RNA CVC 9005 Calibration Verification Controls How to Locate a

RNA CVC 9005
Calibration Verification Controls
How to Locate a Calibration Control Lot
1.
Select Bookmark tab on left:
2.
Locate RNA Kit #:
• Select appropriate kit
• If using the H4 Cartridge, please note sheet specific for H4
3.
To Find a Specific Level and Lot:
• Open file by selecting the “+” icon
• The expanded list of control levels will display
IRMA TruPoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
C V C 9005 Calibration Verification Controls
RNA Medical CVC 9005
Calibration Verification
Control
Kit 0900505
Level 1 Lot 07505
Level 2 Lot 07605
Level 3 Lot 07705
Level 4 Lot 07805
Level 5 Lot 07905
EXP. 10/11
For use with CC, BG, and H3 Cartridges
Expected Values Chart *
LEVEL
1
2
3
4
5
LOT
NUMBER
07505
07605
07705
07805
07905
HCT
%
15.1 – 21.1
20.4 – 26.4
30.9 – 36.9
43.3 – 49.3
57.2 – 63.2
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 1mL syringe, 18 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TruPoint only.
CVC9005_CC_1011
IRMA TruPoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
C V C 9005 Calibration Verification Controls
RNA Medical CVC 9005
Calibration Verification
Control
Kit 0900505
Level 1 Lot 07505
Level 2 Lot 07605
Level 3 Lot 07705
Level 4 Lot 07805
Level 5 Lot 07905
EXP. 10/11
For use with H4 Cartridges
Expected Values Chart *
LEVEL
1
2
3
4
5
LOT
NUMBER
07505
07605
07705
07706
07905
HCT
%
17.4 – 23.4
22.5 – 28.5
32.4 – 38.4
44.2 – 50.2
68.0 – 74.0
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 1mL syringe, 18 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TruPoint only.
CVC9005_H4_0900505_101110
IRMA TruPoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
C V C 9005 Calibration Verification Controls
RNA Medical CVC 9005
Calibration Verification
Control
Kit 1900507
Level 1 Lot 17507
Level 2 Lot 17607
Level 3 Lot 17707
Level 4 Lot 17807
Level 5 Lot 17907
EXP. 12/12
For use with H4 Cartridges
Expected Values Chart *
LEVEL
1
2
3
4
5
LOT
NUMBER
17507
17607
17707
17807
17907
HCT
%
17.0 – 23.0
21.7 – 27.7
32.2 – 38.2
43.2 – 49.2
66.7 – 72.7
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 1mL syringe, 18 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TruPoint only.
CVC9005_H4_1900507_1212
IRMA TRUpoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
C V C 9005 C a l i b r a t i o n V e r i f i c a t i o n C o n t r o l s
RNA Medical CVC 9005
Calibration Verification
Control
Kit 0900506
Level 1 Lot 07506
Level 2 Lot 07606
Level 3 Lot 07706
Level 4 Lot 07806
Level 5 Lot 07906
EXP. 06/12
For use with CC, BG, and H3 Cartridges
Expected Values Chart *
LEVEL
1
2
3
4
5
LOT
NUMBER
07506
07606
07706
07806
07906
HCT (%)
14.2 – 20.2
19.6 – 25.6
30.3 – 36.3
43.0 – 49.0
57.4 – 63.4
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 1mL syringe, 18 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TRUpoint only.
CVC9005_CC_0900506_0612
IRMA TruPoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
C V C 9005 Calibration Verification Controls
RNA Medical CVC 9005
Calibration Verification
Control
Kit 0900506
Level 1 Lot 07506
Level 2 Lot 07606
Level 3 Lot 07706
Level 4 Lot 07806
Level 5 Lot 07906
EXP. 06/12
For use with H4 Cartridges
Expected Values Chart *
LEVEL
1
2
3
4
5
LOT
NUMBER
07506
07606
07706
07806
07906
HCT
%
16.7 – 22.7
22.5 – 28.5
31.4 – 37.4
43.7 – 49.7
67.5 – 73.5
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 1mL syringe, 18 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TruPoint only.
CVC9005_H4_0900506_0612
IRMA TRUpoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
C V C 9005 C a l i b r a t i o n V e r i f i c a t i o n C o n t r o l s
RNA Medical CVC 9005
Calibration Verification
Control
Kit 1900507
Level 1 Lot 17507
Level 2 Lot 17607
Level 3 Lot 17707
Level 4 Lot 17807
Level 5 Lot 17908
EXP. 12/12
For use with CC, BG, and H3 Cartridges
Expected Values Chart *
LEVEL
1
2
3
4
5
LOT
NUMBER
17507
17607
17707
17807
17908
HCT (%)
15.6 – 21.6
20.6 – 26.6
31.4 – 37.4
43.5 – 49.5
56.7 – 62.7
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 1mL syringe, 18 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TRUpoint only.
CVC9005_CC_1900507_1212
IRMA TRUpoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
C V C 9005 C a l i b r a t i o n V e r i f i c a t i o n C o n t r o l s
RNA Medical CVC 9005
Calibration Verification
Control
Kit 2900208
Level 1 Lot 27508
Level 2 Lot 27608
Level 3 Lot 27708
Level 4 Lot 27808
Level 5 Lot 27909
EXP. 06/14
For use with CC, BG, and H3 Cartridges
Expected Values Chart *
LEVEL
1
2
3
4
5
LOT
NUMBER
27508
27608
27708
27808
27909
HCT (%)
15.5 – 21.5
20.4 – 26.4
31.2 – 37.2
43.4 – 49.4
54.7 – 60.7
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 1mL syringe, 18 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TRUpoint only.
Calibration Verification Control 9005 2900208 0614
IRMA TruPoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
C V C 9005 Calibration Verification Controls
RNA Medical CVC 9005
Calibration Verification
Control
Kit 2900208
Level 1 Lot 27508
Level 2 Lot 27608
Level 3 Lot 27708
Level 4 Lot 27808
Level 5 Lot 27909
EXP. 06/14
For use with H4 Cartridges
Expected Values Chart *
LEVEL
1
2
3
4
5
LOT
NUMBER
27508
27608
27708
27808
27909
HCT
%
16.7 – 22.7
21.3 – 27.3
31.8 – 37.8
43.4 – 49.4
66.3 – 72.3
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 1mL syringe, 18 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TruPoint only.
CVC9005_H4_2900208_0614