Lot II -GC analyzer equipment

NL-Petten: The supply, installation, warranty and maintenance of gas
analyzer equipment for Quantitative Gas Analysis at the JRC-IE
Lot II: Gas Chromatography (GC) analyzer equipment
-TABLES -
Page 2:
Table Technical Requirements
Page 5:
Table 7: Gas-A H2 (from reforming of hydrocarbons) gas matrix
Page 6 :
Table 8: Gas-B air gas matrix
Page 7 :
Table 9: Gas-C Fuel gas matrix
Page 8 :
Table 10: Example of further gas impurities
Page 10:
Price table
TABLES LOT II – GC Analyzer Equipment - CONTRACT 104703
1
Table technical requirements
Lot II: Gas chromatography (GC) analyzer Contractor’s Comments/Data
equipment
3.1 EQUIPMENT REQUIREMENTS
The Equipments shall be modular and stand-alone with the
following characteristics:
a. The tables 7 to10 shall be reproduced in documentation
of the offer. The empty cells in these tables shall be
completely filled to inform the Commission about the
characteristics of the Equipment.
b. The layout of the Equipment and its features and
working principles shall be described in detail in the
documentation of the offer. This description shall
contain the expected shifts in Equipment performance
and the wear of its components and parts based on the
intended operational life of the Equipment as specified
in 3.3 a. for the performance of the contract.
c. The documents mentioned in a. and b. and the
maintenance calibration plan shall be part of the offer
AND also be part of the Equipment documentation to
be provided under stage 1b.
d. Equipment spare parts and consumables shall be
available for purchase during the expected life of the
performance of the contract
e. If the Equipment is assembled of more than one
analyzer unit, ease of accessibility shall be assured for
each part.
f. Provided with multiple gas sampling of minimum six
gas streams and a sufficient number of gas calibration
lines, including automated switching via the
Equipment’s operation software
g. Prevent water condensation inside the Equipment in
case of analysis of gas mixtures with residual humidity.
h. Automated Equipment calibration.
j. Computerized data acquisition and operation and
processing software (spectrum visualization, search and
libraries, etc) including computer and flat screen
monitor with GPIB (IEEE 488.2) and ETHERNET
communication interface or equivalent supporting the
TCP/IP network protocol and input for external alarm
signals (i.e. gas & fire alarms, panic buttons, etc).
Operation and analysis software (incl. libraries and
Gaussian/Non-Gaussian de-convolution) licenses shall
apply to the entire period of the performance of the
contract including any possible extended warranty.
TABLES LOT II – GC Analyzer Equipment - CONTRACT 104703
2
k.
The Equipment shall bear the CE mark; approval and
calibration certificates shall accompany Equipment
delivery.
m. Emergency and safety buttons
o. The Equipment shall be safely and easily movable
within the laboratory area (i.e. mobile rack).
p. Liquid’s injection apparatus to allow analysis of
mixtures (see 3.2f.).
q. Equipment sound level in dB at a distance of 1 meter
from the equipment, shall not exceed 70 dB(A) as
measured when in operation.
3.2 MEASUREMENT REQUIREMENTS
a. Quantification (in μmol/mol) of binary gas mixtures A
to H of table 11 subject to the Factory Acceptance Test
(FAT) to be performed in stage 1a.
b. Quantification (in μmol/mol) of ternary gas mixtures A2
to A5, A7, A8, G1, G8, H1 & H7 of table 11.
c. Quantification (in μmol/mol) of quaternary gas mixture
A7-A8 (see table 11).
d. Quantification (in μmol/mol) of binary gas mixtures A
to G of table 12 subject to the Factory Acceptance Test
(FAT) to be performed in stage 1a.
e. Quantification (in mg/l) of binary/ternary liquid
mixtures A and C to E of table 13 subject to the Factory
Acceptance Test (FAT) to be performed in stage 1a
f. Ability for quantification (in mg/l) of binary/ternary
liquid mixtures B and F to J of table 13 subject to the
Factory Acceptance Test (FAT) to be performed in
stage 1a.
g. Time for sampling and analysis of the gas mixtures
mentioned in a. to c. shall not exceed 20 minutes on
average during 3 consecutive sampling intervals in one
hour subject to the Factory Acceptance Test (FAT) to
be performed in stage 1a.
3.3 OPERATIONAL REQUIREMENTS
a. Upon completion of stage 1b, the Equipment shall
operate during the performance of the contract of up to
1,000 hrs per year (corresponding to approximately 8
hours per working day) in performing quantitative and
qualitative analysis.
b. The Equipment shall be operational within 30 minutes
from switch-on.
c. Unattended operation including fail-safe shutdown
d. Ability for Equipment calibration (see 3.1.d). The
Equipment shall have the option of selecting an
automatically calibration performed in intervals
dependent, for example, on number of analyses
TABLES LOT II – GC Analyzer Equipment - CONTRACT 104703
3
performed, elapse of operation hours, analyses value
require so, etc. This information shall be given in the
calibration plan.
e. Handling of gas streams at elevated temperatures (e.g.
up to 180°C) and 1-5 bar pressure.
f. The software shall provide real time multi-tasking
capabilities to support multiple activities concurrently
including programmable measurement and analysis
runs, programmable gas stream selection, spectrum
visualization. The software shall also provide extended
maintenance tools, such as statistical tracking of
analyzer performance, reporting, maintenance logging,
etc.
g. Common data format (i.e. ASCII, Excel, etc) also for
spectrum, libraries and data import and export and
storage.
h. Possibility to use libraries of different origin(s).
3.4 ADVANCED EQUIPMENT PERFORMANCE
FEATURES (Non-mandatory but providing additional
value)
a. Quantification (in μmol/mol) of binary gas mixtures G
and/or H of table 11 with LoDs below the mandatory
requirements (see 3rd column of this table) subject to the
Factory Acceptance Test (FAT) to be performed in
stage 1a
b. Quantification (in μmol/mol) of binary gas mixtures G
and/or H of table 11 with accuracies at the LoDs
concerned below the mandatory requirements (see 4th
column of this table) subject to the Factory Acceptance
Test (FAT) to be performed in stage 1a.
TABLES LOT II – GC Analyzer Equipment - CONTRACT 104703
4
GASES FOR POTENTIAL ANALYSIS
The anticipated gas mixtures used in the tests are listed in the tables below to be completed by tenderer:
Table 7: Gas-A H2 (from reforming of hydrocarbons) gas matrix
Impurities
Gas
concentration
range
Instrument model
type/technique/detector
(µmol/mol)
Water
THC
O2
He; N2; Ar
CO2
CO
CH4
Methanol
HCl
Total
Halogen.
comp.
Formaldehyde
Formic acid
1-5000
1-10000
1-500
1-50000
1-10000
1-10000
1-50000
1-5000
0.1-10
0.1-50
NH3
H2S
20-5000
1-10
Instrument
detection
limit
Precision
(µmol/mol)
%
Relative
detection
limit/lower
detection
limit
(µmol/mol)
Accuracy
(%)
Operating
range
Response
time
Calibration
frequency
(s)
0.1-50
0.1-50
TABLES LOT II – GC Analyzer Equipment - CONTRACT 104703
5
Table 8: Gas-B air gas matrix
Impurities
Gas
concentration
range
Instrument model
type/technique/detector
(µmol/mol)
H2
THC
O2
He; Ar
N2
CO2
CO
CH4
Methanol
SO2
NOx
Formaldehyde
Formic acid
NH3
Total
Halogen.
comp.
Instrument
detection
limit
Precision
(µmol/mol)
%
Relative
detection
limit/lower
detection
limit
(µmol/mol)
Accuracy
(%)
Operating
range
Response
time
Calibration
frequency
(s)
10-10000
1-10000
Up to 100%
1-500
Up to 100%
1-5000
1-1000
1-10000
1-5000
0.5-5000
1-1000
1-500
1-500
0.1-5000
0.1-5000
TABLES LOT II – GC Analyzer Equipment - CONTRACT 104703
6
Table 9: Gas-C Fuel gas matrix
Impurities
Gas
concentration
range
Instrument model
type/technique/detector
(µmol/mol)
H2
THC
O2
He; Ar
N2
CO2
CO
CH4
Methanol
H2S
NH3
Formaldehyde
Instrument
detection
limit
Precision
(µmol/mol)
%
Relative
detection
limit/lower
detection
limit
(µmol/mol)
Accuracy
(%)
Operating
range
Response
time
Calibration
frequency
(s)
Up to 99%
2-100
1-500
5-500
0.1%-50%
2-10
1-10
1-500000
1-5000
0.1-2
0.1-5000
0.1-50
Formic acid
0.1-50
Total
Halogen.
comp.
0.1-50
TABLES LOT II – GC Analyzer Equipment - CONTRACT 104703
7
Table 10: Example of further gas impurities
Impurities
Instrument
type/technique/
detector
Instrument
detection limit
Precision
Relative
detection
limit/lower
detection
limit
Accuracy
Operating
range
Response
time
Calibration
frequency
Unsaturated hydrocarbons
Ethene
Cyclohexene
Aromatic hydrocarbons
Toluene
Benzene
Xylene
Naphtalene
Aliphatic hydrocarbons
Propane
n-hexane
Cyclohexane
Tar
Sulfur containing compounds
COS
Tetrahydrotiophene
Halogen containing
compounds
HCl
HF
Dichloromethane
Trichlorethene
Cis 1,2-Dichloethene
TABLES LOT II – GC Analyzer Equipment - CONTRACT 104703
8
Impurities
Instrument
type/technique/
detector
Instrument
detection limit
Precision
Relative
detection
limit/lower
detection
limit
Accuracy
Operating
range
Response
time
Calibration
frequency
Trifluoromethane
Dichlor Difluoromethane
Si-containing compounds
SiO2
Siloxanes
(Hexamethyldisiloxane,
Trimethylsilanol)
Oxygenated organic
compounds
Ethylenglycol / Propyleneglycol
Isopropanol
Acetone
Methyl-ethylketone
Acetaldehyde
Acetic acid
Butyric acid
MTBE
ETBE
Nitrogen containing
compounds
Pyridine
Acrylates
TABLES LOT II – GC Analyzer Equipment - CONTRACT 104703
9
Price table Lot II GC Analyzer Equipment
STAGE
1a - MANUFACTURE and DELIVERY
Price (€)
1b - INSTALLATION, COMMISSIONING and DOCUMENTATION
1c - TRAINING (based on 8 days)
1d - WARRANTY and PREVENTIVE MAINTENANCE SERVICE
FOR 24 MONTHS
SUBTOTAL STAGE 1 (1a+1b+1c+1d)
2a – PREVENTIVE MAINTENANCE FOR 12 MONTHS
2b – CORRECTIVE MAINTENANCE FOR 12 MONTHS
SUBTOTAL STAGE 2 (2a+2b)
TABLES LOT II – GC Analyzer Equipment - CONTRACT 104703