1 Supplemental Material 2 Additional mass spectrometric methods 3 Instrument configuration 4 All of the measurements described here were made in dual-inlet mode, meaning 5 approximately equal pressures of sample and standard gases are introduced into the 6 bellows of a dual inlet system, and the capillary bleeds and changeover valve block of the 7 Ultra are used to alternately introduce one and the other into the ion source. Typical 8 bellows pressures are ~10 mbar and typical source pressures are ~6x10-8 mbar (both 9 values are background corrected). These values provide a successful compromise 10 between the desire for high ion-beam intensity (so acceptable counting statistics can be 11 reached in a reasonable time) and the need to avoid space-charge effects and uncontrolled, 12 time-varying reactions in the ion source (which are problematic at high source pressures). 13 14 15 Collector configurations The ion beams of interest for our analysis include: 12C2H4+ and 13C12CH4+ (for 16 analysis of ethyl fragments), either 12CH3+ and 13CH3+ (our first target when this method 17 was developed) or 12CH2+ and 13CH2+ (our current preferred target) for methyl fragments, 18 and 12C3H8+ and 12C213CH8+ for the propyl species (i.e., the full molecular ion). These ion 19 beams are measured in three separate acquisitions, each having different Faraday cup 20 configurations. These configurations are summarized in Table 1 and briefly described as 21 follows: 22 23 The MAT 253 Ultra has 7 collector positions, 6 of which are movable, three on either side of the fixed central collector. All collector positions can detect ions using 24 either a Faraday cup, which can be read through an amplifier that varies in gain between 25 107 to 1012, or an electron multiplier. The electron multipliers on the movable collector 26 positions are of the relatively small compact discrete dynode (CDD) design, while that in 27 the central collector position is a larger, conventional secondary electron multiplier 28 (SEM), positioned behind a `retardation lens' (RPQ) for improved abundance sensitivity. 29 Because of the large number of movable collector positions and range of detector 30 sensitivities at each position, it is possible to make three separate measurements of the 31 methyl, ethyl, and propyl fragments simply by changing the magnet current and making 32 minor adjustments of the movable collectors. 33 When 12CH3+ and 13CH3+ are measured, we measure mass 15 on a Faraday cup 34 with a 1012 gain amplifier, on the `L2' collector trolley (Supplemental Figure 1). This 35 collector is positioned so that the high-mass side of the family of mass 15 peaks (i.e., 36 12 37 are excluded. 13CH3+ is simultaneously measured with the SEM detector in the central 38 collector position, with the magnetic field set such that the low-mass side of the family of 39 mass-16 peaks (i.e., 13CH3+) enters the collector and higher-mass, nearly isobaric species 40 (e.g., 12CH2D+ and 12CH4+) are excluded. CH3+) enters the collector, and lower-mass ions (mostly 13CH2+ and a trace of 12CHD+) 41 The alternate approach of measuring CH2+ uses a similar cup configuration: 42 12 43 registered through a 1012 ohm amplifier, and 13CH2+ is measured on the low-mass side of 44 the mass 15 peak, using the center collector SEM. 45 46 CH2+ is measured on the high-mass side of the mass 14 peak using the `L3' Faraday cup When analyzing the ethyl fragment ion, 12C2H4+ is measured on the high-mass side of the mass 28 peak, collected in a Faraday cup read through a 1012 Ohm amplifier 47 on the `L1' collector trolley (Figure 3), and 13C12CH4+ is measured on the low-mass side 48 of the mass 29 peak, using a Faraday cup registered through a 1012 Ohm amplifier on the 49 central (fixed) collector position. 50 The measurement of propyl fragments involves repositioning one of the movable 51 collectors (L1, or cup 4) after making the ethyl fragment measurement. A 1012 Ω resistor 52 is used on cup 5 to measure mass 45, and a 1011 Ω resistor is used on cup L1 to measure 53 mass 44. We generally do not fully resolve 13C12C2H8+ from 12C3H7D+ at mass 45, which 54 requires a mass resolution close to the limit achieved by the prototype Ultra used in this 55 study (the production version of the Ultra achieves ~2x higher mass resolutions and 56 should be capable of cleaning separating these species). Thus, a minor component, 57 typically a few percent, of each beam will contaminate the other. Our solution to this is to 58 simply combine the signals from these species on the Ultra, and then separately measure 59 the D/13C ratio of the sample on a very high-resolution single collector mass spectrometer, 60 a modified version of the Thermo DFS. The DFS is a high-resolution gas source mass 61 spectrometer, having a reverse geometry and a single collector detector. Eiler et al. 2014, 62 2015 have shown that this instrument can be outfitted with a dual inlet system and used to 63 measure ratios of nearly isobaric isotopologues by untangling the shape of the resulting 64 adjacent peaks (Eiler et al., 2013; Eiler et al., 2014). This measurement, combined with 65 some second constraint (like the sum of 13C+D) typically yields an error of ~0.5-2 ‰ in 66 δD, and no significant increase in error in δ13C beyond that created by that second 67 constraint. For our purposes, we only use this measurement to correct for the contribution 68 of D to the 13C+D signal in the Ultra measurement; this variant of our technique is used 69 for only a small subset of the measurements presented here. 70 71 Analyzer tuning and resolution At the beginning of each analytical session, the instrument is configured with its 72 high resolution entrance slit and tuned until it achieves a resolution of ~20,000 mass 73 resolving power (M/∆M, 5/95% definition; see (Eiler et al., 2013) for relevant instrument 74 details). The instrument typically retains a stable high-mass resolution for time scales on 75 the order of 1-2 weeks, with minor adjustments to source and analyzer lens potentials 76 made at the beginning of each day. 77 78 79 Signal intensities and counting statistics limits Given typical source pressures, ion source tuning parameters set to maximize 80 sensitivity, and use of the high-resolution entrance slit, ion currents of the 12CH2+ 81 fragment ion are typically on the order of 550 fA, whereas that for the 13C substituted 82 species, 13CH2+, is typically on the order of 5 fA. When the CH3+ fragment ions are 83 instead used to analyze the methyl fragment ion species, intensities are typically 1050 fA 84 for 12CH3+ and 12 fA for 13CH3+. Similarly, typical intensities for the ion beams used to 85 measure the ethyl fragment are 19 pA for 12C2H4+ and 430 fA for 13C12CH4+. 12C3H8+ has 86 a typical intensity of 11 pA, and 13C12C2H8+ has a typical intensity of 360 fA (Table 1). 87 For comparison, a 1 fA ion current represents 6242 ions/sec, and would require ~160 88 seconds of counting to reach a shot-noise-limited precision of 1‰. 89 90 91 92 Background corrections Given the mass resolutions typically achieved for the instrument and method that we present, all plausibly significant interferences with the target analyte ion beams are 93 formally resolved (i.e., M/∆M values for all interferences of interest are substantially less 94 than the formal 5/95 % resolution measured when the analyzer is tuned). However, the 95 prototype Ultra used in this study has relatively wide exit slits (i.e., detector apertures) to 96 achieve a `flat-topped peak' condition that facilitates high-precision measurements (the 97 production version of this machine possesses one or more narrower exit slits, obviating 98 some of the issues described below). Therefore, in any case where a target analyte is 99 `sandwiched' between closely adjacent higher- and lower-mass interfering peaks, the 100 analyte ion beam must always be collected along with at least one of those interferences. 101 For this reason, it is essential to our method that the intensities of these co-collected 102 neighboring peaks be measured and used to correct the measured intensity of the analyte 103 plus interference(s). Similarly, corrections must be applied for background contributions 104 from scattered ions and for signals associated with detector dark-noise. Generally, all 105 significant corrections of this kind can be made using a single measurement adjacent to 106 the analyte peak. For example, Supplemental Figure 1 shows a scan of the 16 amu peaks, 107 revealing that 13CH3+ can be delivered to the detector without significant contributions 108 from adjacent 12CH4+, but in this case must include contributions from the other closely 109 adjacent background species, 14NH2+. 110 There are two separate background corrections needed for each measurement. The 111 first is specific to the central collector. The magnet is set to a current typically 200-300 112 BDAC units (or, ~ 0.01 amu) lower than that for the targeted peak, and signal at this 113 position is recorded for two minutes. For the first 60 seconds of this period, gas from the 114 sample reservoir is allowed to flow into the ion source; then, the changeover block is 115 switched so that the standard gas flows into the ion source. After the two minute period is 116 complete, we cull the data to exclude the ~15 seconds of unstable gas pressures 117 associated with switching the changeover valve block, and average over a 30-second 118 period in the middle of each ~1 minute of signal from each gas. After the first 119 background measurement is completed, an isotope ratio acquisition is made, and then a 120 second background measurement is made for the lower-mass species being detected on 121 either cup L1 or L3 (depending on which fragment ion is being measured). If the SEM is 122 in use it is first turned off (since the SEM background on the low mass side of the 123 enriched peak was completed first, and the high mass side of the unenriched peak would 124 swamp the sensor), and then the magnet is set to a current 200-300 BDAC units (~0.01 125 AMU) greater than the magnet setting for the measurement; once more, signal is 126 observed for one minute on each bellows, and averaged for the middle 30 seconds of that 127 time. 128 We find that background intensities are strongly correlated with ion source 129 pressure, presumably because the production and/or ionization efficiency of common 130 background species scales with the pressure of analyte in the ion source. When the 131 pressure within the source is within the range that we specify for our default method, 132 background intensities are approximately stable through time (at least, over the course of 133 several acquisitions). However, if source pressure is significantly higher (i.e., >8x10-8 134 mbar) both of the intensities of monitored ion beams and background intensities 135 measured as described above fluctuate too much for precise (sub-per mil) isotope ratio 136 analysis over seconds to minutes. 137 138 139 Figure Caption: 140 Mass scans illustrating the simultaneous positions of two adjacent collectors during the 141 measurement of 13C/12C ratios of methyl (A) or ethyl (B) fragment ions. In both of these 142 figures, the different colored lines represent mass scans across different collectors. In 143 figure A the shoulder where 13CH2 is measured is clearly wide enough that it is easily 144 resolved from 12CHD. In figure B the shoulder where 13C12CH4 is measured is 145 proportionately narrower but still statistically flat for a great enough range to be 146 measured stably. 147 148 149 150 Eiler J., Blake G. A., Dallas B., Kitchen N., Lloyd M. K. and Sessions A. L. (2014) Isotopic Anatomies of Organic Compounds. Goldschmidt Conference. 151 152 153 Eiler J., Clog M., Magyar P., Piasecki A., Sessions A. L., Stolper D., Deerberg M., Schlueter H.-J. and Schwieters J. (2013) A high-resolution gas-source isotope ratio mass spectrometer. International Journal of Mass Spectrometry 335, 45–56. 154
© Copyright 2026 Paperzz