Infra-Red Spectroscopy of Solids and Solutions Experiment This experiment consists of many parts and is intended to familiarize the common sample preparation techniques practiced to obtain IR spectra (in the transmission mode). Obtaining the Spectrum - FTIR. Run the background* (air). Obtain the IR spectrum of the polystyrene film, by placing the polystyrene film in the IR beam. This spectrum can be used to calibrate the instrument. For all other samples, run the appropriate background, set each sample in the IR beam, close the cover and obtain the spectrum after 2 minutes (to ensure the expulsion of atmospheric carbon dioxide). * Run the background of either a) air b) salt plates with no sample or c) the fixed path length cell with the solvent; as needed before obtaining the IR spectrum of the sample. Always allow CO2 free - moisture free air to pass through the cell compartment of the instrument for about 2 minutes with the sample compartment cover closed, before running any data collection for the spectrum generation. IUPAC: Tables of wavenumbers for the calibration of IR spectrometers Wavenumber /cm-1 3027.1 * 2924 2850.7 * 1944.0 * 1871.0 * 1801.6 * 1601.4 * 1583.1 * 1495 1454 1353 1332 1282 1181.4 * 1154.3 * 1069.1 * 1028.0 * 906.7 * 842 752 698.9 * Intensity 9 10 7 3 3 3 9 5 10 10 5 5 3 4 4 6 8 3 3 10 10 * Wavenumbers accurate to better than 1 cm-1 [Butterworths - London (1961)] Sample Preparation: Solids Mulls. In this method the solid material is suspended in oil (e.g. Nujol – a hydrocarbon liquid, or Flourolube 1370-4000 cm-1 – a fluorinated hydrocarbon liquid, depending on the region of interest in the IR region.. The suspension (paste) will then be held between two polished (see page 7) salt plates of KBr or NaCl, as a thin layer. Grind for about 2-3 minutes a dry sample (~0.01g) of the solid (glycine) using an agate mortar and a pestle to a fine powder (would cake at the edge of the pestle and attains a glossy look, 1-2µ particle size < IR radiation wave length). Make a fine, smooth, consistent paste of the powder in the oil by adding a drop of the hydrocarbon oil and further grinding to a paste. Continue grinding for five more minutes until the desired consistency is achieved. Transfer a small amount of this paste on to one salt plate, these plates should be handled at their edges at all times. Cover the ‘paste’ with another plate and press them to remove any air pockets and to spread the sample evenly. Place the plates in a de-mountable cell holder and set the screws to slightly less than finger-tight. Then, tighten the screws to finger tight in a diagonal fashion and in a cyclic fashion. Obtain the IR spectrum using the FTIR spectrometer as described below. If the Nujol peaks overwhelm the set of observed peaks the sample preparation is unsatisfactory or if the peaks are broad the solid sample had not been well grounded and in either case requires a new sample. If the peaks are too strong reduce the sample quantity within the plates; which may be achieved by gently squeezing the plates and/or scraping off material with a clean blade. Reset the cell but do not use screws to squeeze the sandwiched sample. De-mountable cell holder Clean the salt plates with CCl4 moistened paper towel and dry them with lint-free paper towels after use. KBr Pellet: A thin wafer of an intimate mixture of the sample in spectra-grade KBr is employed in this technique. Take a dry glycine sample (100-mg) in a clean and dry agate mortar and grind it vigorously to pulverize the sample to smaller particle size. Transfer 1-5mg of the glycine sample into a plastic vial, add 500mg of dry spectra-grade KBr solid to it and mix well. Transfer this mixture to a die assembly and spread evenly; this sample will eventually be pressed to ~10,000-15,000 pounds under vacuum. Hold the sample at this pressure for about two minutes. The finished KBr pellet must be transparent to visible light. Mount the KBr pellet containing the sample on to a holder to obtain its IR spectrum. Thin Film Preparation: Dissolve ~0.01g of the solid (Vanillin) in about 2mL of ethanol. Evaporate a few drops of the solution on a polished NaCl crystal. Let the volatile ethanol evaporate. Obtain the IR spectrum of the thin film formed. Sample Preparation of Solutions Prepare a ~10% solution of (v/v) benzene or toluene in tetrachloro-ethylene as solvent (by mixing ~100µL of benzene in ~900µL of solvent. Fill a fixed path IR cell with the solvent close the cell with Teflon stoppers; obtain the spectrum of the cell with tetrachloro-ethylene as the background in a fixed path length IR cell. Remove the solvent completely. Introduce the solution of benzene into the fixed path length cell with a clean hypodermic syringe. Close the Teflon stoppers. Run the spectrum as before. Clean the cell immediately with acetone followed by a stream of nitrogen. Store in the proper container and cover it tightly. Obtain the IR of the neat liquid as a capillary film between two NaCl plates. Calibration of the fixed path length IR-Cell – Interference Fringe Method: The interference fringes arising from the constructive and destructive interference of electromagnetic waves is measured in this experiment. Set the output in %T mode. Collect the background file for air. Place the empty cell in the sample compartment and run the ‘IR’ of a fixed path length cell (ZnSe cell, for our purpose). This will generate a fringe pattern. The relationship between the path length t, the number of peaks, n, between two wave numbers say,ν 0 , (the reference peak) and the nth peak (fringe) left of reference peak, say ν n , is (for linear frequency instruments, as the one we have); t= 10n 2(ν n −ν O ) The thickness of the cell spacing is t (in millimeters), and wavenumber is in cm-1. The above equation can be modified to generate a linear plot of wave number vs. fringe number (n). A typical output from an empty cell is shown below. The wave numbers at the minimum transmission positions are shown. Upon rearranging the equation shown above for a series of wave numbers, we get; 5 t ν n = n +ν 0 Pick any peak near the middle of the plot as a ‘reference peak’ on the low frequency side, say it is ν 0 ; and find the 1st peak on the high wavenumber side of the ‘reference peak’, ν 1 . The index for the latter peak is set to 1, i.e. n =1. Continue to find the frequencies for a series of consecutive peaks on the high wavenumber side (n = 2, 3... 8) and record their corresponding, n and ν n values. Plot ν n vs. n; and calculate the path length of the cell and associated uncertainty of the path length from the slope of the best fit line data. FTIR: Instrument Setup The Nicolet FTIR spectrometer is left running continuously to stabilize the system. Open the OMNIC software. On the right and bottom of the OMNIC window there will be a Bench Status indicator. It shows a green check mark indicating the system is ready to be used. If a red cross or a yellow circle shows up it would require the attention of a service person. Open the Experiment Setup window (under the Collect menu). Choose the Collect tab and check that the following settings have been entered before clicking on OK. Number of Scans: 32 Resolution: 4 Final Format: Transmittance Correction: None File Handling: Auto save Background Handling: Collect background after 300 min. Background and Sample Spectrum This spectrometer expects you to collect a background spectrum that it will use for the next five hours. Open the spectrometer sample area by turning the handle in the center of the front face of the spectrometer to insert the background "cell"; if there is no background "cell" employed (i.e. air is the background) leave the door closed. Keep the spectrometer compartment door closed at all times other than that at “cell” exchanges into the sample compartment. Select Collect Background either from the Collect menu, or using the toolbar icon, click OK. As the instrument collects data, the (background) spectrum will appear on the computer screen. At the bottom left of the spectrum a box will show you the progress of scans as the 32 scans are accumulated and averaged. When the experiment is finished, you will be asked whether or not to add the spectrum to window 1; click Yes. Select Collect Sample either from the Collect menu, or using the toolbar icon, click OK. Provide a Title. When the experiment is finished, you will be asked whether or not to add the spectrum to window 1; click Yes. To delete the background spectrum from appearing on the screen, click on the background spectrum, click Edit on the menu bar and select Clear. Click File, select Print. Polystyrene film IR spectrum. Spectrum of Nujol (paraffin, long straight chain alkane, mineral oil) Spectrum of Fluorolube (florinated alkane, mineral oil) Polishing crystals (NaCl or KBr) Precautions: Wear gloves or finger cots. Avoid conditions of excessive humidity. Work in a dust free atmosphere. Keep polishing kit and components clean and free from dust, grit or oil. Alcohol — Felt Method Stretch the felt over one of the ground glass pads and hold it in place with the spring holder. • Pour a small amount of the polishing compound (600) onto the rear of the pad. Moisten the rear third of the pad with ethyl alcohol. Rub the salt surface briskly, figure 8 pattern, over the portion of the felt containing the alcohol and polishing compound, gradually working down into the dry portion of the felt. About thirty strokes should be enough to polish the surface. A dozen strokes or so at the dry end will buff the surface clean of any residual rouge and alcohol. Examine the surface and repeat the procedure if needed. After polishing and buffing, rub the sides of the block on the felt to clean off any residual polishing compound. As polishing proficiency increases, 50:50 water/alcohol may be used. This increases the quality of the polish but requires greater speed. Aqueous Solution — Glass Method Prepare a solution of 25% water in alcohol. Put a few drops on the clean ground surface of one of the glass pads. Rub the salt surface against the glass until the water/alcohol solution has nearly evaporated. Examine the surface and repeat the process, if necessary; usually three times will be sufficient. Buff on dry felt, if necessary. Clean-Up Wipe the agate mortar and pestle with tissue paper. Rinse with acetone or water and wipe again. Wipe the NaCl crystals with tissue paper moistened with alcohol. Return to the container. Store the container in a dessicator. Store the IR grade KBr bottle in a dessicator. Draw any liquid sample remaining from the IR cell, use a syringe. Do not wipe or touch the cell windows. Rinse the interior of the cell with acetone by passing acetone using a syringe. Dry the cell with a stream of nitrogen. Store the cell in a dessicator.
© Copyright 2026 Paperzz