1 1 2 2 INTRODUCTION DISCUSSION & SUMMARY The Omron MicroAir nebulizer (Figure 1, opposite) has been recently developed and its design refined to facilitate hand-held drug aerosol delivery of aqueous aerosol using the latest ultrasonic technology. The MicroAir is designed to transform aqueous drug solutions and suspensions into an aerosol mist in two modes: continuous aerosol generation or manual intermittent (to synchronize with inhalation). The MicroAir employs a mesh having a lot of fine nozzles and horn oscillator that combine to release a fairly fine liquid aerosol with relatively low power consumption. We report an investigation to determine the aerosol output and aerosol size from this unit using testing methods embodied in the new European Standard [1]. 1. The 24 MicroAir nebulizers assessed had notable variation in both aerosol output and aerosol size in the pairs of data obtained for each unit (Table, Figures 2 & 3). 2. The MicroAir units n=1-12 had a markedly slower rate of aerosol output (1 minute output) than units n=13-24 (Figure 2) RESULTS units 1-12 units 13-24 98 µl 156 µl (19) (24) 920 µl 879 µl 83.4 (80.7) 3.19 µm 4.75 µm (0.32) (0.48) aerosol output 1st minut mean µl (std dev) Selection of test nebulizers: 24 MicroAir units were supplied by Omron Japan in early 2002 for testing. These were thought to have aerosol output rates in two distinct categories; low output (units n=112), and high output (U22 units n=13-24). Low output units ( n=1-12) have larger diameter of nozzles in a mesh plate , and high out put units (n=13-24) have smaller diameter of nozzles in a mesh plate. aerosol output total mean µl (std dev) aerosol size mean (MMAD) µm (std dev) 5. Regressing Total Aerosol Output Vs MMAD indicates little or no relationship – i.e. total aerosol output is independant of aerosol size (Figure 4). Fig 2 Aerosol Output from Omron MicroAir Aerosol output 1 minute µg Aerosol Output Total ug 1200 Fig 3 Aerosol Size from Omron MicroAir 6 800 600 MMAD µm Aerosol Output (µg ) 1000 400 6. Regressing Aerosol Output Rate Vs MMAD indicates a clear and direct relationship (Figure 5), ie.e. lower rates of aerosol output related to lower aerosol sizes, and vice versa. This relationship is thought due to evaporation of aerosol solvent as the ’inhaled’ ambient air absorbs aerosol solvent (i.e. water) as it mixes with nebulized aerosol. The amount of solvent absorbed is a constant defined by airflow rate and ambient conditions (oC & %RH). The lower the rate of aerosol output, the greater the influence of evaporative solvent loss on the remaining aerosol particles. Hence, the more drastic reduction in aerosol size. This effect has been observed in other systems (personal observation). 7. By changing the diameter of nozzles selectively within the MicroAir, MMAD of the released aerosol can be controlled. 5 4 3 200 0 0 2 4 6 8 10 12 14 16 18 20 22 2 24 0 MicroAir unit number 2 4 6 8 10 12 14 16 18 20 22 24 MicroAir unit number Fig 4 MicroAir Total Output Vs MMAD Fig 5 MicroAir Aerosol Output Rate Vs MMAD 1500 y = -36.336x + 1045.5 Aerosol Output Rate µg/min aerosol output µg Aerosol Size. Each MicroAir unit was tested as per methods described in the European Standard [1]. Each unit was filled with 2.5 ml of salbutamol and connected to an 8 stage low flow cascade impactor, Marple Series 298X (Westech Instruments Ltd) set to 2.0 L/min flow. The cascade drew the 2 L/min flow at right angles from an overall flow rate of 15 L/min was induced by use of a vacuum pump to simulate patient inhalation. After sampling for 2 minutes, nebulization was suspended and the cascade impaction filters removed for quantification of salbutamol by HPLC. Aerosol size was determined in duplicate from each device. 4. The MicroAir units 1-12 had a lower MMAD compared to units 1324 (Table & Figure 3). The main findings are presented in the Table and Figures below. METHOD Aerosol Output. Each U22 unit tested according to method specified in the European Standard [1]. Each unit was filled with 2.5 ml of salbutamol, connected to a breath simulation device (Figure 2) set at 500ml sinus flow at 15 cycles per minute. Aerosol ‘inhaled’ was collected on a low resistance high capacity 3M filtrete elesctrostatic filter. After collection for 1 minute of nebulization, the collection filter was replaced, and aerosol collection continued to completion (selfindicated by the MicroAir device upon completion of nebulization). Each of these 2 aerosol inhalation filters were then removed for later quantification of salbutamol by HPLC. Aerosol output was measure from each device in duplicate. 3. There was no difference in total aerosol output between units 1-12 and units 13-24 (Figure 2). 2 R = 0.1477 1000 500 0 2 3 4 MMAD µm 5 6 250 REFERENCES y = 32.533x - 2.8478 200 2 R = 0.6111 1. prEN13544-1. European Standard 13544-1. Nebulizing systems and their components. 150 100 50 0 2 3 4 MMAD µm 5 6
© Copyright 2026 Paperzz