Z-Pathfinder™ P1 P2 Protocol

www.zendodirect.com
NON-STERILE
Z-Pathfinder™ P1 P2
Protocol
Precizen (K-file)
up to N°10
Pathfinder_card_rev.indd 1
P1
P2
and/or
WL
WL
Non-sterile product. Please sterilize before use.
14-05-27 11:24 AM
www.zendodirect.com
P1
P2
Establishing the Glide Path
N°12 3%
L 21 mm
L 25 mm
L 29 mm
N°17 3%
L 21 mm
L 25 mm
L 29 mm
• Rotation speed: 250 - 400 rpm
Maxi torque: 1.2 N.cm
Zendo® Direct AG
Manufactured by:
MICRO-MEGA®
5-12, rue du Tunnel
F-25006 Besançon
• If canal constrictions prevent the #10 K file from
R1
R1
easily reaching the EWL use a #08 K file along with
copious irrigation until the #10 file reaches the
EWL. Complete glide path preparation using Zendo
Pathfinder files in continuous rotation. Use P1 and
R2
R2
then P2 to the EWL using a slow and unique downR3
R3
ward movement in a free progression
and without
pressure. In case of too much resistance, withdraw
the file from the canal and restart the movement.
Use a #15 K file to determine Working Length (WL)
electronically and/or radiographically followed by
The ZONE 2.0 for root canal cleaning and shaping.
03/13 Ref 60300724-B
NON-STERILE
This instrument is for single use only and has been designed
for the treatment of only one tooth. Use of a file multiple
times will increase the risk of separation.
Caution: USA Federal law restricts this device to the sale by or on the order of a dentist.
Pathfinder_card_rev.indd 2