SMILE Johns Hopkins University Baltimore, MD USA Rotator/Shaker Maintenance and Calibration- Guidelines Digene Hybrid Capture System Rotary Shaker Maintenance and Calibration Guideline Number Equ10-14-G Effective Date 23 July 2010 Page 1 of 1 Rotator/Shaker Maintenance and CalibrationGuidelines Supersedes New Review by Review date 9-Feb-12 Subject Heidi Hanes SMILE Comments: This document is provided as an example only. It must be revised to accurately reflect your lab’s specific processes and/or specific protocol requirements. Users are directed to countercheck facts when considering their use in other applications. If you have any questions contact your SMILE representative. Audit: Questions pertaining to Rotator/Shaker maintenance and calibration can be found in section Equipment. CAP Accreditation Checklist: Questions pertaining to Rotator/Shaker maintenance and calibration can be found in the Immunology CAP Accreditation Checklist. Background Information: The Hybrid Capture System Rotary Shaker I has been designed for shaking microplates, tube, and microtube racks and other sample container. The shaker can be used in environment from 4°C to 65°C, which allows for cold room use as well as incubated applications. The Hybrid Capture System Rotary Shaker I is compromised of the rotary shaker base and the shaker platform. The shaker platform will hold from one to four standard microplates (96-wells), one tube rack or two microtube racks. The shaker platform has a spring bar allowing it to hold either microplates or tube racks. It also has a non-skid rubber surface to provide stability to sample contains while shaking. Four flat head screws are used to attach the platform to the shaker base and it can hold up to 2 kg of weights. The speed of the shaker is variable, from 200 to 1200 rpm and it operates in a circular motion with an orbit of 3mm. the Hybrid Capture System Rotary Shaker I features a timer with an operating time range from 30 seconds to 30 minutes or continuous shaking when placed on the “hold” positions. Resources 1. College of American Pathologists (CAP) 2006. Commission on Laboratory Accreditation, Laboratory Accreditation Program; Laboratory General Checklist Revised 4/6/2006. 2. Good Clinical Laboratory Practice (GCLP) 2008. 3. Digene Hybrid Capture System Rotary Shaker I Operator’s Manual. Digene Corporation 1999. 81910032 Version#: 1.0 Page 1 of 10 SMILE Johns Hopkins University Baltimore, MD USA Rotator/Shaker Maintenance and Calibration - Policy Rotator/Shaker Maintenance and Calibration Policy Policy Number Equ10-14-P Effective Date 23 July 2010 Approval Signature: Laboratory Director Review Cycle 1 year Supercedes New Policy The laboratory has processes and procedures to effectively ensure that general maintenance and calibration performed on Hybrid Capture System Rotary Shaker I instrument meets the correct standards for laboratory testing. Purpose This policy provides direction for the processes and procedures to effectively ensure that the maintenance and calibration performed on Hybrid Capture System Rotary Shaker I instrument meets the correct standards. Responsibility The Laboratory Medical Director is responsible for (to be determined by Laboratory). The Laboratory Manager is responsible for (to be determined by Laboratory). The Lead Technologists are responsible for (to be determined by Laboratory). Additional Information as required NA Supporting Documents The following processes support this policy: 81910032 Digene DML 2000 instrument plate reader Digene DML 2000 instrument plate washer Maintenance and calibration logs Safety SOP Version#: 1.0 Page 2 of 10 SMILE Johns Hopkins University Baltimore, MD USA Rotator/Shaker Maintenance and Calibration - Process Rotator/Shaker Maintenance and calibration process Policy Number Equ10-14-Pro Effective Date 23 July 2010 Approval Signature e: Laboratory Director Review Cycle 1 year Supercedes New Process for Rotator/Shaker Maintenance and Calibration Who’s Responsible What Happens Instructions General maintenance and calibration of shaker instrument are determined To be determined by the Laboratory Technical SOPs that require the use of shaker General maintenance and calibration is performed on shaker instrument QA Technologist Technicians/Technologists Digene Hybrid Capture System Rotary Shaker I maintenance and calibration SOP 81910032 Version#: 1.0 Page 3 of 10 SMILE Johns Hopkins University Baltimore, MD USA Rotator/Shaker Maintenance and Calibration - SOP Mandana Godard Author(s), Name & Title Document Number Penny Stevens International QA/QC Coordinators Equ10-14-SOP Effective Date 23 July 2010 SMILE Comments: This document is provided as an example only. It must be revised to accurately reflect your lab’s specific processes and/or specific protocol requirements. Name, Title Signature Date Name, Title Signature Date Approved By SOP Annual Review Version # [0.0] Revision Date [dd/mm/yy] Description (notes) Revision History Name (or location) # of copies Name (or location) # of copies Distributed Copies to Associated Forms: 81910032 Version#: 1.0 Page 4 of 10 SMILE Johns Hopkins University Baltimore, MD USA I acknowledge that I have read, understand and agree to follow this SOP. Name (print) 81910032 Signature Version#: 1.0 Date Page 5 of 10 SMILE Johns Hopkins University Baltimore, MD USA Purpose The purpose of this SOP is to describe and ensure the accurate operation, maintenance and calibration of shaker used for variety of laboratory testing. Manufacturers’ recommended instructions are followed for cleaning and disinfecting procedures. Pre-analytic Procedure Refer to the following procedures: Safety SOP Timer Calibration Speed (RPM) calibration The exceptions to these procedures are listed below: NA Definitions: CAP- College of American Pathologists RPM- Revolutions Per Minute SOP- Standard Operating procedure Analytic Procedure All Shaker instruments must be inventoried and uniquely identified. This documentation should at least include: manufacturer, model, serial number, date placed in service, location and current service contact, if any. Equipment and instruments must have written records of service and maintenance for the life of the instrument. Specimen Information: NA Reagents/Media: NA Supplies: Tools Lubricating grease Reflective tape Tachometer 81910032 Version#: 1.0 Page 6 of 10 SMILE Johns Hopkins University Baltimore, MD USA Calibration intervals: It is recommended by the manufacture that the speed (RPM) of the hybrid Capture System Rotary Shaker I be verified every three months. For CAP Accredited labs note that the CAP guidelines state the following: Most serologic centrifuges and timers do not require frequent recalibration. Accuracy of speed and timing must be checked initially, after adjustments/repairs or implementation of new techniques, and periodically thereafter. The frequency of periodic checks should be based on the historical stability of the centrifuge, but at least every 6 months. Equipment Maintenance: Daily maintenance Step Action 1 Clean the shaker with a soft, dry cloth following each use. Any Spills should be cleaned immediately, as stated below. The unit should not be immersed in water or liquid poured over the unit as electrical shock may occur. If Then A spill has occurred. In order to clean the platform it has to be first detached from the Shaker Base using following steps: 1. Remove the center circular cutout in the rubber mat platform, which covers the mounting screws and remove the four flat head screws. 2. Lift Shaker platform up off the base mount. 2 Clean the platform with a cloth damped in a dilute detergent solution. Make sure the platform is completely dry before reinstalling on the shaker base. Prior to use, check that the Shaker is level and on a firm surface. Quarterly Maintenance Step 1 81910032 Action The four plastic bushings supporting the platform should be lubricated with a small drop of the supplied lubricating grease every three months. Version#: 1.0 Page 7 of 10 SMILE Johns Hopkins University Baltimore, MD USA Shaker Speed (RPM) Calibration Step Action 1 Calibrate speed using standard tachometer (Follow the steps bellow a-i). If Then Measured speed is within the designated speed +/- 100 RPM. No adjustment is necessary. If the measured speed is outside of these limit. a. Adjust the shaker speed control knob/dial to desired speed (e.g. 1000 RPM). b. Turn the power to the “ON” position and set the timer to the hold position. c. Apply a 1/16-inch strip of reflective tape to the platform. d. Turn on the shaker by depressing the Start/Stop switch and allow to reach full speed (-5 sec.). e. Set the tachometer for RPM. Start and aim the tachometer beam at the center of the tape strip. (The tachometer must be securely held in place). f. Focus the light by raising or lowering the tachometer until the beam comes into focus. g. Hold the tachometer steady for at least ten seconds. Ensure that the light beam on the tape crosses the beam once every revolution. h. Record RPM and turn off the shaker by depressing the Start/Stop switch. i. After following steps a-h the measured speed is not within the limits. 81910032 Measured speed is within +/- 100 RPM from the shaker speed control setting, speed calibration is verified and no further action required. Remove the speed knob cover by prying off the knob center cap piece with a small screwdriver. Loosen the revealed screw and position the knob to agree with the measured speed. Tighten the screw in this position. Repeat the above steps f-h to confirm that the adjustment made correct. Replace the speed knob center cap piece. Version#: 1.0 Page 8 of 10 SMILE Johns Hopkins University Baltimore, MD USA After following all steps above the measured speed is not within the limits. 2 Take the shaker out of service and contact the manufacture for service. Record all of the results in the log (Appendix C). Quality Control: NA Test Method Instructions: A. Follow the steps in the table below to operate the instrument. Step Action 1 Always make certain that the platform and any hardware are secure before operating the shaker. 2 Turn the power switch to the “ON” position the power indicator will light at this time. 3 Adjust the speed to the desired level using the shaker speed control. 4 Set the desired time interval or turn the knob to the “Hold” position for continuous operation. An audible click is heard when enabling/disabling the “Hold” function. 5 Load microplates(s) onto platform by pulling back on platform tension bar, inserting plate flatly, and slowly releasing tension bar to secure plate (tension bar should be even and tight). 6 Load the shaker symmetrically. If only shaking one microplate, place the microplate in the center of the shaker. 7 To begin shaking, press the Start/Stop button. 8 If Then Microplates, micro tube or tube rack is not being held securely. Set the tension of the bar by adjusting the two screws at each end of the bar. Difficulty pulling the roter bar when loading the microplates, micro tube or tube rack. Set the tension on the bar by adjusting the two screws at each end of the bar. Entire unit is vibrating because of certain load and speed ratios The speed and/or load should be increased or decreased as necessary to eliminate such vibration. Unit vibrates excessively because; a. Unit is placed on 81910032 a. Relocate the unit to a flat even surface. b. Secure each suction feet on the Version#: 1.0 Page 9 of 10 SMILE Johns Hopkins University Baltimore, MD USA an uneven surface. b. Suction feet are loose. c. Shaker platform is loose. d. The four plastic bushings are dry. Power switch not lit when power is placed in the “ON” position. Because; a. Power cord is not plugged in properly. tabletop. c. Securely attach the platform to the base mount by tightening the four screws under the platform center circular cutout. d. Check that the four plastic bushings feet supporting the platform are lubricated. a. Check that the power cord is plugged in properly. b. Check that the power cord is plugged into a known working power surface. c. Check the fuse and replace by the following steps: b. Power source is not functioning. c. Blown out fuse. Remove power cord from power source. Pry open the fuse drawer at the rear of the shaker base unit. Replace only with the same fuse. SMILE recommends that you also include instructions for turning the shaker on and off as established by your lab. References: 1. Clinical Laboratory Standards Institute (CLSI). Clinical Laboratory Technical Procedure Manuals; Fourth Edition. CLSI Document GP2-A4 (ISBN 1-56238-458-9). Clinical and Laboratory Standards Institute, Wayne, PA. 2. Digene Hybrid Capture System Rotary Shaker I Operator’s Manual (1999). Appendices: (Create as needed at your site.) Appendix A: Rotary Shaker: Daily Maintenance Form Appendix B: Rotary Shaker: Quarterly Maintenance Form Appendix C: Rotary Shaker: Quarterly Speed (RPM) Calibration 81910032 Version#: 1.0 Page 10 of 10
© Copyright 2026 Paperzz