Wagner FDIX Memory Feature – Manual Addendum 9/06 H. MEMORY H.1 – DESCRIPTION/OVERVIEW The FDIX Memory feature allows the user to store up to 500 samples for later retrieval. A sample is saved when the user presses the SEND key on the FDIX gauge. Samples can be retrieved and reviewed on the gauge display or sent via the RS232 interface to a computer (PC). DEFINITIONS: Sample Data Point Instantaneous Force Peak Force A single consecutively numbered test performed at one site and at one time. Force values can include: instantaneous, peak compression and peak tension. A numerical force value, here: instantaneous, peak compression and peak tension. A numerical force value that corresponds to the force currently applied to the shaft. A numerical compression or tension force that corresponds to the maximum force applied to the load shaft during one test. Each sample saved contains the following data: 1 - Sample Number – Consecutive: 1-500 samples 2 - Instantaneous force with unit of measure* 3 - Compression Peak force with unit of measure* 4 - Tension Peak force with unit of measure* * Units of force: lbf. ozf. kgf. N. H.2 - MEMORY DISPLAY The following memory items are included on the gauge display (LCD): • Consecutive sample number, instantaneous force and peak compression or peak tension force. • Four triangular tick marks (upper left) indicating by flashing which of the four values is currently being displayed. H.3 - MEMORY SET-UP The FDIX gauge SEND key is used to save samples, send samples via the RS232 port, both or none. The SEND key mode is set up using the gauge setup menu. To set up the SEND key mode, perform the following steps: • Turn off the FDI/FDIX. • Push / Hold ZERO, push ON/OFF. • Release ON/OFF first, then release ZERO after AOFF is displayed. • Press and release SCROLL until the StoreE option is displayed. • Press SELECT: rdATA is displayed. D:\81894293.doc 1 • Press and release SCROLL until the SEnd option is displayed. • Press and release SELECT to enter the SEnd option menu. • Press and release SCROLL until the desired option is selected Available options are: 1) none - SEND key is not used 2) 232 - SEND key sends data to computer via RS232 only 3) dATA - SEND key saves a sample only 4) both - SEND saves sample and sends to computer via RS232 NOTE!!! - If the gauge has been programmed to operate in the gauge initiated continuous mode the memory save feature is disabled. • When the desired option has been selected, press SELECT: done is displayed to indicate that the desired mode has been saved. • Turn the gauge off and on to return to the measurement mode. H.4 - MEMORY SAVE OPERATION A sample is saved when SEND is pressed. When a sample is saved, the gauge will momentarily display SavE, followed by the sample number of the saved data. If the sample memory storage is full, the gauge will display FuLL. H.5 - MEMORY DATA RETREIVAL H.5.1 - Using Gauge Display To enter the sample review mode, perform the following steps: • Turn off the FDI/FDIX. • Push / Hold ZERO, push ON/OFF. • Release ON/OFF first, then release ZERO after AOFF is displayed. • Press and release SCROLL until StoreE is displayed. • Press SELECT: rdATA is displayed. • Press SELECT. Sample number and the first tick mark (left to right) are displayed. The leftmost tick mark will flash, indicating the sample number is displayed. • Press SCROLL. The instantaneous value from this sample is displayed and the second tick mark will flash, indicating the instantaneous sample is displayed. • Press SCROLL. The compression peak value from this sample is displayed and the third tick mark will flash, indicating the compression peak sample is displayed. • Press SCROLL. The tension peak value from this sample is displayed and the fourth tick mark will flash, indicating the tension peak sample is displayed. • Press SCROLL. The sample number for the next sample will be displayed. Continue to press SCROLL to review additional saved samples. D:\81894293.doc 2 NOTE - SEND can be used as a REVERSE SCROLL key. Pressing SEND will result in the display of the previous sample (or sample number). The flashing tick mark will also be updated to indicate what sample (or sample number) is being displayed. The gauge will display End when all values have been displayed. H.5.2 – Using the RS232 Computer (PC) Interface RS232 sample data is returned in the following format: [Sample Number],[Instantaneous force value],[Compression peak value],[Tension peak value] [CR][LF] NOTE: The individual force values are sent in the identical format as detailed in the FDIX User Manual. H.5.2.1 - RS232 Memory Recovery Command Summary Wagner Protocol (Gauge RS232 Functions 1- 4) “MA”<CR> – Sends all samples in memory. ”MZ”<CR> - Goes to beginning of memory, gauge returns "Pass" or "Empty". “MC”<CR> - Sends the number of samples in memory. “MN”<CR> - Sends the next sample record. Gauge sends sample data or "Empty","End" or “Full". “MS”<CR> - Skips the next sample record. Gauge sends sample data or "Empty", "End" or "Full". Chatillon Protocol (Gauge RS232 Functions 5 - 8) “r” – Sends all samples in memory. ”B” - Goes to beginning of memory, gauge returns "Pass" or "Empty". “G” - Sends the number of samples in memory. “M” - Sends the next sample record. Gauge sends sample data or "Empty", "End" or "Full". H.6 - Data Clear H.6.1 - Clear All Data To clear the all saved data, perform the following steps: • Turn off the FDI/FDIX. • Push / Hold ZERO, push ON/OFF. • Release ON/OFF first, then release ZERO after AOFF is displayed. • Press and release SCROLL until the StoreE option is displayed. • Press SELECT: rdATA is displayed. • Press and release SCROLL until the Cl dt (Clear all data) option is displayed. • Press and hold SEND (approximately 5 seconds) until CLrd is displayed. • Turn the gauge off and on to return to the measurement mode. D:\81894293.doc 3 H.6.2 - Clear Last Data Point Only To clear the last sample only, perform the following steps: • Turn off the FDI/FDIX. • Push / Hold ZERO, push ON/OFF. • Release ON/OFF first, then release ZERO after AOFF is displayed. • Press and release SCROLL until the StoreE option is displayed. • Press SELECT: rdATA is displayed. • Press and release SCROLL until the Cl Pt (Clear last data point) option is displayed. • Press and hold SEND (approximately 5 seconds) until CLrd is displayed. • Turn the gauge off and on to return to the measurement mode. D:\81894293.doc 4
© Copyright 2026 Paperzz