MEG II実験 液体キセノンガンマ線検出器に用いる 光検出器MPPCの 実装に向けた最終試験 家城 佳 他 MEG II collaboration + 九大の方々 Introduction 2 LXe detector upgrade upgrade! (CG) inner face 2inch PMT× 216 12x12mm2 MPPC× 4092 3 1m Granularity will improve γ energy resolution & position resolution expected to improve by a factor of 2! Requirements for the MPPC - sensitive to VUV scintillation light - large area (reduce the number of readout ch) MPPC for MEG Hamamatsu S10943-4186(X) Successfully developed a new type of MPPC! - PDE for VUV light > 20% - Large area fall time < 50ns with series connection - 50μm pitch pixel - Four chips - metal quench resister Performance tests done so far: - Small subsample (prototype & sample of final model) tests in LXe - Mass test of prototype MPPCs (600 pcs) in room temperature and in LXe This talk: ① Mass test of final MPPCs (4200 pcs) ② PDE incident angle dependence check 4 Mass test of final MPPCs 5 Mass test Production of MPPCs is ongoing in Hamamatsu. (Production will finish in the end of October. ~2 month delay due to the quality control issue.) 2920 out of 4200 MPPCs are delivered. 2450 are already tested. Setup for the mass test Temperature: 20 deg ~40 MPPCs (~160 chips) are measured at once. relays to change the readout chips We can test ~250 MPPCs / day I-V curve measurement We measure I-V curve of all chips to identify bad chips. We check: I ① Breakdown voltage (Vbd) compare to the data sheet ② Current at Vover = 5.0V (Imeas) ③Any strange shape in the I-V curve? chip0123 Vbd (from linear + quadratic fit) 5V chip0 1 3 2 Imeas (from spline fit) 7 Results Measured values were compared to the data sheet. Operation voltage from data sheet (gain=2.0x106 @ 25℃) Measured breakdown voltage Vhama – Vbd Measured current at V = Vbd+5.0V Current at Vhama from data sheet Imeas – Ihama 8 one outlier (next page) large current chips x ~10 Fraction of outlier chips is very small. Chips with strange I-V curves I-V curves for some of the chips were strange. I I Examples of bad chips strange bump 9 current offset 2 chips 8 chips Some of the strange or large current chips were checked in more detail. Visual check, gain, waveform … nothing strange. Some of them are sent back to Hamamatsu to do more checks. Anyways, the fraction of these suspicious chips is small. PDE (Photon Detection Efficiency) vs. Incident angle 10 In our previous measurement … In the mass test for prototype MPPCs (~600 pcs), we noticed that the PDE is different for the MPPCs placed at different position. PDE vs. incident angle in LXe α α sources 241Am θ MPPCs detected photons PDE = incident photons Angular dependence? (much larger than what we expect from attenuation or reflection.) Also, the absolute value is smaller than our previous measurement (PDE~0.25). 11 Additional measurements We plan to do two different type of tests. Large chamber Small chamber Liquid Xe gas Xe 12 α MPPC (4 chips) • Quick setup, only 4 chips • Confirm angular dependence ↑ Today’s topic α ~20 MPPCs on moving stage α 70cm • Movable stage, 20~30 chips. • Check the angular dependence and MPPC by MPPC variation of PDE Small chamber setup ● Use 1 MPPC (= 4 chips) 2 chips are placed at 30 deg, other 2 chips are placed at 50 deg. ● The effect of reflection from the wall is considered carefully. chips are set very close to α Direct light >> Reflected light 100 <10 α 6mm 27° chip by chip readout 49° ● Relative position of the α and MPPC must be measured precisely. 13 Alignment check We used 3D laser scan to measure the position of the sensors and α. (+ reconstruction from photo image) Scanned data FARO laser scanner 14 α z y x MPPC position accuracy: 50~70μm Basic properties check 1p.e. charge Basic properties such as gain, crosstalk and afterpulse are measured by using LEDs. Crosstalk+afterpulse prob. vs. Vover Gain vs. Vover 0.3 this difference comes from different amplifier that we used. 15 chip 0 1 2 3 0.2 4 6 8 Over voltage [V] Results are consistent with previous measurements with sample MPPC. (This is the first time to use the a final MPPC in LXe.) PDE at different angle PDE vs. Vover PDE reduction chip1,2 chip0,3 angle ~ 30deg angle ~ 50deg PDE reduction expected from prev. meas. 16 parameters for calculation: α energy = 4.78MeV W = 19.6eV PDE angular effect seems to be roughly consistent to what we expected. However, there is a large uncertainty in the correction factor for saturation of pixels. detailed measurement in large chamber Summary • Mass test of final MPPCs – Final MPPCs are being produced in Hamamatsu. – 2920 pcs are already delivered. 2450 are already tested. – Three types of bad chips were found (~20/9800 chips). They are being checked in detail. The fraction of bad chips is small. • PDE vs. incident angle – In our previous test, PDE was measured to be smaller at larger angle. – This effect was reproduced in the new measurement. – Detailed test will be done in large chamber. – Effect of this to MEG II is studied in simulation Next talk 17 Backup 18 MEG MEG II Liquid Xe γ-ray detector stopping rate x ~2 γ resolution & efficiency x 2 e+ μ+ e+ drift chamber & timing counter Gradient magnetic field Sensitivity will improve by factor 10 beam 19 More investigations for bad chips In total, we found ~20 bad chips out of 9800. For some of the bad chips, we did more detailed tests: Waveform check by oscilloscope, basic property measurement with LED etc. • Dark noise rate for the large current chips seemed to be high. • Waveform of each pulse looked OK. • Gain and Vbd were not different from the normal chips. • Nothing strange was found in a visual inspection with microscope. bad chip (large current) normal chip Some of the bad MPPCs are sent back to Hamamatsu for more investigation. 20 Transmittance factor (LXe-Si) 21 Small chamber Incident angle will be different for different pixels in the chips. How different are they? α Angle at center of the chips: 27 deg, 49 deg distributio n of angle 30 deg chips 22 50 deg chips This distribution must be taken into account when we consider this measurement with Kebab. Alignment check adhesive (dropped here Data point distribution by mistake) Position of the α source along the wire is reconstructed by comparing photo image and FARO scan data. Projection along x axis FARO scan data: adhesive, wire holder ring 23 compare Photo image: adhesive, wire holder ring, alpha source accuracy of the position measurement is 50~70 μm. adhesive 𝑦 𝑧 α source 𝑥 on wire 0.5mm
© Copyright 2026 Paperzz