ARTICLE IN PRESS Available online at www.sciencedirect.com Biosystems Engineering (2003) 86 (3), 257–266 doi:10.1016/S1537-5110(03)00138-7 AE}Automation and Emerging Technologies Infrared Radiometry for Measuring Plant Leaf Temperature during Thermal Weed Control Treatment J. Rahkonen; H. Jokela Department of Agricultural Engineering and Household Technology, University of Helsinki, P.O. Box 27, Helsinki FIN-00014, Finland; e-mail of corresponding author: jukka.rahkonen@mmm.fi (Received 1 July 2002; accepted in revised form 14 July 2003; published online 4 September 2003) One side of a plant leaf was exposed to liquefied petroleum gas flames, while the temperature of the other side was measured continuously with an imaging infrared radiometer. Temperature histories of leaves and the performance of the measuring system were studied by flaming over 200 leaves and recording the measurements for off-line analysis. Flaming raised the temperature of the leaves very rapidly. The peak heating rate was typically around 1808C s1. The imaging speed of the radiometer, 50 fields s1, was sufficient to evaluate the temperature histories. Flaming induced marked temperature differences in the flamed leaves, with veins heated to lower temperatures than the thin areas between them. Thermograms with good spatial resolution were essential to characterise the temperature distribution with gradients as steep as 508C over a 1 mm interval. With a factory calibrated radiometer an accuracy of 218C in absolute temperature values for a temperature range of 20–908C was achieved by determining the target emissivity with the radiometer. By calibrating the radiometer for the actual conditions under measurements the error could be reduced to 108C. In conclusion, an imaging infrared radiometer is a superior instrument for measuring fast, spatially distributed temperature changes in plant leaves during thermal weed control treatment. # 2003 Silsoe Research Institute. All rights reserved Published by Elsevier Ltd 1. Introduction Thermal weed control is a collective name for various physical methods to kill weeds by using exceptionally high or low temperatures or electrical fields (Ascard, 1995). In practice, however, thermal weed control usually refers to killing weeds by heat, and methods involving the use of electrical fields or low temperatures are rarely used. For high-temperature treatments, a number of energy sources from focused sun rays to microwave radiation (Diprose et al., 1984) have been tested, but by far the most popular method is flaming with liquefied petroleum gas (LPG) burners. Nonetheless, flaming is not the only method of practical relevance. Infrared radiators (Reifschneider & Nunn, 1965), hot gas stream (Porterfield et al., 1967, Bertram & Meyer, 1996) and hot water (Albrecht, 1985) are all used to some extent for weed control or for an analogous treatment, thermal defoliation. 1537-5110/$30.00 Irrespective of the energy source, all high-temperature weed control methods kill plants by heating a critical number of cells to a lethal temperature. Thus, valuable information about the processes in thermal weed control should be yielded by temperature measurements. However, few studies have been done on plant temperatures in flaming, despite the first ones being performed in the 1960s and 1970s (Thomas, 1967; Porterfield et al., 1971). Yet in other contexts, plant temperatures have been measured intensively at least since 1875 by various investigators, as reviewed by Gates (1980), and different measurement methods and their advantages and disadvantages have been discussed widely. Plant temperatures can be measured either by contact sensors, usually small thermocouples or thermistors, or by infrared temperature meters. Accurate measurements can be obtained with both methods, assuming all possible error sources are understood and eliminated. 257 # 2003 Silsoe Research Institute. All rights reserved Published by Elsevier Ltd ARTICLE IN PRESS 258 J. RAHKONEN; H. JOKELA To ensure reliable temperature measurements with contact sensors, a good thermal contact between the sensor and object is necessary. With plants, this is best achieved by inserting the sensor inside plant tissues. Sensor insertion into thin parts, such as leaves, may be difficult or may cause excessive damage. Leaf temperature is therefore usually obtained by attaching the sensor onto the leaf surface, but great care is needed with this method to assure an adequate thermal connection. Another problem is heat conduction through connection wires when a temperature difference exists between the plant and its surroundings. Beadle et al. (1973) studied the accuracy of leaf temperature measurements and found that a poor thermal contact together with thermal conduction along the thermocouple wires can lead to a marked measurement error. With a relatively large thermocouple junction, the error reached 358C, while the temperature difference between the leaf and surrounding air was no more than 798C. Errors of the same magnitude are reported by Pieters and Schurer (1973), who reported 5–78C measurement errors with a 238C temperature difference between the leaf and air. With a careful positioning of thermocouples, the error could be reduced to 18C. Tarnopolsky and Seginer (1999) postulated, that to reduce conduction error to 1% of the temperature difference, at least 10 mm length of thermocouple wires with insulation removed should be glued to a leaf. Similar results have been given for inserted sensors, which are also prone to conduction error. Cook et al. (1964) got reliable measurements by inserting the thermocouple junction and 10 mm length of thermocouple wires under the leaf epidermal surface. Plant temperature measurement by contact sensors is laborious if measurement errors are to be eliminated. In addition, the sensors themselves may affect the plant temperature (Tanner, 1963). For these reasons, temperature measurement of vegetation by infrared thermometers has become common since the 1960s, when the first commercial models became available on the market. Infrared thermometry is a non-invasive, non-destructive method that has a negligible influence on the target temperature. Infrared thermometers measure the surface temperature, which can sometimes be a drawback. Another drawback of non-imaging IR thermometers is their spatial resolution, typically some millimetres at a minimum, thus limiting their use with very small targets. The accuracy of radiative temperature measurements of plants have been discussed by Fuchs and Tanner (1966), Sutherland and Bartholic (1979), Amiro et al. (1983), Graham (1989), Hipps (1989), van de Griend et al. (1991), and others. Errors in radiative temperature measurements originate from incorrect estimates for emissivity values and background radiation. The lower the emissivity, the greater is the influence of erroneous estimates. Green leaves have very high emissivity values, 095–100 (Gates, 1964; Idso & Jackson, 1968; Willey, 1985; Elwidge, 1988; Hashimoto, 1990), rendering them a favourable target for radiative temperature measurements. Fuchs and Tanner (1966) have shown that the temperature of vegetation can be measured with an infrared thermometer to an accuracy of 01–038C when the emissivity is known. A specialised branch of infrared thermometry is imaging infrared radiometry, or infrared thermography. In thermography, a large amount of point temperatures (i.e. 256 256) are measured over an area and processed to form a thermal map, or thermogram, of the target surface. Thermography with a high spatial resolution is a powerful tool for analysing and visualising targets with thermal gradients. The imaging speed of the instrument is high, e.g. 50–60 images per second, which makes it especially suitable for exploring rapidly changing thermal conditions. Plant leaf temperature patterns have been studied by thermocouple measurements (Cook et al., 1964) and by thermography to determine the relationships between temperature and water stress (Hashimoto et al., 1984) and between temperature and stomatal conductance (Jones, 1999). These investigations have shown that under normal growing conditions, temperature differences of several degrees Celsius can be found within a single leaf. Thus, it could be surmised that in thermal weed control, with conditions far removed from the natural, even wider temperature variations in leaves would be present. With contact sensors, this thermal variation would be very difficult to explore because the high radiation and at least 12008C temperature difference between flames and leaf are likely to introduce radiation and conduction error in measurements. Moreover, careful arranging of several thermocouples on every leaf to be surveyed is both laborious and prone to affect the temperature to be measured. The objective of this paper is to present an experimental design enabling the use of an imaging infrared radiometer for measuring temporal and spatial variation of temperature in plant leaves during thermal weed control and estimate the accuracy of measurements obtained by this method. The measurement data are used for describing the thermal behaviour of true plant leaves treated with LPG flamer and for evaluating the benefits and drawbacks of alternative temperature measurement methods. 2. Materials and methods 2.1. Basic theory Radiometric surface temperature measurements are based on measuring the energy emitted by a target. The ARTICLE IN PRESS MEASUREMENT OF PLANT LEAF TEMPERATURE BY INFRARED RADIOMETRY energy emitted by a blackbody (emissivity e ¼ 1) at any given wavelength l is given by Planck’s distribution law as follows: EðlÞ ¼ 8phcl5 expðhc=klTÞ 1 ð1Þ where: c is the speed of light (3 108 m s1), h is Planck’s constant (663 1034 Js), k is Boltzmann’s constant (138 1023 J K1) and T is the object temperature in K. Integration of Planck’s equation for all wavelengths gives the total energy emitted by the blackbody. The result of the integration is known as Stefan–Boltzmann’s law, R ¼ sT 4 ð2Þ where: R is the flux emitted by unit area of a plane surface into an imaginary hemisphere surrounding it in W m2 and s is Stefan–Boltzmann’s constant (567 108 W m2 K4). With radiometers operating at a limited band width, the Stefan–Boltzmann equation is not applicable, instead Planck’s law must be used. Numerical integration of Planck’s law for a wavelength range of 8–12 mm and a temperature range of 290–360 K is given by R8212 mm ¼ s T 45 ð4Þ Combining Eqns (3) and (4) gives the total amount of thermal infrared radiation entering the radiometer set at a wavelength of 8–12 mm and a temperature of 290– 360 K as follows: RB ¼ es Tl45 þ ð1 eÞs Tb45 2.2. Imaging infrared radiometer Thermal images were obtained by using an Inframetrics 760 E imaging infrared radiometer (Inframetrics, Inc., part of the FLIR Systems company since 1999). Model 760, manufactured in 1990–1998, is a scanning long-wave (8–12 mm) device with a single HgCdTe detector. The imaging speed of the scanner is 50 fields s1 with 240 infrared lines field1 and 194 resolution elements line1 [50% response on Slit Response Function (SRF)]. A spatial resolution of 18 mrad (50% SRF) is obtained by standard optics with a 158 vertical field of view (FOV) and a 208 horizontal FOV. The Inframetrics 760 system converts thermal measurements to an 8-bit (256 level) greyscale video signal that can be recorded by a standard VHS videotape recorder. For the recordings, a regular, new VHS videotape was used. Analysis of the recorded measurements was performed by a model 760 compatible digital thermal image processing system (Thermagram 50 by Thermoteknix Systems Ltd.). 2.3. Experimental arrangement ð3Þ where s* is a constant analogous to Stefan–Boltzmann’s constant. Real objects invariably emit less energy than a blackbody. The ratio of the actual radiation to the hypothetical is indicated by emissivity e. For opaque objects, such as green leaves at wavelengths beyond 8 mm (Gates & Tantraporn, 1952), Eqn (4) is valid for emissivity and reflectivity l: eþl¼1 259 ð5Þ where: RB is the energy flux emitted at wavelength of 8–12 mm in W m2, Tt is the temperature of the target in K and Tb is the background temperature in K that the target is reflecting. Equation (5) addresses the two sources of error in radiative temperature measurements, viz. the estimates of emissivity and background temperature. These errors in connection with plant temperature measurements have been discussed in more detail by Sutherland and Bartholic (1979), Amiro et al. (1983), Hipps (1989) and Svendsen et al. (1990). To investigate the temperature of plants during thermal treatment, the rate of heating and the spatial temperature distribution, one leaf at a time was flamed from one side with a LPG burner, while the temperature of the other side of the leaf was recorded by an imaging infrared radiometer. The experimental configuration is shown in Fig. 1. A vertical plate, made of painted metal, was inserted in front of the infrared scanner at a distance of 70 cm. A fresh leaf, just separated from a plant, was attached to a plate with the underside towards the plate to completely cover a round measuring window (diameter 244 mm) in the middle of the plate. A metal spring was used to smoothly press the margins of the leaf against the plate. The upper surface of the leaf was flamed with a round, horizontally moving LPG burner using gas phase propane as a fuel. The burner was equipped with a 05 mm nozzle and was operated at a pressure of 100 kPa. By changing the burner speed, propane doses 5–180 kg ha1 were achieved. By comparison, under field conditions, LPG consumption over the entire area being flamed is usually 30–80 kg ha1. The infrared scanner was focused to the middle of the measuring window to continuously record the temperature of the underside of the leaf. The total length of the optical path, including a 22 cm internal path inside the scanner, was 92 cm. This, together with a 18 mrad spatial resolution, resulted in a resolution unit size of ARTICLE IN PRESS 260 J. RAHKONEN; H. JOKELA Infrared radiometer Leaf attached on a metal sheet Moving LPG burner Control unit Video tape recorder Fig. 1. Experimental arrangement for measuring temperature of one side of a plant leaf by imaging infrared radiometer while the other side is exposed to liquid petroleum gas (LPG) flames 17 mm by 17 mm on the leaf surface (50% SRF). The total number of resolution units within the measuring window was approximately 160. 2.4. Plant material A primary requirement for a plant was that it be relatively broad-leaved to offer sufficient surface area for investigation of spatial temperature variation. The first test plants were greenhouse-grown rape seeds (Brassica rapa ssp. oleifera) and a benjamin tree (Ficus benjamina), which had until the experiment been cared for as a houseplant. Later, leaves from greenhouse grown field sow-thistle (Sonchus arvensis) were also used. For each plant species, about 100 leaves were flamed at various flaming intensities, the extremes being 5 and 180 kg ha1 LPG. All the measurements were recorded on videotape for later analysis. & Davies, 1972; Pinkley et al., 1977; Zhang et al., 1986; Salisbury & Milton, 1988). The background radiation level was measured by using an aluminium foil inserted near the water surface as a reflector. The radiation levels of a reference surface and a leaf carefully lowered to float on the water were measured. The emissivity of the leaf was calculated using the equation: ðRtarget Rbackground Þ ereference ð6Þ etarget ¼ ðRreference Rbackground Þ The emissivity value was determined to be 098 001 for both Brassica rapa and Sonchus arvensis leaves. 2.6. Thermal image analysis 2.5. Leaf emissivity measurement Several different thermal analyses were performed by using the Thermogram 5.0 digital infrared thermal image processing system. Because of the large amount of measurement data, all of the material was not included in each analysis. The different analyses and the material used in each are listed below. Emissivities of three Brassica rapa and Sonchus arvensis leaves were measured with a reference emittance technique described in the Inframetrics model 760 operator’s manual. A well-stirred water bath at a temperature of about 458C was used as a reference material (Berliner et al., 1984). Temperature of the water bath was controlled by a mercury thermometer accurate to 018C. Emissivity of water (l ¼ 8–12 mm) was assumed to be 098 (Buettner & Kern, 1965; Robinson 2.6.1. Repeatability of image analysis and total noise in recorded signal The manufacturer specifies the accuracy of model 760 radiometers to 28C or 2% of a blackbody target source and the repeatability of measurements to 058C or 05%. These values do not, however, represent the accuracy and repeatability of temperature analysis performed afterwards on the basis of recorded measurements. In the analysis system used in this study, the ARTICLE IN PRESS 261 MEASUREMENT OF PLANT LEAF TEMPERATURE BY INFRARED RADIOMETRY 2.6.2. Spatial variation in temperature Spatial variation in temperature was analysed by capturing thermal images at the moment of peak temperature for every flamed Brassica rapa leaf. These thermal images represent well the temperature patterns over a surface. Exact numbers were obtained by computing histograms of the proportional distribution of point temperatures within a measurement window. Linear temperature gradients were studied by creating line graphs describing temperatures along a line crossing a leaf. 2.6.3. Temporal variation in temperature Temporal variation in temperature was analysed by plotting the average temperature of a selected area of a leaf versus time at a maximum sample rate of 50 Hz. Analysis was performed for all leaves of the three species. The size of the area used for averaging was limited by the computing speed to 1688 image pixels, corresponding to a round surface area of 82 mm in diameter. In addition, simultaneous temperature plots from three circular areas, each with 392 image pixels in a 40 mm diameter, were generated for 50 Brassica rapa leaves. The plot areas were situated on primary veins, secondary veins and the thin areas between veins. Calculations were performed based on actual conditions during measurements. The emissivity value of 098 was assumed to be correct to 001. The measured background temperature in the laboratory was 178C, and in extreme cases the actual radiative temperature was estimated to vary between 12 and 278C. Errors were calculated for the same temperature range as for the numerical integration of Planck’s law, i.e. 17–878C (290–360 K). The influence of emissivity error was also calculated for a case where emissivity is not measured but just assumed to be 098. In this case, the upper limit for actual emissivity is 100. Based on various references for emissivity (l ¼ 8–12 mm) of green leaves, 095 was used as a lower limit (Tanner, 1963; Gates, 1964; Fuchs & Tanner, 1966; Idso & Jackson, 1968; Willey 1985; Salisbury, 1986; Elwidge, 1988; Hashimoto, 1990). The temperature errors caused by the use of incorrect estimates are shown in Fig. 2. When the emissivity value of 098 is accurate to 001, the error in absolute temperature is less than 068C up to 908C and less than 048C below 608C. When temperature differences are considered, the temperatures are read from the same curve and the errors partly cancel each other out. Thus the error at 708C true temperature difference between 20 and 908C is less than 058C. Considerably larger errors can occur if the emissivity value is not determined (the outmost curves), the maximum error ranging from 15 to +108C at 908C. 1.50 1.00 Temperature error, K measurement signal goes through a long chain before the initial radiative measurements are converted to the final temperature values. This chain includes digital to analogue (D/A) conversion to a recordable video signal, videotape recording and playback, capturing of the playback signal, and an A/D conversion before the actual computerised image analysis can take place. As neither the hardware nor the software manufacturer assessed the overall accuracy of the analysis, two tests were performed. In a repeatability test series, at least 10 temperature–time plots were generated from the same measurements. The total noise included in the temperature signal was investigated by creating histograms of single pixel temperatures from a recorded measurement of a large target at a constant temperature. 0.50 0.00 − 0.50 − 1.00 − 1.50 3. Accuracy of measurements 3.1. Error induced by uncertain target emissivity and background temperature values The image processing system makes an automatic background radiation correction and calculates the target temperature by using the given emissivity and background temperature values as parameters. Equation (5) can be used for calculating the temperature errors arising from faulty parameters. − 2.00 270 290 310 330 350 True target temperature, K 370 390 Fig. 2. Temperature errors in using values 098 and 290 K for emissivity and background temperature, respectively, instead of true values; true values: &}&, 100 and 300 K; } , 099 and 300 K; n}n, 097 and 285 K; +}+, 095 and 285 K; the outermost curves represent maximum errors when true emissivity varies between extreme values for green leaves, i.e. from 100 to 095, while the curves in between represent maximum errors when emissivity value is measured to accuracy of 001, which was the case in this study ARTICLE IN PRESS 262 J. RAHKONEN; H. JOKELA 3.2. Other error sources 3.2.1. Accuracy of the instrument The Inframetrics model 760 radiometer is factory calibrated to an accuracy of 28C or 2%, which is valid over an ambient temperature range of 15 to 508C. This is also the accuracy of the instrument used in calculations. Lacking possibility for blackbody calibration and without knowing the actual calibration curves no better accuracy could be assumed over the whole measurement range of the instrument, even though according to the manufacturer the typical accuracy at a constant ambient temperature of 238C is 18C or 1% (Inframetrics, Inc., 1991). At a limited target temperature range of 20–508C, the radiometer was calibrated against a mercury thermometer while the plant leaf emissivity was measured, and observed to be accurate to 18C. 3.2.2. Total signal noise The total signal noise of the image processing system was measured by analysing thermal images obtained from a painted metal plate at a uniform temperature. Thermal images showed Gaussian distribution for single pixel temperatures. The standard deviation was 0858C at an average temperature of 2288C (sample size 19 008 pixels). At a 95% confidence level, the accuracy of a single pixel measurement is 188C. To obtain 018C accuracy, the result must be the average of at least 320 pixels. Pixel number 1688, used for time–temperature plots, yields an accuracy of 0048C at a 95% confidence level. with 1688 image pixels were plotted over 20 s at a maximum sample rate of 50 s. The average of the peak temperatures was 6198C, with a minimum of 6158C and a maximum of 6218C. Standard deviation was 0178C, therefore, at a 95% confidence level, the accuracy was 048C. The average temperatures on time–temperature plots were compared with temperatures computed for the same areas on sequential images captured from live video. When the frame capture was successful, these two temperatures were the same to 028C. This was, however, not always the case. Occasionally, frame capture failed totally or, in the worse case, the temperature information on the captured image was corrupted, leading to temperature error of several degrees. 3.3. Total error The total error for independent errors showing normal distributions can be calculated as follows: ffi qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi Xn 2 Dm ð7Þ Dmtotal ¼ i i¼1 where: Dmtotal is the total error and Dmi . . . Dmn are the independent error terms. A summary of the errors caused by individual error terms and the total error of measurements is given in Table 1. 4. Results 4.1. Spatial temperature variation in leaves 3.2.3. Repeatability of live video temperature plotting and video frame capturing The reliability of live video temperature plotting was tested by repeating the plotting 15 times for the same video sequence. Average temperatures of a round area Thermal images captured at the moment of peak temperature demonstrated a marked spatial variability in leaf surface temperature. A typical example is presented in Fig. 3. A greyscale image is a thermograph Table 1 Examples of independent error terms in temperature measurement by imaging infrared radiometer and the total error of measurement Source of the error Magnitude of the error Conditions (a) Wrong estimates (b) Wrong estimates (c) Calibration error (d) Calibration error 058C 15108C 28C or 2% 18C (e) Sample error (f) Plotting error 018C 048C At 608C, emissivity measured At 908C, emissivity estimated At ambient temperature range of 15 to 508C At ambient temperature range of 208C and target temperature range of 20–508C Averaged from 320 pixels At 608C Total error at 20–508C Total error at 608C Total error at 908C 128C 218C 25 to 238C ða þ d þ e þ f Þ ða þ c þ e þ f Þ ðb þ c þ e þ f Þ ARTICLE IN PRESS 263 MEASUREMENT OF PLANT LEAF TEMPERATURE BY INFRARED RADIOMETRY with a whitish colour denoting higher temperatures. Beneath the image, a line graph presents the temperatures along a horizontal line marked on the image. A histogram gives the distribution of temperatures of the 19 984 image pixels within the measuring window. The average temperature of the leaf at the moment of image acquisition was 6408C. However, almost 40% of the leaf area was either colder than 508C or hotter than 808C, the minimum and maximum temperatures being 246 and 9708C. (a) 4.2. Temporal temperature variation in leaves 100 90 Temperature, °C 80 70 60 50 40 30 20 0 5 10 15 Distance, mm (b) 20 25 400 350 Number of pixels 300 250 200 Different intensities in flaming treatments were achieved by changing the burner velocity. This can be seen in Fig. 4, which describes the average temperatures of Sonchus arvensis leaves (seven replicates) flamed at different intensities. Initially the temperature of the leaves rises at a fairly equal speed in all treatments, but with heavier flamings, the leaves are heated for longer periods and reach higher temperatures. The maximum rates of heating were 57, 160, 187 and 1808C s1 for treatments 20, 30, 70 and 100 kg ha1 LPG, respectively (calculated over a 01 s duration). The applicability of the 50 Hz sample rate to characterise the temperature changes was verified by visually checking the ascending part of temperature curves. An example, given in Fig. 5, presents temperature rise in a Brassica rapa leaf receiving 70 kg ha1 LPG treatment. Spatial temperature variation in leaves was also changing with time as can be seen in Fig. 6. The figure presents temperature histories of three distinct areas within a single Brassica rapa leaf flamed at intensity of 70 kg ha1 LPG. The variation was as largest immediately after the thermal treatment and faded gradually, but temperature differences of several degrees Celsius 150 100 100 90 50 21 30 40 48 57 64 72 Temperature, °C 80 87 94 Fig. 3. (a) Typical thermograph showing temperature variation at the moment of peak temperature in a Brassica rapa leaf flamed at 70 kg ha1; (b) curve presenting temperatures on the line shown in the thermograph; (c) histogram presenting the distribution of pixel temperatures within the circle shown in the thermograph Temperature, °C 0 (c) 100 kg/ha 80 70 kg/ha 70 30 kg/ha 60 50 20 kg/ha 40 30 20 10 0 1 2 3 4 5 Time, s of a Brassica rapa leaf just flamed at 70 kg ha1 LPG. Veins, which remain relatively cool, can be clearly distinguished as dark lines from the thin middle areas Fig. 4. Temperature histories of Sonchus arvensis leaves flamed at intensities of 20, 30, 70 and 100 kg ha1 LPG; each curve is an average of seven replicates ARTICLE IN PRESS 264 J. RAHKONEN; H. JOKELA 70 Temperature, °C 60 50 40 30 20 10 1.4 1.5 1.6 1.7 1.8 1.9 Time, s Fig 5. Temperature rise in a Brassica rapa leaf receiving 70 kg ha1 LPG flaming treatment; the points present momentary temperatures measured at 50 Hz sample rate 100 90 Temperature, °C 80 1 70 2 60 3 50 40 30 20 10 0 1 2 3 4 5 Time, s Fig. 6. Simultaneous time–temperature plots from three circular areas, each with 392 image pixels in a 40 mm diameter, situated on: 1, the thin area between veins; 2, secondary vein; and 3, primary vein of a Brassica rapa leaf flamed at intensity of 70 kg ha1 LPG existed for many seconds. Figure 6 is also presenting an extremely fast temperature rise: thin area of the leaf was warmed up from 249 to 8698C within 01 s corresponding to heating rate of 6208C s1. 5. Discussion Three distinct factors introduce error to temperature measurement with imaging infrared radiometers: incorrect estimates of target emissivity and background radiation, inaccuracy of the instrument itself, and discrepancies over the signal processing chain between native measurements and final results. For leaf temperature measurement during LPG flaming using one particular measurement and analysis system, the magnitudes of these three errors were 058C, 208C and 048C, respectively, at 608C over an area of 320 image pixels. Combining these independent errors yields an overall measurement accuracy of 218C. The first error, caused by the faulty estimates of emissivity and background, is similar for all infrared radiometers, with slight differences according to the wavelength range utilised. To reduce this error, the estimates should be determined with an accuracy better than 001. However, radiative temperature measurement of green leaves is relatively insensitive to erroneous parameters since the emissivity of leaves is very high (e ¼ 095–100), and thus, more accurate estimates would only yield a small improvement to the overall accuracy. Furthermore, because of natural variation, some uncertainty in the emissivity of leaves is likely to remain. The second error term, the accuracy of the radiometer itself, is instrument-specific. The accuracy of the radiometer used in this study was specified by the manufacturer as 28C for an ambient temperature range of 15 to 508C. Further, according to the manufacturer, the typical accuracy at an ambient temperature of 238C is 18C, but because calibration data was lacking, this information was not deemed reliable. Had this data been available, the accuracy under standardised ambient conditions could be improved to 058C, the value specified by the manufacturer for repeatability of measurements. As calibration error is the biggest error term, its reduction would greatly improve overall accuracy. A 078C calibration error would yield an overall accuracy of 108C if other conditions remain unchanged. The third error term, also system-specific, comprises noise from various sources in the image analysis system as well as the error caused by the step-wise response of digital systems. The system used in this study, with an eight-bit dynamic range and an analogue VHS videotape as a bulk data storage medium, gave an error of 048C. Some of this error might be eliminated by using the new imaging infrared system with a 12-bit dynamic range and all digital data storage, but again the potential improvement to overall accuracy would be relatively small. However, the benefit of using digital data storage would be considerable if the problems occasionally encountered in data retrieval with the analogue system could be avoided. Both the temporal and the spatial variations in temperature of the leaves under flaming treatment were great. The sample rate of 50 Hz obtained by the radiometer was high enough to characterise the temporal variation, as can be seen in Fig. 5, where both the rising part of the slope and the moment of peak ARTICLE IN PRESS MEASUREMENT OF PLANT LEAF TEMPERATURE BY INFRARED RADIOMETRY temperature are well covered with several sample points. In this measurement, a sample rate of 25 Hz can be considered as a minimum to represent the shape of the temperature curve correct. In a thermogram, each point temperature is an average temperature from a larger area called a resolution unit. Temperatures from points with intervals equal to or greater than the size of the resolution unit are independent, otherwise they are averaged from areas partly covering each others, thus leading to underestimated temperature differences. Therefore, the size of the resolution unit should suit the minimum distance over which accurate temperature differences are to be measured. In this study the size of the resolution unit was 17 mm by 17 mm. Figure 3 shows temperature differences are as high as 508C over 1 mm distance. This indicates that reducing the size of the resolution unit would have been advantageous. Standard optics of the radiometer make used in this study allow a resolution unit size of 09 mm by 09 mm, but with alternative optics even the size of 0004 mm by 0004 mm is achievable. Limiting factor in practice is the distance between the leaf and the radiometer needed to protect the valuable instrument from flames possible penetrating through the measuring window. On the basis of our study, point temperatures with at most 2 mm intervals should be measured at 25 Hz sample rate at minimum in order to characterise the spatial and temporal temperature variation in plant leaves during thermal weed control properly. In addition, the measuring device should have very fast response time and it should be insensitive to thermal disturbance caused by the thermal treatment. In order to satisfy these demands a high-grade imaging infrared radiometer with fast imaging speed and good optics is a superior tool. Non-imaging infrared thermometers, based on the same principles of radiative temperature measurements, can offer as good measurement accuracy as imaging radiometers, but because they are limited to measure single-point temperatures no information about spatial variation of temperature can be gained. Infrared thermometers has also too slow response times to properly record the very fast temperature changes we found to occur during thermal treatment. By using contact sensors, because of the tough conditions during a thermal treatment it would be very hard to achieve measurement accuracy comparable with radiative temperature measurement methods even in single-point measurements. When not performed with special care, measurement by thermocouples has been reported to give a 5–78C temperature error with a temperature difference of 238C between the leaf and 265 surroundings (Pieters & Schurer, 1973). Higher accuracy can be obtained if good thermal contact exists between the plant and the thermocouple, and conduction and radiation errors are eliminated. In practice, this can be achieved, for example, by gluing at least a 10 mm length of thermocouple wires onto a leaf (Tarnopolsky & Seginer, 1999). This will reduce the conduction error to 1% of the temperature difference. When it comes to thermal weed control by flaming, where the temperature difference between flames and air can be as high as 15008C, this 1% will result in a 158C error in measured leaf temperature. One way to further reduce the conduction error is to prevent flames from entering the side of the leaves where the thermal sensors are located. The experimental design would then be similar to that used in this study, where flaming treatment was only directed to one side of leaves to block out direct thermal radiation from flames. However, gluing even a single thermocouples on a leaf would change its thermal properties, and to install thermocouples with 2 mm intervals to measure the spatial variation in temperature is not reasonable. 6. Conclusions The temporal and spatial variations in temperature of leaves during flaming are possible to measure with good accuracy by using an imaging infrared radiometer and an experimental setup described in this paper. In this study, an accuracy of 218C in absolute temperature values was achieved, and for temperature differences the error can be reduced further to 108C for a temperature range of 20–908C. The temperature of the leaves rose rapidly during flaming, typically around 1808C s1. To record the temperature changes successfully a sample rate of 25 Hz or higher was needed. The radiometer with an imaging speed of 50 fields s1 had no problems in recording the temporal variation in temperature. The high spatial resolution of the imaging infrared radiometer, 17 mm by 17 mm expressed as the size of the resolution unit, was essential to reveal the large spatial variation in temperature of leaves during flaming. The steepest temperature gradients measured reached 508C mm1. In conclusion, plant leaf temperature during thermal weed control treatment can be measured with success by using an imaging infrared radiometer. The overall accuracy of the infrared radiometer is at least as good as obtainable with other temperature measurement methods, and both the sample rate and spatial resolution of high-grade radiometers are good enough to characterise the temperature patterns in leaves. ARTICLE IN PRESS 266 J. RAHKONEN; H. JOKELA References Albrecht E (1985). Umweltneutrale Massnahmen zur Besaitigung von unerwunschtem Bewuchs auf befestigten Fl.achen. [Environmentally neutral measurements to remove undesired vegetation from paved areas.] PhD Thesis, Technical University of Berlin, Berlin, Germany Amiro B D; Thurtell G W; Gillespie T J (1983). A small infrared thermometer for measuring leaf temperatures in leaf chambers. Journal of Experimental Botany, 34, 1569–1676 Ascard J (1995). Thermal weed control by flaming: biological and technical aspects. PhD Thesis, Department of Agricultural Engineering, Swedish University of Agricultural Sciences, Alnarp, Sweden Beadle C L; Stevenson K R; Thurtell G W (1973). Leaf temperature measurement and control in a gas-exchange cuvette. Canadian Journal of Plant Science, 53, 407–412 Berliner B; Oosterhuis D M; Green G C (1984). Evaluation of the infrared thermometer as a crop stress detector. Agricultural and Forest Meteorology, 31(3/4), 219–230 Bertram A; Meyer J (1996). Verfahrentechnische Optimierung der thermischen Unkrautbek.ampfung. [Optimization of thermal weed control.] Zeitschrift fur Planzenkrankheiten und Pflanzenschutz, Sonderheft XV. pp 407–416 Buettner K J K; Kern C D (1965). The determination of infrared emissivities of terrestrial surfaces. Journal of Geophysical Research, 70(6), 1329–1337 Cook G D; Dixon J R; Leopold A C (1964). Transpiration: its effect on plant leaf temperature. Science, 144, 546–547 Diprose, M F; Benson F A; Willis A J (1984). The effect of externally applied electrostatic fields, microwave radiation and electric currents on plants and other organisms, with special reference to weed control. The Botanical Review, 50, 171–223 Elwidge C G (1988). Thermal infrared reflectance of dry plant materials: 25–200 mm. Remote Sensing of Environment, 26, 265–285 Fuchs M; Tanner C B (1966). Infrared thermometry of vegetation. Agronomy Journal, 58, 597–601 Gates D M (1964). Leaf temperature and transpiration. Agronomy Journal, 56, 273–277 Gates D M (1980). Biophysical Ecology. Springer-Verlag, New York Gates D M; Tantraporn W (1952). The reflectivity of deciduous trees and herbaceous plants in the infrared to 25 microns. Science, 115, 613–616 Graham M E D (1989). Calibration of a small infrared sensor for measuring leaf temperature in the field: non-steady state conditions. Agricultural and Forest Meteorology, 44, 295–305 Hashimoto Y (1990). Leaf temperature based on image processing. In: Measurement Techniques in Plant Science (Hashimoto Y; Kramer P J; Nonami H; Strain B R, eds), Academic Press, New York, pp 373–386. Hashimoto Y; Ino T; Kramer P J; Naylor A W; Strain B R (1984). Dynamic analysis of water stress of sunflower leaves by means of thermal image processing system. Plant Physiology, 76, 266–269 Hipps L E (1989). The infrared emissivities of soil and Artemisia tridentata and subsequent temperature corrections in a shrub–steppe ecosystem. Remote Sensing of Environment, 27, 337–342 Idso S B; Jackson R D (1968). Significance of fluctuations in sky radiant emittance for infrared thermometry. Agronomy Journal, 60, 388–392 Inframetrics, Inc. (1991). Model 760 IR imaging radiometer operator’s manual Jones H G (1999). Use of thermography for quantitative studies of spatial and temporal variation of stomatal conductance over leaf surfaces. Plant, Cell & Environment, 22(9), 1043–1055 Pieters G A; Schurer K (1973). Leaf temperature measurements I: thermocouples. Acta Botanica Neerlandica, 22(5), 569–580 Pinkley L W; Sethna P P; Williams D (1977). Optical constants of water in the infrared: influence of the temperature. Journal for Optical Society of America, 67(4), 494–499 Porterfield J G; Batchelder D G; Bashford L; McLaughlin G (1971). Two stage thermal defoliation. Proceedings of the Eighth Annual Symposium on Thermal Agriculture, Dallas, Texas, pp 32–34 Porterfield J G; Batchelder D G; Budo D P (1967). Thermal defoliation progress. Proceedings of the Fourth Annual Symposium on Thermal Agriculture, Kansas City, Missouri, pp 70–72 Reifschneider D; Nunn R C (1965). Infrared cotton defoliation or desiccation. Proceedings of the Second Annual Symposium, Use of Flame in Agriculture, St. Louis, Missouri, pp 25–29 Robinson P J; Davies J A (1972). Laboratory determinations of water surface emissivity. Journal of Applied Meteorology, 11, 1391–1393 Salisbury J W (1986). Preliminary measurements of leaf spectral reflectance in the 8–14 mm region. International Journal of Remote Sensing, 7(12), 1879–1886 Salisbury J W; Milton N M (1988). Thermal infrared (25 to 135 mm) directional hemispherical reflectance of leaves. Photogrammetric Engineering and Remote Sensing, 54(9), 1301–1304 Sutherland R A; Bartholic J F (1979). Emissivity correction for interpreting thermal radiation from a terrestrial surface. Journal of Applied Meteorology, 18, 1165–1171 Svendsen H; Jensen H E; Jensen S E; Mogensen V O (1990). The effect of clear sky radiation on crop surface temperature determined by thermal thermometry. Agricultural and Forest Meteorology, 50(3), 293–243 Tanner C B (1963). Plant temperatures. Agronomy Journal Notes, 55, 210–211 Tarnopolsky M; Seginer I (1999). Leaf temperature error from heat conduction along thermocouple wires. Agricultural and Forest Meteorology, 93(3), 185–194 Thomas C H (1967). The analysis of temperature distributions in living plants following a short exposure to flame. Proceedings of the Fourth Annual symposium on Thermal Agriculture, Kansas City, Missouri, pp 56–57 van de Griend A A; Owe M; Groen M; Stoll M P (1991). Measurement and spatial variation of thermal infrared surface emissivity in a Savanna environment. Water Resources Research 27(3), 371–379 Willey R R (1985). An instrument to measure spectral emittance from 2 to 20 micrometers. SPIE Infrared Technology and Applications, 590, 248–252 Zhang Y-W; Zhang C-G; Klemas W (1986). Quantitative measurements of ambient radiation, emissivity, and truth temperature of a greybody: methods and experimental results. Applied Optics, 28(20), 4482–4486
© Copyright 2026 Paperzz