KVAC663_InfoVAC Trouble SS-cnd:Layout 2 9/9/10 3:21 PM Page 1 ® InfoV.A.C. Therapy Alarm Troubleshooting Quick reference for the delivery of prescribed therapy. Detecting and resolving common InfoV.A.C.® Therapy alarms. To be used in conjunction with device User Manual.* System Alarm Alarm Condition Low Pressure Alarm Unit has not reached the selected therapy setting Action/Resolution • Check tubing for closed clamps, kinks, crimps or blockages User Tips • Ensure a quarter-sized (2.5cm) hole has been cut in the drape • Ensure dressing and drape has not shifted and blocked SensaT.R.A.C.® Pad’s outer lumens • Lower therapy unit and tubing to or below wound level Blockage Alarm Unit has detected a potential blockage • Inspect to ensure a quarter-sized (2.5cm) hole has been cut in the drape • Ensure the dressing and drape have not shifted underneath, blocking the SensaT.R.A.C.® Pad’s outer lumens Leakage Alarm Unit has detected a significant negative pressure leak • Use leak detection procedures and the Seal Check™ tool to help find and repair leak • Lower therapy unit and tubing to or below wound level • Ensure pad is located in a flat area of the body, avoiding a skin fold • For larger, highly exudating wounds, adjust intensity level to highest level to ensure a faster draw down and immediate seal Therapy unit and tubing should be at or lower than wound surface KVAC663_InfoVAC Trouble SS-cnd:Layout 2 9/9/10 3:21 PM Page 2 InfoV.A.C.® Leak Alarm Reset Quarter-Sized is the Right Size Following an alarm condition, the InfoV.A.C.® Therapy System requires a manual reset of the leak alarm screen after the condition has been addressed. Resetting IMPORTANT! To optimize performance of the SensaT.R.A.C.® Pad, the unit and resuming prescribed therapy is described below: you must cut a quarter-sized (2.5cm) hole in the drape. A smaller hole, an “x” or a slit may cause blockage, pressure fluctuations and other issues. This can prevent delivery of prescribed therapy and cost caregivers valuable time. After alarm has been corrected, you must hit RESET. The unit will resume normal operations. Sensing Lumens NOTE: If these steps are skipped, therapy will shut down after 3 minutes. For additional questions or concerns, call your local KCI representative or contact us at 1-800-668-5403 or visit www.kci-medical.com Important: Indications, Contraindications, Warnings, Precautions and Other Important Safety Information are contained in the V.A.C.® Therapy General Safety Information Sheet. Please consult a physician and product instructions for use prior to application. Caution: Federal law restricts this device to sale/rental by or on the order of a physician. ©2008 KCI Licensing, Inc. All rights reserved. The V.A.C.® (Vacuum Assisted Closure®) System is a registered trademark of KCI, its affiliates and licensors. The V.A.C.® System and most KCI products are subject to patents and/or pending patents. 09/08 Lit. #29-B-218
© Copyright 2026 Paperzz