Spatial Resolution Performance Tests of a Multiple-Wavelength NIR Light Transmission Scanner N.M. Uzunov1,4, M. Bello4, G. Baldazzi3, G. Moschini2,4, P. Rossi2,5, R. Pani6,7 1 Department of Natural Sciences, Shumen University, Shumen, Bulgaria. 2 Department of Physics, Padova University, Padova, Italy. 3 Department of Physics, Bologna University, Bologna, Italy. 4 INFN, Laboratori Nazionali di Legnaro, Legnaro (Padova), Italy. 5 INFN Padova Section, Padova, Italy. 6 Department of Experimental Medicine, “La Sapienza” University, Roma, Italy. 7 INFN Roma Section, Roma, Italy. INTRODUCTION In our previous publication we have shown some preliminary results and images obtained using a multiplewavelength NIR scanning device designed in the Laboratory for Radiopharmaceuticals and Molecular Imaging (LRIM) at the National Laboratories of Legnaro, INFN [1]. The device described there has been designed for simple straight-line scanning of the NIR light transmitted through the scanned object. In this article we present the test results from our conceptually new NIR scanning device realized for smallobject and tissue imaging in XY plane. EXPERIMENTAL SETUP The NIR transmission scanner is realized on the basis of 256 pixels InGaAs array detector Hamamatsu G9203256D, placed in a socket of a NIR detector board. Unlike the model used in [1] however, we have changed the mode of scanning fixing the panel with NIR light emitting diodes and NIR detector with the detector board to the scanner compartment. The scanning is performed while the scanned object, placed in a platform with a thin CaF2 window is moving in XY plane (figure 1). A stepping motors mechanical system, based on bipolar stepping motors model 4018L Liengineering with optical encoder controlled by a stepper board system of Robot Factory, has been constructed to realize movement in XY plane. The scanner is able to scan the whole optical window with dimensions of 50 mm x 80 mm in two directions in steps as small as 10-3 mm. A set of five NIR light-emitting diodes fixed to the compartment containing the NIR detector and the NIR detector board is placed at a distance of 25 cm above the scanned object. The stepping motors guidance, the NIR diodes emission and the image acquisitions are fully controlled by a PCbased multifunction intelligent board National Instruments FPGA 7831R. A graphical user interface based on the developing environment LabView 8.5 was created to control the scanner functions as well as to visualize and process the acquired images. Detector signals of the NIR light transmitted through the object at five different wavelengths are collected at each step by switching consecutively the NIR light-emitting diodes. The set of NIR LED employs the following wavelengths: 940 nm, 1070 nm, 1200 nm, 1300 nm and 1550 nm. The scanning procedure produces five images of the NIR light transmitted through the biological objects analyzed. EXPERIMENTAL RESULTS Fig. 1. Block-diagram of multiple-wavelength contact NIR transmission scanner. /1/$QQXDO5HSRUW Each of NIR diodes emits light in a cone with an opening angle of 150. It is important to adjust the light cones emitted by the diodes with their axis of symmetry perpendicular to the scanned surface in order to have a homogeneously illuminated scanned surface. The diode’s orientation in the LED panel is adjusted according its signal in the NIR detector created at a fixed value of the diode current. The signal level read by the NIR detector should be almost equal in all detector cells. Measurements of the NIR diodes current causing a certain light intensity and respectively a certain response in NIR detector have been conducted. Coefficients of the polynomial fit through the measured values have been used to calculate the diodes’ currents to equalize the light intensity produced in the NIR sensor by the five diodes. The spatial resolution of the scanner has been $SSOLHGDQG,QWHUGLVFLSOLQDU\3K\VLFV,QVWUXPHQWDWLRQ determined along two perpendicular directions using USAF 1951 resolution standard, placed at different distances from the detector. For convenience the scan direction perpendicular to the detector sensible array was assigned X and the direction parallel to the detector array was assigned Y. The dependence of the spatial resolution in directions X and Y versus the distance to the scanned object has been measured for each of the five NIR wavelengths. The overall spatial resolution dependence of the system calculated from the measured five wavelengths of NIR light versus the distance along X direction is shown in figure 2 (detector without FOP). The corresponding overall spatial resolution dependence of the system versus the distance along Y direction is shown in figure 3 (detector without FOP). 400 Spatial resolution ( μm) 350 300 250 200 150 100 Sensor without FOP Sensor with FOP 50 0 0 5 10 15 20 25 30 35 40 Sensor-to-object distance ( mm) Fig. 2. Overall spatial resolution along the direction X calculated from the spatial resolution at five wavelengths of NIR light versus the distance to the scanned object. CONCLUSIONS The analysis of the spatial resolutions of the NIR detector measured at 5 NIR light wavelengths shows that the partial spatial resolutions do not differ significantly. As it is seen from figure 2 and figure 3, the spatial resolution of the scanner along the two directions of the scanned surface, provisionally assigned X and Y, are not equal. This is due to the difference in the pixel dimensions of Hamamatsu G9203-256D: the pixel size is 50 μm x 500 μm [4] oriented with its longer size along the X axis. The use of FOP slightly deteriorates the spatial resolution along both directions, resulting in a shift of about 50 μm in both graphs. This is due to the fact that the FOP is compounded by optical fibers with a diameter of 6 μm each with optical isolation between the fibers. Nevertheless the spatial resolution of the detector is satisfactory good for imaging of thin layers of tissues in transmitted NIR light and of a tissue bioluminescence as well. ACKNOWLEDGEMENTS One of the authors (N. Uzunov) would like to acknowledge the support received from the ICTP, Trieste, Italy and in particular the Program for Training and Research in Italian Laboratories (TRIL). 350 300 Spatial resolution ( μm) A fiber optical plate (FOP) model FOP(D35T20.5) containing ordered optically isolated fibers with diameter of 6 μm has been optically coupled to the NIR detector in order to apply the so-called tilting-collimator technique [2,3]. It has a cylindrical shape with diameter of 35 mm and height of 20.5mm. The spatial resolution of the system thus formed (NIR detector with FOP) has been measured scanning the USAF 1951 resolution standard at several distances from the FOP surface for the five NIR wavelengths. The dependence of the overall calculated spatial resolution along directions X and Y versus the distance to the scanned object, is shown in figure 2 and figure 3 respectively (detector with FOP). 250 200 150 100 Sensor without FOP Sensor with FOP 50 0 0 5 10 15 20 25 30 35 40 Sensor-to-object distance (mm) Fig. 3. Overall spatial resolution along the direction Y calculated from the spatial resolution at five wavelengths of NIR light versus the distance to the scanned object. /1/$QQXDO5HSRUW [1] N. M. Uzunov et al., LNL Annual Report (2008) 141. [2] N. M. Uzunov et al., Physics in Medicine and Biology, 51 (2006) N199. [3] N. M. Uzunov et al., LNL Annual Report (2008) 139. [4] http://jp.hamamatsu.com/products/sensorssd/pd101/pd105/pd109/ G9203-256D/index_en.html. $SSOLHGDQG,QWHUGLVFLSOLQDU\3K\VLFV,QVWUXPHQWDWLRQ
© Copyright 2026 Paperzz