Group Assignment 5: The Quality Assurance Checks 1. Coherence Function Check 1.1 Shaker test coherence check Figure 1: Driving point FRF at point 5. Figure 2: node number on the wood plate The coherence defines the proportion of the response which is completely accounted for by a linear response to the measured values lie between 0 and l, with a poor coherence [1]. From the above figures, we can find that the hammer test at point 4 has very poor coherence plot, but the most spikes happen near the resonance region. Also, the coherence of shaker test is much better than that of hammer test. The poor coherence in our case may be caused by instrumentation noise on one or both of the force or unmeasured extraneous excitation. Another possibility is that there are some non-linearities of the wood plate itself. 1.2 Hammer test coherence check Figure 3: Driving point FRF&Coherence at point 4 Figure 4: Driving point FRF&Coherence at point 6 2. Reciprocity check Figure 5: Reciprocity between point 7 and point 9 Figure 6: Difference between FRFs The check is performed by measuring two transfer mobility FRFs at point 7 and point 9, swapping the locations of the driving point and the accelerometer between the two measurements. The FRFs are then compared. Also, the change of FRFs between two measurements is plotted as shown in the above figure. Based on the comparison, even there are some relative high differences at certain frequency range; we can consider the linearity check acceptable. 3. Symmetric check Figure 7: FRFs of point 4 and point 6 Figure 8: FRFs at point 7 with two levels of force The symmetric check was made by comparing the driving point FRFs at point 4 and point 6. The result of symmetric check is poor because the wood plate is not fully fixed to the table. Using only a couple clamps may introduce unsymmetrical boundary condition at the base. 4. Homogeneity check For the homogeneity check, we impacted the plate at point 7 with two different levels of force. It can obviously be seen that the measurement made using low-level excitation is poorer than that made using high-level excitation. However, there were no large discrepancies between the location and magnitudes of the peaks corresponding to modes of vibration of this structure and, therefore, the check were deemed to be satisfactory. 5. FRF shape check using shaker test data Figure 9: Shaker test FRF and Phase Figure 11: Hammer test FRF and Phase at point 7 Figure 10: Zoom in frequency from 200~800 Hz Figure 12: Repeatability check at point 7 For both hammer test and shaker test, we found the FRF shape check results are satisfactory to meet the following two statements. -Anti-resonance should always exist between adjacent resonances. -There should be a negative phase shift through resonances and a positive phase shift through antiresonances. For point mobility FRFs, the value of phase should always be between 0 and - 180 degrees. [1] 6. FRF repeatability check From the repeatability check, we found that there is little difference between two FRFs. We can conclude that the environmental noise to the hammer test won’t have a large effect on the modal properties of the structure. Reference [1] Quality assurance procedures for the modal testing of building floor structure.
© Copyright 2026 Paperzz