GE Water & Process Technologies Analytical Instruments Technical Bulletin The Sievers* Inorganic Carbon Remover (ICR) In measuring Total Organic Carbon (TOC), all analyzers must deal with the presence of Inorganic Carbon (IC). IC is defined to be carbon from CO2, HCO3-, and CO3-2. Sources of IC include dissolved limestone and absorption of CO2 from air. Almost all water samples contain both organic and inorganic carbon. Combined, these components are commonly referred to as Total Carbon (TC). Total Carbon (TC) = Organic Carbon (TOC) + Inorganic Carbon (IC) Because it is not analytically possible to measure only the organic carbon when inorganic carbon is also present, most TOC analyzers work by measuring the TC content of the sample, the IC content of the sample, and then calculate the difference as TOC. Total Carbon (TC) - Inorganic Carbon (IC) = Organic Carbon (TOC) Measured Measured Calculated Alternatively, a TOC analyzer may operate by sparging out the inorganic carbon, and the subsequent carbon measurement is assumed to be 100% free of inorganic carbon. In this case, Total carbon is measured and is reported as the TOC of the sample. TC = TOC Some TOC analyzers can both measure and remove IC, giving the operator flexibility depending on the level of IC in the sample. When IC is relatively small compared to the TOC level of a sample, accurate measurements can be obtained without removing IC; it can simply be measured and subtracted from the TC value. Figure 1 imagination at work However, when IC is very high and TOC is low (e.g., IC = 10x TOC), TOC measurements can become unstable if IC is not removed or reduced. In the example where an instrument is measuring both TC and IC to calculate TOC, both the TC and IC values will be very large (since IC is a component of TC), and the stated instrument error in measuring TC and IC can be a significant percentage of the final calculated TOC value. In this case, improved performance will be realized by first eliminating or reducing the IC prior to conducting the analysis. Take for example a sample with 100 ppb of TOC and 1900 ppb of IC. We will assume that the instrument measures TC and IC with 2% accuracy. In one case IC is not removed, in the other it is reduced to 100 ppb (Figure 1). In cases where IC is high, and TOC is low, removal or reduction of IC will result in improved analytical performance. In general, for Sievers* TOC analyzers, we recommend reduction/removal of IC when it becomes 10 times higher than the expected TOC value. The 900 Series TOC Analyzers automatically detect this situation, and can inform the user when IC removal is recommended. IC Removal and Reduction Methods Some TOC analyzers remove the IC by sparging the sample with gas, and what remains is organic carbon, which is then measured. While sparging can be an effective means of IC removal, there are several concerns: 1. Purity of the sparge gas (so as to avoid contamination of the sample with organics) 2. Loss of volatile organics 3. If IC removal is not 100%, the remaining IC may be reported as TOC, introducing a bias into the analytical system. 4. Sparge gases increase the cost of ownership 5. Increases sample preparation/analysis time. In the proposed EPA TOC Method 415.3, “Determination of Total Organic Carbon and Specific UV Absorbance at 254 nm in Source Water and Drinking Water,” the USEPA specifies a sparging time of 20 minutes, with 100-200ml/min of gas flow. The Sievers technology utilizes a gas-free ICR (Inorganic Carbon Remover) to reduce IC content. This method is patented and approved for use for USEPA compliance monitoring. ICR Theory of Operation IC is removed by first acidifying the sample to force all IC into the form of CO2. IC exists in ionic and non-ionic forms; the ionic forms are carbonate and bicarbonate, and the non-ionic form is CO2. The ratio of ionic to non-ionic forms is pH-dependent. By acidifying the sample, all IC is forced into the form of CO2, which can then be degassed. CO2 Low → ← HCO3pH → ← CO3-2 High Acidification is accomplished automatically by the Analyzer which detects the presence of the Inorganic Carbon Remover unit, and using the same acid that is used for normal TOC analysis, so no additional reagents are necessary. The acidified sample is then passed through a small membrane module in the ICR; a vacuum is created on the outside of the membrane, and the inorganic carbon (CO2) is pulled out of the sample. Air that is passed across the membrane is first 'scrubbed' by a builtin chemical trap that removes any organics from the air and ensures no sample contamination. The IC removal process is approximately 95-99% efficient. One hundred percent IC removal is not required since Sievers TOC analyzers still measure any remaining IC and subtract that value from the TC value to calculate TOC. Because the IC level is greatly reduced, improved analytical performance is realized. There are several advantages to this method of IC reduction/removal. 1. No sparge gases are necessary, keeping costs low and IC removal simple 2. The sample is degassed inline with the analysis sample stream, so there is no additional time required for IC reduction / removal. 3. Volatile Organic Carbon is not lost during the process; this can be important as lost VOC from influent samples may result in lower TOC removal efficiency calculations between influent and effluent samples. 4. Because the process is automated by the analyzer, no operator action is necessary. The ICR can be easily bypassed via its switching valve, allowing a fast and easy transition to normal monitoring modes when IC removal is neither required nor desired. Application Recommendations An ICR should be considered when IC levels exceed 10x TOC. This can be common in raw surface and ground waters. In some cases, finished drinking water samples may also benefit from IC reduction / removal. For online, continuously monitoring applications, the ICR should be turned on and remain on for all samples. The ICR is housed within the Sievers 900 On-line TOC Analyzer's case for maximum environmental protection; status lights and the bypass valve are on the outside of the 900's case. For laboratory monitoring with an autosampler, a standalone ICR is used, and can be controlled by the autosampler software to be utilized only on those samples where IC removal is recommended or necessary. * Trademark of General Electric Company; may be registered in one or more countries. For more information, visit www.GEinstruments.com. Find a sales partner near you through the “Contact Us” Section. USA GE Analytical Instruments 6060 Spine Road Boulder, CO 80301-3687 USA T +1 800 255 6964 T +1 303 444 2009 F +1 303 444 9543 [email protected] www.geinstruments.com Europe Unit 3 Mercury Way Mercury Park Estate Trafford Park Manchester, UK M41 7LY T +44 (0) 161 866 9337 F +44 (0) 161 866 9630 [email protected] ©2005, General Electric Company. All rights reserved. 300 00109 Rev A MC05-194
© Copyright 2026 Paperzz