Effect of processing time delay on the dose response of Kodak EDR2 film Nathan L. Childressa) and Isaac I. Rosen Department of Radiation Physics, The University of Texas M. D. Anderson Cancer Center, 1515 Holcombe Boulevard, Unit 94, Houston, Texas 77030 共Received 6 April 2004; revised 19 May 2004; accepted for publication 28 May 2004; published 23 July 2004兲 Kodak EDR2 film is a widely used two-dimensional dosimeter for intensity modulated radiotherapy 共IMRT兲 measurements. Our clinical use of EDR2 film for IMRT verifications revealed variations and uncertainties in dose response that were larger than expected, given that we perform film calibrations for every experimental measurement. We found that the length of time between film exposure and processing can affect the absolute dose response of EDR2 film by as much as 4%– 6%. EDR2 films were exposed to 300 cGy using 6 and 18 MV 10⫻10 cm2 fields and then processed after time delays ranging from 2 min to 24 h. An ion chamber measured the relative dose for these film exposures. The ratio of optical density 共OD兲 to dose stabilized after 3 h. Compared to its stable value, the film response was 4%– 6% lower at 2 min and 1% lower at 1 h. The results of the 4 min and 1 h processing time delays were verified with a total of four different EDR2 film batches. The OD/dose response for XV2 films was consistent for time periods of 4 min and 1 h between exposure and processing. To investigate possible interactions of the processing time delay effect with dose, single EDR2 films were irradiated to eight different dose levels between 45 and 330 cGy using smaller 3⫻3 cm2 areas. These films were processed after time delays of 1, 3, and 6 h, using 6 and 18 MV photon qualities. The results at all dose levels were consistent, indicating that there is no change in the processing time delay effect for different doses. The difference in the time delay effect between the 6 and 18 MV measurements was negligible for all experiments. To rule out bias in selecting film regions for OD measurement, we compared the use of a specialized algorithm that systematically determines regions of interest inside the 10⫻10 cm2 exposure areas to manually selected regions of interest. There was a maximum difference of only 0.07% between the manually and automatically selected regions, indicating that the use of a systematic algorithm to determine regions of interest in large and fairly uniform areas is not necessary. Based on these results, we recommend a minimum time of 1 h between exposure and processing for all EDR2 film measurements. © 2004 American Association of Physicists in Medicine. 关DOI: 10.1118/1.1774111兴 Key words: film dosimetry, IMRT treatment plan verification, quality assurance, radiation therapy INTRODUCTION Radiographic film allows for fast and convenient twodimensional dose distribution measurements. However, its high silver halide content overresponds to low energy photons because of an increase in photoelectric effect interactions. Kodak’s extended dose range 2 共EDR2兲 film has a significantly lower silver halide content than earlier films,1 and is a commonly used film for intensity modulated radiotherapy verifications.2–7 Its altered chemical formulation and smaller grain size allow it to measure higher doses without becoming saturated and to produce results that exhibit good agreement with other dosimeters.8 –14 However, these changes have resulted in a film that has a much different chemical structure than older, well-characterized15–21 radiographic films. In our clinical use of EDR2 film for full patient plan intensity modulated radiotherapy 共IMRT兲 verifications, we noticed variations and uncertainties in dose response that were larger than expected, given that we perform film calibrations for every experimental measurement. We discovered that the length of time between exposure and processing alters EDR2 2284 Med. Phys. 31 „8…, August 2004 film’s absolute dose response. Thus, an IMRT verification may have varying absolute dose responses depending on the time elapsed between exposure and processing of both patient film共s兲 and calibration film共s兲. We quantitatively characterized the relationship between the absolute dose response of EDR2 film and the time between its exposure and processing. METHODS AND MATERIALS All films were processed using an RP X-Omat Model M6B film processor 共Eastman Kodak Company, Rochester, NY兲 with Polymax RT developer 共Eastman Kodak Company兲 at 33 °C. A 16-bit VXR-16 Dosimetry Pro film scanner 共VIDAR Systems Corporation, Herndon, VA兲 was used to scan all films. Optical density response versus time delay, single dose Kodak EDR2 films 共Eastman Kodak Company兲 were radiated in a solid water phantom at 95 cm source to surface distance 共SSD兲 with 10⫻10 cm2 fields of 6 and 18 MV pho- 0094-2405Õ2004Õ31„8…Õ2284Õ5Õ$22.00 © 2004 Am. Assoc. Phys. Med. 2284 2285 N. L. Childress and I. I. Rosen: Effect of processing time delay on EDR2 film tons from a Varian Clinac 2100EX 共Varian Medical Systems, Inc., Palo Alto, CA兲. Every film was from the same batch. An ion chamber placed 5 cm below the film recorded relative doses at the center of the field to allow the variations in machine output to be taken into consideration. The relative dose to the film was computed based on this measurement. Films were irradiated perpendicular to the beam’s central axis at 5 cm depth in their original jackets to approximately 300 cGy. They were processed following time delays of 2, 4, and 30 min and 1, 3, 6, 12, and 24 h. One film was used for the 2 and 4 min intervals, two films were used for the 30 min time interval, and three films were used for all other time intervals. The average optical density 共OD兲 per ion chambermeasured dose of the film was calculated for each time interval. These values were normalized to the time period with the highest OD/dose response for each energy. Assuming that the uncertainties in OD measurement and ion chamber measurement are independent of the time delay, a single standard deviation representing the overall reproducibility of the experiment was computed and applied to all time periods. The overall standard deviation for each photon energy was calculated by combining the ratio of each individual OD/dose measurement to its corresponding mean for all data with a processing time delay of at least 1 h. Optical density response versus time delay, multiple doses The single-dose experiment was designed to explore a wide range of processing time delay effects and so it did not determine whether the processing time delay effect is dose dependent. A limited number of time delay values were studied at different clinically relevant dose levels to see if processing time delays affect all OD levels of the film equally. Eight dose levels were delivered to a single film using a multileaf collimator-generated eight box pattern.22 Experiments were done with 6 and 18 MV beams. Processing time delays of 1, 3, and 6 h were used to investigate the dose dependence of the processing time delay effect. Three films in their original jackets were placed perpendicular to the beam’s central axis at a depth of 5 cm in a solid water phantom at 95 cm SSD and were exposed at each time interval. While this type of dose pattern is typically used for singlefilm calibrations, it was used in this experiment to deliver many dose regions to a single piece of film. It was not possible to use an ion chamber to measure relative doses during the delivery because multiple dose levels were delivered simultaneously to each piece of film. The average OD of each dose region and the background OD were used for analysis. Results for both photon energies were normalized to their average OD at 6 h. Optical density response versus time delay, multiple film batches The processing time delay effect was also explored using three EDR2 film batches that differed from the batch used in the previous two experiments. We also performed measurements with one batch of Kodak XV2 film for comparison Medical Physics, Vol. 31, No. 8, August 2004 2285 purposes. For each batch of film, the average OD/dose of three pieces of film with a 1 h processing time delay was compared to the average OD/dose of three pieces of film with 2, 4, and 6 min processing time delays. A one-tailed nonpaired heteroscedastic t-test was performed to determine whether the means of the two samples were statistically different for each batch of film. The order in which the films were scanned and processed was randomized. The experimental methods of the single dose experiment were used to expose the film and to measure relative doses with an ion chamber. Only 6 MV beams were used, and only 100 cGy was delivered to XV2 films. Manual selection of OD region versus automatic analysis As the magnitude of the processing time delay effect is on the order of a few percentage points, it was crucial to eliminate as many sources of error as possible. An in-house film dosimetry program was used to calculate the average OD of a manually selected rectangular region in each of the scanned film images. While the 10⫻10 cm2 exposure areas were relatively uniform, differences in the sizes, shapes, and locations of these user-defined areas could add uncertainty to the final results. For comparison, we developed a program that automatically and systematically determined a region of interest and computed its average value. The program detected the edges of the radiation field in order to calculate a region of interest that comprised the central 25% of the total area of the square. Both methods were used to determine the average values of the OD regions. Finally, the effect of rotational errors was explored by analyzing films with and without a 2° rotation. RESULTS AND DISCUSSION Optical density response versus time delay, single dose Figure 1 displays the processing time delay effect on OD/ dose response over 24 h for 6 and 18 MV photon beams. The overall standard deviation of the measurements from each processing time delay group’s average was 0.25% for all points with at least 1 h between processing and exposure. The maximum deviation in machine output was 1.1% for the 6 MV data and 0.6% for the 18 MV data. The 6 MV 2 min through 6 h data points were measured a second time to verify that these results were not anomalous 共Fig. 2兲. Both experiments produced similar results. The normalized OD/ dose value of the 2 min data point is very dependent on the exact length of time between exposure and processing, as it lies in a high-gradient portion of the curve. While this value was difficult to accurately determine, it was 4%– 6% lower than the stable OD/dose value for both experiments. The OD/dose response stabilized after 3 h. The film’s OD/dose was within 1% of its stable value after 1 h, and within 2% of its stable value after 30 min. It appears that there may be latent image fading for 6 MV exposures 共lower OD/dose as processing time delays increase兲 for delays greater than three 2286 N. L. Childress and I. I. Rosen: Effect of processing time delay on EDR2 film 2286 FIG. 2. EDR2 film OD per dose time delay effect for 6 MV. Each data point represents a single film. The OD/dose values were normalized to the average OD/dose at 6 h. This experiment was performed to verify the more complete results displayed in Fig. 1. measurements to correlate the OD results for each time period. These figures confirm that the average OD at 1 h is approximately 1% less than the average OD at 3 and 6 h, and there is no significant change in these percentages with dose levels. Furthermore, the background of the film did not significantly vary with processing time delay. As EDR2 film’s processing time delay effect does not show a dose depen- FIG. 1. EDR2 film OD per dose response at various processing time delays for 共a兲 6 MV and 共b兲 18 MV exposures. Each data set was normalized to its highest OD/dose response. Processing EDR2 films immediately after exposure results in 4 – 6% less OD/dose compared to processing films after a 3– 6 hour time delay. Error bars represent one standard deviation. hours. Although the difference between the 6 and 24 h data points of 0.6% is marginally significant, the 18 MV data does not show as clear of a trend. All 18 MV measurements past 3 h agree within one standard deviation. Overall, there was no significant difference between the 6 and 18 MV time delay responses. Optical density response versus time delay, multiple doses Figure 3 shows the dose dependence of the processing time delay for 6 and 18 MV exposures. Again, no significant energy dependence was observed. The 18 MV exposures result in higher doses owing to the beam’s lower attenuation in the solid water. The 210/230 cGy 共6/18 MV兲 data points consistently had a higher standard deviation than the other dose levels, thus their abnormal OD response is more likely due to their lower reproducibility than to a trend in EDR2 film’s dose dependence. Although not shown in the figures, the standard deviations of the reproducibility of these data points are generally 0.4%. This standard deviation excludes the possible machine output variation that could not be measured, as there were no unifying ion chamber absolute dose Medical Physics, Vol. 31, No. 8, August 2004 FIG. 3. EDR2 film normalized OD response to 3⫻3 cm2 squares with several different doses, for 共a兲 6 MV and 共b兲 18 MV exposures. There was no clear processing time delay dose dependence. While not shown in the figures, the standard deviation for each of the data points is approximately 0.4%. 2287 N. L. Childress and I. I. Rosen: Effect of processing time delay on EDR2 film FIG. 4. Processing time delay results for different film batches. All EDR2 film batches were different than the batch used to generate Figs. 1–3. The processing time delay effect was statistically significant for all three EDR2 batches, but not for XV2 film. Error bars indicate one standard deviation. dence at 1 h, it is safe to use a 1 h period between exposure and processing for clinical verifications without experiencing additional discrepancies due to the range of dose levels. Optical density response versus time delay, multiple film batches The differences in films with an average 4 min processing time delay and 1 h delay for three batches of film are shown in Fig. 4. All three batches of film exhibited similar behavior. The maximum deviation in machine output was 0.2%. The difference between the two processing time delay groups was statistically significant in all batches, with a maximum p-value of 0.007. The standard deviation of the 4 min delay OD/dose values was slightly higher than the 1 h values, as the 2– 6 min delay points are in a high gradient region of the time delay effect curve. The mean of all 9 EDR2 films in the 2– 6 min processing time delay group was 0.962⫾0.009 共one standard deviation兲, significantly lower than the 1 h group’s 1.000⫾0.002 OD/dose response (p⬍0.0001). The batch of XV2 film did not have a statistically significant processing time delay effect for the 4 min and 1 h time periods studied. These results confirm this effect is fairly consistent among different batches of film and is statistically significant for EDR2 films. Manual selection of OD region versus automatic analysis The maximum difference between the average OD/dose of user-defined areas and automatically generated areas was an insignificant 0.07%. The average difference was 0.02% ⫾0.02% 共one standard deviation兲. Changing the film’s rotation by 2° produced similar results. It can be concluded that a specialized algorithm to determine region averages is not needed for large exposure areas, the authors are remarkably skilled at accurately and consistently selecting regions of interests, or both. CONCLUSIONS example, if calibration film共s兲 are exposed following several patient-specific quality assurance film exposures and then all films are processed together, the time delay for the calibration film共s兲 would be shorter than for the patient films. Consequently, the film calibration would show an OD/dose that is too low compared to the patient films and the patient doses would be overestimated. Since fluence verifications typically require only a few minutes to deliver, their doses would be uniformly high. Full-patient plan verifications, which require between 10 and 45 min to deliver, could exhibit different dose responses for individual beams. Thus, the beam that was delivered first can have a 4%– 6% greater OD/dose response than the beam that was delivered last if the film is processed immediately after exposure. This results in planning target volume dose normalizations that can differ from one verification to another, depending on the length of time elapsed between calibration and patient films and the order in which they were exposed. This effect may partially explain why some clinics routinely observe measured to calculated dose normalizations that are not equal to 1.23 Clearly, not compensating for the processing time delay effect in clinical verifications could lead to confusing results for both individual verifications and analyses that compare the dose normalization for a large number of patients. All films should be processed at least 1 h after exposure when performing measurements with Kodak EDR2 film. This guarantees that the OD/dose measurements of the films have reached 99% of their stable values. The OD/dose response for XV2 films is consistent for time periods of 4 min and 1 h between exposure and processing. It is not known if the processing time delay effect varies with different film processors or developing solutions. If a film calibration is performed before an IMRT verification and both films are processed immediately, selected beams from the IMRT film may exhibit an OD/dose response up to 4%– 6% lower than the same dose in the calibration film共s兲. Other beams may show a 0%– 6% decrease in the expected OD/dose, resulting in confusing patient-specific IMRT verification results. These effects would be reversed if the calibration film共s兲 were exposed after the patient film共s兲. Further research focusing on the chemical reaction rates in EDR2 film would be necessary to determine the cause of this effect. As there was no observed dose dependence, any time period between exposure and processing is acceptable if one does not need to correlate the OD/dose response between films with short exposure durations. ACKNOWLEDGMENTS This research was partially supported by Philips Medical Systems and the American Legion Auxiliary Fellowship. We would also like to thank Jeff Byng from Kodak Canada for his assistance and suggestion to research the processing time delay effect. a兲 Electronic mail: [email protected] X. R. Zhu, S. Yoo, P. A. Jursinic, D. F. Grimm, F. Lopez, J. J. Rownd, and M. T. Gillin, ‘‘Characteristics of sensitometric curves of radiographic films,’’ Med. Phys. 30, 912–919 共2003兲. 1 The sequencing of film exposure and processing could produce 4%– 6% errors in EDR2 film’s optical densities. For Medical Physics, Vol. 31, No. 8, August 2004 2287 2288 N. L. Childress and I. I. Rosen: Effect of processing time delay on EDR2 film 2 P. Cadman, R. Bassalow, N. P. Sidhu, G. Ibbott, and A. Nelson, ‘‘Dosimetric considerations for validation of a sequential IMRT process with a commercial treatment planning system,’’ Phys. Med. Biol. 47, 3001– 3010 共2002兲. 3 N. Dogan, L. B. Leybovich, A. Sethi, and B. Emami, ‘‘Automatic feathering of split fields for step-and-shoot intensity modulated radiation therapy,’’ Phys. Med. Biol. 48, 1133–1140 共2003兲. 4 D. Georg, B. Kroupa, P. Winkler, and R. Potter, ‘‘Normalized sensitometric curves for the verification of hybrid IMRT treatment plans with multiple energies,’’ Med. Phys. 30, 1142–1150 共2003兲. 5 A. Kapulsky, E. Mullokandov, and G. Gejerman, ‘‘An automated phantom-film QA procedure for intensity-modulated radiation therapy,’’ Med. Dosim 27, 201–207 共2002兲. 6 P. N. McDermott, T. He, and A. DeYoung, ‘‘Dose calculation accuracy of lung planning with a commercial IMRT treatment planning system,’’ J. Appl. Clin. Med. Phys. 4, 341–351 共2003兲. 7 S. L. Richardson, W. A. Tome, N. P. Orton, T. R. McNutt, and B. R. Paliwal, ‘‘IMRT delivery verification using a spiral phantom,’’ Med. Phys. 30, 2553–2558 共2003兲. 8 P. M. Charland, I. J. Chetty, S. Yokoyama, and B. A. Fraass, ‘‘Dosimetric comparison of extended dose range film with ionization measurements in water and lung equivalent heterogeneous media exposed to megavoltage photons,’’ J. Appl. Clin. Med. Phys. 4, 25–39 共2003兲. 9 I. J. Chetty and P. M. Charland, ‘‘Investigation of Kodak extended dose range 共EDR兲 film for megavoltage photon beam dosimetry,’’ Phys. Med. Biol. 47, 3629–3641 共2002兲. 10 N. Dogan, L. B. Leybovich, and A. Sethi, ‘‘Comparative evaluation of Kodak EDR2 and XV2 films for verification of intensity modulated radiation therapy,’’ Phys. Med. Biol. 47, 4121– 4130 共2002兲. 11 N. Dogan and G. P. Glasgow, ‘‘Surface and build-up region dosimetry for obliquely incident intensity modulated radiotherapy 6 MV x rays,’’ Med. Phys. 30, 3091–3096 共2003兲. 12 J. Esthappan, S. Mutic, W. B. Harms, J. F. Dempsey, and D. A. Low, ‘‘Dosimetry of therapeutic photon beams using an extended dose range film,’’ Med. Phys. 29, 2438 –2445 共2002兲. Medical Physics, Vol. 31, No. 8, August 2004 13 2288 A. J. Olch, ‘‘Dosimetric performance of an enhanced dose range radiographic film for intensity-modulated radiation therapy quality assurance,’’ Med. Phys. 29, 2159–2168 共2002兲. 14 X. R. Zhu, P. A. Jursinic, D. F. Grimm, F. Lopez, J. J. Rownd, and M. T. Gillin, ‘‘Evaluation of Kodak EDR2 film for dose verification of intensity modulated radiation therapy delivered by a static multileaf collimator,’’ Med. Phys. 29, 1687–1692 共2002兲. 15 M. Bucciolini, F. B. Buonamici, and M. Casati, ‘‘Verification of IMRT fields by film dosimetry,’’ Med. Phys. 31, 161–168 共2004兲. 16 S. E. Burch, K. J. Kearfott, J. H. Trueblood, W. C. Sheils, J. I. Yeo, and C. K. Wang, ‘‘A new approach to film dosimetry for high energy photon beams: lateral scatter filtering,’’ Med. Phys. 24, 775–783 共1997兲. 17 C. Danciu, B. S. Proimos, J. C. Rosenwald, and B. J. Mijnheer, ‘‘Variation of sensitometric curves of radiographic films in high energy photon beams,’’ Med. Phys. 28, 966 –974 共2001兲. 18 S. G. Ju, Y. C. Ahn, S. J. Huh, and I. J. Yeo, ‘‘Film dosimetry for intensity modulated radiation therapy: dosimetric evaluation,’’ Med. Phys. 29, 351–355 共2002兲. 19 C. Martens, I. Claeys, C. De Wagter, and W. De Neve, ‘‘The value of radiographic film for the characterization of intensity-modulated beams,’’ Phys. Med. Biol. 47, 2221–2234 共2002兲. 20 J. L. Robar and B. G. Clark, ‘‘The use of radiographic film for linear accelerator stereotactic radiosurgical dosimetry,’’ Med. Phys. 26, 2144 – 2150 共1999兲. 21 N. Suchowerska, P. Hoban, M. Butson, A. Davison, and P. Metcalfe, ‘‘Directional dependence in film dosimetry: radiographic and radiochromic film,’’ Phys. Med. Biol. 46, 1391–1397 共2001兲. 22 N. L. Childress, L. Dong, and I. I. Rosen, ‘‘Rapid radiographic film calibration for IMRT verification using automated MLC fields,’’ Med. Phys. 29, 2384 –2390 共2002兲. 23 A. J. Olch, ‘‘Dosimetric accuracy of the ITP inverse treatment planning system,’’ Med. Phys. 29, 2484 –2488 共2002兲.
© Copyright 2026 Paperzz