User Manual Ball Anchor Angulated Abutment

nt
te ing
Paend
P
MP-UI026 Rev.4, Sep. 2014
Extensive insertion path
Maximum angulations for varying
insertion path:
User Manual
Ball Anchor Angulated Abutment
Recommended tightening
moment 30 Ncm.
(MB-AN152/MB-AN252)
^
The Ball anchor 15° angulated
abutment allows a maximum
30° insertion path.
^
30°
^
^
^
50°
^
^
The Ball anchor 25° angulated
abutment allows a maximum
50° insertion path.
^
15° 15°
Recommended tightening
moment 30 Ncm.
^
^
^
^
25° 25°
MIS Implants Technologies Ltd.
P.O.Box 7, Bar Lev Industrial Park, 2015600, ISRAEL
Website: www.mis-implants.com
Sterilization
- The device is not sterile.
- The device must be sterilized before use by
autoclave, at a temperature of 134ºC (273ºF)
during 6 minutes. Do not exceed 134ºC
during sterilization.
®
Key to codes used:
Attention, see
instructions for use
Batch code
EC
Catalogue number
Manufacturer
0483
REP
MIS Implants Technologies GmbH
Simeonscarre 2, 32423 Minden, Germany
Tel: +49 571-972-7620
Email: [email protected]
By prescription only
© MIS Corporation. All rights Reserved.
®
A definition of the product.
Operating Procedure.
The device is intended as an intra-oral abutment secured
to implants onto which a multiple unit prosthetic restoration
is screw-retained ■ Standard impression procedure ■
Standard restorative components ■ Low profile, allowing
a wide spectrum of insertion possibilities for correction of
up to 50° ■ Advanced solution for rehabilitation on implants
in total edentulous patients ■ Use of supplemental standard
MIS ball attachment system components.
3
The Ball anchor angulated abutment is now positioned in place
and secured to the implant.
1
In cases where ball anchor angulated abutment is needed, select
abutment according to the implant's angulations (15°/25°).
The Ball Anchor Angulated
abutment assembly components:
Ball Ø2.25
Ø4.6
Abutment.
MB-AN152
MB-AN152- Ball anchor 15°
angulated abutment.
15°
3mm
1.7mm
or
Ø4
Ball Ø2.25
MB-AN252
MB-AN252- Ball anchor 25°
angulated abutment.
Characteristics
.6
The Ball anchor angulated abutments are made of titanium
Ti6Al4V ELI ■ The Ball anchor angulated abutment can be connected
in one of 6 possible positions on the fixture head.
2
25°
3.5mm
1.7mm
Place the Ball anchor angulated abutment (15°/25°) on
the implant by the gripping bar. To screw the abutment
to the implant, position the Hand Hex. Driver (MT-HHR13)
through the gripping bar’s opening. Recommended
tightening moment 30 Ncm.
Gripping Bar.
MM-PG250
The gripping bar indicates the
direction when positioning
the ball anchor angulated
abutment and supports the
abutment when screwed onto
the implant.
A.
B.
C.