Sensor Technologies Measurement & Control Instrumentation Ozone Detector-Transmitter 0 E2608-O3 Features Wall-mount or duct-mount version Industrial housing, attached or remote sensor Two analog outputs settable to 4-20 mA or 0-10 V Two relays for alarm / ventilation control RS485 Modbus RTU digital interface 82 55 0...5 ppm O3 (peak 50 ppm) Resolution 0,03 ppm O3 Repeatability < 2 % of signal Signal update every 1 second Response time <60 seconds Maintenance interval 6 months Sensor lifetime 2 years Self-diagnostics full functionality check at start-up Warm-up time ≤ 1 min Power supply 11...30 VDC or 90...265 VAC Power consumption <2W Digital interface RS485, Modbus RTU protocol Analog outputs 2 × 4-20 mA / 0-10 V, user settable Output relays 2 × SPST relays (closing contact), 250 VAC / 30 VDC, 5 A max Default alarm set-points low 0,3 ppm O3 high 1 ppm O3 Enclosure light grey ABS plastic, wall mount, protection class IP65 Dimensions H85 × W82 × D55 mm Cable options for remote probe shielded cable, 2,5 or 5 m Operating environment industrial indoor and outdoor locations Operating conditions -20...+45 °C, 15...90 %RH Ø16 36 3 Detection range 27 diffusion 115 electrochemical Sampling method 80 Sensing method Ø34 max 120° Ø65 Ozone 35 Calibration 77 Ozone O3 27 Detected gase 85 55 Specifications Ø35...36 Hole J3 - return to factory settings 4-20 mA / 0-10 V 4-20 mA / 0-10 V 0V / GND Fieldbus 24 VDC OUT1 OUT2 0V A RS485 B +U = 11...30 VDC L 90...265 VAC N RE1 RE2 S J1 - OUT1 type J2 - OUT2 type Evikon MCI OÜ • Riia 185, Tartu, 51014 Estonia • Tel +372 7336310 • [email protected] • www.evikon.eu Ø50