Titanium High Speed Air Turbine Handpiece
X500K/X600K/X700K
X500KL/X600KL/X650KL/X700KL (For KaVo® MULTIflex® LUX)
X500SL/X600SL/X700SL (For Sirona® Quick Coupling)
X500WL/X600W/X600WL/X700WL (For W&H® RotoQuick®)
X500BL/X600BL/X700BL (For Bien-Air® Unifix® L)
OM-T0263E 001
Intended to Use
This handpiece is designed only for dental treatment use.
Caution
· When operating the handpiece always consider the safety of the patient.
· Do not attempt to disassemble the handpiece nor tamper with the mechanism.
· Check for vibration, noise and overheating outside the patient’s mouth. If any abnormalities are found do
not operate. Contact authorized dealer for service.
· Should the handpiece runs abnormally, cease operation immediately and return the handpiece to an
authorized dealer for repair.
· Do not exceed the rotation speed shown in Specifications.
· Depressing the push-button while handpiece bur is in rotation will result in OVERHEATING of the
handpiece head. Special caution must be exercised during use to keep cheek tissue AWAY from the
push-button of the handpiece. Contact with cheek tissue may cause the push-button to depress and
injury to the patient may occur.
· Do not allow any impact on to the handpiece. Do not drop the handpiece.
· Always keep bur shanks clean. Dirt or debris in a chuck could cause poor bur concentricity or low chuck
retention force.
· Remove the bur or handpiece only after it has completely stopped rotating.
· Do not use long surgical burs. Do not use burs longer than 26mm.
· Do not wipe nor immerse the handpiece in acidic water or acidic solutions.
· Perform periodical maintenance checks.
· If the handpiece has not been used for a long period, check for noise, vibration, and overheating before
use.
· Refer to the coupling manual for connecting it to the handpiece hose.
1. Specifications
Model
X500KL/X500SL/X500WL/X500BL X500K X600KL/X650KL/X600SL/X600WL/X600BL X600K/X600W X700KL/X700SL/X700WL/X700BL X700K
Head Type
Miniature
Standard
Torque
Rotation Speed
390,000 ~ 450,000min-1
380,000 ~ 440,000min-1
300,000 ~ 360,000min-1
Drive Air Pressure
0.25MPa (2.5kgf/cm2) ~0.3MPa (3.0kgf/cm2)
Spray Type
Quatro Spray
Chuck Type
Ultra Push
ISO1797-1 Type3 ø1.59~1.60
Bur Size
Short Shank
Standard Shank
Optics
Glass Rod
—
Glass Rod
—
Glass Rod
—
Head Diameter
ø10.3mm
ø11.2mm
ø13.2mm
Head Height
11.7mm
13.6mm
13.6mm
*KaVo® and MULTIflex® are registered trademarks of Kaltenbach & Voigt GmbH & Co., Germany.
Sirona® is a registered trademark of Sirona Dental Systems GmbH, Germany.
W&H® and Roto Quick® are registered trademarks of W&H Dentalwerk Bürmoos GmbH, Austria.
Bien-Air® and Unifix® are registered trademarks of Bien-Air Dental S.A., Switzerland.
2. Connection / Disconnection the Handpiece
First make sure the manufacture’s coupling is firmly attached to the handpiece tubing and the drive air
pressure is correct.
For detail, refer to the coupling manual.
Caution
Check tightness at the connection between the coupling and the handpiece.
3. Insertion/Removal of the Bur
3-1 To Insert the Bur
3-2 To Remove the Bur
1
2
Press the
Push
Button
2
Insert the
Bur until it
is secure
Remove the Bur
1
Press the
Push
Button
Caution
Test that the bur is secure by gently pulling and pushing the bur without depressing the Push Button.
4.
Cautions for Bur Usage
· Do not use non-standard burs. The ISO standard shank diameter is ø1.59 - ø1.60 mm.
· Do not use bent, worn, damaged, or non-concentric burs.
· Always use within the bur manufacturer's recommended allowable speed.
· Always use clean burs. Uncleaned burs may allow debris to build up inside chuck, causing weak bur
retention or premature chuck wear.
· Insert the bur all the way into the chuck until it stops with the push-button fully depressed.
· Do not use short shank burs in Standard or Torque head handpiece.
5. Lubrication
All handpieces can be cleaned and disinfected with a thermo
disinfector.
All handpieces can be autoclavable up to Max.135°C.
Manual
2
3
1
Spray Nozzle
Insert the Spray Nozzle to lubricant
spray and handpiece backend
Spray for approximately 2 - 3
seconds until the oil comes
out of the handpiece head
Shake the
can 3 - 4
times
PANA Spray plus
/ PANA Spray
Automated
Clean and lubricate handpieces with NSK Care3 Plus. Be sure to use a right adaptor for each handpiece.
Refer to the Care3 Plus Operation Manual for detail.
Weekly Maintenance -Chuck Cleaning
2
1
Arrowed Head Spray Nozzle
Insert the Nozzle to lubricant spray and handpiece head
Shake the
can 3 - 4
times.
PANA Spray plus
/ PANA Spray
Arrowed Head Spray
Nozzle
3
A
Keep pressing the push
button to open the chuck
and lubricate chuck directly
thorough the hole for bur.
Caution
· Be sure to hold the handpiece firmly to prevent the handpiece
from slipping out by the spray pressure when lubricating.
· Keep the NSK PANA SPRAY Plus / PANA SPRAY can upright.
· Use of lubricants other than NSK’s Spray can cause excessive
heat generation and possible handpiece damage.
· Perform Chuck Cleaning at least once a week.
View from A
Hole for bur
6. Sterilization
• Autoclave sterilization is recommended.
• Sterilization is required first time you use and after each patient as noted below.
Autoclave Procedure
1) Scrub dirt and debris from the handpiece, and wipe clean with alcohol-immersed cotton swab or
cloth. Do not use a wire brush.
2) Lubricate the handpiece. (Refer to 5. Lubrication)
3) Insert into an autoclave pouch. Seal the pouch.
4) Autoclavable up to max. 135˚C.
ex.) Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C.
5) Keep the handpiece in the autoclave pouch to keep it clean until you use it.
Caution
· Do not heat or cool the handpiece quickly. Rapid change in temperature could break the glass rod
give abnormal strain to other metals.
· The color may be slightly faded by repeated sterilization. (X650K L)
7. Replacing the optic illumination bulb (X WL,X BL Series)
1
2
Loosen the Taper Ring
and remove it
Taper Ring
Pull out the Joint and remove the Bulb Holder and replace the
new Bulb
Bulb Holder
Bulb Holder
Bulb
Bulb
Joint
Joint
<X WL Series>
3
<X BL Series>
Returen the Bulb Holder to the handpiece and Tighten the Taper Ring.
Optional Bulb : TA Bulb (Pack of 3) Order No.Y900132
Caution
· Do not touch the glass section of new Bulb.
· Screw the Taper Ring tightly. A loose fitting Taper Ring could result in water and air leakage. It is
easier to screw in the Taper Ring when the bulb holder is lightly pressed in by hand.
8. Cleaning the Celluar Glass Rod Optic
Wipe clean the Cellular Glass Rod Optic Entry and Exit Point with an alcohol-immersed cotton swab.
Cotton Swab
Cotton Swab
Cellular Glass Optic Rod Entry Point
Cellular Glass Optic Rod
Exit Point
Caution
Do not use a sharp tool to clean the cellular glass optic rod. It could damage the glass and reduce the
light transmission. If illumination becomes dim please contact your dealer.
The EU directive 93/42/EEC was applied in the design and production of this medical device.
2013.00.20 00 S