Gas Chromatography Another method for separation of organic compounds is gas chromatography (also called gas-liquid chromatography, gc, and glc). This is an instrumental method that requires that the compounds to be separated are volatile and do not decompose with heating. Just as in extraction, this methodology requires the organic compound to distribute itself between two “layers” or phases. One phase is a stationary liquid and the other is a moving vapor phase. Stepwise, the procedure goes like this. The compounds to be separated are injected, by means of a syringe with a very small bore, through a septum in the injection port and into a moving stream of carrier gas. The compounds are then carried into the column, which is the heart of the separation. The column is either walled coated with the stationary phase or packed with what inert beads of diatomaceous earth which have been coated in the liquid phase. As the compounds to be separated reach the column they will distribute themselves between being in the gas phase and being dissolved in this liquid phase. Of course, that suggests that the nature of liquid is very important to the separation, as “like dissolves like.” That which does not dissolve immediately moves on farther down the column where it encounters additional liquid phase. That which does dissolve will soon be re-distributed between “clean” carrier gas the liquid phase. This repetitive distribution creates a separation. Things that are the most soluble in the liquid phase will cling to the stationary phase longer and take longer to move down the column. Things that are not as soluble in the liquid phase will spend more time in the mobile gas phase and move down the column faster. In either case, all compounds will eventually exit the column. How do we know when they exit? Since most organic compounds do not have color, we need something other than visible monitoring. One of the most common methods is a detector called a thermal conductivity detector. The detector “senses” the presence of the now separated compounds in the carrier gas and sends an electric signal to some type of recording device as long as there is anything other than carrier gas in the detector. The sample size is very small, even for most packed columns. For the 1/8 inch diameter packed columns in the instrument you will be using, the suggested volume of sample is 0.5 microliter. In other words, one drop of liquid from an ordinary pipette or burette would contain enough for about 50 injections. The injection port must be heated to ensure that the entire sample goes into the column in the gas phase. If there is any material in the injection that will not vaporize, the instrument may be fitted with a glass sleeve inside the injection port which will eventually plug-up with the residue but can be replaced. The carrier gas must be pure, inert, and preferably have a low molecular weight. The carrier gas for this lab is helium which is chromatographic grade (pure – dry) which has a molar mass of 4.0026m/mole. The only possible carrier gas of lower molecular weight is hydrogen (molar mass = 2.0159gm/mole), and it is not completely inert as it is known to react with pi bonds in the presence of metal catalysts (most columns are made of copper metal). The importance of molar mass will become clearer when we discuss the detector. The column, as described above, has two working components – the liquid phase and the inert beads. The beads should be inert, mechanically strong, and be of similar sizes. While diatomaceous earth is neither inert nor pure, it can be processed to meet our needs. If it is acid washed, metallic impurities have been removed but there are still some reactive –OH groups present. These –OH groups are chemically converted to more inert groups which also create a greater physical strength in the beads. This treated and washed diatomaceous earth is then sieved to ensure a fairly narrow range of diameters to ensure even packing. The liquid phase, if we were to observe it at room temperature, might appear more like a wax or very viscous axle grease. Remember, however, that the column is heated and as temperature increases, viscosity decreases. To prepare the packing for a column, one decides what percentage by weight of the packing should be the liquid phase. Since the compounds to be separated are dissolved in the liquid for part of their time in the column, the greater the percent by weight of liquid, the longer the compounds spend in the column. The amount of liquid is sometimes referred to as the percent liquid loading. What are our criteria for these liquids? First, they should have a fairly high boiling point so that they will not boil off the column while you are running the separation. Therefore, most columns come with instructions that they are not to be heated over a certain temperature for that very reason. Second, the liquid should be such that the compounds you wish to separate have a different solubility in that liquid. If we use a non-polar stationary liquid for two polar compounds, they will probably spend very little dissolved time and, therefore, will exit the column (elute) at very similar times. However, if we use a stationary liquid that is moderately polar (such that one compound is more soluble than the other is) the compound which is the more soluble will spend the most time dissolved in the stationary phase and elute last. Therefore, the liquid components of the column and the liquid loading are generally written on a tag attached to the column so that we can select what we need. There are two basic types of simple detectors. The universal detector is the thermal conductivity detector which responds to how efficiently the carrier gas cools a heated filament. The sample is not destroyed and can be collected at the column exit (if you are very patient). Another class of simple detectors is called ionizing detectors. They convert a portion of the compounds eluting into ions and measure the flow of ions between charged plates at the exit of the column. Some ionization detectors create ions by burning the sample in a flame and others by passing the carrier gas over a radioactive source, which provides electrons to be “stuck” onto the sample molecules. These two types of detectors simply create a signal when any compound other than the carrier gas passes through the detector. They do not give any information about the chemical nature of these compounds. More modern equipment can include detectors such as a mass spectrometer, which not only create a signal when the compound passes but also give structural information about the compound. How does a thermal conductivity (or TC) detector work? A TC detector is a Wheatstone bridge composed of 4 resistors in a circuit. Two resistors have carrier gas flowing over them at all times. The other two have the gas coming off the column passing over them. Initially, the circuit is balanced so that the resistance on both sides of the detector ( carrier gas only and carrier gas + eluants) is equal and the amount of current passing through each side is equal. Physics tells us that resistance changes as temperature changes; therefore, as long as both sides have only carrier gas moving over the filaments, the resistance remains equal. However, the smaller a molecule is, the more efficiently it cools the filament. Therefore, when a compound other than He enters one side of the detector, the filament heats up, and it’s resistance changes. This change in resistance causes a different amount of electricity to pass through one side as opposed to the other. This difference in electric current can be measured and an electric signal sent to a recording device. Clearly not all compounds cool the filament equally well. Therefore, equal amounts of compounds A and B may give a different size signal from the detector. In order to quantify the amount of material represented by a signal, one simply runs the same kind of calibration curve ( mass vs. magnitude of signal) you have run for the Beer-Lambert Law. For this lab, the electronic signal is being sent to an integrator which will not only give us a plot of millivolts vs time, but will also do the integration of the peaks to give the area under the peak. It is the area under the peak which is correlated with the amount of compound, not peak height. Each component will be identified by its retention time – the amount of time between the injection of the sample and the maximum detector response (top of the peak). Retention time is usually given in minutes or seconds, but when it is measured from a chart record, it can also be given in units of distance i.e. cm or mm. That is because the chart paper is moving at a constant rate. Therefore, a retention time of 1 minute could also be expressed as 1 cm if the chart paper is moving at 1 cm per minute. Since we could use several different stationary liquid phases, we sometimes need a standard measure of how well two compounds are separated on a given column. The measure of the efficiency of separation is called resolution. Resolution is measured for two peaks of comparable size as follows: Resolution = (the distance between the peak maximum in mm or cm) ½ [ width of peak 1at the base + width of peak 2 at the base] If the distance between the peak maxima is equal to the average of the peak width, we say that there is baseline separation and the resolution is 1. Resolutions of less than 1 may require changing the column. Resolutions of more than 1 are certainly acceptable, but you may choose to change the column so as to reduce the resoltion number and spend less time per injection. As we use columns, they change. Sometimes very small amounts of the liquid phase are slowly baking off the packing. Sometimes compounds we inject stick and don’t come off. When things like this happen, the efficiency of a column changes; it’s ability to separate compounds changes. When this change reaches some specified value (pre-determined by the user and procedure), the column must be replaced. In order to measure the “health” of the column, chromatographers frequently measure the number of theoretical plates. For now let’s define a theoretical plate as one equilibrium between being in the liquid phase and being in the gas phase. That’s not exactly right, but it will suffice for now. If we inject a known compound at a given carrier gas rate and temperature, we can measure the Theoretical Plates available from the column as follows: Theo. Plates = 16 * [ retention time / width of peak at the base]2 While a distillation such as you ran would probably represent less than 100 plates, gas chromatography frequently will exhibit several hundred to several thousand plates. Another related technique called high performance liquid chromatography (HPLC) which uses a second liquid phase rather than the inert carrier gas, frequently has hundreds of thousands of plates.
© Copyright 2026 Paperzz