INSTRUCTION OF CARBON DIOXIDE (CO2) Carbon Dioxide (CO2) ASSAY KIT BY ENZYMATIC METHOD [ PRODUCT NAME ] Carbon Dioxide (CO2) Assay Kit by Enzymatic Method [ Packing specifications ] Basing on the bottle type, product model can be classified into 7170, 7060, 7020, Beckman, Toshiba, Mindray, Dupont, Innova, Siemens, KHB, Abbott, General type, etc. R:60ml×4 R:20ml×4 R:30ml×1 R:20ml×6 R:90ml×3 R:1×170T R:50ml×4 R:2×170T R:100ml×1 R:4×170T R:60ml×3 R:6×170T R:40ml×4 R:1×510T R:60ml×1 R:3×510T R:60ml×6 R:4×510T R:20ml×1 R:6×510T R:20ml×2 [ Intended Use ] This product is used to determination CO2 in human serum. [ PRINCIPLE ] Catalyzed by phosphoenolpyruvate carboxylase (PEPC), HCO3− in serum reacts with phosphoenolpyruvate (PEP) to generate oxaloacetic acid and phosphoric acid. Then oxaloacetic acid becomes malic acid and NADH is oxidized into NAD + at the same time under the catalysis of malic dehydrogenase (MDH). At wavelength 405 nm, the declining rate of absorbance is proportional to the carbon dioxide concentration in sample. Carbon dioxide concentration in the sample can be calculated by comparing the calibrator result with same processing. PEPC+Mg2+ PEP+HCO3− → Phosphoric acid Oxaloacetic MDH Oxaloacetic acid +NADH+H+ → [ Composition ] R1:Detection Reagent 1 Tris buffer PEP MgSO4 NADH MDH PEPC Preservatives Calibrator & Control Calibrator:1.0ml×1 acid + malic acid +NAD+ 50.0mmol/L 0.8g/L 0.65g/L 0.4g/L 1KU/L 2 KU/L 1g/L Optional [ Storage And Stability ] Stored for up to 12 months at 2-8°C, protect from light. After opening, the reagent remains stable for 7 days at 2-8°C, protect from light. [ Applicable Instrument ] This assay kit is suitable for automatic or semi-automatic biochemical analyzer with 405nm wavelength. [ Sample Requirements ] Serum without hemolysis. Samples testing should be completed at the same day after collected. Otherwise, the samples should be cryopreserved and avoid repeated freeze-thaw cycles. CO2 in the samples remain stable for 7 days at 2-8°C and for 3 months at cryopreservation condition [ TEST METHOD ] 1. BASIC PARAMETERS: Method: two-point rate assay Temperature: 37℃ Primary wavelength: 405nm Secondary wavelength: none Sample: 3μl R:300μl Response direction: Negative Reaction time: 10 min Calibration mode: Two points calibration 2. Assay Procedure Blank(B) Sample(U) Calibrator(Ci) ddH2O(μl) 3 - - Sample(μl) - 3 - Calibrator(μl) - - 3 300 300 Reagent R(μl) 300 Mix well , 37 ℃ for 2 minutes. Measure the absorbance value A1 at 405 nm and measure the absorbance value A2 after 2 minutes. Calculate ΔA/min= (A2-A1)/2min ASSAY PROCEDURE SUMMARY: ←Detection→ Set blank well Time: 10 min Temperature:37℃ Primary wavelength 405nm Secondary wavelength none ↑ R:300μl Sample:3μl ↑ 3. CALCULATIONS Based on two points calibrator concentration and the corresponding absorbance change rate ΔA/min, the two point linear calibration model is used to define the working curve, sample absorbance change rate on the working curve with the corresponding concentration is concentration of sample. [ Reference Range ] 22.0 mmol/L-29.0 mmol/L Recommendations: Each laboratory should establish its own reference range. [ Explanation for the test result ] 1. The test results reflect only the status at the sampling time. Clinicians need to be combined with clinical data and other relevant test results to make judgment. 2. There is no significant effect on the test result when the sample contains Ascorbic acid≤0.5g/L; Bilirubin≤0.5g/L; Hemoglobin≤5.0g/L; Triglycerides≤20g/L. [ Test Method Limitations ] Dilute with physiological saline and multiply the result by the dilution factor when the concentration of CO2 is over 50.0mmol/L. [ PERFORMANCE ] 1. DETECTION RANGE : 2.0mmol/L-50.0mmol/L, r≥0.990. 2. PRECISION:Intra-assay CV≤8%, Inter-assay CV≤10% 3. ACCURACY:Inaccuracy≤10%。 4. Blank absorbance:Absorbance values ≥0.900 at 405nm wavelength and optical path 10mm. [ PRECAUTIONS ] 1. For in vitro diagnostic use only。 2. Reagents become cloudy or blank absorbance values< 0.9, should be discarded. 3. The reagent and sample amount can be changed with same ratio according to needs. 4. Reagents and components of different batches are not interchangeable. 5. The test equipment must be clean to avoid contamination. 6. The waste solution generated by the test and decomposition difficultly packaging materials should be collected and sent to local waste treatment station. [ Reference ] Yingwu Ye et al. National Guide to Clinical Laboratory Procedures (Third Edition). Southeast University Press. 2006: 379-381. [ MANUFACTURER ] Wuhan Life Origin Biotech. Co., Ltd. Wuhan Hi-tech Medical Devices Park, Building B11, #818 Gaoxin Road, Donghu Hi-Tech Development Area, Wuhan, Hubei Province 430206, P.R. China Tel:027-87926888 | Fax:027-87196150 Website: http://en.szybio.com Email: [email protected] [ Medical Devices manufacturer license number ] Hubei SFDA No. 20100488 [ Medical Devices Registration Certificate Number ] Hubei SFDA 2013 No. 2401572 [ Product standards] YZB / E 0919-2013 [Specification approval date ] January 06, 2014
© Copyright 2026 Paperzz