IEEE SENSORS JOURNAL, VOL. 9, NO. 9, SEPTEMBER 2009 1103 A Small and Low-Cost 3-D Tactile Sensor for a Wearable Force Plate Tao Liu, Member, IEEE, Yoshio Inoue, and Kyoko Shibata Abstract—In this paper, a new 3-D tactile sensor is proposed for measuring triaxial ground reaction force (GRF) distribution. A pressure-sensitive electric conductive rubber (PSECR) and compact pectinate circuits were used to design the sensing cells of the sensor, making it possible to implement a low-cost and compact system without a complex 3-D structure. Moreover, to tailor the application for measuring human GRF, we adopted the use of elastic rubber as the contact interface of the sensor in order to realize a comfortable human–sensor interface. Calibration and test experiments were conducted to characterize the developed sensor, and a small triaxial force sensor (Tec Gihan, Japan) as well as a six-axial force sensor (Nitta Corporation, Japan) were used as verification measurement devices. Coupling effect tests were performed to calculate cross-sensitivity of the sensor. The experimental results of repeatability, nonlinearity, hysteresis, and dynamic tests indicate that the sensor is feasible for implementing 3-D tactile measurement. Index Terms—Conductive rubber, force plate, ground reaction force, tactile sensor. I. INTRODUCTION T he use of a force plate has been successfully employed to measure ground reaction force (GRF) during gait in the laboratory environment [1], [2]. However, this device requires a sizeable operating space and expensive signal processing devices, and moreover, not more than one stride at a time can be measured during a trial. Therefore, it is difficult to apply it to measurements in real life human environments, which require the device to be small, noninvasive, reliable, easy to use, and low-cost. Winter has pointed out that “A suitably instrumented lightweight ‘force-plate shoe’ is needed to give us the ground reaction forces step by step” [3]. Goran et al. state that the existing methods do not accurately reflect individual physical activity levels, and that further methodological development of physical activity tools should be a high priority for research [4]. Recently, many wearable sensor systems have been developed for the applications of GRF measurements as an alternative to the traditional force plate. Pressure sensors have been used to estimate the distributed normal forces in [5]–[7], but Manuscript received April 02, 2009; accepted June 09, 2009. Current version published August 14, 2009. This paper was presented at the Seventh IEEE Sensors Conference and was published in its proceedings. The Associate Editor coordinating the review of this paper and approving it for publication was Prof. Gerald Gerlach. The authors are with the Department of Intelligent Mechanical Systems Engineering, Kochi University of Technology, Kochi 782-8502, Japan (e-mail: [email protected]; [email protected]; [email protected]). Color versions of one or more of the figures in this paper are available online at http://ieeexplore.ieee.org. Digital Object Identifier 10.1109/JSEN.2009.2026509 Fig. 1. Research background about wearable force sensors. the transverse or frictional force could not be measured. Some flexible force sensors designed by using new materials such as silicon or polyimide and polydimethyl-siloxane have been proposed to measure the normal and transverse forces [8], [9], but force levels of these sensors using these expensive materials were limited to the measurements of small forces (about 50 N). By placing two universal six-axial force sensors at the front and rear boards of a special shoe, researchers have developed some wearable sensor systems for ambulatory GRF measurement during uninterrupted walking in trials [10], [11], but the increased height of the instrumented shoes necessarily affects normal human gait. A method was introduced to estimate the complete GRF by using pressure insoles during walking [12], in which a complex pressure insole measurement system with 99 pressure sensors was constructed and the ensuing data were fed into regression models. By measuring the human body’s center of gravity and foot pressure, Miyawaki et al. proposed an indirect measurement method to estimate the GRF vector [13]. However, high-precision measurement of the human center of gravity is difficult. As shown in Fig. 1, we are presently concentrating on developing and applying some wearable force sensors to measure GRF during gait, and our research on sensors has been divided into four phases. In the first phase, the flexi-tactile sensor system 1530-437X/$26.00 © 2009 IEEE Authorized licensed use limited to: Kochi University of Technology. Downloaded on September 1, 2009 at 02:08 from IEEE Xplore. Restrictions apply. 1104 IEEE SENSORS JOURNAL, VOL. 9, NO. 9, SEPTEMBER 2009 Fig. 2. Three-dimensional ground reaction force measurement system. F-scan (Nitta Corporation, Japan), which can measure only the distribution of normal force, was integrated into an insole to estimate the normal components of GRF. In our second phase, we developed a multiaxial force sensor to measure the triaxial GRF and coordinates of the center of pressure when fixed under a specially designed shoe [14]. However, its hard interface and the weight load on the foot affected the normal human gait according to our experimental test. In the last two phases, we want to develop a thin and light force plate based on 3-D tactile sensors and using lower cost materials. A sensor matrix will be constructed to directly perform 3-D force measurement [15], which also can be integrated into a flexible material, so this sensor system will be able to measure the 3-D GRF via what is a comfortable interface for the human body (see Fig. 2). A new design for a low-cost 3-D tactile sensor that can be easily constructed and miniaturized is presented in this paper. II. MATERIALS AND METHODS Fig. 3. Material test. (a) Test platform. (b) Test results. A. Sensing Material The sensing cell in the sensor was designed using pressuresensitive electric conductive rubber (PSECR) provided by the Pongpara Codan Rubber (PCR) Technology Company, Japan [16]. The PSECR is a composite of an elastomer and specially treated carbon particles, and is available in gray–black flexible sheet form, 0.5 mm in thickness. The PSECR can be glued onto the surface of pectinate electrodes to make a resistor, and the PSECR can be covered with hard-wearing rubber as a protective coat. When a normal force is applied to the surface of a PSECR film, the resistance in the pectinate circuit will decrease. As shown in Fig. 3(a), we made a sensing cell prototype (size: 10 mm 10 mm), and designed a test platform to carry out repeatability, nonlinearity, and hysteresis tests of the sensing cell. The imposed normal loading ranged from 0 to 100 N, and could be accurately measured using a reference force sensor IFS-67M25A50-L40 (Nitta Corporation). The resistance of the pectinate circuit were measured using a multimeter. When the sensing cell was tested under the two conditions of increasing load from 0 to 100 N and decreasing load from 100 to 0 N, we calculated its characteristic parameters, which included: sensitivity of 3.2 k /N, repeatability of 1.5% rate output (R.O.), nonlinearity of 2.7% R.O., and hysteresis of 2.4% R.O. A representative result of the hysteresis test experiments is given in Fig. 3(b). copper-film substrate, because electrode lead lines could be constructed on the same substrate. We designed a 3-D tactile sensor by integrating four sensing cells, each composed of a PSECR film and a fan-shaped pectinate circuit. A PSECR film was glued to the pectinate electrodes, and then, an elastic dome as a spherical interface was glued to the PSECR film. The basic mechanism of the 3-D sensor is that changes in the normal distribution and the amount of force measured on the four separated pressure sensors are used to estimate tridirectional forces imposed on the sensor. As shown in Fig. 4(a), the and axes of the sensor cell were defined along the transverse direction, and the axis was chosen as the vertical axis. A point force vector exerted on the spherical interface of the sensor can , be decomposed into tridirectional components including , and . The measurements of four sensing cells were used to estimate the normal component force ( ) and coordinates ( , and ) of the center of pressure (COP) on the – plane defined by the two transverse axes ( axis and axis) using the following equations: (1) (2) (3) B. Structure and Mechanism of Sensor The pectinate electrodes made it possible that multiple sensing cells made of PSECR could be constructed on a , and are the normal forces measured where using the four sensing cells. We can estimate the four normal Authorized licensed use limited to: Kochi University of Technology. Downloaded on September 1, 2009 at 02:08 from IEEE Xplore. Restrictions apply. LIU et al.: A SMALL AND LOW-COST 3-D TACTILE SENSOR FOR A WEARABLE FORCE PLATE 1105 Fig. 4. Mechanism of the sensor. forces from the voltage signal outputs of the four sensing cells, so (1)–(3) could be given as new forms as (4) is defined as a When the constant coefficient and the subscript indicates the label single parameter of the sensing cells, in order to estimate the tridirectional components of the force vector, we can rewrite (4), (7), and (8) as follows: (5) (9) (10) (11) (6) , and are the calibration coefficients for the where four sensing cells in the 3-D sensor, and the subscripts (1, 2, 3, and 4) indicating the label of the sensing cells are shown in Fig. 4(d). As shown in Fig. 4(a), the force vector’s orientation can be estimated by the coordinates of the COP, and the two transverse and ) can be calculated by the vector force components ( orientation and the normal component ( ) of the force vector. Since the positions of the COP projected on the two transverse axes ( axis axis) are connected with the direction of the imposed force vector [see Fig. 4(b) and (c)], we can estimate the and ) of the force vector extwo transverse components ( erted on the dome by the following equations: where is the calibration coefficients for estimating the normal component ( ) of the applied force vector. Since the manufacturing and assembling of the three parts including the pectinate electrodes, PSECR film, and elastic dome can change the boundary condition of each sensing cell in the sensor, we esti, and ) by mate the calibration coefficients ( an experimental calibration method. As shown in Fig. 5, four pectinate circuits and condition circuits for the sensing cells were constructed on a copper-film substrate, and a prototype of the 3-D sensor (diameter height: 10 mm 10 mm, weight: 10 g) was developed for our experimental test. When a force vector is applied to the sensor, the changes in resistance of the PSECR film are converted into four voltage outputs using the pectinate circuits and operation amplifiers. III. EXPERIMENTAL STUDY (7) (8) A. Static Calibration of Sensor To calibrate the developed sensor, a calibration experiment system composed of a small triaxial force sensor USL06-H5500N-C (Tec Gihan, Japan) and a 3-D stage (Micro Mill) were constructed to provide triaxial reference forces (see Fig. 6). For Authorized licensed use limited to: Kochi University of Technology. Downloaded on September 1, 2009 at 02:08 from IEEE Xplore. Restrictions apply. 1106 IEEE SENSORS JOURNAL, VOL. 9, NO. 9, SEPTEMBER 2009 Fig. 7. Experimental results of sensor calibration. (a) x axial reference load. (b) Output signals of sensing cells. Fig. 5. Three-dimensional tactile sensor prototype. Fig. 6. Calibration experiment system. the first step, we needed to obtain calibration coefficients for the two transverse directions referred to as and axes. In the static calibration experiment, a low-pass filter (cutoff frequency: 10 Hz) was applied on the small triaxial force sensor for measuring input forces. As shown in Figs. 7 and 8, we applied, respectively, the two directional transverse forces ( and axes) to the sensor and the imposed forces were measured using the small triaxial force sensor. A group of results including the four sensing cells’ voltage outputs and the imposed two directional reference forces ( axis: 0, 6.1, 11.8, 18.1, 24.3, 30.1, , , , , and N; axis: 0, 6.0, 12.1, , , , , and N) 18.3, 23.9, 30.0, were obtained. This calibration experiment provided estimates , and of the calibration coefficients including in (9) and (10). We completed the multiple regression analysis of the sampled sensor data using a statistical software package Statistical Package for the Social Sciences (SPSS) 11.0 J. The Fig. 8. Experimental results of sensor calibration. (a) y axial reference load. (b) Output signals of sensing cells. TABLE I RESULTS OF THE MULTIPLE REGRESSION ANALYSIS results of the multiple regression analysis are given in Table I, column of calibration coefficients were used for the and the next step of the vertical direction force (z-axial force) calibration experiment. By substituting the results of the first step of the calibration of the transverse forces into (11), a linear regression analysis was used for obtaining the calibration coefficient ( ) and calculating the axial force. Fig. 9 presents the graphs of the data, which were imported into the linear regression analysis in SPSS 11.0 J. Authorized licensed use limited to: Kochi University of Technology. Downloaded on September 1, 2009 at 02:08 from IEEE Xplore. Restrictions apply. LIU et al.: A SMALL AND LOW-COST 3-D TACTILE SENSOR FOR A WEARABLE FORCE PLATE 1107 Fig. 9. Experimental results of sensor calibration. (a) z axial vertical reference load. (b) Output signals of four pressure sensing cells. The results of the linear regression analysis show a low standard . error of 0.09 for B. Characterization of Sensor Based on the calibration experiment system, we completed repeatability, nonlinearity, hysteresis tests, and a load range evaluation of the developed sensor. According to the PSECR material test results, the maximal normal load capacity of the sensor was limited to 100 N, and we calculated that the maximal static friction force or transverse force on the sensor surface was limited to 35 N when the 100-N normal load was imposed. N for Therefore, load ranges of the sensor are rated as and axes, and 100 N for axis. To analyze repeatability of the sensor on the three axes, we applied -directional reference forces (0, 10, 20, and 30 N), -directional reference forces (0, 10, 20, and 30 N), and -directional reference forces (0, 30, 60, and 90 N), respectively. We repeated loading on the three directions five times, and repeatability of the sensor in measuring -, -, and -directional forces is 3.2%, 3.5%, and 2.1% R.O., respectively. By using the measurement results of the sensor outputs in the repeatability test, we could calculate the nonlinearity of the sensor, and the nonlinearity in measuring -, -, and -directional forces was 5.4%, 4.3%, and 5.0% R.O., respectively. In the hysteresis test, the sensor was tested for the three directions under two conditions; first, of increasing load from zero to rate load, and second, of decreasing load from rate load to zero load. The hysteresis of the sensor in the -, -, and -directions was 5.2%, 5.0%, and 4.1% R.O., respectively. Fig. 10. Test experiment system. (a) Schematics of drag equipment. We adopted this mechanism to produce vertical and horizontal reference forces. (b) Experimental equipments picture. Fig. 11. Dynamic sensitivity of the sensor. C. Dynamic Test of the Sensor When the developed sensor is used for dynamic measurements, errors of dynamic response can occur because of the different frequency ranges of the load imposed on the sensor. The dynamic sensitivity (DS) of the force sensor could be calculated , in which is the sensor output based on the using denotes the static calibration of the developed sensor and imposed dynamic force detected accurately using the reference sensor IFS-67M25A50-L40 (Nitta Corporation). As shown in Fig. 10, a special test system was developed based on the static calibration and the test system to investigate the dynamic behavior of the sensor. Using a DSP control system (DS1104 controller, dsPACE), we controlled a dc motor using a gear system and a rotational encoder (3863H012C-2016-IE2-512, Koshin Denki Kogoyo Company, Ltd.) to produce a sinusoidal load Authorized licensed use limited to: Kochi University of Technology. Downloaded on September 1, 2009 at 02:08 from IEEE Xplore. Restrictions apply. 1108 IEEE SENSORS JOURNAL, VOL. 9, NO. 9, SEPTEMBER 2009 TABLE II RESULTS OF THE COUPLING EFFECT TEST Fig. 13. Verification experiment in a triaxial force measurement. Fig. 12. Loading response of the four sensing cells. (a) z axial force input. (b) x axial force input. (c) y axial force input. on the normal and transverse directions. Fig. 11 shows the frequency characteristics of the normal force for the developed and . The sensor, and the DS of the transverse forces sensor shows acceptable frequency characteristics up to 100 Hz in all the tridirectional tests. This operational frequency range can meet the requirements of human dynamics analysis for measuring GRF during gait [17]. D. Coupling Effect Test After obtaining the calibration parameters, we can calculate triaxial force from the normal force measurements on the four sensing cells, but the coupling effect, which is a common problem in the design of multiaxial sensors, should be addressed. Coupling effect tests were performed to calculate the cross-sensitivity of the sensor using equipment specifically developed for the purpose. The cross-sensitivity can be expressed as the forces measured on the sensors, which are normal to the testing direction loading force [18]. The four sensing cells’ outputs, in terms of voltage change versus loading force, in responding to loads in three different directions were plotted in Fig. 12. The coupling effect of the sensor was evaluated according to the results of the cross-sensitivity test. To produce a transverse force on the interface of the developed sensor when testing the - and -directional loads, a constant pressure load ( N) was applied to the sensor. When the sensor was tested in the -direction, the cross-sensitivity for the - and -directions was calculated as 5.10% and 2.33%, respectively. While the tests were being carried out in the - and -directions, the cross-sensitivity was calculated as 3.33% and 3.13%, and 1.45% and 1.21%, respectively (Table II). E. Verification Experiments of Triaxial Force Measurement In order to evaluate the precision of the sensor in measuring GRF, 3-D measurement tests were performed using the developed sensor test system composed of a triaxial force sensor and a 3-D stage. Since the transverse components of GRF are less than 10% of the normal component of GRF during gait, we produced a dynamic load with a large normal component to stimulate human GRF in walking. The correlation coefficient was used as a measure of the association between two results of the two sensor systems, and the correlation coefficient ( ) is defined as Authorized licensed use limited to: Kochi University of Technology. Downloaded on September 1, 2009 at 02:08 from IEEE Xplore. Restrictions apply. LIU et al.: A SMALL AND LOW-COST 3-D TACTILE SENSOR FOR A WEARABLE FORCE PLATE , where is the force measured by the developed is the force measured by the reference sensor, and sensor, denotes the number of the sample data [19]. Moreover, the root of the mean of the square differences (rms) was used to compare the closeness in amplitude of the two sensor measurement results. The percent error (PE) was calculated as the ratio between the rms errors and the average peak-to-peak amplitude of the measurements of the reference force sensor. As shown in Fig. 13, when the developed sensor and the reference sensor simultaneously measured the input force vector, data from the two sensors were sampled at the same frequency (100 Hz), and were compared with the tridirectional reference loads. The correlation coefficients for the vertical force ( axis) and two-axial transverse force ( and axes) were 0.92, 0.98, and 0.93, respectively. In the measurement experiment, the PEs of , , and axial outputs were 1.2%, 2.0%, and 4.5%, respectively. IV. DISCUSSION AND CONCLUSION PSECR and pectinate circuits were used to design sensing cells in a 3-D tactile sensor, making it possible to implement a low-cost, compact, and light sensor system without a complex 3-D structure. We constructed an experimental test system using a triaxial force sensor to measure reference loads, and a calibration experimental test of the sensor was completed through the application of arbitrary forces with known magnitude and direction. The verification experiment results indicate that the sensor can measure the triaxial force with high precision. Coupling effect tests were performed to examine the cross-sensitivity of the sensor, and we found that the change of vertical load may affect the precision in measuring the transverse forces because of the deformation of the elastic interface under a high pressure force. For the next step of the application, if abundant sensors are constructed under a plate (see Fig. 2), it is desired that the distributed pressure may produce less effect on the elastic interface; whereas, if a single sensor is used to measure large normal force (over 100 N) and transverse force, a hard interface is necessary. N for and Load ranges of the sensor were rated as axes (transverse direction), and 100 N for axis (the normal direction); so, it follows that if we integrate more than eight sensors, a wearable force plate can be implemented for GRF measurement during gait of a normal human subject. Moreover, the small size (diameter height: 10 mm 10 mm) and light weight (10 g) of the sensor mean that it can be easily fixed beneath shoes with a low impact on the weight and height of the shoes, so the sensor will correspond with the need for a method of ambulatory gait assessment in clinical applications, a setting where GRF measurement is required in daily life. REFERENCES [1] M. Brandes, R. Schomaker, G. Mollenhoff, and D. Rosenbaum, “Quantity versus quality of gait and quality of life in patients with osteoarthritis,” Gait Posture, vol. 28, pp. 74–79, 2008. 1109 [2] V. M. Thorup, B. Laursen, and B. R. Jensen, “Development of a more robust tool for postural stability analysis of laparoscopic surgeons,” Surg. Endosc., vol. 22, pp. 1087–1092, 2008. [3] D. A. Winter, “Calculation and interpretation of mechanical energy of movement,” Exerc. Sport Sci. Rev., vol. 6, pp. 183–201, 1978. [4] M. I. Goran and M. Sun, “Total energy expenditure and physical activity in prepubertal children: Recent advances based on the application of the doubly labeled water,” Amer. J. Clin. Nutr., vol. 68, no. 4, pp. 944–949, 1998. [5] R. Liu, Y. L. Kwok, Y. Li, T. T. H. Lao, X. Zhang, and X. Q. Dai, “Objective evaluation of skin pressure distribution of graduated elastic compression stockings,” Dermatol. Surg., vol. 31, pp. 615–624, 2005. [6] D. T.-P. Fong, Y.-Y. Chan, Y. Hong, P. S.-H. Yung, K.-Y. Fung, and K.-M. Chan, “A three-pressure-sensor (3PS) system for monitoring ankle supination torque during sport motions,” J. Biomech., vol. 41, no. 11, pp. 2562–2566, 2008. [7] K. Zhang, M. Sun, D. K. Lester, F. X. Pi-Sunyer, C. N. Boozer, and R. W. Longman, “Assessment of human locomotion by using an insole measurement system and artificial neural networks,” J. Biomech., vol. 38, no. 11, pp. 2276–2287, 2005. [8] P. Valdastri, S. Roccella, L. Beccai, E. Cattin, A. Menciassi, M. C. Carrozza, and P. Dario, “Characterization of a novel hybrid silicon three-axial force sensor,” Sens. Actuators A, Phys., vol. 123–124, pp. 249–257, 2005. [9] E.-S. Hwang, Y.-R. Yoon, H.-R. Yoon, T.-M. Shin, and Y.-J. Kim, “Flexible contact force sensing device using metal/polymer multilayer structure for robotic applications,” Sens. Mater., vol. 20, no. 2, pp. 55–69, 2008. [10] P. H. Veltink, C. Liedtke, E. Droog, and H. Kooij, “Ambulatory measurement of ground reaction forces,” IEEE Trans. Neural Syst. Rehabil. Eng., vol. 13, no. 3, pp. 423–527, Sept. 2005. [11] C. Liedtke, S. A. W. Fokkenrood, J. T. Menger, H. Kooij, and P. H. Veltink, “Evaluation of instrumented shoes for ambulatory assessment of ground reaction forces,” Gait Posture, vol. 26, pp. 39–47, 2007. [12] D. T.-P. Fong, Y.-Y. Chan, Y. Hong, P. S.-H. Yung, K.-Y. Fung, and K.-M. Chan, “Estimating the complete ground reaction forces with pressure insoles in walking,” J. Biomech., vol. 41, no. 11, pp. 2597–2601, 2008. [13] K. Miyawaki, T. Iwami, G. Obinata, and Y. Shimada, “Estimate of floor reaction force vector using foot-pressure sensors,” Trans. Jpn. Soc. Mech. Eng., vol. 74, no. 3, pt. C, pp. 749–751, 2008. [14] T. Liu, Y. Inoue, and K. Shibata, “Wearable force sensor with parallel structure for measurement of ground-reaction force,” Measurement, vol. 40, pp. 644–653, 2007. [15] T. Liu, Y. Inoue, and K. Shibata, “Design of low-cost tactile force sensor for 3D force scan,” presented at the 7th IEEE Conf. Sensors, (in presented at the) Lecce, Italy, Oct. 22–29, 2008. [16] [Online]. Available: http://pcr.lar.jp/index.html. 2007 [17] H. M. Schepers, H. F. J. M. Koopman, and P. H. Veltink, “Ambulatory assessment of ankle and foot dynamics,” IEEE Trans. Biomed. Eng., vol. 54, no. 5, pp. 895–902, May 2007. [18] S. Yaldiz and F. Ünsacar, “A dynameter design for measurement the cutting forces on turning,” Measurement, vol. 39, pp. 80–89, 2006. [19] J. K. Taylor and C. Cihon, Statistical Techniques for Data Analysis. Reading, MA: CRV Press Co., 1996, pp. 206–209. Tao Liu (M’08) received the B.Eng. and M. Eng. degrees in mechanical engineering from the Harbin University of Science and Technology, Harbin, China, in 2003 and the Doctorate degree in engineering from Kochi University of Technology, Kochi, Japan, in 2006. From 2003 to 2006, he was a Research Assistant at the Kochi University of Technology, where during 2007, he was a Postdoctoral Fellow in the Department of intelligent Mechanical Systems Engineering. He has been an Assistant Professor in the Department of Intelligent Mechanical Systems Engineering, since 2008. His current research interests include wearable sensor systems, multiaxial force sensors, human motion analysis, and rehabilitation robots. Dr. Liu is a member of the Japan Society of Mechanical Engineers. Authorized licensed use limited to: Kochi University of Technology. Downloaded on September 1, 2009 at 02:08 from IEEE Xplore. Restrictions apply. 1110 IEEE SENSORS JOURNAL, VOL. 9, NO. 9, SEPTEMBER 2009 Yoshio Inoue received the B.Eng. degree in mechanical engineering, the M.Eng. degree in precision engineering, and the Doctorate degree in engineering from Kyoto University, Kyoto, Japan, in 1970, 1972, and 1993, respectively. From 1989 to 1996, he was the Head of the Vibration and Sound Laboratory, Kobe Steel, Ltd., Kobe, Japan. During 2007, he was a Special Director with Kobe Steel, Ltd. He is currently a Professor in the Department of Intelligent Mechanical Systems Engineering, Kochi University of Technology, Kochi, Japan. His current research interests include vibration analysis, human dynamics, and robotics. Dr. Inoue is a member of the Robotics Society of Japan, the Society of Automotive Engineers of Japan, the Society of Instrument and Control Engineers, the Japan Society for Design Engineering, and the Sports Engineering, Editorial Board, and a Fellow of the Japan Society of Mechanical Engineers. He is a recipient of the merit commendation of the 110th Japan Society of Mechanical Engineers Anniversary in 2008 and the Director-General Prize of the Science and Technology Agency of Japan government in 1984. Kyoko Shibata received the B.Eng. degree in measurement engineering, the M.Eng. degree in measurement engineering, and the Doctorate degree in engineering from Seikei University, Tokyo, Japan, in 1994, 1996, and 2001, respectively. In 1996, she joined Sony System Design, Ltd., Japan. From 1998 to 2002, she was a Research Associate in the Department of Measurement Engineering, Seikei University. During 2003, she was an Assistant Professor in the Department of Intelligent Mechanical Systems Engineering, Kochi University of Technology, Kochi, Japan, where she is currently an Associate Professor. Her current research interests include medical and welfare robots and human friendly robotics. Dr. Shibata is a member of the Japan Society of Mechanical Engineers, the Japan Society for Precision Engineering, the Society of Instrument and Control Engineers, the Institute of Electrical Engineers, and the Institute of Electronics, Information and Communication Engineers. Authorized licensed use limited to: Kochi University of Technology. Downloaded on September 1, 2009 at 02:08 from IEEE Xplore. Restrictions apply.
© Copyright 2026 Paperzz