Electron – Positron Annihilation e+ + e- = 2 * 511keV gammas at 180 degrees Positioning the event Ring of detectors Coincidence detection – i.e. Electronic collimation “Standard” tomographic imaging Imaging ring Detector gamma ray 1 Point of positron emission Event into image gamma ray 2 Detector Coincidence circuit First Second 2D 3D Detector blocks Consider for PET Random events Scatter Attenuation Resolution Sensitivity Detector properties Random Events Finite time window photons from two separate decays may be detected as coincident events Scattered events “blurring” of the image Random Events Co-incidence time window length selected to match properties of the detector material Fast detectors allows shorter windows – hence random events are reduced Short window time with a slow detector - reduce sensitivity Random Events Random rate = 2 t R1R2 (where R1 and R2 are the count rates of the two detectors individually) As the real concidence rate increases the proportion of randoms increases more quickly until it becomes dominant Dealing with Random Events Adding a delay into one arm of the coincidence circuitry can detect just randoms. In theory these can be subtracted from the full data set to leave the reals. Poor count statistics means that data quality deteriorates rapidly once randoms exceeds the number of real events Scatter Scattering in the patient will be result in events being detected as real events but in the wrong position Scattered events will cause image degredation Scatter In the gamma camera scattered events are rejected by good pulse height analysis i.e. only accept gammas in the photo peak (e.g 140+-10% for Tc 99m on a GC) Unfortunately -the detectors used in PET have poor energy resolution (e.g BGO) Energy windows are typically set at 350-650keV and thus accept a lot of scatter Scatter Correction Often ignored for 2D imaging but some correction required for 3D Various approaches for approximate scatter correction images obtained from energy windows set to collect scatter can be subtracted from the main image Scatter Correction-iterative More complex approaches can be used to give more accurate correction Model the physics of the whole system using an attenuation map, the emission images and the characteristics of the detector geometry. As emission images contain scatter an iterative approach is required Attenuation is a major problem in a PET 2 photons need to be detected for a real event i.e if either is attenuated the event is lost. Thus attenuation is significantly greater than in single photon detection Half value thickness in tissue is ~7cm at 511keV Attenuation can be as high as 50% in large patients Attenuation Attenuation can be corrected for Attenuation Correction Attenuation is independent of the source distribution Need to establish the distribution of the attenuating properties of the tissue in the FOV to apply suitable correction factors 2 approaches: Transmission source of isotope CT Scan on PET/CT system Resolution Finite range of the positron in tissue Annihilation photons are not exactly at 180 Resolution limit 1.5mm-2.5mm theoretically In Practice technological limitations - image resolution is ~ 5-8mm Sensitivity As there is no in-plane collimation – sensitivity increased over single photon imaging by ~50 With 3D imaging there is no physical collimation – further increase in sensitivity by a factor of ~6 Thus image noise is greatly reduced as compared with standard NM to th higher count densities Detector materials Initially NaI but inefficient at stopping 511keV photons (Standard gamma cameras were used to perform PET imaging but the poor count statistics & resolution at 511keV rapidly lead to the procurement of PET imaging systems) Bismuth germanate (BGO) – commonly used Newer - Lutetium Oxyorthosilicate (LSO) - shorter decay time – essential for 3D Ideal Scintillator for PET High light output in photons/event ie good scintillation efficiency Short decay time for fast light output (helps for short coincidence times) High stopping power at 511keV Refractive index close to glass to get good impedance matching with PMTs Robust, easy to manufacture and cheap Typical Detector Characteristics at 511keV light output (%NaI) decay constant attenution coefficient (ns) ( per cm) NaI 100 230 0.34 BGO 15 300 0.92 LSO 75 40 0.87 PET- advantages over standard NM imaging Higher sensitivity with resultant lower Poisson noise Attenuation correction is easier Can model PSF over the field of view for Scatter correction methods readily available Theortically accurate quantification of uptake is possible Radionuclides used in PET Cyclotron produced e.g. 12C + 2H = 13N + n Nuclide 18F 11C 13N 15O 1/2life 110m 20.4m 10m 122s Common PET Radiopharmaceuticals and applications 15 O2 oxygen extraction 13N-ammonia blood flow measurement 11C-N-methylspiperone receptor studies 18FDG (fluorodeoxyglucose) tissue metabolism “Accuracy” of PET in Oncology Investigation Ring-PET CT Staging Lung Cancer 92% 70% Detecting Breast Cancer 85% 67 % (Mammography Staging Ovarian Cancer 87% 70% Summary- Advantages of PET Advantages High Sensitivity – reduces Poisson noise Potentially Absolute Quantification - modelling Physiological Isotopes – radio chemistry Disadvantages Cost - both capital & running costs Requires on site cyclotron (or limited to only FDG 18) Currently 5 PET cameras in Scotland Real Events; Random & Scattered events
© Copyright 2026 Paperzz