Carbon Monoxide - Health and Safety Laboratory

Biological Monitoring
Guidance Values
Guidance sheet for:
Method for Monitoring Carbon
Monoxide by Breath Carbon Monoxide
BMGV= 30ppm CO in end-tidal breath
Hazardous Substance
Carbon monoxide, CO
CAS No 630-08-0
Workplace Exposure
Limit
= 30 ppm
Short-term Exposure Limit
= 200 ppm
The Health and Safety Laboratory is an
Agency of the Health and Safety Executive
Biological Monitoring
Guidance Values
Method for Monitoring Carbon Monoxide byBreath Carbon Monoxide
Biological Monitoring
Guidance Value:
Guidance Value = 30 ppm end-tidal
breath carbon monoxide (CO)
Sample Collection
Time: Breath samples should be taken at the end
of the shift
Description of Suggested Method
The OESs for CO have been established at
values which maintain blood carboxyhaemoglobin
(COHb) levels below 5%. The good relationship
between breath CO and COHb means that
measurement of CO in end-tidal breath is a
reliable non-invasive approach to biological
monitoring. A number of portable, direct reading
carbon monoxide monitors for breath analysis are
commercially available. The CO monitors are
invariably based on electro chemical sensors with
a display of CO concentration or blood COH
equivalents.
Analytical Evaluation
Drift - less than 2% a month
Detection Limit - 2-3 ppm
Calibration Range - typically 0-500 ppm
Analytical Interferences - None likely to be encountered in breath.
Negligible effect from organic solvents.
Environmental CO exposure can influence the
measurement; it is
recommended to carry out breath analysis in an
environment removed
from external CO sources.
Other Information
Elimination half-time
- Elimination half-life in breath after CO exposure is 3-5 hours
Confounding Factors
- Carbon monoxide exposure from confined vehicle exhaust emissions
and tobacco smoke. Some dihalomethanes
(e.g. dichloromethane and
bromochloromethane) are metabolised to CO.
Passive smoking will not
significantly influence breath CO levels.
Biological Monitoring
Guidance Values
Method for Monitoring Carbon Monoxide byBreath Carbon Monoxide
Quality Assurance
Quality assurance for breath sampling is not
practical. However, regular calibration of these
instruments is recommended using standard gas
mixtures containing CO concentrations at
appropriate levels (50 ppm). Technical data with
specific instruments will suggest calibration
intervals, but are usually between three to six
months. If the CO monitor needs zeroing between
individual readings it is important that this is
performed in a low CO contaminated atmosphere.
Outdoor atmospheres without excessive vehicle
emissions are generally suitable for this purpose
with CO levels of 2 ppm or less. Before zeroing
indoors, sources of CO from the work-process,
heating appliances or tobacco smoke need to be
considered.
Other Guidance Values
The ACGIH BEI is 20ppm CO in end exhaled air
or 3.5% carboxy haemoglobin in blood collected
at the end of the shift. The DFG BAT is 5%
carboxy haemoglobin in blood collected at the
end of the shift.
Interpretation
Exposure to the OES of 30ppm CO in a
non-smoker leads to an end-of-shift breath CO
level of 30 ppm equivalent to 5% COHb
Unexposed Levels
- in non-smokers <5ppm.
- in light smokers <20ppm.
- in heavy smokers >20ppm.
(DoH guidance 1983)
Therefore smoking during the workshift will
substantially reduce the value of the end-of-shift
breath measurement as a measure of
occupational CO exposure. Since the elimination
half-life is 3-5 hours, if the worker has not
smoked during the shift, its confounding
influence is reduced. Although smoking during
the shift can complicate the interpretation of the
BMGV for CO for that individual, assessment of
biological monitoring on a group basis may still
be useful in determining the effectiveness of
control.
Biological Monitoring
Guidance Values
Method for Monitoring Carbon Monoxide byBreath Carbon Monoxide
Links
EH40 List of Approved Workplace Exposure
Limits http://www.hse.gov.uk/coshh/table1.pdf
Biological Monitoring at HSL
http://www.hsl.gov.uk/online-ordering/analyticalservices.aspx
References
EH64: Summaries criteria for occupational
exposure limits - 1998 supplement, HSE Books
Irving et al (1988) Evaluation of a portable
measure of carbon monoxide.
Preventative Medicine, 17, 109-115.
Jarvis et al. (1986) Low cost carbon monoxide
monitors in smoking assessment. Thorax, 41,
886-887.
Kuster & Sherwood. (1996) Determination of
carbon monoxide in breath.
A study of portable techniques. Occupational
Hygiene, 3, 243-254.