SPACE TELESCOPE SCIENCE INSTITUTE Operated for NASA by AURA Instrument Science Report COS 2013-03(v1) Second COS FUV Lifetime Position: Verification of Aperture and FUV Spectrum Placement (FENA2) Charles R. Proffitt1, 2, Steven V. Penton1,3, David J. Sahnow1, Cristina M. Oliveira1, Derck Massa1, Rachel A. Osten1, Steven N. Osterman3, and Alessandra Aloisi1 1 2 Space Telescope Science Institute, Baltimore, MD Science Programs, Computer Sciences Corporation 3 University of Colorado, Boulder, CO 8 February 2013 ABSTRACT CAL/COS proposal 12795 was used to determine the exact updates for the Science Instrument Aperture File (SIAF) and for the COS aperture block positioning needed to implement routine science operations at COS FUV Lifetime Position 2. It also verified that the location of each grating’s spectrum on the FUV XDL detector met the requirements to deliver good quality science data. The data obtained showed that a +3.5” offset in the cross-dispersion direction, which corresponds to a +41 pixel shift on the detector, would provide the desired positioning on the detector. To keep the COS aperture centered at this location requires shifting the aperture block by −73 steps in the cross-dispersion direction and adjusting the SIAF locations of the Primary Science Aperture (PSA) at the new position by +3.5” in the cross-dispersion direction and by −0.05” in the dispersion direction. Instrument Science Report COS 2013-03(v1) Page 1 Contents 1. Introduction 2 2. Goals and Design of 12795 2.1 Goals, Design, and Execution of Visit 1 2.2 Goals, Design, and Execution of Visit 2 6 6 8 3. Results 3.1 Placement of the New Lifetime Position 3.2 Aperture Centering at the New Lifetime Position 10 10 17 4. Summary 21 Acknowledgements 22 Change History for COS ISR 2013-‐03 22 References 22 1. Introduction The gain of the COS/FUV Cross-Delay-Line (XDL) microchannel plate detector declines with usage, eventually becoming too low to reliably detect and centroid photons incident on the photocathode. To extend the useful life of the COS FUV detector, it is necessary to periodically shift the spectrum to another location on the detector. In preparation for the first such move, a series of exploratory programs were executed in mid-2011 to gather the information needed to pick the location and sequencing of subsequent lifetime positions. By December 2011, the results of these programs led to the decision that for Lifetime Position 2 the FUV spectra should be shifted by approximately +3.5” in the cross-dispersion direction (see Oliveira et al. 2013b). However, before this new lifetime position could be implemented for routine use, a number of parameters needed to be measured and optimized. To do these a series of four FUV enablement (FENA) programs were implemented (see Osten et al. 2013 for an overall description of the lifetime move enablement and calibration programs and requirements). FENA2 was designed to address two of the specific high-level requirements of the overall FENA effort: - FENA requirement FUV2: The placement of the FUV spectra at the detector location corresponding to the new lifetime position for the PSA shall be verified. FENA requirement FUV3: The PSA aperture shall be located so that the target is well centered in the aperture when the spectra are placed at the new lifetime position. The choice of the new lifetime position was primarily driven by two competing considerations – maximizing resolution and keeping the spectra at the new position Instrument Science Report COS 2013-03(v1) Page 2 away from the gain-sagged regions produced by observations at the original position. The FUV Exploratory program 12678 (Sahnow et al. 2013) had shown that the spectral resolution degraded with increasing cross-dispersion offset from the original position, but did so more slowly in the +Y direction. The best possible resolution would in principle be obtained by shifting the spectrum by ≈ +1.5” in the crossdispersion direction. This behavior was supported by analysis of ray-trace models (Osterman 2010, private communication). In addition, keeping the 2nd lifetime position as close as possible to the original one would have the additional benefit of maximizing the detector area available to use for subsequent lifetime positions. FUV Exploratory program 12676 (Massa et al. 2013 ) used deuterium lamp spectra to map the gain and flat field characteristics over the accessible region of the FUV detector, clearly illustrating the sagged regions that need to be avoided at the new lifetime position. The most intense illumination and the most rapid microchannel plate damage results from the geocoronal Lyα illuminating the 2.5” diameter COS PSA aperture during G130M observations. Each of the 4 FP-POS positions of each of the 5 original G130M CENWAVE settings illuminates a different portion of the B segment (FUVB) of the FUV XDL detector, leading to 20 total spots (see Figure 1). Two of the four similar spots that result on the A segment (FUVA) from the G140L 1105 setting are easily visible in the gain map near the left edge of the segment. Each of these full-aperture spots spans the cross-dispersion range of well-centered point source spectral locations for observations with all three gratings. Consequently it is necessary to ensure that the new location will keep all FUV spectra sufficiently clear of these previously damaged regions. Instrument Science Report COS 2013-03(v1) Page 3 Figure 1: Gain maps for FUVA (above) and FUVB (below) of the COS FUV XDL detector based on internal deuterium lamp spectra obtained in July 2011 as part of CAL/COS program 12676. The regions of the detector where the gain has dropped as a result of incident photons are evident as dark bands across the middle of each segment. A number of more deeply sagged regions are visible on FUVB at the locations where the geocoronal Lyα projects onto the detector. Two fainter spots from the G140L 1105 geocoronal Lyα can be seen near the left edge of FUVA near columns 2300 and 2600. Other localized low gain regions are also visible. For the Instrument Science Report COS 2013-03(v1) Page 4 second lifetime position, the most significant of these is the region near column 8000 on segment A. Under the assumption that usage patterns at the new position would match those at the original position, analysis of the 12676 data concluded that shifting the location by +35 pixels would be adequate to avoid overlap of the spectra at the new location with the previously sagged regions. Initial plate scale estimates suggested that this was equivalent to a 3.5” shift, and the ray-trace models suggested that the decrease in spectral resolution with this shift would be 10 to 15%, which was judged to be acceptable. However, there were still a number of complications associated with translating the desired pixel shift into the required operational parameters. One issue is uncertainties in the “Y-walk” correction. The apparent position of events on the COS FUV detector varies with gain due to the geometry of the micro-channel plate pores (Sahnow et al. 2011). For planning and analysis of this program, it was assumed that events are shifted downward on the detector below their true position by 0.45 pixels for each unit decrease in the event’s gain. Inaccuracies in the Y-walk correction may lead to the mis-correction of the position of events. An improved correction is being developed by Kriss et al. (2013a). Another difficulty at the time this program was planned was that the effective plate scales for each grating were not known with sufficient accuracy. The available data gave inconsistent results, perhaps due to inadequate Y-walk corrections or the variations in the shape of the spectral cross-dispersion profile with position. Each of the three COS FUV gratings, (G160M, G130M, and G140L), falls at a slightly different position in the cross-dispersion direction, and, for consistently centered point-sources, only the wings of the cross-dispersion profiles overlap. This had the advantage of spreading the photon accumulation over a wider area of the detector, and to some extent mitigates the effect of microchannel plate damage resulting from use of each grating on observations done with the other gratings. Early in the lifetime of COS, target acquisitions for many of the calibration programs, including the routine sensitivity monitors, were done using FUV dispersed light target acquisitions. It was only subsequently realized that the increasing y-walk resulting from the on-going gain-sag meant that for observations that relied on such target acquisitions, the location of the spectra on the detector were slowly shifting over time. For our analysis, we would ideally like to compare our measured spectral locations with reference spectra taken at the original lifetime position prior to the onset of significant gain sag. To ensure that the new and the old spectra are consistently positioned, we will choose reference spectra that used NUV imaging target acquisitions done with the bright object aperture (BOA); BOA imaging acquisitions were also done for the visits in this program. This will minimize spurious shifts due to gain-sag and walk effects. The routine sensitivity monitor observations were switched to using NUV BOA imaging acquisitions in March of 2010, and so for our comparison spectra we will select G140L 1280 and G130M 1291 observations from program 11897 taken on March 3, 2010 and G160M 1600 observations from the same program taken on March 31, 2010. The cross-dispersion profiles for these observations, collapsed over the dispersion direction, are shown in Figure 2. In addition to the external source spectrum, the G130M FUVB 1291 image also shows a strong full aperture Ly-α emission line that is locally much brighter and wider than the stellar spectrum. We omitted this geo-coronal emission from the Instrument Science Report COS 2013-03(v1) Page 5 collapsed cross-dispersion profile for the G130M shown in Figure 2, and instead.separately show the profile for the core of the geocoronal line to illustrate how it overlaps with the profiles of the other gratings. The detailed shape of the crossdispersion profile for each grating varies significantly with wavelength and CENWAVE setting, even within a single segment, but the collapsed profiles used here will serve as a reference for measuring the overall shift of each grating as a function of the adopted pointing offset in the cross-dispersion position. Figure 2: The collapsed cross-dispersion profiles at the original lifetime position for selected sensitivity monitor observations taken in March 2010. Results for FUVA are shown on the left and for FUVB on the right. The FUVB G130M cross-dispersion profile shown here excludes the geocoronal Lyα photons. (For this setting, XCORR values between pixels 9000 and 9170). The cross-dispersion profile for the core of the Lyα line in this exposure (XCORR values between 9050 and 9125) is shown separately by the dotted line. 2. Goals and Design of 12795 2.1 Goals, Design, and Execution of Visit 1 The primary purpose of visit 1 was to take one spectrum with each grating at the new lifetime position in order to evaluate whether the placement of the spectra on the detector was acceptable. It was also desirable to provide estimates of the local plate scales for the FUV gratings at the new position and to measure the offsets between the science and wavecal spectra. These latter quantities are needed during execution of PEAKXD acquisitions to determine how to center the target and these vary significantly with grating, position, and gain level. Because FENA 2 executed prior to FENA4, measurements of these quantities from FENA2 were used as initial estimates for the FENA4 tests. Exploratory work for choosing the new lifetime position (Oliveira et al 2013b) had concluded that the new lifetime position should be shifted from the old one by about 35 pixels. Originally it was assumed that this was equivalent to a +3.5” shift, but by the time that the next COS lifetime position was selected, it was already appreciated that the nominal FUV cross-dispersion plate scales of 10 pixels/arc-sec that had been adopted during the initial discussions were not correct. Analysis of data taken at an offset of +3” as part of the exploratory program 12678 (Sahnow et al. 2013), suggested that for G130M at least, the actual plate scale was closer to 11.905 Instrument Science Report COS 2013-03(v1) Page 6 pixels/arc-sec. Shifting the spectrum by 35 pixels would then require a spacecraft pointing offset of only +2.94”, rather than the nominal +3.5” that had been previously discussed. However, at the time visit 1 was being planned, not only were there uncertainties in the geometric and walk corrections that made interpretation of previous data difficult, but the strategy for managing detector gain at the new position was still under development, and the initial gain levels and starting HV levels were not yet set. In addition, only G130M data was available near the new position. Using as small an offset as possible would improve spectrum resolution and leave more room for subsequent lifetime positions, but this would allow the data to be more affected by gain sag from the original position. It was decided that visit 1 should obtain data for all three gratings with both a +2.94” offset (~35 pixel shift) and a 3.74” offset (~45 pixel shift). A final decision on the exact placement of the lifetime position could then be deferred until after the visit 1 data were examined and the effects of gain sag could be directly determined. As will be discussed below, ultimately it was decided to adopt a +3.5” offset (~41 pixels) for the new lifetime position, and that value was used for the visit 2 exposures. For this visit, the DB WD standard star WD0308-565 was selected as it has a flatter spectral energy distribution than most other WD standards, allowing for more uniform exposure times and S/N vs. wavelength for the different gratings. The target acquisition was done using an ACQ/IMAGE exposure with the BOA aperture and MIRRORA. At the time of execution, the default HV levels for the COS FUV detector at the initial lifetime position were 169 steps for FUVA and 175 steps for FUVB. The final HV values that would eventually be adopted for use at the second lifetime position had not yet been determined. While the 169/175 settings gave the best results at the gain sagged region of the detector near the original position, using that setting at the higher gain regions near the new lifetime position would result in misregistered events due to the excessively high gain, (Sahnow et al. 2011). To avoid this artifact, the FUVB voltage was lowered to 167 for all exposures in both visits of this program. Note that the reference exposures discussed above from program 11897 in March 2010 also used the 169/167 step HV settings for FUVA and FUVB. Note that there was no need for FENA2 to use the exact voltages that would be adopted for operations at the second lifetime position. This allowed FENA2 to proceed without waiting for results from any other enablement programs, such as the HV sweep done as part of FENA1 (Kriss et al. 2013b). A single G130M observation was taken at the original position, and then one exposure with each of the G130M, G160M, and G140L gratings was taken at a crossdispersion offset of +2.94” (targeting a shift of ~ 35 pixels), and then again at +3.78” (targeting ~ 45 pixels). The local plate scale estimates provided by using each grating at two positions were intended to not only provide the initial estimates for use by FENA4, but also to allow fine adjustment of the exact offset. The inputs for the ALIGN/APER command used to move the aperture block to nonstandard positions are specified in units of motor steps relative to the default position. For this program this default aperture position was the standard Lifetime Position 1 Prime Science Aperture (PSA) position. It was assumed that commanding an offset of −21 XAPER steps is equivalent to moving the aperture +1” on the sky in Instrument Science Report COS 2013-03(v1) Page 7 the POSTARG2 direction, while +21 YAPER steps is equivalent to moving the aperture +1” in the dispersion direction. Note that this aperture coordinate system uses the detector coordinates that were devised early in instrument development and these differ from those used in standard CALCOS pipeline products, where the +x is in the direction of increasing wavelength and +y is in the +POSTARG2 direction. This change was made to allow the APT interface and the pipeline products for COS to be consistent with those of STIS and the other HST instruments. All exposures at nonstandard aperture positions used the special command ELNOAPMAIN to avoid automatic relocation of the aperture back to its normal PSA position. To improve the S/N for the measurement of the WCA aperture locations, the lamp flash times were increased to 36 s for the G130M and G160M and to 21 s for the G140L. This resulted in the sequence of seven FUV exposures shown in Table 1. Table 1: FUV exposures in visit 1 of 12795. The target for visit 1 exposures was the DB star WD 0308−0565. The visit executed on February 21, 2012. Setting Exposure Exp. Time POSTARG2 Flash XAPER name (s) (“) (s) G130M 1291 LBX401H9Q 54 0 36 0 G130M 1291 LBX401HBQ 54 +2.94 36 −62 G160M 1600 LBX401HDQ 163 +2.94 36 −62 G140L 1280 LBX401HFQ 40 +2.94 21 −62 G130M 1291 LBX401HHQ 54 +3.78 36 −79 G160M 1600 LBX401HJQ 163 +3.78 36 −79 G140L 1280 LBX401HLQ 40 +3.78 21 −79 2.2 Goals, Design, and Execution of Visit 2 The primary purpose of visit 2 was to verify the aperture block centering at the new lifetime position in both the dispersion and cross-dispersion direction. This visit also served to confirm the final revisions to the cross-dispersion offset defining the new lifetime position to be at +3.5”, (about +41 pixels), that were made after the initial evaluation of the visit 1 data (see discussion in section 3.1 below). The aperture center in principle could be determined either by keeping the aperture at a fixed position and scanning the HST pointing via POSTARGs, or by keeping the pointing fixed and scanning the COS PSA aperture across the target. Moving the spacecraft has the disadvantage that the location of the spectrum on the detector changes at each POSTARG position, and the detected spectrum would then vary due to changes in PSF shape, shifts with respect to the detector fixed pattern noise, and for changes in the dispersion direction, variation in the exact wavelength range falling on each segment. As a result, it was decided to keep the target fixed while moving the aperture to better isolate the vignetting caused by the aperture edges. During these aperture scans, POSTARG1 was held at 0 and POSTARG2 at a value of +3.5 arcseconds. The vignetting of an external target as a function of the target’s position within the COS aperture was previously studied by Ghavamian et al. 2010. For the G130M grating, the throughput as a function of centering offset is relatively flat in the central ±0.4” and falls off to about 40% of the central value at the 1.25” aperture radius. For this program we chose a 13 point pattern of aperture motions extending to ±29 motor Instrument Science Report COS 2013-03(v1) Page 8 steps or 1.381” from the central positions. Spacings of 4 to 6 motor steps between exposures were used, with slightly coarser centering being used near the center and in the far wings. The resulting sequence of exposures is documented in Table 2. The images of geocoronal lines will move on the detector as the aperture moves, and these are expected to vary in brightness as the sight line with respect to the earth changes and as HST moves between orbital day and night. To allow easier removal of these effects, visit 2 used the sensitivity monitor target WD 0947+0857 as the relatively broad stellar Lyα line in this DA white dwarf allows easier isolation of the stellar and geocoronal flux. For these observations the G130M 1309 CENWAVE setting was used as it puts the brightest of the geocoronal O I lines in the gap between the detector segments. Table 2: The FUV Exposures in visit 2 of 12795. The target for visit 2 exposures was the DA star WD 0947+857. This visit executed on March 12, 2012. For clarity, in this table values repeated from the previous row are in a grey font. The LAMP column lists which Pt/Ne lamp and current setting was used for each exposure. XAPER and YAPER give the offsets, for the cross-dispersion and dispersion directions respectively, of the aperture block in motor steps from their default values for the first lifetime position. The quantities δ2 and δ1 give the equivalent offset of the aperture block in arc-seconds, assuming a plate scale of 21 motor steps per arc-second. The sign convention on δ1 and δ2 is chosen so setting POSTARG1 to δ1 or adding δ2 to the actual POSTARG2 parameter would give the offset that would have allowed the target to follow the aperture during the sweep. XAPER YAPER Setting Exposure Exp POSTARG2 LAMP δ2 δ1 counts (“) Name (s) (“) (“) A G130M 1309 LBX402BGQ 25 0.0 LINE1, MED 0 0 0 0 Cross-‐dispersion Aperture Scan G130M 1309 LBX402BIQ 50 +3.5 LINE1, MED -‐102 +1.381 0 0 45836 G130M 1309 LBX402BKQ 50 +3.5 LINE1, MED -‐96 +1.095 0 0 102432 G130M 1309 LBX402BMQ 50 +3.5 LINE1, MED -‐91 +0.857 0 0 129156 G130M 1309 LBX403BOQ 50 +3.5 LINE1, MED -‐87 +0.667 0 0 146464 G130M 1309 LBX402BQQ 50 +3.5 LINE1, MED -‐83 +0.476 0 0 161434 G130M 1309 LBX402BSQ 50 +3.5 LINE1, MED -‐79 +0.285 0 0 165072 G130M 1309 LBX402BUQ 50 +3.5 LINE1, MED -‐73 0 0 0 166609 G130M 1309 LBX402BWQ 50 +3.5 LINE1, MED -‐67 -‐0.285 0 0 165094 G130M 1309 LBX402BYQ 50 +3.5 LINE1, MED -‐63 -‐0.476 0 0 159852 G130M 1309 LBX402C0Q 50 +3.5 LINE1, MED -‐59 -‐0.667 0 0 142244 G130M 1309 LBX402C2Q 50 +3.5 LINE1, MED -‐55 -‐0.857 0 0 124864 G130M 1309 LBX402C7Q 50 +3.5 LINE1, MED -‐50 -‐1.095 0 0 99958 G130M 1309 LBX402CFQ 50 +3.5 LINE1, MED -‐44 -‐1.381 0 0 42629 Along Dispersion Aperture Scan G130M 1309 LBX402CHQ 50 +3.5 LINE2, LOW -‐73 0 +29 +1.381 66782 G130M 1309 LBX402CJQ 50 +3.5 LINE2, LOW -‐73 0 +23 +1.095 114745 G130M 1309 LBX402CLQ 50 +3.5 LINE2, LOW -‐73 0 +18 +0.857 135332 G130M 1309 LBX402CNQ 50 +3.5 LINE2, LOW -‐73 0 +14 +0.667 150439 G130M 1309 LBX402CPQ 50 +3.5 LINE2, LOW -‐73 0 +10 +0.476 163334 G130M 1309 LBX402CRQ 50 +3.5 LINE2, LOW -‐73 0 +6 +0.285 166219 G130M 1309 LBX402CTQ 50 +3.5 LINE1, MED -‐73 0 0 0 166936 G130M 1309 LBX402CVQ 50 +3.5 LINE2, LOW -‐73 0 -‐6 -‐0.285 164479 G130M 1309 LBX402CXQ 50 +3.5 LINE2, LOW -‐73 0 -‐10 -‐0.476 154292 G130M 1309 LBX402CZQ 50 +3.5 LINE2, LOW -‐73 0 -‐14 -‐0.667 141511 G130M 1309 LBX402D1Q 50 +3.5 LINE2, LOW -‐73 0 -‐18 -‐0.857 127658 G130M 1309 LBX402D3Q 50 +3.5 LINE2, LOW -‐73 0 -‐23 -‐1.095 103090 G130M 1309 LBX402D5Q 50 +3.5 LINE2, LOW -‐73 0 -‐29 -‐1.381 51280 Instrument Science Report COS 2013-03(v1) Page 9 counts B 69013 148572 185654 211382 233375 239643 242522 241906 233356 209185 186096 151277 75807 95876 165608 195515 217498 234763 238045 241133 236836 223733 204080 185263 149725 74293 Confirmation of final G160M Position G160M 1600 LBX402D7Q 50 +3.5 LINE1, MED -‐73 0 0 0 It has been shown by Oliveira et al. (2013a) that for sufficiently large offsets of the aperture block in the cross-dispersion direction lamp light can leak through the flatfield calibration aperture (FCA). This can result in excessive illumination from ordinary wavelength calibration lamp flashes. To guard against the possibility that a similar light leak might result from offsets of the aperture block in the dispersion direction, for all exposures with non-zero YAPER values, the LINE2 lamp at a LOW current setting was used with 8 s lamp flashes with 20 s spacing in place of the standard LINE1 at the MEDIUM setting. This would reduce the expected count rate, even in the event of a substantial light leak, to safe levels. These lamp flashes are too faint to use for ordinary wavelength calibration. For those exposures, the offsets found for YAPER=0 can be used to define the wavelength scale zero point adjustment for spectral extraction. Since HST and the COS detector and optics are being held in a fixed position with respect to the external target, there is no need to adjust this wavelength zero-point for the motion of the aperture block. In any case, as our analysis will be done using XCORR and YCORR coordinates, the zero point of the wavelength scale is unimportant for our purposes. Those visit 2 exposures done with YAPER=0 (i.e., no dispersion direction offset) used the standard lamp LINE1 and medium current setting, and, as in visit 1, the flash duration was set to 36 s. As we will discuss below, the overlap of the G160M spectral location with the gain sagged regions at the original lifetime position was the critical factor that controlled the minimum distance between the old and the new position, so a single G160M spectrum at +3.5” was included in visit 2 as a final verification of this overlap at the +3.5” lifetime position. 3. Results 3.1 Placement of the New Lifetime Position Since the COS FUV cross-dispersion profiles can be asymmetric, it is necessary to clearly define the algorithm used for measuring the profile center. Unless otherwise noted, we will use the YCORR locations of events that have been corrected for thermal and geometric distortion and for Y-walk effects, but not for Doppler effects, (which are small and apply only in the dispersion direction), or corrections determined from the wavelength calibration exposure. For the external spectrum profiles, we will adopt a flux weighted centroid as a measure of the overall position in the cross-dispersion. YC = Σ CYYCORR/ΣCY, where CY is the total number of counts between YCORR−1/2 pixel and YCORR+1/2 pixel, and the sum is taken over the region that includes significant contribution from the spectrum of the external source. As a measure of the upper and lower bounds of a cross-dispersion profile, we will use the Y positions at the 5th (Y05) and 95th (Y95) percentile of the cumulative profile in the cross-dispersion direction. The wavecal spectrum is fainter and has a more symmetrical cross-dispersion than the external spectrum, so we will define the Y location of the wavecal spectrum, WC, as the center of the best fit Gaussian. Instrument Science Report COS 2013-03(v1) Page 10 These position measures are given in Table 3 for those FUV spectral images from this program where the aperture was centered on the target. This includes all exposures from visit 1 plus the exposures from visit 2 taken with the aperture block at the XAPER=-73 and YAPER=0 position. We also include sensitivity monitor observations from March 2010 taken at the CENWAVE values used in visits 1 and 2. For each of the exposures from 12795, we also list the shift relative to the location in the March 2010 reference image for both the centroid of the external spectrum, δYC = YC(new) – YC(reference) and wavecal location, δWC = YC(new) – YC(reference). These numbers show that relative to the 2010 exposures, the +2.94” shift moved the external spectra by about 34 YCORR pixels on the detector, the +3.5” shift by ~ 41 pixels, and +3.78” by ~ 43 pixels. In each case the wavecal spectrum moved by 1 to 3 pixels less than the external spectrum. This discrepancy may be due to errors in the walk correction, inaccuracies in the geometric distortion, or differences in the illumination angle of lamp vs. external light. Any miscentering of the external target in the aperture would also contribute to the offset between the target and wavecal spectra. Table 3 Cross-dispersion locations of external spectra (YC) and wavelength calibration spectra (WC) on the detector. For 12795 visit 2 we only include here exposures taken with the aperture centered on the target. Grating Seg CENWAVE POSTARG2 YC Y05 Y95 WC (“) (pixels) (pixels) (pixels) (pixels) Reference exposures from program 11897, March 2010, FUVA, target WD0947+0857 G140L A 1230 0 493.501 485.924 499.096 595.412 G130M A 1309 0 485.527 478.589 491.779 587.443 G130M A 1291 0 486.192 477.664 493.898 587.994 G160M A 1600 0 478.788 472.964 483.306 581.381 V01 12795, FUVA, target WD0308−565 G130M A 1291 0 484.963 476.766 492.556 587.823 G130M A 1291 2.94 520.537 512.367 528.187 620.795 G160M A 1600 2.94 512.681 507.766 517.162 613.201 G140L A 1280 2.94 527.309 520.651 532.515 628.005 G130M A 1291 3.78 529.615 521.866 536.898 629.241 G160M A 1600 3.78 521.803 516.733 526.433 622.069 G140L A 1280 3.78 535.846 529.433 541.643 636.424 V02 12795, FUVA, target WD0947+0857 G130M A 1309 0 483.805 477.129 490.133 587.029 G130M A 1309 3.50 526.069 519.348 532.034 625.624 G130M A 1309 3.50 526.530 519.693 532.582 626.099 G160M A 1600 3.50 519.505 514.348 524.487 619.641 Reference exposures from program 11897, March 2010, FUVB, target WD0947+0857 G140L B 1230 0 551.739 540.292 561.489 ... G130M B 1309 0 545.652 538.032 551.908 647.921 G130M B 1291 0 546.239 536.949 553.951 648.270 G160M B 1600 0 537.436 532.433 541.868 640.823 V01 12795, FUVB, target WD0308−565 G130M B 1291 0 546.171 536.354 554.508 647.928 G130M B 1291 2.94 580.856 571.458 589.427 681.176 G160M B 1600 2.94 571.922 566.783 576.101 672.920 G140L B 1280 2.94 585.438 575.396 595.360 ... G130M B 1291 3.78 589.552 581.281 597.435 689.780 G160M B 1600 3.78 581.072 576.094 585.162 681.829 G140L B 1280 3.78 594.211 584.473 605.062 ... V02 12795, FUVB, target WD0947+0857 G130M B 1309 0 544.977 537.315 551.380 647.492 G130M B 1309 3.50 585.952 579.046 591.970 686.565 G130M B 1309 3.50 586.251 579.205 592.497 686.910 G160M B 1600 3.50 578.909 573.957 583.002 679.537 Instrument Science Report COS 2013-03(v1) Page 11 Exposure Name δYC (pixels) δWC (pixels) LBB908N5Q LBB908NEQ LBB908NGQ LBB9X1MSQ 0 0 0 0 0 0 0 0 LBX401H9Q LBX401HBQ LBX401HDQ LBX401HFQ LBX401HHQ LBX401HJQ LBX401HLQ −1.23 34.35 33.89 33.81 43.42 43.01 42.35 −0.17 32.80 31.82 32.59 41.24 40.69 41.01 LBX402BGQ LBX402BUQ LBX402CTQ LBX402D7Q −1.72 40.54 41.00 40.72 −0.41 38.18 38.66 38.26 LBB908N5Q LBB908NEQ LBB908NGQ LBB9X1MSQ 0 0 0 0 ... 0 0 0 LBX401H9Q LBX401HBQ LBX401HDQ LBX401HFQ LBX401HHQ LBX401HJQ LBX401HLQ −0.07 34.62 34.49 33.70 43.31 43.64 42.47 −0.34 32.91 32.10 ... 41.51 41.01 ... LBX402BGQ LBX402BUQ LBX402CTQ LBX402D7Q −0.67 40.33 40.60 41.47 −0.43 38.64 38.99 38.71 Collapsed cross-dispersion profiles for the +2.94” and +3.78” shifts of visit 1, and the +3.5” shift used in visit 2 are compared with the equivalent reference exposures from March 2010 in Figure 3, while in Figure 4 we show the modal gain values as a function of the XCORR position for G130M and G160M exposures from visit 1. The region over which modal gain shown in these plots is measured is determined by sorting the events within ±50 YCORR pixels of each spectrum’s center by their XCORR location and dividing these events into bins along the dispersion direction so that each XCORR bin includes the same number of events. The pulse height distribution within each bin was then fit with a Gaussian function; the gain at the peak of the Gaussian fit is taken as the measure of the modal gain. Figure 3: The collapsed cross-dispersion profiles at each of the offset positions used in 12795 (solid lines) are compared with cross-dispersion profiles at the original lifetime position (broken lines). Line colors have the same meaning as in Figure 2, the orange lines show the G160M 1600 profile, blue lines show G130M, and green G140L 1280. (No G140L exposures were taken at the +3.5” position used for visit 2 of 12795). The black line shows the cross-dispersion profile of the Instrument Science Report COS 2013-03(v1) Page 12 core of the G130M geocoronal Lyα line. Note that the G130M observations in the middle panels above (+3.5” and the comparison) were taken during visit 2 and used the 1309 CENWAVE position, while those in the other panels show observations taken during visit 1 that used the 1291 CENWAVE setting for the G130M exposures. Figure 4: Modal gain of the events observed in the G130M 1291, (upper panels), and G160M 1600 (lower panels) exposures from visit 1 of program 12795 are shown as a function of XCORR pixel, as is the gain for the comparison exposures from March 2010. Note that the gain of the G160M spectra on the B segment, (green curve), taken at the +2.94” spectrum are noticeably impacted at several locations between pixels 8000 and 10000 by the upper edge of the G130M gain-sag holes from the original position. Dips at the location of the G140L gain sag holes on the FUVA are also visible in the G160M spectrum at the +2.94” position (green curve) at two locations near pixels 2000-2500. The very deep feature on FUVA near XCORR = 8000 corresponds to a low gain region that has been present since before launch. This corresponds to the dark line seen in the middle of the upper part of the upper panel in Figure 1 at the same XCORR location. It is apparent that for the +2.94” shift of about 34 pixels, the G160M spectrum is significantly impacted by the gain sag holes of the original lifetime position on both the FUVA and FUVB segments. Figure 5 shows a more detailed view of the G160M modal gain vs. pixel position in the dispersion direction for the spectra taken at the +2.94” and +3.78” positions used in visit 1, as well as at the +3.50” location used for visit 2. In Figure 6, we show the locations of these spectra overlaid on images of gain maps constructed from the 12676 data, and it is again clear that at the +2.94” position the wings of the old gain sag holes substantially impact the modal gain of the extracted spectrum. At the +3.5” position, there is still some measurable impact on the G160M spectra, but only by about 1 PHA bin or less over very localized regions. Instrument Science Report COS 2013-03(v1) Page 13 Figure 5: Details of the modal gain vs. position along the dispersion direction for the G160M exposures taken in this program. Note that most, but not quite all, of the impact of the preexisting Lyα holes on the modal gain seen at the +2.94” position, (green curves), can be avoided by moving to the +3.5” position, (blue curves). Instrument Science Report COS 2013-03(v1) Page 14 Figure 6: Details of the FUVA (upper) and B (lower) gain maps near the Lyα holes are compared with the locations spanned by G160M spectra at a variety of offset positions. The solid lines show the YC central position for POSTARGs of 2.94”, 3.5”, and 3.78” (green, blue, and orange solid lines respectively), while the short dashed lines show the Y05 positions and the long dashed lines the Y95 positions that mark the upper and lower bounds containing 90% of the total encircled energy. This image uses a linear brightness scale running from PHA=0 (black) to PHA=18 (white). Note that the lower edge of the +2.94” POSTARG region is significantly impacted by the gain sag holes on both FUVA and FUVB. Initial estimates based on the deuterium lamp flats had suggested that a +35 pixel shift was adequate to meet the lifetime requirements at the new position. These estimates had been done by shifting 20 pixel high extraction boxes to a variety of positions relative to the default spectral positions tabulated in the XTRACTAB reference file. For G160M on FUVB this tabulated location is about 2 pixels higher on the detector than we found for the actual location in our March 2010 reference exposures. For a 35 Instrument Science Report COS 2013-03(v1) Page 15 pixel shift, the test extraction region on FUVB would still have spanned YCORR pixels values from about 564 through 584, which extends below the Y05 lower edge value of 566.783 found for the G160M 1600 position with a +2.94” POSTARG (green dashed line in lower panel of Figure 6). So the overlap of the older gain sag region with the new G160M location at a shift of +35 pixels was properly treated in the analysis of Massa et al. (2013). However, this initial analysis had been done assuming that operations would begin at the 169/167 HV levels for FUVA and FUVB respectively, and that it was only necessary to keep the modal gain above a value of 3 as long as possible. Subsequently it was decided that starting at lower initial voltage levels would give substantial lifetime benefits, and also that noticeable effects on data quality can begin once the modal gain drops below 5. This made smaller starting gain values less acceptable than had been initially assumed. After review of the data from visit 1 and detailed comparison with the gain maps from program 12676, it was decided that a position of +3.5” (a shift of about 41 pixels in the cross-dispersion direction) provided the best compromise between avoidance of the gain sag holes while still maximizing spectral resolution and preserving fresh area near the top of the detector for a subsequent lifetime position. This position was then adopted as the baseline for visit 2. The measurements listed in Table 3 can also be used to estimate values for the crossdispersion plate scale. Several measures of this are compared in Table 4. From the data within visit 1, we can derive a measure of the local plate scale by comparing the change in the vertical position, δYC, between the two POSTARG2 positions used for each grating near the 2nd lifetime position. We can also compare these positions with the positions measured for the reference exposures to obtain an estimate of the plate scale between the 2nd and 1st lifetime positions. The latter values give significantly smaller pixel sizes than the former, suggesting that the current geometric distortion solution does not correct pixels to a uniform spacing on the sky, although changes in the optical distortion of each grating’s PSF as a function of position will also play a role. Table 4: Values for the COS FUV cross-dispersion plate scale are compared. Results given here are derived from geometrically, thermally, and y-walk corrected positions. YCORR Plate scale results from this program Grating/segment Local scale 2.94 – Ref 3.50 – Ref 3.78 – Ref (3.78 – 2.94) Visit 1 Visit 2 Visit 1 (“/pixel) (“/pixel) (“/pixel) (“/pixel) G130M/FUVA 10.81 11.68 11.65 11.49 G160M/FUVA 10.93 11.51 11.63 11.38 G140L/FUVA 10.16 11.50 … 11.20 G130M/FUVB 10.35 11.77 11.56 11.46 G160M/FUVB 10.89 11.73 11.85 11.54 G140L/FUVB 10.44 11.46 … 11.23 Instrument Science Report COS 2013-03(v1) Page 16 Plate scale results from ISR 2010-09 at 1st Lifetime Position Grating Published values Within ±0.5” (“/pixel) (“/pixel) G130M 9.72 11.88 G160M 11.33 11.44 G140L 10.97 11.95 We can also compare with the plate scale measures of Ghavamian et al (2010) for the original lifetime position. His values differ significantly, especially for the G130M. These were, however, based on measurements of individual emission lines from an external source moved using POSTARG commands while the PSA aperture was held at a fixed position. As the external source approaches the edge of the aperture, vignetting from the aperture edge will cut off one wing of the point spread function, leading to a reduced apparent motion of the target on the detector. If we instead consider only the Y position values from Ghavamian et al. (2010) for positions within 0.5” of the aperture center, results are closer to the estimates from our new data. The most appropriate plate scale values to use will depend on the application. For the target acquisition parameters used for PEAKXD exposures, the centroid of the position needs to be measured in uncorrected RAW coordinates using the same algorithm that the flight software uses, and the effects of the edge vignetting on the effective plate scale need to be considered. This is discussed in greater detail by Penton et al. (2013). Observers interpreting exposures of multiple or extended objects within the aperture will also need to evaluate how such edge effects will bias their interpretations. For planning the location of future lifetime positions, the data taken during the lifetime exploratory program 12678, “COS/FUV Characterization of Optical Effects”, where the aperture was moved to follow the target, will remain the best resource. 3.2 Aperture Centering at the New Lifetime Position As described in section 2.2, for determining the positioning of the source in the PSA at the new lifetime position, the pointing of HST was held fixed while the aperture block was scanned across the detector. The relative throughput was then estimated by summing all counts over the region of the detector containing counts from the source while taking care to keep the geocoronal Lyα photons and lamp light masked out at all positions. This was done by using the subarray [1135:1526,45:590] for XCORR and YCORR values on FUVA, and the two subarrays [800:6980,500:630] and [7285:15045,500:630] on FUVB. These choices exclude the lamp light at all positions in the cross-dispersion aperture scans, and the geocoronal Lyα light from the G130M 1309 CENWAVE setting used here at all positions of the along-dispersion scan. The number of counts measured for each of the exposures in these subarrays are listed in the last two columns of Table 2, and their relative values, normalized to the counts measured at the central position, are plotted in Figure 7 and Figure 8. Comparison with dark exposures taken the same month as these observations shows that detector background contributions to these subarrays during a 50 s observation should be between 230 to 415 counts on FUVA, and 160 to 360 counts on FUVB. As this is Instrument Science Report COS 2013-03(v1) Page 17 never more than about 1% of the observed counts from the external source, this small background contribution has been neglected. Figure 7: The relative throughput of the COS PSA aperture as a function of the aperture block offset in the cross-dispersion direction is shown for observations done as part of visit 2 of program 12795. The dashed line shows measures for FUVA, and the solid for FUVB. Positions marking the nominal center, +/-0.4”, and +/-1.25” are marked with vertical dotted lines assuming a plate scale of 21 XAPER steps per arc-second. Figure 8: The equivalent of Figure 7 for the aperture block scan in the direction along the spectral dispersion. The profile in the cross-dispersion direction appears well centered on the nominal −73 XAPER position relative to the original position (Figure 7). However, the best Instrument Science Report COS 2013-03(v1) Page 18 centering found by the dispersion direction scan appears to be significantly displaced in the positive YAPER direction (Figure 8). The definition of the “best” center for an asymmetric profile is influenced by a number of factors. In addition to simply maximizing the throughout, it is also desirable to avoid asymmetric clipping of the line-spread function, and to also ensure consistency between the pointing resulting from NUV imaging acquisitions, such as was done for this visit, with the results of dispersed light acquisitions. COS PEAKD acquisitions in particular function by measuring the flux at a variety of dwell points to find the aperture center, and this is more sensitive to throughput variations near the aperture edges than to the smaller throughput variations near the aperture center. To quantify the centering, we will first calculate the bisectors of the curves shown in Figure 7 and Figure 8, (that is the mid-point between points of the same height on opposite sides of the curve). Values between the measured points were linearly interpolated. We then compare the position of this bisector to the half-width of the curve at that height (Figure 9). For the cross-dispersion scan, the offsets are mostly less than 0.02”, show little trend as a function of the profile width, and appear to be of opposite sign for the different segments. Since 1 motor step of the aperture block corresponds to about 0.05”, this centering is about as good as could have been expected. For the scan along the dispersion direction, the results for the two segments are in close agreement, but there appears to be a somewhat larger offset determined from the center of the profile than from the wings. Recommended dwell point spacing for COS PEAKD acquisitions are 1.3” when 3 positions are used and 0.9” for the preferred 5-position sequence. To allow the centering produced by COS FUV PEAKXD observations at the new lifetime position to be consistent with the centering produced by COS NUV acquisitions, an offset of 0.05” was adopted as our best determination of the aperture centering in the dispersion direction. Figure 9: Offsets of bisectors of the aperture throughput curves as a function of the half-width of the throughput curves at the same height as the bisector points are shown for each segment for the aperture scan perpendicular to the dispersion axis (left) and parallel (right). This shows the offset in the mid-point found for different sample spacings. Since the desired offset in the dispersion direction is just about 1 aperture motor step, the centering at the new lifetime position could have been adjusted by moving the aperture block. However, this would have required a mechanism motion in the dispersion direction when switching between NUV and FUV exposures. Currently all COS aperture positions used operationally differ only in the cross-dispersion direction and movements in the dispersion direction are only needed for special engineering Instrument Science Report COS 2013-03(v1) Page 19 tests. To avoid the need for such moves and the associated overheads during routine operations, it was decided instead to correct for the dispersion offset in the SIAF definition of the new lifetime position, i.e., by slightly repointing the telescope to center the target in the dispersion direction. So for the second COS FUV lifetime position, the FUV aperture positions need to be shifted from their original position by the equivalent of a −0.05” POSTARG1 in the dispersion direction and by a +3.5” POSTARG2 in the cross-dispersion direction. This changes the FUV aperture positions in the V2, V3 coordinate system used in the Science Instrument Aperture File (SIAF) from the original values of 232.6560, −237.4450 to new values of +235.0810, −234.9400. The NUV aperture continues to use the original position. Note that due to an error in the original analysis, the SIAF update was initially delivered with the wrong sign for the shift in the dispersion direction (i.e., +0.05, +3.50 instead of −0.05, +3.50). Fortunately, as it was recognized that sign errors of this kind can be easy to make and difficult to track down, PEAKD exposure sequences included in visit 2 of FENA4 (program 12797) were examined to verify the aperture centering in the dispersion direction. These checks revealed the sign error and once the source of the error was properly understood, a corrected SIAF file was delivered and then tested in visit 5 of FENA4 prior to the start of routine operations at the new lifetime position. In this report we have presented only the corrected analysis. In Table 5 and Table 6 we list the relative throughput as a function of the equivalent POSTARG that would move the target towards the same edge of the aperture as did the ALIGN/APER command. This has the opposite sign from the POSTARG that would follow the aperture motion. For the dispersion direction case, we include the 0.05” offset found above. Observers may find these tables useful for estimating throughput losses for mis-centered targets. Table 5: Fractional throughput as a function of target offset perpendicular to the dispersion direction. Note that the sign convention given here for the POSTARG value is the reverse of that given in Table 2, as there we were specifying the POSTARG needed to follow a moving aperture, while here we are giving the POSTARG necessary to put the target off-center in a fixed aperture. POSTARG2 (“) -1.381 -1.095 -0.857 -0.667 -0.476 -0.286 +0.000 +0.286 +0.476 +0.667 +0.857 +1.095 +1.381 XAPER -29 -23 -18 -14 -10 -6 0 +6 +10 +14 +18 +23 +29 Relative throughput FUVA FUVB 0.275 0.285 0.615 0.613 0.775 0.766 0.879 0.872 0.969 0.962 0.991 0.988 1.000 1.000 0.991 0.997 0.959 0.962 0.854 0.863 0.749 0.767 0.600 0.624 0.256 0.313 Instrument Science Report COS 2013-03(v1) Page 20 Table 6: Fractional throughput as a function of target offset along the dispersion direction. POSTARG1 (“) -1.331 -1.045 -0.807 -0.617 -0.426 -0.236 +0.050 +0.336 +0.526 +0.717 +0.907 +1.145 +1.431 YAPER +29 +23 +18 +14 +10 +6 0 -6 -10 -14 -18 -23 -29 Relative Throughput FUVA FUVB 0.400 0.398 0.687 0.687 0.811 0.811 0.901 0.902 0.978 0.974 0.996 0.987 1.000 1.000 0.985 0.982 0.924 0.928 0.848 0.846 0.765 0.768 0.618 0.621 0.307 0.308 In Figure 10 we compare our results from the cross-dispersion scan at the new position with those of the SMOV program 11490 at the original position. The SMOV program used rather different procedures for measuring the variation in aperture throughput as a function of target centering. The aperture centering was adjusted using POSTARGs to adjust the spacecraft pointing rather than by moving the aperture block, and the target used was an emission line source, which concentrated most of the detector counts onto a few limited areas of the detector. These differences made the SMOV program more vulnerable to localized differences in detector throughput and flat fielding and so it is not surprising that the measured throughput curves from 11490 are not as smooth or as symmetrical as those found here. This is particularly noticeable for the central +/−0.5”. Figure 10: Results of the cross-dispersion scan done at the new +3.5” lifetime position (solid lines) in 12795 are compared with a G130M scan of symbiotic star LIN 139 done in 2009 at the original lifetime position as part of program 11490 (dashed lines) and reported in Ghavaniam et al. 2010. This comparison assumes 21 XAPER motor steps are equivalent to a 1” POSTARG. Results for FUVA are shown on the left and FUVB on the right. The vertical dotted lines mark the +/1.25”radius of the COS PSA aperture. 4. Summary The results presented here demonstrate that offsetting the COS FUV Lifetime Position 2 from Lifetime Position 1 by the equivalent of a −0.05, +3.50 POSTARG Instrument Science Report COS 2013-03(v1) Page 21 move while shifting the aperture block by −73 steps in the XAPER direction moves the position of FUV spectra on the detector up by about 41 YCORR pixels in the cross-dispersion direction while keeping the target well centered in the aperture. This yields adequate separation from the gain-sagged regions at the original position, while staying close enough to that position to keep the predicted degradation in resolution within the desired margin of 10 to 15%. Acknowledgements Change History for COS ISR 2013-03 Version 1: 03 April 2013 – Original Document References Ghavamian, P., et al., 2010, Instrument Science Report COS 2010-09(v1), “A description of the External Spectroscopic Performance of the FUV Channel on COS” Kriss, G. et al., 2013a, Instrument Science Report, in preparation, “Characterization of X- and Y-Walk on the COS FUV Detector” Kriss, G. et al., 2013b, Instrument Science Report, in preparation, “Choosing the Operating High Voltage at the COS FUV New Lifetime Position” Massa, D. L., et al., 2013, Instrument Science Report, in preparation, “Characterization of the COS FUV Detectors” Oliveira, C. et al., 2013a, Instrument Science Report COS 2013-02(v1), “COS/FUV Mapping of Stray PtNe Lamp Light Through the FCA” Oliveira, C. et al., 2013b, Instrument Science Report, in preparation, “Initial Exploratory Work for Defining the COS FUV Lifetime Positions” Osten, R., et al., 2013, Technical Instrument Report, in preparation, “Requirements and Preparations for a COS FUV Lifetime Position Move” Penton, S. V., et al., 2013, Instrument Science Report, in preparation, “FENA4: Enabling Target Acquisition At the 2nd COS Lifetime Position” Sahnow, D. J. et al. 2011, "Gain Sag in the FUV detector of the Cosmic Origins Spectrograph", Proc. SPIE 8145, 81450Q Sahnow, D. J. et al. 2013, Instrument Science Report, in preparation, “COS/FUV Characterization of Optical Effects at Potential Lifetime Positions” Instrument Science Report COS 2013-03(v1) Page 22
© Copyright 2026 Paperzz