Chapter-5 Measurement of linear attenuation coefficient of irregular shaped FaL-G (flyash-lime-gypsum) samples using ‘Simplified Two media’ method 5.1 Introduction Measurement of linear and mass attenuation coefficient of regular and irregular shaped FaL-G (flyash-lime-gypsum) samples employing ‘Simplified Two media’ method are presented in this Chapter. Seven different liquid materials plus air have been used as media to measure attenuation coefficient of these samples. Obtained results have been compared with those for regular shaped samples. Experimental values have also been compared with theoretical values calculated from FFAST (Chantler et al., 2005) and XCOM (Berger and Hubbell, 1987). A good agreement has been observed between experimental and theoretical values. Gamma-ray transmission method utilizes the application of Lambert-Beer law for the measurement of the linear attenuation coefficient ( of the sample under investigation. But the accuracy of the method demands that the sample thickness should to be known precisely. However many samples of archaeological type (such as rocks, construction material, flyash materials etc.) have irregular shapes. The precise thickness of such samples can not be measured with a micrometer or a vernier calliper, which, in Part of this work has been published in Appl. Radiat. Isotopes 69 (2011) 1516-1520 120 turn, limits the direct application of Lambert-Beer law for the determining of their linear attenuation coefficient. The ‘Two media’ method (Silva and Appoloni, 2000) provides a solution to this problem. This method utilizes standard Lambert-Beer law in such a way that thickness of sample under study is not required. However, Silva and Appoloni (2000), in their conclusions suggested that the accuracy of the method depends upon the difference in the values of linear attenuation coefficient of the pair of media used. Thus larger the different in the values of linear attenuation coefficient, greater the accuracy of method. Secondly, the media should be preferably homogenous. Keeping in view of above suggestion, Elias (2003) theoretically proposed a simple procedure to introduce some new combinations by considering air as one of the media. This modified ‘Two media’ method is called ‘Simplified Two media’ method. This procedure not only simplifies experimental work, but at the same time, it also allows a greater number of repetitions as well as introduces larger difference in the values of attenuation coefficient of the pair of media used. Further, the resulting linear attenuation coefficient value is an absolute value and not a relative one referred to the linear attenuation values of the media. In present work, attempt has been made for the first time to check experimental reliability and accuracy of this method by measuring the linear attenuation coefficient of both regular as well as irregular shaped FaL-G (fly ash, lime and gypsum) bricks. Seven different materials namely water, glycerin, mobile oil, machine oil, cottonseed oil, ethanol and methanol plus air have been used as media in the procedure of measurements. 121 5.1.1 FaL-G: An eco-friendly building material Increasing use of thermal power plants for electricity production in developing countries like India, results in production of huge amount of fly ash, disposal of which poses significant challenges for the power plants. Production of building materials, particularly FaL-G (flyash-lime-gypsum) bricks is considered to be one of the effective solutions to the ever increasing fly ash disposal problem. Besides this, it offers a viable, energy efficient and environment friendly alternative over traditional burnt bricks used for construction. As production of FaL-G bricks does not involve sintering, thus eliminating the burning of fossil fuels required in the clay brick production process and ultimately contributes to the reduction of greenhouse gas emissions. Therefore, increasing use of flyash materials as a substitute of burnt bricks in construction section demands proper knowledge of gamma ray spectroscopic parameter such as linear and mass attenuation coefficient. This forms the basis of selecting FaL-G bricks as target samples in present investigation. 5.2 Theoretical formulation of ‘Simplified Two media’ method A collimated beam of gamma-ray having initial intensity Io is attenuated in absorber of thickness ‘x’ of absorber according to the Lambert-Beer law: I I o e x (5.1) Where I is transmitted bean intensity of unaffected primary photons. ‘Simplified Two media’ method utilizes this standard Lambert-Beer law in such a way that it is possible to measure linear attenuation coefficient of any shaped sample without using its thickness. In this method, the gamma-ray transmission intensity through the sample under study is measured by immersing it, turn wise, into two different media with known linear attenuation coefficients. Figure 5.1 illustrates 122 ‘Simplified Two media’ method. Irregular shape sample of unknown thickness of which linear attenuation coefficient is to be measured is placed inside an acrylic box of known internal dimensions. The empty space inside the box and around sample is filled with some medium of known attenuation coefficient. Gamma ray beam intensity is measured through that medium. This procedure is repeated for at least two media with known but different attenuation coefficients. Following mathematical expression is obtained for resultant transmitted beam intensity, when the sample under study was immersed in medium 1 which is air in present case: I1 I o (e x e a a ) (5.2) Where I 1 represents transmitted beam intensity by the assembly of sample, medium 1 and acrylic box, Io is incident beam intensity. and a represent linear attenuation coefficients of sample and acrylic box respectively. While ‘a’ represents total thickness of acrylic box. Now expression of resultant beam intensity without sample becomes: I1 I o (e a a ) (5.3) I1 is transmitted beam intensity of medium 1 and acrylic box Similarly by immersing the sample in medium 2 we get: I 2 I o (e 2 ( D x ) e x e a a ) (5.4) Where I 2 represents transmitted beam intensity by the assembly of sample, medium 1 and acrylic box, 2 is the linear attenuation coefficient of medium 2, ‘D’ is the internal dimension of acrylic box and other parameters have the same meaning as describes above. In the absence of sample expression (5.4) can be rewritten as: I 2 I o (e 2 D e a a ) 123 (5.5) a1 a2 a= a1+a2 D=d1+d2 Io d1 d2 I X Media Sample Acrylic Box D - Internal width of acrylic box a - total width of walls of acrylic box X - Unknown thickness of sample Fig. 5.1: Schematic diagram of ‘Simplified Two media’ method. 124 where, I2 is transmitted beam intensity of medium 2 and acrylic box Performing the proper substitutions of above equations, we get the following equation which determine linear attenuation coefficient of irregular shape sample of unknown thickness Where C1 2 ln(C 2 ) 1 ln(C1 ) (cm-1) (5.6) I I1 and C2 2 I2 I1 Thus, we get the equation that determines the linear attenuation coefficient of sample even without the knowledge of its thickness. In present measurement, values of linear attenuation coefficient for both irregular as well as regular shaped samples of FaL-G(flyash ,lime and gypsum) were obtained using expression (5.6). 5.2.2 Error Linear attenuation coefficient measured using ‘Simplified Two media’ method using expression (5.6) is a function of different variables such as C1 and C2 i.e. f ( 2 , C1 , C2 ) . Thus estimated error associated with is the resultant of uncertainties involved with the measurement of each variable is given by following relation (Singh et. al, 2008): V 2 C2 (5.7) where Variance term V is given by 2 2 2 V 22 C21 C22 2 C1 C2 2 125 where 2 corresponds to the independent uncertainty over each variable and the Covariance term C is given by ' COV ( I1 I1' ) 2 ' COV ( I 2 I 2' ) C 2 2 I1 I1 I 2 I 2 Where COV(x,y) corresponds to the covariance for the variable x and y, which measures the correlation among the uncertainties of each one of them and it is defined as: COV ( x, y ) 1 n ( xi x )( yi y ) n i 1 (5.8) While the maximum error linked with linear attenuation coefficient measured employing standard transmission geometry has been calculated from the errors associated with different physical parameters such intensities without sample (Io), intensities with sample (I) and thickness (x) using the following relation: 1 I o I I o x ln x I I I x 2 2 2 2 (5.9) where I o , I and x are errors linked with the intensities Io, I and thickness of the samples respectively. 5.3 Sample preparation Two types of FaL-G (flyash-lime-gypsum) samples has been used in present study; regular shaped as well as irregular shaped. Both types of samples have been prepared in laboratories of Physics Department of Sant Longowal Institute of Engineering and Technology, Longowal, Punjab, India. Flyash used in samples has been procured from Guru Hargobind Thermal Plant Lehra, Bathinda, India. While lime and gypsum has been procured from LOBA CHEMIE and MERCK chemical Pvt. Ltd. 126 Table 5.1: Sample Code Chemical composition, thickness and density of regular as well as irregular samples. Category Thickness (cm) Density (g/cm3) Percentage by weight(%) composition of flyash-lime-gypsum Flyash Lime Gypsum RS-1 Regular 0.60 1.15 70 20 10 RS-2 Regular 0.60 1.16 65 20 15 RS-3 Regular 0.60 1.18 60 20 20 RS-4 Regular 0.60 1.19 55 20 25 RS-5 Regular 0.60 1.21 50 20 30 IRS-1 Irregular unknown 1.17 70 20 10 IRS-2 Irregular unknown 1.18 65 20 15 IRS-3 Irregular unknown 1.20 60 20 20 IRS-4 Irregular unknown 1.19 55 20 25 IRS-5 Irregular unknown 1.22 50 20 30 RS-regular shaped sample of FaL-G IRS-irregular shaped sample of FaL-G 127 India respectively. In order to standardized ‘Simplified Two media’ method for measuring linear attenuation coefficient of odd shaped samples, initially the results obtained for attenuation coefficient of regular shaped samples employing this method has to be compared with those incurred using standard narrow beam transmission method. For this, five regular discs shaped samples have been formed using the procedure as given below: Appropriate amount of flyash, lime and gypsum were weighted using an electronic balance having an accuracy of 0.0001g. The materials were then thoroughly mixed in pestle mortar for half an hour and sieved through mesh (-200) to achieved particle size homogeneity. Some water was added to the mixture and the paste so formed was poured into a specially designed disc shaped cylindrical mould and pressed gently. Disc shaped samples so obtained were taken out from the mould and were kept in sunlight for about 25-30 hours to completely remove the moisture content. Further to check the validity and reliability of method for irregular shaped samples, five odd shaped samples were prepared using the same procedure as described above except the mould was of irregular shape. The detailed description of all the regular as well as irregular shaped samples has been given in Tables 5.1. 5.4 Experimental details and method of measurement 100 mCi 241 Am (59.54 keV) point source along with Ortec (3x3-in) NaI scintillation detector coupled to EG&G Ortec multichannel analyzer has been used in present measurements. Three collimators of aperture 4mm, 3mm and 2mm have been used as source, sample and detector collimators respectively. Schematic diagram of the 128 experimental setup used in present investigation has been shown in figure 5.2. To avoid the scattered radiations from reaching the detector, it was properly surrounded with lead. The ray diagram (figure 5.3) represents the total scatter acceptance angle (θsc) which is the sum of the incidence beam divergence (θin) and angle subtended by exit collimation (θout). For present dimensions the total scatter acceptance angle (θsc) was 2.79o, which is 30. Therefore for this acceptance angle, scattered radiations reaching the detector can produce a ray sum error of (0.5-1.0) %, which is within acceptable limit (Midgley, 2006). Incident (Io) and transmitted beam intensity (I) for each sample were measured for sufficiently large fixed preset time so that the statistical uncertainty could be kept below 0.5%. Stability and reproducibility of the procedure was tested before and after each run. Acrylic box of dimensions of 5cm × 5 cm × 9 cm has been used to hold sample and liquid media in ‘Simplified Two media’ method. Experimental procedure for measuring linear attenuation coefficient of aforesaid samples has been divided into three phases. In first phase, linear attenuation coefficient of regular shaped FaL-G samples as well as aforesaid liquids used as media has been measured by using standard transmission method by adopting narrow beam geometry as shown in figure 5.2. In second and third phase of experiment, linear attenuation coefficient of both regular and irregular shaped samples has been measured by using ‘Simplified Two media’ method adopting the procedure described as follow: Firstly the sample under study has been immersed in medium 1 which is air in present case. Transmitted beam intensity I1 has been measured through the whole assembly (sample, medium 1 and acrylic box). Then the transmission beam intensity I 1 129 10 mm 10 mm C1 4 mm 10 mm C2 3 mm C3 a 2 mm D S 270 mm Amplifier ORTEC 272A C1, C2 and C3- Collimators a-Acrylic Box Multi Channel Analyzer ORTEC A64-BI S-Source H.V Power Supply ORTEC 556 D-Detector Fig. 5.2: Experimental setup of standard transmission geometry. 130 S in C T sc C T-Target S-Source C-Collimator D-Detector out D Fig. 5.3: Schematic diagram representing the incident beam divergence and scattered radiations reaching the detector. 131 without sample has been measured. Similarly transmission beam intensities I 2 and I 2 have been measured for medium 2 with same sample. Then by applying expression (5.6), linear attenuation coefficients of given sample has been obtained for that particular pair of media (i.e. air plus liquid media). Same procedure has been repeated for all other media as well as for regular and irregular shaped samples. 5.5 Result and discussion In first phase of experiment linear attenuation coefficients of seven different materials; glycerin, water, cottonseed oil, Mobile oil, machine oil, ethanol and methanol have been measured employing narrow beam transmission geometry. Similarly linear attenuation coefficients of regular shaped samples of FaL-G have been measured by using same geometry and have been shown in Table 5.2 along with theoretical values obtained from XCOM (Berger and Hubbell, 1987) and FFAST (Chantler et. Al.,2005). It has been observed from tables that measured values were in good agreement with theoretical values within experimental uncertainties, which is ~2%. In second phase of experiment, linear attenuation coefficients of same regular shaped samples of FaL-G were measured employing ‘Simplified Two media’ method. For this, following combinations of seven different materials with air have been used; Glycerine-air, water-air, ethanol-air, methanol-air, cottonseed oil-air, machine oil-air and mobile oil-air. Each of these above pairs acts as two media in ‘Simplified Two media’ method. For each pair of media, a set of five different values of linear attenuation coefficient of a given sample have been obtained. Same procedure has been repeated with all other combinations of seven media and samples. Average value of linear attenuation coefficient thus obtained for each sample has been depicted in Table 5.3. For standardization of this method, results have also been compared with those 132 obtained from standard transmission geometry as well as from FFAST (Chantler et al., 2005) and XCOM (Berger and Hubbell, 1987). The maximum experimental error associated with linear attenuation coefficient measured using ‘Simplified Two media’ method is ≤ 3%. Measured values have been found to be in good agreement with theoretical values within experimental uncertainties. In third phase of experiment, linear attenuation coefficients of irregular shaped samples of FaL-G of same composition (as of regular shape samples) were measured employing ‘Simplified Two media’ method. The procedure of measurement is same as described in second phase. Obtained experimental results along with theoretical values calculated from programme FFAST (Chantler et al., 2005) and XCOM (Berger and Hubbell, 1987) have been shown in Table 5.4. It can be observed from the Table that the measured values are in good agreement with calculated values within experimental uncertainties. To compare the results of regular and irregular shaped samples directly obtained using ‘Simplified Two media’ method, their corresponding mass attenuation coefficient has been measured. This is because; all mass attenuation coefficients are independent of the actual density and physical state (gas, liquid or solid). To obtain the values of mass attenuation coefficient of both regular and irregular shaped FLaG samples, linear attenuation coefficient values of samples measured using Simplified Two media method have been divided by their corresponding densities. Densities of regular shaped samples have been measured using standard mass/volume formula, whereas densities of irregular shaped samples have been measured using standard Archimedes principle formula adopting the same procedure as describe elsewhere (Kirdsiri et al., 2009). Table 5.5 shows a the comparison of mass attenuation coefficient values of irregular shaped samples obtained using ‘Simplified Two media’ 133 method and of regular shaped samples obtained using standard transmission geometry. A good agreement between the values obtained using both methods has proved that the ‘Simplified Two media’ methodology has a potential application for the determination of the linear and mass attenuation coefficient of irregularly shaped samples. 134 Table 5.2: a b Comparison of measured values of linear attenuation coefficient cm-1) of regular shaped samples obtained employing standard transmission geometry with calculated values. S.No. Sample code Experimental values Theoretical values 1. RS-1 0.399 0.397a 0.393b 2. RS-2 0.405 0.404a 0.400b 3. RS-3 0.410 0.414a 0.410b 4. RS-4 0.419 0.420a 0.416b 5. RS-5 0.429 0.431a 0.427b XCOM ( Berger and Hubbell,1987) values FFAST (Chantler et al., 2005) Values 135 Table 5.3: Comparison of the average experimental value of linear attenuation coefficient (cm-1) of regular shaped samples measured employing ‘Simplified Two media’ method with calculated values. S.No. Sample Code 1. RS-1 2. Average experimental values Theoretical Values Experimental values c 0.401 0.397a 0.393b 0.399 RS-2 0.406 0.404a 0.400b 0.405 3. RS-3 0.413 0.414a 0.410b 0.410 4. RS-4 0.424 0.420a 0.416b 0.419 0.428 0.431a 0.427b 0.429 5. RS-5 a XCOM ( Berger and Hubbell,1987) values FFAST ( Chantler et al., 2005) Values c Obtained using linear transmission geometry b 136 Table 5.4: a Comparison of the average experimental value of linear attenuation coefficient (cm-1) of irregular shaped samples measured employing ‘Simplified Two media’ method with calculated values. Average experimental values S.No. Sample Code 1. IRS-1 0.404 0.403a 0.400b 2. IRS-2 0.410 0.411a 0.406b 3. IRS-3 0.418 0.421a 0.416b 4. IRS-4 0.421 0.420a 0.417b 5. IRS-5 0.430 0.434a 0.431b XCOM ( Berger and Hubbell,1987) values FFAST ( Chantler et al.(2005) Values b 137 Theoretical Values Table 5.5: Comparison of the average experimental values of mass attenuation coefficient / (gm/cm2) of irregular and regular shaped FaL-G (flyash-lime-gypsum) samples obtained using both the ‘Simplified Two media’ method and standard transmission geometry respectively Average experimental values Sample Code of mass attenuation a a Measured using ‘Simplifiedcoefficient Two media method’. Sample S.No Code b a b Mass attenuation coefficient value b Measured using standard transmission geometry. 1. IRS-1 0.345 RS-1 0.347 2. IRS-2 0.347 RS-2 0.349 3. IRS-3 0.348 RS-3 0.347 4. IRS-4 0.354 RS-4 0.352 5. IRS-5 0.352 RS-5 0.355 Obtained using ‘Simplified Two media’ method. Obtained using standard transmission geometry. 138
© Copyright 2026 Paperzz