the use of a refrigerated autosampler, and in particular conducting duplicate or triplicate measurements using separate vials, may improve such results. Accuracy may be improved by running standards spaced periodically throughout a run, to account for any changes in Isolink extraction and ionization efficiency within the mass spectrometer. Tests were performed to determine if replicate injections on the same vial would improve precision. Although there was little difference in the isotopic precision between single injections and multiple injections (Table 2), the precision of peak areas, and thus quantification of DIC concentrations, was improved by having up to three or four replicate injections (Table 2). Another advantage of replicate injections is that there occasionally appear to be injections with too small a peak area or other problems (as noted above), and having several peak injections provides the investigator the ability to statistically discard the erroneous peak. In practice, any peak with an area diverging by >5% from the mean should be considered suspect. The disadvantages of multiple peak injections are time and autosampler wear. Sample runs with one injection per vial can require as little as 200 s, plus an additional 40 Fig. 2. Performance tests. (A), Effect of sample size on isotopic value. Circles are samples injected using autosampler at nominal (300/50) flow conditions. Size adjustments were achieved by changing injection volumes. Triangles are values obtained using the Isolink manual injector and changing loop volumes. Squares are values obtained using the autosampler at a flow rate of 400 µL min–1 mobile phase and 10 µL min–1 acid. Error bars represent 1 SD. (B), Effect of overfilling sample loop. Autosampler loop has a nominal volume of 25 µL. (C), Peak area (total sum of mass 44, 45, and 46) versus concentration of Na2CO3 standard. 734 Brandes Rapid isotopic analysis of DIC Table 4. Effect of different mobile-phase and acid flow rates on DIC peak area, peak width, and isotopic value. Total flow, µL min–1 300 300 325 350 350 350 350 350 350 350 400 400 400 450 450 500 500 Mobile-phase flow rate Acid flow rate Area, % standard Isotope, ‰ difference from standard Peak width, % standard 250 290 300 250 300 300 325 335 340 345 300 350 390 400 440 400 450 50 10 25 100 50 50 25 15 10 5 100 50 10 50 10 100 50 103 104 103 100 100 100 101 100 101 100 94 96 97 93 92 88 88 0.08 0.08 0.07 –0.01 0 –0.01 0.02 0.01 –0.01 –0.07 –0.2 –0.17 –0.13 –0.24 –0.24 –0.33 –0.34 108 106 103 101 100 100 100 99 99 100 98 95 94 92 90 90 89 Areas and widths are relative to those found for standard conditions of 350/50 mobile-phase/acid flow, whereas isotopic values are those referenced to values obtained at the standard conditions. as the density profile at this location was homogeneous. A plot of δ13C-DIC versus salinity (Fig. 5) shows a near-linear relationship between these two parameters. However, a two end-member mixing model (Fry 2002) for δ13C-DIC indicates that the observed trend cannot be described simply as an abiotic process. Savannah River DIC concentrations and δ13CDIC were determined at the head of tides to be 570 µM and –10.0‰, respectively, and using these plus the seawater endmember at stations S0 and W0 results in the mixing curve shown in Fig. 5. Sample δ13C-DIC values deviate rapidly from this line, indicating the input of significant amounts of organic carbon remineralization within the estuary. Although many estuarine systems are assumed to be conservative with respect to DIC (Fry 2002), a number of studies have shown deviations of >1‰ in δ13C-DIC values from conservative mixing lines (Chanton 1999; Coffin and Cifuentes 1999; Kaldy et al. 2005). With rapid, precise measurement of δ13C-DIC, such “small” signals may be used to examine the contributions of organic matter remineralization within these systems. Although a calibration of DIC concentrations was not performed at the time of these estuarine measurements, most of the riverine apparent DIC concentrations were above those predicted from conservative mixing and sample salinity values. This is not unexpected, as the Savannah River is highly turbid and passes through a major urban area. Furthermore, the Wilmington (and Savannah, through mixing) is heavily influenced by marsh respiration and the consequent addition of DIC from this source (Cai and Wang 1998; Cai et al. 2003). Future work to better constrain the seasonal cycling of C within this system could involve much s between runs for peak centering routines on the mass spectrometer. This 200-s limit is also set by the autosampler; attempts at analyzing samples faster than this rate led to occasional skipped injections. Likewise, the replicate injection rate was set to 220 s to avoid skipped injections during multiple injection analyses. Sample ocean application The ability to rapidly and with little effort analyze large numbers of samples for δ13C-DIC opens up several avenues for research. One example is the study of estuarine carbon cycling. Samples were collected in fall 2008 along two transects within the Savannah and Wilmington River estuaries (Fig. 3). Samples were collected by Rosette into 10-L Niskin bottles, subsampled into 60-mL prerinsed plastic bottles, and filtered through 0.2-µm pore size nylon syringe filters into 2mL vials as noted above. Samples were collected roughly every 2 m from surface to 1 m above the bottom. The LC method used here completed the entire sequence of analyses in less than 1 day including duplicate analyses of each bottle sample. Samples collected on the South Atlantic Bight shelf were isotopically homogeneous, as expected from density profiles (Fig. 4). However, δ13C-DIC values were rapidly influenced by remineralization at locations very close to the entrance of both rivers. The Savannah River has a much larger drainage basin than the Wilmington (which is essentially a tidal estuary), and the influence of this input is reflected in a stronger surface layer at upstream stations. There is the suggestion of a benthic respiration signal at station W4 in the bottommost sample. This was not caused by mixing of another water mass, 735 Brandes Rapid isotopic analysis of DIC Fig. 3. Map of Georgia coast sampling locations. Fig. 4. DIC δ13C values versus depth for stations indicated in Fig. 3. Error bars for isotopic data are within the width of the points in each figure. 736 Brandes Rapid isotopic analysis of DIC such as the XYZ autosamplers found on GC and headspace analyzers, may also avoid this problem. Second, care should be taken to avoid the entrapment of bubbles in the 2-mL autosampler vials, as even small bubbles pose the potential for isotopic exchange with the gaseous phase. Third, replicate (duplicate or triplicate) vials should be analyzed wherever possible for high-precision applications. This is especially critical because the autosampler appears to occasionally mis-inject or skip injections on random vials, and re-analyzing such vials at a later time introduces the possibility of outgassing or other contamination. The second primary concern when attempting to measure small isotopic signals is accuracy. Although the system can be calibrated to secondary standards, which are in turn calibrated to primaries, there exists no widely available DIC isotopic standard. Although use of solid standards can calibrate measured 13C-DIC values to within 0.1‰ or better with sufficient care, there exist difficulties in comparing the isotopic values of solid standards with those found in solutions (Gillikin and Bouillon 2007). Given the small signals discussed above for anthropogenic signals, better and consistent standardization is needed. Ideally, an analog to the VSMOW/SLAP/GISP water isotope standards should be produced and made available to the scientific community, with a range in δ13C and concentrations (possibly characteristic surface and deep oceanic waters). These could then be used in individual laboratories to calibrate secondary DIC solution isotopic values. Having a range of concentrations is critical, as at the highest precision requirements the linearity correction (~0.04‰ V–1) becomes significant even between surface and deep ocean samples. Such a set of standards will aid in improving the accuracy of the reported concentrations as well. Fig. 5. DIC δ13C values versus salinity for stations indicated in Fig. 3. Filled circles are stations S0–S4 and open squares are W0–W4. Line represents conservative mixing curve between seawater and riverine endmembers. Inset shows range of model isotopic values including river endmember at –10‰. more intensive time-series sampling from shore points (docks, bridges) within the instrument time constraints of this method. Discussion and recommendations Several authors (Tans et al. 1993; Quay and Stutsman 2003; Quay et al. 2007) have noted the potential for using 13C-DIC as a tracer for oceanic uptake of anthropogenic CO2. However, the requirements for such a tracer are significant. Anthropogenic signals are on the order of 0.025‰ year–1 in the subtropics to 0.01‰ year–1 in high-latitude oceans (Sonnerup et al. 1999; Tanaka et al. 2003; Tagliabue and Bopp 2008). The precision of the LC-IRMS method reported here (±0.04‰) would therefore dictate that it could be used to observe changes taking place over roughly a decade (subtropics) to 20 years (high latitude) timescales (S/N = 5). To obtain the highest precision from this technique, the following recommendations should be employed. One primary source of poor precision appears to be sample bottle pressurization. The evidence for this is that samples collected from near–room temperature waters and from lab standards have routinely low peak area and isotopic variability, whereas samples collected from colder waters exhibit larger variabilities, on the order of 1% and 0.6‰, respectively. Therefore, either first warming samples before subsampling (if bubble formation can be avoided) or the use of a refrigerated autosampler tray should act to reduce this problem. It may be possible that other autosampler types, Comments The Isolink interface provides a new avenue for δ13C-DIC analyses, with high precision, low sample preparation requirements, rapid analyses, and small sample volume as primary advantages. Challenges include sample storage methods for small volumes, isotopic linearity in lower concentration samples, and calibration standards. The operating costs and student/technician effort required for this application are low, consisting of 2-mL vials + septa, a syringe + filter (reusable with rinsing), and a few minutes for sample filling and capping. The lack of an oxidant allows for the use of lowcost deionized water for reagents, mobile phase, and wash solutions. If samples are not poisoned, then no hazardous waste is generated other than an easily neutralizable weak H3PO4 solution. The automated nature of the analyses means that long runs are possible, exceeding those from other IRMS interfaces such as elemental analyzers. Although Isolink interfaces are uncommon, this technique opens up their use for high-throughput analyses with no modification required, extending their utility when not being used for organic compound analyses. 737 Brandes Rapid isotopic analysis of DIC References matography/isotope ratio mass spectrometry by removing excess oxygen. Rapid Comm. Mass Spectrom. 21:4135-4141. Hu, X. P., and D. J. Burdige. 2007. Enriched stable carbon isotopes in the pore waters of carbonate sediments dominated by seagrasses: Evidence for coupled carbonate dissolution and reprecipitation. Geochim. Cosmochim. Acta 71:129-144. Kaldy, J. E., L. A. Cifuentes, and D. Brock. 2005. Using stable isotope analyses to assess carbon dynamics in a shallow subtropical estuary. Estuaries 28:86-95. Knohl, A., R. A. Werner, H. Geilmann, and W. A. Brand. 2004. Kel-F™ discs improve storage time of canopy air samples in 10-mL vials for CO2-delta C-13 analysis. Rapid Comm. Mass Spectrom. 18:1663-1665. Kroopnick, P. 1974. Dissolved O2-CO2-13C system in Eastern Equatorial Pacific. Deep-Sea Res. 21:211-227. Krummen, M., A. W. Hilkert, D. Juchelka, A. Duhr, H. J. Schluter, and R. Pesch. 2004. A new concept for isotope ratio monitoring liquid chromatography/mass spectrometry. Rapid Comm. Mass Spectrom. 18:2260-2266. McCorkle, D. C., and G. P. Klinkhammer. 1991. Porewater cadmium geochemistry and the porewater cadmium-delta 13C relationship. Geochim. Cosmochim. Acta 55:161-168. Nelson, S. T. 2000. Sample vial influences on the accuracy and precision of carbon and oxygen isotope ratio analysis in continuous flow mass spectrometric applications. Rapid Comm. Mass Spectrom. 14:293-297. Quay, P., and J. Stutsman. 2003. Surface layer carbon budget for the subtropical N. Pacific: Delta C-13 constraints at station ALOHA. Deep-Sea Res. Part I Oceanogr. Res. Pap. 50:1045-1061. ———, R. Sonnerup, J. Stutsman, J. Maurer, A. Körtzinger, X. A. Padin, and C. Robinson. 2007. Anthropogenic CO2 accumulation rates in the North Atlantic Ocean from changes in the C-13/C-12 of dissolved inorganic carbon. Global Biogeochem. Cycles 21 [doi:10.1029/2006GB002761]. Quay, P. D., B. Tilbrook, and C. S. Wong. 1992. Oceanic uptake of fossil fuel CO2: Carbon-13 evidence. Science 256:74-79. Salata, G. G., L. A. Roelke, and L. A. Cifuentes. 2000. A rapid and precise method for measuring stable carbon isotope ratios of dissolved inorganic carbon. Mar. Chem. 69:153-161. Sonnerup, R. E., P. D. Quay, A. P. Mcnichol, J. L. Bullister, T. A. Westby, and H. L. Anderson. 1999. Reconstructing the oceanic C-13 Suess effect. Global Biogeochem. Cycles 13:857-872. ———, A. P. McNichol, P. D. Quay, R. H. Gammon, J. L. Bullister, C. L. Sabine, and R. D. Slater. 2007. Anthropogenic delta C-13 changes in the North Pacific Ocean reconstructed using a multiparameter mixing approach (MIX). Tellus Ser. B Chem. Phys. Meteorol. 59:303-317. Tagliabue, A., and L. Bopp. 2008. Towards understanding global variability in ocean carbon-13. Global Biogeochem. Cycles 22 [doi:10.1029/2007GB003037]. Taipale, S. J., and E. Sonninen. 2009. The influence of preservation method and time on the delta C-13 value of dis- Butler, P. G., J. D. Scourse, C. A. Richardson, A. D. Wanamaker, C. L. Bryant, and J. D. Bennell. 2009. Continuous marine radiocarbon reservoir calibration and the C-13 Suess effect in the Irish Sea: Results from the first multi-centennial shell-based marine master chronology. Earth Planet. Sci. Lett. 279:230-241. Cai, W. J., and Y. Wang. 1998. The chemistry, fluxes, and sources of carbon dioxide in the estuarine waters of the Satilla and Altamaha Rivers, Georgia. Limnol. Oceanogr. 43:657-668. ———, Z. H. A. Wang, and Y. C. Wang. 2003. The role of marsh-dominated heterotrophic continental margins in transport of CO2 between the atmosphere, the land-sea interface and the ocean. Geophys. Res. Lett. 30(16) [doi:10.1029/2003GL017633]. Chanton, J. P. 1999. Plankton and dissolved inorganic carbon isotopic composition in a river-dominated estuary: Apalachicola Bay, Florida. Estuaries 22:575-583. Coffin, R. B., and L. A. Cifuentes. 1999. Stable isotope analysis of carbon cycling in the Perdido Estuary, Florida. Estuaries 22:917-926. Coplen, T. B. 1996. More uncertainty than necessary. Paleoceanography 11:369-370. Feely, R. A., T. Takahashi, R. Wanninkhof, M. J. McPhaden, C. E. Cosca, S. C. Sutherland, and M.-E. Carr. 2006. Decadal variability of the air-sea CO2 fluxes in the equatorial Pacific Ocean. J. Geophys. Res. Oceans 111 [doi:10.1029/2005JC003129]. Fry, B. 2002. Conservative mixing of stable isotopes across estuarine salinity gradients: A conceptual framework for monitoring watershed influences on downstream fisheries production. Estuaries 25:264-271. Gehlen, M., A. Mucci, and B. Boudreau. 1999. Modelling the distribution of stable carbon isotopes in porewaters of deepsea sediments. Geochim. Cosmochim. Acta 63:2763-2773. Gillikin, D. P., and S. Bouillon. 2007. Determination of delta O-18 of water and delta C-13 of dissolved inorganic carbon using a simple modification of an elemental analyzer-isotope ratio mass spectrometer: An evaluation. Rapid. Comm. Mass Spectrom. 21:1475-1478. Gruber, N., C. D. Keeling, and T. F. Stocker. 1998. Carbon-13 constraints on the seasonal inorganic carbon budget at the BATS site in the northwestern Sargasso Sea. Deep-Sea Res. Part I Oceanogr. Res. Pap. 45:673-717. ———, and others. 1999. Spatiotemporal patterns of carbon13 in the global surface oceans and the oceanic Suess effect. Global Biogeochem. Cycles 13:307-335. Hedges, J. I., and R. G. Keil. 1995. Sedimentary organic-matter preservation: An assessment and speculative synthesis. Mar. Chem. 49:81-115. Hettmann, E., W. A. Brand, and G. Gleixner. 2007. Improved isotope ratio measurement performance in liquid chro738 Brandes Rapid isotopic analysis of DIC Tans, P. P., J. A. Berry, and R. F. Keeling. 1993. Oceanic C-13/ C-12 observations: A new window on ocean CO2 uptake. Global Biogeochem. Cycles 7:353-368. Torres, M. E., A. C. Mix, and W. D. Rugh. 2005. Precise delta C-13 analysis of dissolved inorganic carbon in natural waters using automated headspace sampling and continuous-flow mass spectrometry. Limnol. Oceanogr. Methods 3:349-360. solved inorganic carbon in water samples. Rapid Comm. Mass Spectrom. 23:2507-2510. Takahashi, T., S. C. Sutherland, R. A. Feely, and R. Wanninkhof. 2006. Decadal change of the surface water pCO2 in the North Pacific: A synthesis of 35 years of observations. J. Geophys. Res. Oceans 111 [doi:10.1029/2005JC003074]. Tanaka, T., Y. W. Watanabe, S. Watanabe, S. Noriki, N. Tsurushima, and Y. Nojiri. 2003. Oceanic Suess effect of delta C-13 in subpolar region: The North Pacific. Geophys. Res. Lett. 30(22) [doi:10.1029/2003GL018503]. Submitted 9 June 2009 Revised 15 September 2009 Accepted 25 September 2009 739
© Copyright 2026 Paperzz