DANIEL ULTRASONIC
METER T-2 TO T-11
TRANSDUCER UPGRADE
INSTRUCTIONS
__________________________________________
DANIEL MEASUREMENT AND CONTROL, INC.
AN EMERSON PROCESS MANAGEMENT COMPANY
HOUSTON, TEXAS
Part Number 3-9000-751
Revision A
DECEMBER 2005
IMPORTANT INSTRUCTIONS
Because these measurement instruments are sophisticated technical products, you must install, use
and maintain them in accordance with Daniel’s guidelines to ensure that they operate within the
range specified on the equipment nameplate. The following instructions must be adhered to and
integrated into your safety program when installing, using and maintaining Daniel Products.
•
•
•
•
•
•
•
•
•
•
•
Read all instructions prior to installing, operating and servicing the product. If this
instruction manual is not the correct manual, call 1-713-467-6000 (24-hour response number
for both Service and Sales Support) and the requested manual will be provided. Save this
instruction manual for future reference.
If you do not understand any of the instructions, contact your Daniel representative for
clarification.
Follow all warnings, cautions and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation and maintenance of
the product.
Install your equipment as specified in the installation instructions of the appropriate
instruction manual and per applicable local and national codes. Connect all products to the
proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program
and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts
specified by the manufacturer. Unauthorized parts and procedures can affect the product's
performance and place the safe operation of your process at risk. Look-alike substitutions
may result in fire, electrical hazards or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when
maintenance is being performed by qualified persons, to prevent personal injury.
ALWAYS READ AND FOLLOW THE DANIEL ULTRASONIC METER T-2 TO T11 TRANSDUCER UPGRADE INSTRUCTIONS AND ALL PRODUCT WARNINGS
AND INSTRUCTIONS.
Use of this equipment for any purpose other than its intended purpose may result in property
damage and/or serious personal injury or death.
Before opening the flameproof enclosure in a flammable atmosphere, the electrical circuits
must be interrupted.
This page intentionally left blank.
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
D E C 2005
DANIEL MEASUREMENT AND CONTROL, INC.
DANIEL ULTRASONIC METER T-2 TO T-11
TRANSDUCER UPGRADE INSTRUCTIONS
NOTICE
DANIEL MEASUREMENT AND CONTROL, INC. ("DANIEL") SHALL NOT BE LIABLE FOR TECHNICAL OR
EDITORIAL ERRORS IN THIS MANUAL OR OMISSIONS FROM THIS MANUAL. DANIEL M AKES NO
W ARRANTIES, EXPRESSED OR IM PLIED, INCLUDING THE IM PLIED WARRANTIES OF
M ERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE W ITH RESPECT TO THIS
M ANUAL AND, IN NO EVENT, SHALL DANIEL BE LIABLE FOR ANY SPECIAL OR CONSEQUENTIAL
DAM AGES INCLUDING, BUT NOT LIM ITED TO, LOSS OF PRODUCTION, LOSS OF PROFITS, ETC.
PRODUCT NAM ES USED HEREIN ARE FOR M ANUFACTURER OR SUPPLIER IDENTIFICATION ONLY
AND M AY BE TRADEM ARKS/REGISTERED TRADEM ARKS OF THESE COM PANIES.
THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND
W HILE EVERY EFFORT HAS BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE
CONSTRUED AS W ARRANTIES OR GUARANTEES, EXPRESSED OR IMPLIED, REGARDING THE
PRODUCTS OR SERVICES DESCRIBED HEREIN OR THEIR USE OR APPLICABILITY. W E RESERVE THE
RIGHT TO M ODIFY OR IMPROVE THE DESIGNS OR SPECIFICATIONS OF SUCH PRODUCTS AT ANY TIME.
DANIEL DOES NOT ASSUME RESPONSIBILITY FOR THE SELECTION, USE OR MAINTENANCE OF ANY
PRODUCT. RESPONSIBILITY FOR PROPER SELECTION, USE AND M AINTENANCE OF ANY DANIEL
PRODUCT REMAINS SOLELY W ITH THE PURCHASER AND END-USER.
DANIEL AND THE DANIEL LOGO ARE REGISTERED TRADEMARKS OF DANIEL INDUSTRIES, INC. THE
EMERSON LOGO IS A TRADEMARK AND SERVICE MARK OF EMERSON ELECTRIC CO.
THE DANIEL ULTRASONIC GAS FLOW METER INCLUDING THE M ARK III ELECTRONICS IS COVERED
BY UNITED STATES PATENTS 5,983,730 AND 4,646,575 AND PATENTS PENDING.
COPYRIGHT © 2005
BY DANIEL MEASUREMENT AND CONTROL, INC.
HOUSTON, TEXAS, U.S.A.
All rights reserved. No part of this work may be reproduced or
copied in any form or by any means - graphic, electronic or
mechanical - without first receiving the written permission of
Daniel Measurement and Control, Inc., Houston, Texas, U.S.A.
PREFACE
i
D E C 2005
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
WARRANTY
1. LIM ITED W ARRANTY: Subject to the limitations contained in Section 2 herein and except as otherwise expressly
provided herein, Daniel Measurement and Control, Inc. ("Daniel") warrants that the firmware will execute the
programming instructions provided by Daniel, and that the Goods manufactured or Services provided by Daniel will be
free from defects in materials or workmanship under normal use and care until the expiration of the applicable warranty
period. Goods are warranted for twelve (12) months from the date of initial installation or eighteen (18) months from
the date of shipment by Daniel, whichever period expires first. Consumables and Services are warranted for a period
of 90 days from the date of shipment or completion of the Services. Products purchased by Daniel from a third party for
resale to Buyer ("Resale Products") shall carry only the warranty extended by the original manufacturer. Buyer agrees
that Daniel has no liability for Resale Products beyond making a reasonable commercial effort to arrange for procurement
and shipping of the Resale Products. If Buyer discovers any warranty defects and notifies Daniel thereof in writing during
the applicable warranty period, Daniel shall, at its option, promptly correct any errors that are found by Daniel in the
firmware or Services, or repair or replace F.O.B. point of manufacture that portion of the Goods or firmware found by
Daniel to be defective, or refund the purchase price of the defective portion of the Goods/Services. All replacements or
repairs necessitated by inadequate maintenance, normal wear and usage, unsuitable power sources, unsuitable
environmental conditions, accident, misuse, improper installation, modification, repair, storage or handling, or any other
cause not the fault of Daniel are not covered by this limited warranty, and shall be at Buyer's expense. Daniel shall not
be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon in writing
in advance by an authorized Daniel representative. All costs of dismantling, reinstallation and freight and the time and
expenses of Daniel's personnel for site travel and diagnosis under this warranty clause shall be borne by Buyer unless
accepted in writing by Daniel. Goods repaired and parts replaced during the warranty period shall be in warranty for the
remainder of the original warranty period or ninety (90) days, whichever is longer. This limited warranty is the only
warranty made by Daniel and can be amended only in a writing signed by an authorized representative of Daniel. Except
as otherwise expressly provided in the Agreement, THERE ARE NO REPRESENTATIONS OR W ARRANTIES OF
ANY KIND, EXPRESSED OR IMPLIED, AS TO M ERCHANTABILITY, FITNESS FOR PARTICULAR PURPOSE,
OR ANY OTHER MATTER W ITH RESPECT TO ANY OF THE GOODS OR SERVICES.
It is understood that
corrosion or erosion of materials is not covered by our guarantee.
2. LIM ITATION OF REM EDY AND LIABILITY: DANIEL SHALL NOT BE LIABLE FOR DAMAGES
CAUSED BY DELAY IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF
W ARRANTY HEREUNDER SHALL BE LIMITED TO REPAIR, CORRECTION, REPLACEMENT OR REFUND
OF PURCHASE PRICE UNDER THE LIMITED W ARRANTY CLAUSE IN SECTION 1 HEREIN. IN NO EVENT,
REGARDLESS OF THE FORM OF THE CLAIM OR CAUSE OF ACTION (W HETHER BASED IN CONTRACT,
INFRINGEMENT, NEGLIGENCE, STRICT LIABILITY, OTHER TORT OR OTHERW ISE), SHALL DANIEL'S
LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXCEED THE PRICE TO BUYER OF THE SPECIFIC
GOODS MANUFACTURED OR SERVICES PROVIDED BY DANIEL GIVING RISE TO THE CLAIM OR CAUSE
OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL DANIEL'S LIABILITY TO BUYER AND/OR ITS
CUSTOMERS EXTEND TO INCLUDE INCIDENTAL, CONSEQUENTIAL OR PUNITIVE DAM AGES. THE
TERM "CONSEQUENTIAL DAMAGES" SHALL INCLUDE, BUT NOT BE LIMITED TO, LOSS OF
ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE AND COST OF CAPITAL.
ii
PREFACE
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
D E C 2005
TABLE OF CONTENTS
1.0
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
2.0
BEFORE YOU BEGIN! . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.1
Tools and Materials Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
3.0
KEY POINTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
4.0
TRANSDUCER UPGRADE PROCEDURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
APPENDICES
APPENDIX A
REMOVAL AND RE-INSTALLATION OF WRITEPROTECTION JUMPER ON MARK II ELECTRONICS . . . . . A-1
TABLE OF CONTENTS
iii
D E C 2005
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
This page intentionally left blank.
iv
TABLE OF CONTENTS
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
1.0
INTRODUCTION
1.1
General
D E C 2005
Welcome to the Daniel Ultrasonic Meter T-2 to T-11 Transducer Upgrade Instructions Manual.
This manual has been designed to provide you with a step-by-step set of instructions for upgrading
Daniel JuniorSonic™ and SeniorSonic™ meters from T-2 to T-11 transducers.
Please read the ‘Before You Begin’ section to make sure you have all the components
necessary to perform the upgrade before taking the JuniorSonic™ and SeniorSonic™ meters
out of service.
INTRODUCTION
1-1
D E C 2005
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
This page intentionally left blank.
1-2
INTRODUCTION
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
2.0
D E C 2005
BEFORE YOU BEGIN
This section lists all of the parts and tools required to perform the upgrade. Ensure that all of these
items are available before taking the JuniorSonic™ and SeniorSonic™ meters out of service.
SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
If the meter cannot be depressurized, use the ultrasonic extraction tool to remove and reinstall the transducer assemblies.
Failure to follow the Ultrasonic Extractor Tool Operation Manual (3-9000-729) procedures
may result in serious personal injury or death and may cause equipment damage.
If the meter cannot be depressurized, then the Daniel Ultrasonic Extractor Tool will be
required to remove the transducer assembly while the meter and associated piping remains
under pressure. If you do NOT have this tool, it can be purchased or rented from Daniel
Measurement Services.
2.1
Required Tools and Materials
1.
2.
3.
4.
5.
6.
7.
8.
9.
1¼" wrench or ratchet wrench c/w 1¼" socket
1/4" flat-blade screw driver
1/16" Allen wrench (2 or more recommended)
Heat gun (type used for heat shrink tubing)
Dow Corning® III silicone grease or equivalent
Nickel anti-seize compound (P/N 3-9960-134)
Sharpie® (or equivalent) permanent ink marker pen
Laptop computer complete with Daniel CUI 3.10 or later installed
Zero Flow Calibration Data Sheet (transducer assembly dimensions)
BEFORE YOU BEGIN
2-1
D E C 2005
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
This page intentionally left blank.
2-2
BEFORE YOU BEGIN
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
3.0
D E C 2005
KEY POINTS
EQUIPMENT DAMAGE
DO NOT proceed with the transducer upgrade if the original component lengths are not
available. Obtain the Zero Flow Calibration Data Sheet from Daniel. The original
component lengths of all equipment being replaced must be known in order to calculate
the new path length for each chord.
Exchanging any component in the transducer assembly may alter the length of the chord.
Document each component exchanged to ensure that the changes are correctly entered into the
meter configuration. Failure to do so, or entry of erroneous data, may impact meter accuracy.
•
It is recommended that Daniel CUI 3.10 or later be used for this procedure. Although older
versions will work, the latest version is always recommended.
All the screen images in this document are from Daniel CUI 3.10. Daniel CUI updates are
available at the following website: http://www.emersonprocess.com/daniel/.
•
Transducers are always replaced in pairs, and it is recommended they be installed one pair
at a time to avoid accidental interchange of components, leading to erroneous dimensions.
When replacing the transducers, it is important to note their position (upstream or
downstream) and replace them in the correct location to ensure proper low-velocity
performance. Each transducer pair is marked with either a number 1 or 2 that relates to the
meter location (i.e. A1, A2, B1, B2, etc). The number “1” transducers must be installed in
a location such as A1, B1, C1 or D1 and the “2” transducer must be installed in the
corresponding A2, B2, C2 or D2.
•
The transducer kit, P/N 2-9-3400-847, includes four pairs of transducers and is designed for
the SeniorSonic™ meter. If a JuniorSonic™ meter is being upgraded, this kit can be used,
and the remaining transducers can be kept as spares. If the spares are not needed, contact
Daniel Measurement Services for a list of components and part numbers.
KEY POINTS
3-1
D E C 2005
•
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
Older transducer assemblies (transducer, stalk, and holder) did not have the lengths etched
on the individual component. New ones do. These lengths are required for the Daniel CUI
3.10 software to calculate the new chord length.
To ensure that the required information is available, it is highly recommended that the original
Zero Flow Calibration Data Sheet be obtained from Daniel prior to replacing the transducer
pair.
The data sheet, shown in Figure 4-13, provides all of the serial numbers (S/N) and length data
pertaining to the meter’s original transducer assembly components. This data sheet can be obtained
by contacting Daniel Measurement Services at 713-827-6314.
•
Prior to replacing the transducers, it is imperative that each of the transducer pairs be
installed in the correct location (i.e. A1, A2, etc) since changing the transducers, or any
components of the assembly, will impact the chord length. Mark the location on the
Transducer Pair Data Sheet (see Figure 4-13, the data sheet included in the upgrade kit for
new transducers), and also mark the location on the side of the transducer with the marker
pen (see Figure 4-5 for these components). This will prevent any confusion when the
transducers are installed and the meter’s configuration parameters are updated. The values
on the Transducer Pair Data Sheet relate to a unique pair of transducers. Do not mix
transducers from different pairs.
•
It is recommended that only one pair (i.e. A1 & A2 or B1 & B2 etc.) of transducer assemblies
be removed at a time to avoid installing the transducer assemblies in the wrong position.
When this isn’t practical, before removing the transducer assemblies from the mount, ensure
that the transducer location (i.e. A1, A2, etc) is marked on one of the flats of the 1¼" holder
head (see Figure 4-4). Ensure the marking will not be accidentally erased during removal
of the old transducer.
•
When the transducers are replaced, the applicable chord parameters must be updated. Use
the Transducer Swap-out Wizard in Daniel CUI (see Figures 4-6 through 4-9 for screen shots
of this wizard). This procedure requires writing changes to the meter. If the write-protection
jumper (JP2 on the DFI Board) is installed, it will need to be removed.
3-2
KEY POINTS
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
D E C 2005
4.0
TRANSDUCER UPGRADE PROCEDURE
1.
Prior to starting the transducer replacements, a minimum of a two-minute maintenance log
file should be collected.
2.
Use the Edit/Compare Configuration Wizard to read the configuration from the meter.
3.
Name the configuration file clearly to indicate it is before the transducer upgrade (by editing
the Daniel CUI automated file name shown in the Edit/Compare Configuration Wizard
screen (e.g. "Meter Name, Meter config, T-2 Transducers, 3-3-2005 4-48-58 PM.cfg").
4.
When the upgrade procedure is complete and all of transducers are replaced and the data
updated, save the configuration file named "Meter Name, Meter config, T-11 Transducers,
3-3-2005 5-10-07 PM.cfg".
The chart in Figure 4-1 shows a typical SOS (Speed of Sound) comparison chart for a meter at zero
flow. It can be seen on the Charts sheet in the maintenance log file.
Figure 4-1. Typical SOS Comparison Chart
TRANSDUCER UPGRADE PROCEDURE
4-1
D E C 2005
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
If the meter cannot be depressurized, use the Ultrasonic Extraction tool to remove and
re-install the transducer assemblies.
Failure to follow the extractor tool manual procedures may result in serious personal injury or
death and may cause equipment damage.
If the meter cannot be depressurized, refer to the Daniel Ultrasonic Extractor Tool Operation
Manual (P/N 3-9000-729). The most recent copy of this manual may be downloaded from the Daniel
website (http://www.emersonprocess.com/daniel/). Follow the steps in Section 4 of the Ultrasonic
Extractor Tool Operation Manual for removing the transducers under pressure, instead of steps 5
through 8 below.
5.
Depressurize the meter and isolate the meter run piping, and vent the gas to atmosphere.
Verify that the pressure is 0 psig before proceeding.
6.
Prior to removing the holder cover, put an alignment mark on the edges of both the holder
cover and the mount cover using the marker pen. This will help facilitate rotational
alignment during the re-assembly. Remove the holder cover (see Figure 4-2) by completely
loosening the two captive screws. Gently pull back the cover and cable, unplug the cable and
expose the mount cover assembly.
Figure 4-2. Side and End View of M-Mount Transducer Assembly
4-2
TRANSDUCER UPGRADE PROCEDURE
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
7.
D E C 2005
After labeling the mount cover location (A1, A2, etc.), loosen the two screws in the mount
cover as shown in Figure 4-3, below. Note that these two screws may or may not be captive.
If they are not captive, remove them carefully. With the two screws fully unscrewed, pull the
mount cover off of the hex head of the holder exposing the mount and holder assembly.
Figure 4-3. Mount Cover
8.
Loosen the T-Slot transducer assembly holder with the 1¼" wrench or socket (see Figure 44). Once it has been loosened, it should be possible to unscrew it by hand.
Unless the meter/piping has a vent open to the atmosphere, pressure may build up in the
meter/piping due to minor valve leakage.
Ensure there is no pressure buildup before completely unscrewing the holder. Carefully pull
out the transducer assembly (Figure 4-5).
If the hex head of the holder does not clearly indicate the location of the transducer assembly
(A1, A2, etc. as shown in Figure 4-4), mark it now with a Sharpie® marker pen (or equivalent)
to ensure the assembly is re-installed in the correct location.
TRANSDUCER UPGRADE PROCEDURE
4-3
D E C 2005
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
Figure 4-4. Mount and T-Slot Transducer Assembly
9.
Loosen the three 6-32 set screws that hold the transducer in place. The screws do not need
to be completely removed. On some of the older assemblies, Loc-Tite™ is used on the set
screws. If the set screw is very tight and does not appear to come out without damage to the
wrench or the set screw, try heating the area around the set screw with a heat gun. Often this
will help in releasing the bond and allows the screw to be removed without damage. Use
caution when heating the stalk area around the set screw. Overheating may cause damage
to the stalk.
If the inside of the set screw is stripped due to Allen wrench slippage and the screws cannot be
extracted, it may be necessary to loosen the set screws by holding the stalk in place. If this is
the case, a new stalk must be obtained and installed. The length of the replacement stalk will
be required to update the meter configuration.
4-4
TRANSDUCER UPGRADE PROCEDURE
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
D E C 2005
Figure 4-5. Transducer Assembly
10.
After all three set screws have been loosened, as shown in Figure 4-5, remove the transducer
from the holder assembly.
11.
Replace the O-rings and back-up rings on the transducer holder. It is recommended that they
be replaced while the transducer is removed from the holder/stalk to prevent damage to the
transducer.
12.
Install the new transducer designated for that location. The parts are keyed so that they only
fit together one way. Tighten the three set screws equally to lock the new transducer into
place. DO NOT OVER TIGHTEN. Confirm the serial number of the transducer, and
installation location (i.e. A1, A2, etc.) previously marked on the Transducer Pair Data Sheet.
See annotated data sheet in Figure 4-13.
13.
Clean the transducer ports, mounts, and T-Slot transducer before re-installing the new
transducer pair.
TRANSDUCER UPGRADE PROCEDURE
4-5
D E C 2005
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
14.
Lubricate the O-rings/back-up rings with a small amount of Dow Corning® III silicon grease
(or equivalent).
15.
Apply a small amount of Nickel anti-seize compound (P/N 3-9960-134 or equivalent) to the
threads on the transducer holder.
SERIOUS PERSONAL INJURY OR DEATH POSSIBLE
If the extractor tool was used to remove the transducer assemblies while under pressure,
use the tool to re-install the transducers. Follow the procedures in Section 5 of the
Ultrasonic Extractor Tool manual, instead of step 16 in this procedure. Once the
transducer has been installed, continue with step 17.
Failure to follow the extractor tool manual procedures may result in serious personal injury or
death and may cause equipment damage.
16.
Carefully insert the assembly into the assigned transducer port. With the 1¼" wrench or
socket, tighten the holder in the mount, until it is snug and securely seated (specification
torque = 35 lb-ft). DO NOT OVER TIGHTEN.
17.
Repeat the entire procedure for each of the remaining transducer pairs being replaced.
18.
With the system sealed, apply minimal pressure to the meter. Using soapy water, or other
leak check method, check for leakage around the transducer mounts and holder. Fix any leak
before proceeding. If no leaks are detected, fully pressurize the meter to line conditions (or
at least 200 psi).
19.
Replace the mount covers in their original locations. Position them so the cover slips over
the 1 ¼” hex head screw of the transducer holder and the screw slots on the side allow access
to the screw-hole in the transducer mount underneath. Observe the alignment marks applied
in step 6. The metal locking plate on the surface of the cover can be rotated approximately
¼" to engage the plate on the 1¼" hex head of the holder. While holding the body of the
cover, rotate the locking plate by pushing on the tabs with the thumb and finger while
pushing down (see Figure 4-3). Once the locking plate is engaged and the cover is fully
seated on the mount, install the two screws that attach the cover to the mount.
4-6
TRANSDUCER UPGRADE PROCEDURE
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
D E C 2005
20.
Align the cable connector plug with the transducer connector pins in the holder, and plug in
the cable. The internal connector is keyed to align with the holder. Observe the alignment
marks applied in step 6 to align the screws in the holder cover with the holes in the mount
cover. Also, note that there are small “V” shaped marks on the sides of the mount cover to
align it with the holder cover.
21.
Once the transducers have been replaced, the meter configuration must be updated for the
new chord length, Delay times and Delta Delay times. This is most easily done using the
Transducer Swap-out Wizard in Daniel CUI 3.10.
22.
Prior to making any changes to the meter’s configuration, open the communications
connection to the meter and ensure that all the transducers are operating correctly. The SOS
difference graph will show different values, but all chords should be operational. If any chord
has failed, troubleshoot and repair before proceeding.
The following steps discuss configuration changes required to maintain the meter’s accuracy. If the
jumpers (Mark II electronics) are configured for write-protection, they must be removed (see
Appendix A for detailed procedures).
23.
If the configuration has not yet been read from the meter and saved, do so now. Figure 4-6
shows an example of the Edit/Compare Configuration wizard.
TRANSDUCER UPGRADE PROCEDURE
4-7
D E C 2005
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
Figure 4-6. Edit/Compare Configuration Screen
24.
Use the Daniel CUI Tools menu path and select Transducer Swap-out. The Test Meter dialog
box shown in Figure 4-7 displays. Select all the chords by clicking the check box beside the
chord name in the lower left-hand corner. A check mark appears in the box as shown in the
figure.
25.
Click the Next button to continue the configuration process (see Figure 4-8).
4-8
TRANSDUCER UPGRADE PROCEDURE
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
D E C 2005
Figure 4-7. Chord Parameters Adjustment Dialog
26.
From this dialog (Figure 4-8) select the components that have been replaced by clicking the
radio buttons beside the applicable component. Note that only one chord will be entered at
a time. The example in Figure 4-8 shows that only the transducer on Chord A was replaced
(chord indicated at the top of the Select Components dialog).
TRANSDUCER UPGRADE PROCEDURE
4-9
D E C 2005
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
Figure 4-8. Select Components for Chord A Dialog
27.
Click the Next button to display the Select Components for Chord A data entry dialog (see
Figure 4-9). Enter the lengths of the “Components Removed” (from original Zero Flow
Calibration Data Sheet), and the “Components Added”. The Daniel CUI software
automatically calculates the new chord length based on the data input.
28.
Next enter the Delay and Delta times for the new transducer pair (from data sheet sent with
each transducer pair).
4-10
TRANSDUCER UPGRADE PROCEDURE
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
D E C 2005
Figure 4-9. Update Chord A Data Entry Dialog
29.
Click the Write to Meter button to apply the new values to the meter configuration.
30.
Repeat this process for each of the remaining chords. Again, this process only permits
changes to the component information one chord at a time.
31.
Read the configuration from the meter. As suggested in step 1, give this file a name which
clearly indicates that it is the configuration after the transducers have been replaced (i.e.
"Meter Name, Meter config, T-11 Transducers, 3-3-2005 5-10-07 PM.cfg").
TRANSDUCER UPGRADE PROCEDURE
4-11
D E C 2005
32.
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
Perform a file comparison to verify the changes to the meter’s configuration. To perform this
comparison, use the Tools>Edit/Compare Configuration menu path.
(a) In the File section at the bottom of the screen click the Open button. Select the
directory and file folder in which the original configuration (created in step 1) was saved.
(b) Open the configuration file. The Daniel CUI configuration window displays the
configuration.
(c) Next, click the Compare button at the bottom of the screen. Select the directory and file
folder in which the new configuration (created in step 31) was saved.
(d) Open the configuration file. The Compare Configurations window displays. An example
of this comparison is shown in Figure 4-10. Only the "L" dimensions (Lx, Registers 312
- 318), Delay (AvgDlyx, Registers 332 - 338) and Delta times (Dltdlyx, Registers 340 346) should be different.
Figure 4-10. Compare Configurations Window
(e) Collect a maintenance log file for at least two minutes while the meter is flowing faster
than 3 ft./sec. Verify that the SOS of each chord is within 2 ft./sec. of the others. If a SOS
was calculated using AGA10 and a known gas composition, verify that the meter average
SOS is within 2 ft./sec. of the AGA10 SOS.
4-12
TRANSDUCER UPGRADE PROCEDURE
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
D E C 2005
Figure 4-11. SOS Comparison after Transducer Replacement
(f) If the security jumpers were removed, power down the meter and re-install the jumpers
to implement the write-protection. When complete, restore power to the meter and check
the meter to ensure proper operation using Daniel CUI.
(g) Disconnect the communication to the meter via Daniel CUI and install the end covers on
the meter head.
TRANSDUCER UPGRADE PROCEDURE
4-13
D E C 2005
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
Figure 12. Ultrasonic Flow Meter Zero Flow Calibration Report
4-14
TRANSDUCER UPGRADE PROCEDURE
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
D E C 2005
Figure 4-13. Transducer Pair Data Sheet
TRANSDUCER UPGRADE PROCEDURE
4-15
D E C 2005
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
This page intentionally left blank.
4-16
TRANSDUCER UPGRADE PROCEDURE
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
A.1
D E C 2005
REMOVAL AND RE-INSTALLATION OF WRITE-PROTECTION JUMPER ON
MARK II ELECTRONICS
If the jumper has been installed on JP2 on the DFI Board, then the meter configuration will be writeprotected. The effect of this jumper is to prevent changes to the values in the meter’s configuration.
When the transducers or any component of the transducer assembly are replaced, the values related
to the chord length “L”, the delay and the delta delay times will need to be revised in order to
maintain the meter’s accuracy. It is then necessary to remove the jumper, change the applicable
configuration values and re-install the jumper.
APPENDIX A
A-1
D E C 2005
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
Figure A-1. Mark II Electronics
A-2
APPENDIX A
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
D E C 2005
Figure A-2. View of Mark II Electronics Board Stack from End of Enclosure
1.
From Daniel CUI use the Meter pull-down menu and select Meter Information. The Meter
Information dialog box appears (see Figure A-2). The S/W Ver. shows the firmware version
installed in the meter electronics. If the version is 3.34 or earlier, then jumpers JP2 on the
DFI Board and JP9 on the CPU must be removed in order to modify the meter’s
configuration. If the firmware version is 3.62 or later, only jumper JP2 must be removed to
apply the changes. Once the firmware version has been determined, click the Close button
and exit Daniel CUI..
APPENDIX A
A-3
D E C 2005
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
Figure A-3. Daniel CUI Meter Information Dialog
2.
Turn off power to the electronics.
3.
Disconnect the cable originating from the Field Connection Board by unplugging the
connector on the Power Supply Board (see Figure A-3). Loosen the screws completely to
unplug the connector.
4.
Completely loosen the four captive screws that hold the Electronics Board stack in the
enclosure. With the cable held to the side, grasp the screws and pull the assembly straight out
of the enclosure. The three-board assembly consisting of the Power Supply Board, the DFI
Board and the CPU Board are removed as a single unit.
A-4
APPENDIX A
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
5.
D E C 2005
As shown in Figure A-4, jumper JP2 is located near the outer edge of the DFI Board. Access
to the jumper is possible without disassembling the three boards. Using needle nose pliers,
carefully lift the jumper from the contact pins. One side of the jumper can be installed on
one of the pins so that the jumper is not lost. If the firmware version is 3.62 or later, then the
electronics can be re-installed so that the configuration can be changed. If the firmware
version is 3.34 or earlier, then jumper JP9 on the CPU Board (see Figure A-5) will also need
to be removed. As with jumper JP2 on the DFI Board, jumper JP9 on the CPU Board can
also be carefully removed by using a pair of needle nose pliers.
Figure A-4. DFI Board Showing Jumper JP2
APPENDIX A
A-5
D E C 2005
DANIEL T-2 TO T-11 TRANSDUCER UPGRADE INSTRUCTIONS
Figure A-5. CPU Board Showing Jumper JP9
6.
Align the electronics assembly and re-install it in the enclosure, ensuring that the connection
between the electronics and the enclosure is correctly engaged. Tighten the four captive
screws and reconnect the cable from the Field Connection Board.
7.
Power up the electronics. With the write-protection jumper removed, it is now possible to
change values, which influence the measurement function, in the meter configuration.
8.
After the values in the configuration have been changed, and the meter is operating within
the expected tolerances, re-install jumper JP2 on the DFI Board, and jumper JP9 on the CPU
Board (if removed) using the same procedure as outlined above, except that the jumper needs
to be re-installed to the position it was originally found, bridging both jumper contact pins.
A-6
APPENDIX A
NOTES
NOTES
WARRANTY CLAIM PROCEDURES
To make a warranty claim, you, the Purchaser, must:
1.
Provide Daniel with proof of the Date of Purchase and proof of the Date of Shipment of the
product in question.
2.
Return the product to Daniel within twelve (12) months of the date of original shipment of
the product, or within eighteen (18) months of the date of original shipment of the product
to destinations outside of the United States. The Purchaser must prepay any shipping charges.
In addition, the Purchaser is responsible for insuring any product shipped for return, and
assumes the risk of loss of the product during shipment.
3.
To obtain Warranty service or to locate the nearest Daniel office, sales, or service center call
(713) 827-6314, Fax (713) 827-6312 write to:
Daniel Measurement Services
11100 Brittmoore Park Drive
Houston, Texas 77041
Or contact Daniel via the following site:
www.emersonprocess.com/daniel
4.
When contacting Daniel for product service, the Purchaser is asked to provide information
as indicated on the following page entitled "Customer Repair Report".
5.
For product returns from locations outside the United States, it will be necessary for you to
obtain the import consignment address so that Daniel’s customs broker can handle the
importation with the U.S. Customs Service.
6.
Daniel Measurement Services offers both on call and contract maintenance service designed
to afford single source responsibility for all its products.
7.
Daniel reserves the right to make changes at any time to any product to improve its design
and to insure the best available product.
DANIEL MEASUREMENT AND CONTROL, INC.
CUSTOMER REPAIR REPORT
FOR SERVICE, COMPLETE THIS FORM, AND RETURN IT ALONG WITH THE AFFECTED EQUIPMENT TO CUSTOMER
SERVICE AT THE ADDRESS INDICATED BELOW.
COMPANY NAME:
TECHNICAL CONTACT:
REPAIR P. O. #:
PHONE:
IF WARRANTY, UNIT S/N:
INVOICE ADDRESS:
SHIPPING ADDRESS:
RETURN SHIPPING METHOD:
EQUIPMENT MODEL #:
S/N:
FAILURE DATE:
DESCRIPTION OF PROBLEM:
WHAT WAS HAPPENING AT TIME OF FAILURE?
ADDITIONAL COMMENTS:
REPORT PREPARED BY:
TITLE:
IF YOU REQUIRE TECHNICAL ASSISTANCE, PLEASE FAX OR WRITE THE CUSTOMER SERVICE DEPARTMENT AT:
DANIEL FIELD SERVICES
ATTN: CUSTOMER SERVICE
11100 BRITTMOORE PARK DRIVE
HOUSTON, TEXAS 77041
FOR FASTEST SERVICE CONTACT DANIEL VIA OUR WEBSITE:
www.emersonprocess.com/daniel
PHONE: (713) 827-6314
FAX: (713) 827-6312
The sales and service offices of Daniel Measurement and Control are located
throughout the United States and in major countries overseas.
Please contact Daniel Measurement Services at
11100 Brittmoore Park Drive, Houston, Texas 77041, or phone (713) 467-6000
for the location of the sales or service office nearest you.
Daniel Measurement Services offers both on-call and contract
maintenance service designed to provide single-source
responsibility for all Daniel products.
Daniel Measurement and Control, Inc., and Daniel Measurement Services, Inc.
Divisions of Emerson Process Management reserves the right to make changes to any of its products or services
at any time without prior notification in order to improve that product or service and to supply
the best product or service possible.
www.emersonprocess.com/daniel
© Copyright 2026 Paperzz