NanoSpective TEM Techniques Summary

TEM Techniques Summary
The TEM is an analytical instrument in which a thin membrane (typically < 100nm) is placed in the
path of an energetic and highly coherent beam of electrons. Typical operating voltages are between
100-300kV. As the electrons transit the thin membrane, they interact with the atoms of specimen
giving rise to a number of observable phenomena or signals. The resultant signals may be collected
by the appropriate detector and analyzed to obtain information about the morphology, microstructure,
composition, and chemistry of the specimen. The TEM may be used to form images of the specimen,
produce electron diffraction patterns, or collect spectroscopic data. The TEM may be operated in a
number of different modes, conventional imaging (CTEM), scanning transmission electron microscopy
(STEM), diffraction mode, either selected area diffraction (SAD), or convergent beam diffraction
(CBED), energy filtered imaging (EFTEM) and spectroscopy mode, X-ray energy-dispersive
spectrometry (X-EDS) and electron energy loss spectrometry (EELS).
The contrast observed in conventional TEM images is the result of three mechanisms; diffraction
contrast, phase contrast, and mass-thickness contrast. Diffraction contrast occurs in crystalline
specimens when the crystallographic planes that are oriented nearly parallel to the incident beam act
as minute diffraction gratings that divert the post specimen electrons away from the optic axis.
Diffraction contrast in a bright field (BF) TEM image causes the strongly diffracting regions of the
image to appear dark in the image. BF diffraction contrast is enhanced by inserting an aperture into
the back focal plane of the objective lens. Diffraction contrast is also the operative mechanism in dark
field (DF) TEM imaging. A centered dark field (CDF) TEM image of a crystalline specimen is created
by tilting the incident beam so that the incident beam approaches the specimen at an angle equal and
opposite to a selected diffraction angle. The action of tilting the incident beam causes the transmitted
beam to be laterally shifted off the optic axis of the TEM while selected diffracted beams are
simultaneously directed down the optic axis. If an aperture is concurrently inserted into the back focal
plane of the objective lens, only the rays passing through the central opening of the annulus will
contribute to the image. By precisely varying and recording the tilt and rotation angles of the incident
beam, reflections from specific crystallographic planes can be directed down the optic axis and thus
contribute to the dark field images. Crystals with lattice planes that are diffracting down the optic axis
at a given tilt angle will appear bright on the image. The second contrast mechanism, phase contrast,
occurs due to the wave nature of the electrons. As the beam passes through various portions of the
specimen the relative phase of the electron waves is modified. The superposition or recombination of
the post specimen waves causes spatially varying regions of constructive and destructive interference
that results in an observable intensity deviation or contrast. The third contrast mechanism, massthickness contrast occurs due to variations in specimen thickness and/or atomic composition.
Electron scattering increases with increasing specimen thickness and increasing atomic number (Z).
Thus, regions that are thicker or are composed of higher Z elements will strongly scatter the beam
electrons and will appear darker on the image. Mass thickness contrast is an important contrast
mechanism in amorphous specimens.
STEM imaging is performed when the incident electron probe is focused to a few nanometers in
diameter and is rastered over the specimen. The image is formed digitally as the intensity recorded at
the each of the raster coordinates is mapped to the corresponding pixels in the image. Both bright
field and dark field STEM imaging is possible. Diffraction contrast and mass-thickness contrast are
observed in STEM images. Contrast that arises primarily due to atomic composition (Z contrast) is
due to Rutherford scattering. Rutherford scattering is characterized by a large angular deviation from
the optic axis. By using a high angular dark field detector (HAADF) and concurrently selecting a small
camera length, Z-contrast becomes the dominant contrast mechanism in the observed HAADF STEM
image.
NanoSpective, Inc. Ÿ 3267 Progress Drive Ÿ Suite 137 Ÿ Orlando, Florida 32826
407 249 1440
www.nanospective.com
Energy filtered EFTEM imaging, spectrum imaging, and electron energy loss spectroscopy EELS are
possible because certain electron beam/atomic interactions cause the incident electrons to lose
energy. Such interactions are collectively called inelastic scattering processes. As the incident highenergy electrons pass through the thin membrane they have the potential to interact with the core
electrons of the constituent atoms. If an incident electron causes the ejection of a core electron from
the specimen, then the incident electron loses a quantum of energy equal to the characteristic
ionization energy of the ejected core electron. Thus, by using an electron spectrometer, which
consists of several lenses, a magnetic prism and an energy-selecting slit, the emerging electrons can
be filtered according to their post specimen energy.
Energy filtered imaging and EELS are closely related techniques both based on the inelastic
interactions between the electron beam and the specimen. The same system, the Gatan Image Filter
(GIF), acts as both a parallel EEL-spectrometer as well as an energy-filtering system. The difference
between the two techniques is a function of user-selected microscope operating parameters. The
back focal plane (BFP) of the last projector lens in the TEM column serves as the object plane for the
spectrometer. Selecting TEM lens settings that will place a diffraction pattern in the object plane and
consequently the dispersion plane of the spectrometer allows the post slit lens assembly of the
spectrometer to form an image on the square 1024 x 1024 pixel charged coupled device (CCD)
camera. The microscope is said to be in image mode but the spectrometer is diffraction coupled. If
the energy-selecting slit is inserted into the dispersion plane of the spectrometer then an energy
filtered (EFTEM) image is formed. By adjusting the strength of the magnetic prism and the width of
the energy-selecting slit (typically 10-50eV), only electrons of a specifically determined energy are
allowed to contribute to the image. A zero loss EFTEM image is formed when only electrons that
have lost no energy while transiting the specimen are selected to contribute to the image. If the
spectrometer is operated as described above with no energy slit inserted then an unfiltered image is
formed. Although an unfiltered image is collected with the GIF camera, the primary contrast
mechanisms are the same as in any BF TEM image.
The GIF system acts as a spectrometer when the TEM lens settings are adjusted so that an image is
present in the object and dispersion planes of the spectrometer. The post slit lens assembly then
projects a dispersed energy spectrum on a linear 1024 x 1 array CCD. The microscope is said to be
in diffraction mode but the spectrometer is said to be image coupled. Because the entire spectrum is
collected simultaneously, the technique is specifically referred to as parallel electron energy loss
spectroscopy (PEELS). A PEELS spectrum from a single spot may be collected either with the TEM
in conventional imaging mode, diffraction mode or in STEM mode. A PEELS line scan or PEELS
areal scan is only possible when operating in STEM mode. A PEELS areal scan is more accurately
referred to as a spectrum image because a complete spectrum is recorded from each (x,y) coordinate
in the raster. Thus, a data cube is generated and elemental maps can be subsequently generated
from the stored data.
NanoSpective, Inc. Ÿ 3267 Progress Drive Ÿ Suite 137 Ÿ Orlando, Florida 32826
407 249 1440
www.nanospective.com
Figure 1. Schematic diagram of a Gatan imaging filter energy (GIF) showing collection planes for an energy loss spectrum
J(E) and an energy filtered image, respectively.
PEELS and EFTEM imaging are tools for qualitative and quantitative elemental analysis in very thin
specimens. Detection limits can be as low as 0.1% by weight. Because PEELS is characterized by
exceptionally high energy resolution it can with very careful parameterization be used to obtain
chemical or bonding information. Both EFTEM and PEELS provide a means for elemental mapping
based on the characteristic ionization energies of the elements.
NanoSpective, Inc. Ÿ 3267 Progress Drive Ÿ Suite 137 Ÿ Orlando, Florida 32826
407 249 1440
www.nanospective.com
Inelastic scattering events occur when the electrons transiting the specimen lose energy due to
interactions with the atoms of that specimen. Such events are completely characterized by recording
the scattering intensity as a function of the variables J(x, y, z, qx, qy, E), where x, y and z refer to
position in real space, qx, qy are scattering vectors and E is energy. It would be cumbersome to
record and manipulate this large amount of data simultaneously. Thus, as a matter practicality, the
energy loss data acquisition is restricted into categories.
An energy loss spectrum J(E) which is a single PEELS spectrum where the number of electrons
(intensity) having lost a given amount of energy is plotted as a function of energy.
Figure 2. An energy loss spectrum J(E)
NanoSpective, Inc. Ÿ 3267 Progress Drive Ÿ Suite 137 Ÿ Orlando, Florida 32826
407 249 1440
www.nanospective.com
A line plot J(y, E) or J(qy, E) is created as the electron beam is advanced along a spatially defined
path on the specimen. A single spectrum is collected at each position of the beam. Post acquisition,
the individual spectra are analyzed and the number of electrons (intensity) having lost energy within a
range specified by the analyst is plotted as a function of position on the specimen. The energy ranges
selected for analysis typically correspond to known characteristic energy loss ranges corresponding to
specific elements.
Figure 3. A single energy loss spectrum J(E) in which an energy range has been defined by the green energy window. A
line spectrum J(y, E) (figure 4) is generated point by point from integrated intensity within the limits of the energy window for
each spectrum collected along the path of the beam. The intensity is plotted as a function of beam position on the TEM
specimen.
Figure 4a. High angle annular dark field image with red line showing path of the beam along the TEM specimen; b line plot
showing intensity of carbon signal as a function of beam position.
NanoSpective, Inc. Ÿ 3267 Progress Drive Ÿ Suite 137 Ÿ Orlando, Florida 32826
407 249 1440
www.nanospective.com
Figure 5. Line plot shown in figure 4 overlaid on high angle annular dark field image.
NanoSpective, Inc. Ÿ 3267 Progress Drive Ÿ Suite 137 Ÿ Orlando, Florida 32826
407 249 1440
www.nanospective.com
An energy selected or EFTEM image J(x, y,) or filtered diffraction pattern J(qx, qy,) is created when
the lens configurations are adjusted so that an image is projected on the GIF camera. In a fashion
that is conceptually similar to the placement of the energy window over a desired range in the line
spectrum shown above, an energy range is selected by placing a physical slit in the path of the
spectrum and the subsequent image is formed with only electrons that have passed through the slit,
i.e., electrons that have lost a specified amount of energy.
Figure 6a. EFTEM image formed with only the electrons that have undergone no energy loss to elastic scattering events;
b. EFTEM image formed with only the electrons that lost energy via momentum transfer with the nitrogen K-shell electrons;
b. EFTEM image formed with only the electrons that lost energy via momentum transfer to the nitrogen K-shell electrons;
c. EFTEM image formed with only the electrons that lost energy via momentum transfer to the titanium L-shell electrons;
d. EFTEM image formed with only the electrons that lost energy via momentum transfer to the oxygen K-shell electrons
e. multicolor overlay of EFTEM images each formed with only the electrons that lost energy via momentum transfer to the
elements listed in the key.
NanoSpective, Inc. Ÿ 3267 Progress Drive Ÿ Suite 137 Ÿ Orlando, Florida 32826
407 249 1440
www.nanospective.com
A spectrum image J(x, y, E) an energy loss spectrum is recorded at each pixel which generates a
cube of data. From this cube either a PEELS spectrum or EFTEM image may be retrieved post
analysis.
Figure 7. Cube of data resulting from spectrum imaging. The x and y axes represent the position coordinates on the TEM
specimen. The ΔE axis represents the distribution of energy lost by the incident electrons while transiting the TEM
specimen, where the energy scale is increasing from top to bottom. Each xy plane contains an energy filtered image J(x,y)
where the energy loss to be mapped is a coordinate on the vertical energy axis. Each vertical column contains a PEELS
spectrum J(E) where the xy coordinates correspond to the collection location on the TEM specimen.
NanoSpective, Inc. Ÿ 3267 Progress Drive Ÿ Suite 137 Ÿ Orlando, Florida 32826
407 249 1440
www.nanospective.com