Application Note: 30217 1000 Proteins per Hour [pph] Identified from an E. coli Digest Maximizing Protein Identification on the LTQ Orbitrap Velos Kai Scheffler, Thermo Fisher Scientific, Dreieich, Germany Eugen Damoc, Thomas Möhring, Thermo Fisher Scientific, Bremen, Germany Introduction Key Words •LTQ Orbitrap Velos •CID, ETD, HCD •DDDT, MS/MS • E. coli •Proteome Discoverer •Protein Identification Enormous improvements in mass spectrometry technology over the last few years have allowed for ever deepening analysis of complex proteomes. However, many low abundance proteins remain undetected in current large-scale proteomic analyses due to the limits on the rate at which MS/MS spectra can be acquired. To solve this undersampling problem, we developed the Thermo Scientific LTQ Orbitrap Velos, a new hybrid linear ion trap-Orbitrap mass spectrometer that provides faster MS/MS scanning in the ion trap (up to 10 peptide sequences per second), thereby increasing the depth of analysis of complex protein mixtures. This application note demonstrates and discusses the ability of the newly developed hybrid instrument to identify more proteins, with increased sequence coverage and confidence. Different strategies regarding chromatographic run times and utilizing the instruments various dissociation techniques are compared. Instrument Innovation The instrument evaluated in this application note is the Thermo Scientific LTQ Orbitrap Velos mass spectrometer equipped with Electron Transfer Dissociation (ETD). It is a hybrid instrument which includes a novel dual-pressure linear ion trap combined with an Orbitrap™ detector as shown in Figure 1. Significant technological improvements have been implemented in both parts of the hybrid instrument. Changes implemented in the front-end of the hybrid instrument, the Thermo Scientific LTQ Velos ion trap, include (a) a progressive stacked-ring ion guide (S-Lens) with 5-10x increased ion transmission;1 (b) a dual-pressure differentially pumped linear ion trap with a higher-pressure cell for improved ion trapping, isolation, and CID efficiencies, and a lower pressure cell for improved resolution and/or scan speed and (c) predictive automatic gain control (AGC) for increased scan speed.2 Electrospray Ion Source S-Lens Square Quadrupole Octopole Improvements in the rear portion of the instrument, the Orbitrap mass analyzer section, are: (d) an HCDcollision cell with axial field combined with the C-trap and (e) improved vacuum in the Orbitrap detector chamber for improved analysis of intact proteins.3 Experimental Sample Preparation The lyophilized soluble fraction of an E. coli whole cell lysate was solubilized in 50 mM ammonium bicarbonate buffer followed by reduction of disulfide bonds with dithiothreitol (DTT) at 56 °C for 1 hr, and alkylation of cysteines with iodoacetamide for 30 min at room temperature in the dark. The sample was then enzymatically digested for 14 hours with a Lys/Arg specific protease. The digest was split in aliquots and stored at -20 °C until use. Samples were diluted in 0.1% formic acid prior to analysis by reversed-phase nanoHPLC-MS/MS. LC-MS Analysis The proteolytic digest of E. coli was separated by online reversed-phase chromatography for each run using a Thermo Scientific Surveyor MS Pump Plus and Micro AS autosampler with a reversed-phase peptide trap (100 μm inner diameter, 2 cm length) and a reversed-phase analytical column (75 μm inner diameter, 10 cm length, 3 μm particle size, both NanoSeparations, NL), at a flow rate of 250 nl/min. The chromatography system was coupled on-line with an LTQ Orbitrap Velos mass spectrometer equipped with an ETD source. High Pressure Cell Low Pressure Cell Quadrupole Mass Filter C-Trap HCD Collision Cell Transfer Multipole Reagent Ion Source e- Orbitrap Mass Analyzer Reagent 1 Heated Inlet Figure 1: Schematic of the Thermo Scientific LTQ Orbitrap Velos mass spectrometer with ETD. Reagent 2 Heated Inlet Instrument Configuration Pump Thermo Scientific Surveyor MS Pump Plus Autosampler Thermo Scientific Micro AS Mass spectrometer Thermo Scientific LTQ Orbitrap Velos with ETD Ionization source Nanospray I Trapping column 20 mm in length, 100 µm ID Analytical column 100 mm x 75 µm ID 3 µm particle size, 300 Å pore size Flow rate 250 nl/min at 50% B Gradient 2-30% acetonitrile in 0.1% formic acid (either 60 or 90 minutes) Table 1: Instrument configuration. Mass Spectrometer Settings Sourcenano-ESI Capillary temperature 250 °C S-Lens RF level 50% Source voltage [kV] 1.5 Max. injection times Full MS 500ms MSn 50ms (CID, ETD) MSn 200ms (HCD) Full MS mass range 380 - 2000 [m/z] Details for the chromatographic and mass spectrometric settings are listed in Table 1 and 2. Dilutions of the injected sample amounts were performed to evaluate the effects of decreasing sample quantity in regards to sensitivity of the instrument as judged upon the total number of proteins identified as well as their sequence coverage. Data-dependent tandem MS acquisition methods were used for all experiments. Different fragmentation techniques, namely CID, ETD and HCD and combinations thereof by using the data dependent decision tree (DDDT) acquisition method were compared.4 All runs were performed at least in duplicates. In all experiments the Full MS scans were acquired over a mass range of 380-2000 m/z with detection in the Orbitrap mass analyzer at a resolution setting of 30,000. MS/MS spectra were acquired in the ion trap except for fragment ions produced using higher-energy collision-induced dissociation (HCD) taking place in the collision cell positioned in axial position next to the C-trap. Fragment ion spectra produced via HCD were acquired in the Orbitrap mass analyzer with a resolution setting of 7,500. For data-dependent acquisition, the method was set to analyze the top twenty most intense ions when fragmentation was performed using CID (Table 3). Resolution settings Full MS 30.000 MSn (HCD) 7.500 Lock mass (all runs) 445.120024 Table 2: Mass spectrometer parameter settings. Fragmentation Type CID ETD HCD DDDT Isolation width [u]2222 Collision energy30n.a.4030 (CID) Activation time [ms]101000.110 (CID) / 100 (ETD) Supplemental activationn.a.yesn.a.yes (ETD only) Pred. AGC enabledyesyesyesyes No. of microscans for Full MS2212 Target value Full MS (FT)1.00E+061.00E+061.00E+061.00E+06 Detector for MSn spectraIon TrapIon TrapOrbitrap Ion Trap Target value MSn5.00E+035.00E+035.00E+045.00E+03 Top n MS/MS, with n =20101010 Charge state dep. ETD time enabledn.a.yesn.a.yes Dynamic exclusion enabledyesyesyesyes General Settings Dynamic exclusion settings repeat count 1 exclusion list size 500 exclusion duration 70 s early expiration disabled exclusion mass width low/high 10 ppm Charge state screening on: +1 and unassigned charge states rejected Monoisotopic precursor selection enabled Table 3: Method parameter settings. 2 The top ten most intense ions per scan cycle were fragmented in case of ETD spectra and also when the DDDT method was applied. Parameters for the CID/ETD data dependent decision tree were set to perform ETD when the precursor charge state was z=3 and m/z of the precursor < 650 u, with charge state z=4 and m/z < 900 u and charge state z=5 with m/z < 900 u. In all other cases peptide fragmentation was triggered with CID. For data-dependent acquisition of HCD spectra, the top ten most intense ions were selected for fragmentation in each scan cycle and Full MS as well as fragment ion spectra were detected in the Orbitrap mass analyzer. Exclusion conditions were optimized according to observed chromatographic peak width. Due to the greater ion transmission of the ion source, a lower injection time was selected for all MS/MS spectra detected in the ion trap: 50 ms compared to 100-200 ms on previous instruments. The activation time for resonance CID fragmentation was 10 ms compared to 30 ms used on the Thermo Scientific LTQ XL linear ion trap, taking advantage of the increased fragmentation efficiency in the higher-pressure cell of the ion trap. For HCD experiments a lower target value of 5e4 ions was used compared to 2e5 on previous instruments, taking the advantage of better transmission of fragment ions from the axial field collision cell. Database Search The raw data files were searched using Thermo Scientific Proteome Discoverer software v. 1.2 with Mascot™ v. 2.2.1 search engine (Matrix Sciences Ltd., London, UK). The search parameters applied in the database searches are listed in Table 4. The reverse database search option was enabled in Mascot, and all data were filtered to satisfy a false discovery rate (FDR) of 1% or better. All datasets were processed using the workflow feature in Proteome Discoverer™ software. The workflow used to process datasets containing ETD spectra included the node called non-fragment ion filter which aims at the removal of intact precursors and its charged reduced species since they are very often the most abundant peaks in the fragment ion spectra und thus compromise the score. Protein Annotation All identified proteins from the duplicate runs from CID, ETD, HCD and DDDT methods were exported from the Proteome Discoverer software as ProtXML format and submitted to the Thermo Scientific ProteinCenter data interpretation software to retrieve relevant biological information from publicly available protein databases including gene ontology references for function, biological process and subcellular location. Software Platform Search engine Proteome Discoverer Mascot v 2.2.1 DatabaseSwissProt v 56.0 TaxonomyEscherichia coli Maximum number of missed cleavages1 Precursor mass tolerance 5 ppm Fragment mass tolerance 0.8 Da (CID, ETD), 20 mmu (HCD) Protein massunrestricted Reverse database search enabled Target false discovery rate 0.01 strict/0.05 relaxed Instrument typeDefault Modifications static: carbamidomethyl cysteines (C) Scoring Peptide cut off score: 10 MUDPIT scoring Protein relevance threshold: 20 Protein relevance factor: 1 DTA Generation conditions Total intensity threshold: 100 Minimum peak count: 8 S/N threshold: 2 Precursor mass range: 760 - 10.000 Da Non-fragment filter (for ETD spectra only) Mass window offset for precursor peak: 4 Da Mass window offset for charge reduced precursors: 2 Da Mass window offset for neutral losses from charge reduced precursors: 2 Da Remove only known masses: false Maximum neutral loss mass: 120 Da Result Filters Confidencehigh Min. number of peptides per protein 0/2 FDRfixed to 1% Table 4: Data analysis parameter settings. Results and Discussion The LTQ Orbitrap Velos mass spectrometer identifies more than 1,000 proteins per hour [pph] Figure 2 illustrates the base peak chromatograms from analyses of 1,000 ng, 100 ng and 20 ng of a proteolytic digest of E. coli whole cell lysate, separated over 60 min via reversed-phase chromatography on the LTQ Orbitrap Velos mass spectrometer with a split flow from a Surveyor MS pump. The peak intensities in the chromatograms correlate very well with the sample amount analyzed and also demonstrate the moderate complexity of the sample compared to more complex lysates from other organisms such as C. elegans2 or human HeLa cells.3 3 514.30 Relative Abundance 100 442.95 80 652.40 421.76 562.28 449.74 607.82 701.37 519.93 15 20 25 30 35 849.75 806.45 45 50 Time (min) 55 718.38 544.29 60 65 70 75 601.97 442.95 688.82 100 ng 586.33 423.78 609.79 594.26 40 652.40 562.28 463.91 449.74 20 15 607.82 538.79 20 25 30 35 40 637.61 688.82 609.79 20 806.45 60 20 ng 748.43 423.78 701.37 25 30 35 75 652.40 607.82 609.69 543.81 735.90 803.45 624.37 20 70 NL: 8.80E6 Base Peak F:ms MS 445.12 15 65 594.32 443.20 562.28 55 483.02 594.26 60 449.74 823.91 706.73 586.33 601.97 80 859.11 983.48 45 50 Time (min) 514.30 100 40 NL: 3.26E7 Base Peak F:ms MS 594.32 514.30 60 0 609.35 748.43 595.04 80 0 40 982.02 983.48 613.32 20 100 Relative Abundance 1 µg 859.11 40 NL: 2.89E8 Base Peak F:ms MS 706.73 60 0 Relative Abundance 601.97 688.82 706.73 594.32 586.33 423.78 40 45 50 Time (min) 55 723.40 60 445.12 65 445.12 445.12 70 75 Figure 2: Base peak chromatograms of 1 µg (top trace), 100 ng (middle trace) and 20 ng (bottom trace) of E. coli proteolytic digest loaded on a 75 µm C18 reversed phase column and separated via a 60 min gradient (2-30% ACN). For a typical and reproducible 60-minute separation the LTQ Orbitrap Velos mass spectrometer is able to identify over 1,000 proteins from 1 µg of a proteolytic digest of E. coli whole cell lysate using a ddTOP20 CID method with acquisition of the full scan in the Orbitrap detector and the fragment ion spectra in the ion trap. Reducing the sample amount by a factor of 10 to a final 100 ng applied on column, the number of identified proteins is only reduced by about 10%. Reducing the sample amount again by a factor of 5 to 20 ng, the number of identified proteins is again reduced by about 10%, resulting in 844 proteins on average as depicted in Figure 3a. (a) 1200 1046 1054 1000 800 Proteins Proteins with ≥ 2 Peptides 942 789 859 829 809 683 600 947 689 577 567 400 200 0 4 1000 100 Sample Amount Loaded [ng] 20 The total number of proteins identified in the duplicate runs for the three sample concentrations are very close and the overlap of the identified proteins was 90% across all concentrations indicating high reproducibility. Identification rate on the peptide levels demonstrates a success rate of identified peptides relative to the total number of CID spectra acquired of ~53% for all concentrations. The percentage of unique peptides identified versus all identified peptides varies between 70% (1 µg load) and 90% (20 ng load) as displayed in Figure 3b. (b) 10000 9000 8000 7000 6000 5000 4000 3000 2000 1000 0 All Peptides Unique Peptides 9833 9257 6424 6740 6504 5356 6432 5322 4282 4165 3736 1000 100 Sample Amount Loaded [ng] 3793 20 Figure 3: Identification of unique proteins (a) and identification of unique peptides (b) from 1 µg/100 ng/20 ng of E. coli proteolytic digest using the 60 min ddTOP20 CID method. For a discussion on: “False discovery rates of protein identification: A strike against the two-peptide rule.” see7. The LTQ Orbitrap Velos offers great acquisition speed The high numbers of identified peptides and proteins using the LTQ Orbitrap Velos system is strongly based on the high acquisition speed that the instrument offers. Figure 4 highlights the speed at which spectra are acquired. Every vertical line in the diagrams represents a scan resp. a spectrum in the chromatogram. The green lines represent the Full MS scans, whereas the orange lines represent the MS/MS spectra. In the ddTOP20 CID method (Figure 4, top) the time elapsed for the acquisition of a Full MS spectrum is 1.13 sec due to the acquisition with 2 µscans and a resolution setting of 30k. The time elapsed to acquire 20 CID MS/MS spectra is 2.36 sec resulting in 3.49 sec for a full scan cycle comprising of 1 Full MS spectrum (Orbitrap detector) and 20 MS/MS spectra (ion trap detector). RT: 29.488 - 29.600 0.10 0.08 0.06 0.04 0.02 0.00 Relative Abundance 29.507 0.10 0.08 0.06 0.04 0.02 0.00 1.13 sec 29.50 29.51 29.52 29.53 29.54 29.55 Time (min) 45.597 1.0 0.5 29.56 29.585 45.591 Full MS/MS trace TOP20 CID 29.57 29.59 29.60 Full MS trace 30k, 1 µscan 2.29 sec 45.657 45.607 45.613 45.620 45.628 45.635 45.609 45.616 45.624 45.632 45.640 1.0 Full MS/MS trace TOP10 HCD, 7.5k 0.5 0.0 45.57 29.58 45.645 HCD 0.64 sec Relative Abundance Full MS trace 30k, 2 µscans 2.36 sec 29.526 29.563 29.561 29.528 29.559 29.529 29.556 29.531 29.554 29.533 29.552 29.535 29.550 29.537 29.549 29.539 29.547 29.541 29.545 29.542 RT: 45.570 - 45.680 1.5 0.0 1.5 29.565 CID 29.504 29.49 In higher energy collision induced dissociation (HCD), a dissociation technique that has proven to be especially useful for de novo sequencing as well as for quantitation using chemical labeling, all scans are detected in the Orbitrap mass analyzer at high mass accuracy and a high resolution setting of choice between 7,500 and 100,000.5,6 The cycle time of a ddTOP10 HCD (Figure 4, bottom) is 2.93 sec, comprising of 0.64 sec for the Full MS acquired with 1 µscan and 2.29 sec for 10 MS/MS HCD spectra recorded at a resolution setting of 7,500. These two examples show the higher speed at which CID and HCD spectra can be acquired compared to the LTQ Orbitrap XL mass spectrometer, which is due to the implemented hardware improvements.3 45.58 45.59 45.60 45.61 45.62 45.63 Time (min) 45.64 45.65 45.66 45.67 45.68 Figure 4: Demonstration of scan speed using a ddTOP20 CID method (top) compared to a ddTOP10 HCD method (bottom). With CID 1.13 sec are spent on the acquisition of the Full MS scan followed by 2.36 sec for 20 MS/MS spectra acquired in the ion trap. With HCD 0.64 sec are spent on the Full MS scan followed by 2.29 sec on 10 HCD MS/MS spectra acquired at a resolution of 7,500. omparing different fragmentation techniques for maxiC mizing protein ID The LTQ Orbitrap system offers various different fragmentation techniques respectively methods. Whereas CID, PQD (pulsed Q Dissociation) , ETD and HCD are fragmentation techniques, DDDT is an instrument method. The choice which of the available fragmentation methods is most suitable for the analysis very much depends on the complexity and the amount of sample available for repetitive runs, the protease used for digestion, and last but not least the aim of the analysis such as purely protein ID or identification combined with quantitation, de novo sequencing or the analysis of post-translation modifications. In this current study the aim was to identify the maximum number of proteins using a standard digest with a K,Rspecific endoprotease. Figure 5 summarizes the results regarding peptide and protein identification resulting from four of the possible fragmentation modes run as ddTOP20 CID, ddTOP10 DDDT, ddTOP10 ETD and ddTOP10 HCD. All methods were run in duplicate runs with 1 µg and 100 ng sample amount each with a separation of the peptide mixture over the course of 90 min. The numbers presented in the diagram are the average of the duplicate runs. All methods applied show a very high number of unique peptides as well as unique proteins identified with CID leading the field with an additional ~5% of proteins identified compared to the DDDT method. Overlap of proteins identified with all four methods is about 80% (Figure 6). 5 (b) Unique Peptides 8000 (a) Proteins 7000 1200 6000 1000 800 600 5000 1157 995 1095 957 1017 806 914 832 7601 5858 7009 5603 6344 4759 5577 4595 1 µg 100 ng 1 µg 100 ng 1 µg 100 ng 1 µg 100 ng CID DDDT HCD ETD 3000 400 2000 200 0 4000 1000 1 µg 100 ng 1 µg 100 ng 1 µg 100 ng 1 µg 100 ng CID DDDT HCD ETD 0 Figure 5: Identification of unique proteins (a) and identification of unique peptides (b) from 1µg and 100 ng E. coli digest using 90 min LC separation and different methods: ddTOP20 CID, ddTOP10 DDDT, ddTOP10 HCD and ddTOP10 ETD (a) Total of 8825 Unique Peptide IDs at < 1% FDR and < 5ppm (a) Total of 1268 Protein IDs CID 158 CID/HCD 134 HCD 520 HCD 39 CID/HCD/ETD 851 CID/HCD/ETD 4397 HCD/ETD 348 HCD/ETD 14 CID/ETD 26 CID/HCD 1378 CID 1338 CID/ETD 523 ETD 46 (b) Total of 1306 Protein IDs (b) Total of 9069 Unique Peptide IDs at < 1% FDR and < 5ppm CID/DDDT 169 CID/DDDT 1431 CID 123 DDDT 77 CID 1285 CID/ETD/DDDT 857 CID/ETD 21 ETD 321 CID/ETD/DDDT 4493 ETD/DDDT 17 ETD 42 CID/ETD 427 DDDT 764 ETD/DDDT 365 ETD 304 Figure 6: Venn diagrams summarizing the results from three 90 min nanoLC-MS/MS runs of 1 µg E. coli whole cell lysate digest using three different fragmentation methods. (a) CID, ETD and HCD and (b) CID, ETD and DDDT. All numbers displayed are from single runs. The method using HCD for fragmentation also provided more than 1,000 proteins for 1 µg of sample, demonstrating a much improved success rate of identifying peptides and proteins compared with the LTQ Orbitrap XL mass spectrometer, which is also due to an improved HCD cell resulting in a higher spectral quality. 6 The number of protein IDs with ETD is also very high yet lower compared to all other methods applied. This was as expected considering the nature of the sample consisting mainly of smaller peptides resulting in predominantly doubly charged peptides which provide better fragment ion spectra quality under CID and HCD conditions. The location and biological function of all proteins identified from the proteolytic digest of E. coli whole cell lysate in this study is detailed in Figure 7. (a) Molecular Functions Molecular Functions Antioxidant activity Catalytic activity DNA binding Enzyme regulator activity Metal ion binding Motor activity 9 1021 211 26 382 1 Nucleotide binding 394 Protein binding 330 Receptor activity RNA binding Signal transducer activity Structural molecule activity Transcription regulator activity Translation regulator activity 16 132 44 72 121 19 Transporter activity 175 Unannotated 161 (b) Cellular Component Cellular Component Cell surface Chromosome Cytoplasm Cytoskeletone Cytosol Endoplasmic reticulum 1 13 657 7 53 6 Extracellular 43 Membrane 312 Mitochondrion23 Nucleus Organelle lumen Proteasome Ribosome Unannotated 32 3 1 67 558 (c) Biological Processes Cell communication Biological Processes Cell death Cell differentiation 7 8 Cell division 42 Cell growth 2 Cell motility 2 Cell organization and biogenesis Cell proliferation Cellular homeostasis 159 3 33 Coagulation 5 Conjugation 4 Defense response 6 Development Metabolic process Regulation of biological process Reproduction Figure 7: Gene Ontology annotation of all proteins identified from E. coli digest by CID, ETD, HCD, and DDDT: (a) molecular functions, (b) cellular component, (c) biological processes. 87 1 1131 225 20 Response to stimulus 178 Transport 254 Unannotated 205 7 Conclusion Acknowledgements The identification of proteins from complex mixtures requires high sensitivity as well as high acquisition speed. The technological advancements implemented in the LTQ Orbitrap Velos mass spectrometer with more than twice the scan speed of an LTQ Orbitrap XL mass spectrometer, help to overcome undersampling, increasing the depth of the analysis of complex proteomes. The enhanced sensitivity and improved duty cycle of the new instrument enables a dramatic increase of the number of unique peptide IDs, which translates into an increased number of proteins identified with higher sequence coverage and confidence. The LTQ Orbitrap Velos mass spectrometer We would like to thank Christian Ravnsborg for providing the results from the ProteinCenter™ data interpretation software and Eduard Denisov for his support and helpful discussions. •provides an increased scan speed in MS and MS/MS mode due to its higher transmission ion source, dual pressure ion trap and improved HCD cell. •identifies more than 1,000 proteins per hour from a proteolytic digest of E. coli whole cell lysate. •provides a higher HCD scan speed and much improved spectral quality opening up new perspectives for de novo sequencing, quantitation of chemically labeled samples, and analysis of modifications which require high mass accuracy. 1 Wouters, E.R., Splendore, M., Senko M.W., Syka, J.E.P., Dunyach, J.-J., Design of a progressively stacked ring ion guide for improved ion transmission at high pressure, ASMS 2008. 2 Pekar Second, T., Blethrow, J.D., Schwartz, J.C., Jennifer E. Merrihew, J.E., MacCoss, M.J., Swaney, D.L., Russell, J.D., Coon, J.J., and Zabrouskov, V., Dual-Pressure Linear Ion Trap Mass Spectrometer Improving the Analysis of Complex Protein Mixtures, Anal. Chem. 2009, 81, 7757– 7765. 3 Olsen J.V., Schwartz J.C., Griep-Raming J., Nielsen M.L., Damoc E., Denisov E., Lange O., Remes P., Taylor D., Splendore M., Wouters E.R., Senko M., Makarov A., Mann M., Horning S., A dual pressure linear ion trap - Orbitrap instrument with very high sequencing speed. Mol Cell Proteomics. 2009, 8, 2759-69. 4 Swaney D.L., McAlister G.C., Coon J.J., Decision tree-driven tandem mass spectrometry for shotgun proteomics. Nat Methods. 2008, 5, 959-64. 5 McAlister G.C., Phanstiel D., Wenger C.D., Lee M.V., Coon J.J., Analysis of tandem mass spectra by FTMS for improved large-scale proteomics with superior protein quantification. Anal Chem. 2010, 82, 316-22. 6 Köcher T., Pichler P., Schutzbier M., Stingl C., Kaul A., Teucher N., Hasenfuss G., Penninger J.M., and Mechtler K., High precision quantitative Proteomics using iTRAQ on an LTQ Orbitrap: A new mass spectrometric method combining the benefits of all. J. Prot. Res. 2009, 8, 4743-4752. 7 Gupta N., and Pevzner P.A., False Discovery Rates of Protein Identifications: A Strike against the two-peptide rule. P.A., J. Prot. Res. 2009, 8, 4173-4181. www.thermoscientific.com Legal Notices: © 2010 Thermo Fisher Scientific Inc. All rights reserved. MASCOT is a trademark of Matrix Sciences Ltd., UK. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Part of Thermo Fisher Scientific offices, Thermo Fisher Scientific maintains a network of represen tative organizations throughout the world. References •offers an increase in experimental throughput, confidently identifying more peptides and proteins than any other mass spectrometer currently on the market. In addition to these Africa-Other +27 11 570 1840 Australia +61 3 9757 4300 Austria +43 1 333 50 34 0 Belgium +32 53 73 42 41 Canada +1 800 530 8447 China +86 10 8419 3588 Denmark +45 70 23 62 60 Europe-Other +43 1 333 50 34 0 Finland/Norway/ Sweden +46 8 556 468 00 France +33 1 60 92 48 00 Germany +49 6103 408 1014 India +91 22 6742 9434 Italy +39 02 950 591 Japan +81 45 453 9100 Latin America +1 561 688 8700 Middle East +43 1 333 50 34 0 Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Russia/CIS +43 1 333 50 34 0 South Africa +27 11 570 1840 Spain +34 914 845 965 Switzerland +41 61 716 77 00 UK +44 1442 233555 USA +1 800 532 4752 Thermo Fisher Scientific (Bremen) GmbH Management System Registered to ISO 9001:2008 Thermo Fisher Scientific, San Jose, CA USA is ISO Certified. AN30217_E 08/10C
© Copyright 2025 Paperzz