HPLC Procedure for Testing Sugars Philyaw; Senior Lab Project Spring 2004 1. First step is to prime the reference cell. This can be done by locating the outlet for the mobile phase. The outlet is noted by the metal screw type cap found on one of the lines connected to the refractive index unit. The reference cell lines are both capped with plastic ends. Remove the metal screw cap from the outlet line. This can be done without tools. Now remove both of the plastic caps from the reference lines and replace one of them with the metal screw cap. In the top of the metal screw cap is a plastic plug. Remove this plug by unscrewing it and set it aside. A special syringe will screw into the top of this metal cap where you removed the plastic plug. Locate and fill the syringe with a small amount of the mobile phase, then screw it into the metal cap. Flush until the fluid comes out of the reference cell outlet. Most mobile phases are toxic so be sure to have some kind of vessel to catch the excess fluid. Once this is done, cap the outlet end with the plastic cap removed earlier and replace the plastic plug removed from the metal screw cap to prevent evaporation. If many samples need to be analyzed, the reference cell may need to be primed again after several hours of running. 2. Next, verify that all units have power Make sure the power strip is on. Turn on the refractive index unit by pushing the power button in. A light should illuminate the button noting the unit is on. Next, plug in the column. There is no on or off button so, once the column is plugged in, it is on. The column is attached to a heater. To set the desired temperature of the column, lift up the box unit the column sits in and flip it over so you can see the back. *Note: the plastic cover for the column is magnetized and may fall off when flipping the unit over. On the back of the unit, there are arrows that allow the user to change the temperature up and down as necessary. Once the temperature is set, then place the unit back in its holder. Now, set the pump flow to 0.5 ml/min and flip the switch to on. *Note: be sure that a vessel in place to catch waste mobile phase before turning the pump on! Fluid should start coming out of the outlet line with in a minute of starting the pump. If fluid does not come out after a minute, there may be a clog or a leak. Refer to the manufacture’s information if the filter needs to be replaced or leaks occur. 3. Flushing the system Now that all the units are running, let the mobile phase flow through the system for a minimum of 30 minutes before running the first sample. This will ensure that anything left in the column from previous experiments is cleaned out, in addition to giving the refractive index unit time to properly warm up. 4. Setting up the computer program Log on to the computer sitting next to the HPLC unit with the username and password specified on the computer tower. On the desktop is an icon called DIAMIR. Double click on the icon to open the program. A prompt will ask you for a password, type in the same password to log on to the computer. Once the program is open, a method must be created. Follow the DIAMIR file instructions located next to the computer. The method allows the user to specify sample run time in addition to sample information. Note if the retention time of the samples is not known, it is better to have a longer run time to ensure elution of the sample. The user can manually stop the program at anytime, but cannot extend the time frame once the program is running without going back into the method. 5. Sample preparation Samples are first filtered with 0.2-micrometer syringe filters to ensure all bacteria, yeast, or any other foreign particles are captured. Note that these particles can clog and/or damage the column and must be removed before injecting the sample. Rinse the injecting syringe at least three times with the sample to ensure no residual remains from the previous sample. Draw a slightly larger amount of the sample into the injecting syringe and gently push out the excess. Dab the tip of the syringe on a Kimwipe® to dry the excess and make sure no bubbles exist in the syringe. 6. Preparing the software for acquisition, baseline monitoring and sample runs. Before running samples and after the equipment has properly warmed up, the baseline must be checked to ensure proper detection of samples. At the top of the DIAMIR, program select acquisitionmonitoring baseline. A chromatograph should appear. If it does not appear, look on the lower left corner there should be three tabs. Choose the system tab. At the top left should now be a title that says Waters or HPLC. Click on the plus sign next to this title, and then click in the boxes below that say status and chromatograph so that they now have a check beside each. Once the chromatograph is visible, change the range so that the baseline is visible by choosing the far left icon. This will open a menu containing two tabs. Choose the bottom tab, uncheck the boxes next to the range, and close. Now the baseline should be visible on a larger scale. If the baseline is not trending on zero, then turn the optical zero knob on the refractive index unit while you watch the baseline until it is stable and close to zero. Clockwise turns of the optical zero knob makes the baseline trend down and counterclockwise turns trends the baseline up. It does not have to be perfectly on zero but the closer the better. Most important is that the baseline is steady and not drifting up or down. Now to set up to run samples: At the top of the DIAMIR, program select acquisition quick start. A drop down box will appear, choose the method created earlier. Next a sample identification menu will appear. The first line will note the name of the file, so make it specific to the sample. Other things to note are flow rate, sample concentration, mobile phase concentration, etc. Once this is complete, hit enter. This will bring up a new page that should say waiting for injection. If this page does not appear, look on the lower left corner there should be three tabs. Choose the system tab. At the top left should now be a title that says Waters or HPLC. Click on the plus sign next to this title then click in the boxes below that say status and chromatograph. Now the page that says waiting for injection should be visible. The program will automatically start once the sample has been injected. 7. Injecting the sample Now that the DIAMIR program is waiting for injection, make sure the switch on the injection port is on load and not inject. Now draw the sample into the injection syringe as described in step 5, and then release the injection port by turning the lever down. Pull out the round silver plug and place it on a clean Kimwipe® on top of the pump unit. These next steps should be done quickly. This will only come with practice so don’t be alarmed. Inject the sample into the port, place the round plug back into the port and lock it by turning the lever up until it is tight. IF YOU DO NOT TIGHTEN THE PLUG, IT CAN SHOOT OUT WITH THE FORCE EQUAL TO THE BACK PRESSURE! THIS IS VERY DANGEREOUS SO DON”T FORGET THIS STEP! Next, turn the switch from load to inject. You will notice that DIAMIR will start as soon as you switch it to inject. Wait until you see 0.3 minutes on the DIMAIR timer and turn the switch back to load. The user may manually stop the program when all the components have eluted (usually by 12 minutes) by clicking on the red icon located at the top of the chromatograph window containing an X. 8. Printing your Chromatograph In the DIAMIR program window choose file open Chromatograph. A window will appear with a list of files; choose the particular file by double clicking on it. It should open revealing the chromatograph. Now choose the results option just below the method string located at the top left menu box. Retention times, area percent, and start/stop times should be listed below the chromatograph. Injection Port Pump Luna Column Refractive Index Unit Figure 5 - HPLC apparatus DIAMIR Program
© Copyright 2025 Paperzz