Quick Start Guide D-7010-0052, Rev A May 2016 Rosemount CT5100 Continuous Gas Analyzer Safety Precautions Operators, maintenance personnel, and authorized users must observe the following safety precautions and warnings. DANGER! ELECTRIC SHOCK The CT5100 Continuous Gas Analyzer (CT5100 CGA) operates using mains voltage, which may cause death or serious injury to personnel. Make sure that the circuit breakers are set to Off and locked out and tagged out off before removing the top cover or opening the front cover. Death, personal injury, and/or damage to persons and/or property may result if this is not observed. WARNING! FIRE, BURN, AND OPTICAL RADIATION EXPOSURE HAZARD Operators and service personnel do not have access to the laser/electrical and the upper cell compartments for maintenance or service. Electrical shock or thermal burns may occur. WARNING! FLAMMABLE SUBSTANCES Some parts of the CT5100 CGA may reach temperatures of 190 °C (374 °F) and may present an ignition source. Exercise care when using oil, paint, cleaning rags, or other flammable substances near the CT5100 CGA. A fire may result if this precaution is not observed. The interior of a CT5100 CGA is always hot unless it has been switched off and allowed to cool down. WARNING! EXPLOSION HAZARD Always lock out the gas handling system when shutting down the CT5100 CGA. Unauthorized performance on the CT5100 CGA or its associated pipes/hoses may result in highly flammable gas being released, causing fire or explosion. Failure to lock out gas handling system may cause death. WARNING! EXPLOSION HAZARD The sample gas in the system must be vented to prevent fire or explosion during maintenance and to prevent damage to the CT5100 CGA during startup. The sample gas in the pipes leading to the CT5100 CGA must be purged to prevent hazards to personnel during maintenance. You must purge the sample gas in accordance with the safe working procedures for this site. Allow the CT5100 CGA and the system for returning the sample gas to run for five minutes to allow any sample gas in the instrument to be returned to the exhaust. Failure to vent sample gas may cause death. WARNING! BURNS Some parts of the CT5100 CGA may be heated to 190 °C (374 °F). To prevent burns, do not touch any of the hot parts. All parts of a CT5100 CGA are always hot unless it has been switched off and allowed to cool down. Before fitting, removing, or performing any maintenance on the CT5100 CGA, make sure that it has been switched off and allowed to cool for at least two hours. Before performing any maintenance on, or in the vicinity of, the analysis cell, allow the CT5100 CGA to cool for at least twelve hours as the analysis cell is insulated against heat loss. When handling the CT5100 CGA, always use suitable protective gloves. Personal injury and/or damage to property may result if these precautions are not observed. These precautions are particularly important when working at heights. If a burn is received, seek medical treatment immediately. WARNING! OPTICAL RADIATION EXPOSURE HAZARD The CT5100 CGA contains lasers. Opening the CT5100 CGA and attempting to perform adjustments and procedures other than those specified in this manual may result in hazardous optical radiation exposure. All lasers used within the CT5100 CGA are Class 1. The combined laser powers are sufficiently low at the first accessible aperture that the unprotected eye will not be damaged. This class is eye safe under all operating conditions. Failure to follow the correct procedures may cause damage to the eye. Do not look at the lasers with any kind of magnifier or optical measuring device. There are three types of laser that may be included in the CT5100 CGA: Quantum Cascade Lasers (QCLs), Interband Cascade Lasers (ICLs), and diode lasers. The lasers within the CT5100 CGA are Class 1. The characteristics of the lasers contained within the CT5100 CGA are given in the table below. Parameter QCL ICL Diode Operation mode Pulsed Pulsed Pulsed Lasers per system 1-6 1-6 1-6 Wavelength 4 - 10 µm 2- 5 µm Approximately 760 nm Power < 5 mW < 5 mW < 5 mW Pulse duration < 1 µs < 1 µs < 5 µs Pulse repetition frequency < 100 kHz < 100 kHz < 100 kHz Duty cycle < 5% < 5% < 25% Comment Maximum of 6 lasers per system Combined power of QCL at first accessible aperture: < 9.62mW The combined power of the QCL, ICL, and diode lasers at the first accessible aperture is < 9.62 mW. The CT5100 Continuous Gas Analyzer has warning labels in appropriate positions according to USA 21 CFR 1040.10. The location of laser safety labels on the CT5100 Continuous Gas Analyzer is specified in Section 4.3. The use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure. WARNING! HAZARDOUS SUBSTANCES The CT5100 CGA may contain hazardous substances. Always handle CT5100 CGA assemblies and components with extreme caution. Wear personal protective equipment (PPE) when handling the equipment. Gas handling components within the CT5100 CGA contain particulate matter residue from the sample gases. Over the life of the CT5100 CGA, the concentration of particulate matter will become enriched within the gas handling components. When performing repairs and maintenance on the CT5100 CGA: • Handle used gas handling components with extreme caution. • Avoid direct skin contact with used gas handling components. • Do not smoke, drink, or eat in the work area. • Wear goggles or eye shields. • Wear a suitable face mask to protect against inhalation of particulate matter. • Keep food and beverages away from the CT5100 Continuous Gas Analyzer. • Do not wet fingers, eyes, or any exposed skin. • Pack used gas handling components for disposal in sealed packaging and label them Contaminated. • Dispose of contaminated items as hazardous material in accordance with applicable local, national, or international health and safety regulations and pollution regulations. WARNING! TRANSPORTATION HAZARD Handle the CT5100 CGA with caution during unpacking, installation, maintenance, and transport to prevent crushing of hands, feet, or other body parts. The CT5100 CGA weighs 50.3 kg (111 lb.) and should always be lifted and moved using suitable lifting/moving equipment. Wear suitable protective gloves and protective footwear. When preparing the CT5100 CGA for transport by air, road, or rail, safeguard the CT5100 CGA against movement or break-away during transport by securely strapping it in place. WARNING! HAZARDOUS GAS The product stream that the CT5100 CGA is examining may be hazardous even at low concentrations. Special care should therefore be taken to ensure that the sample gas return port either returns the sample gas to the product stream or discharges the sample gas to a location that will not cause a hazard. WARNING! HIGH PRESSURE GAS AND AIR The calibration gas supply and compressed air supply operate at a pressure that can cause injury, e.g., damage to eyes and skin punctures from debris blown by the high pressure gas or compressed air. Always lock off or tag out the calibration gas supply and compressed air supply when shutting down the CT5100 CGA. CAUTION! EQUIPMENT DAMAGE Do not power up or try to operate the CT5100 CGA unless it is physically secure and all electrical and pneumatic connections to the CT5100 CGA are in place. Before commencing the startup process, it is important to ensure that electrical power, sample gas handling facilities, and any calibration gases that are required are available to the CT5100 CGA. Failure to perform pre-system startup checks may cause damage to equipment. CAUTION! EQUIPMENT DAMAGE Always follow the startup procedure. Damage to the CT5100 CGA may result from a failure to follow this procedure. CAUTION! EQUIPMENT DAMAGE Always follow the shutdown procedure. Damage to the CT5100 CGA may result from a failure to follow this procedure. CAUTION! LASER RADIATION Invisible laser radiation. Do not stare into beam. Class 1 laser. Failure to follow the correct procedures may cause damage to the eye. CAUTION! EMC This is a Class A product. In a domestic environment, this product may cause radio interference, in which case the user may be required to take adequate measures. Contents Contents Chapter 1 Plan ..................................................................................................................................1 1.1 1.2 1.3 Chapter 2 Install ...............................................................................................................................9 2.1 2.2 2.3 Chapter 3 System overview ............................................................................................................................ 9 Gas inputs and outputs ................................................................................................................ 10 Electrical/electronic inputs and outputs ....................................................................................... 12 Navigating the Displays and Indicators ........................................................................... 15 3.1 Chapter 4 Unpacking the Emerson CT5100 Continuous Gas Analyzer (CGA) .................................................. 1 Mounting the CT5100 Continuous Gas Analyzer .............................................................................3 1.2.1 Tools required ................................................................................................................. 6 Detailed system specifications ....................................................................................................... 6 Controls and indicators ................................................................................................................ 15 3.1.1 Front panel controls and indicators ................................................................................15 3.1.2 Display controller .......................................................................................................... 17 3.1.3 Gas sensor main screen ................................................................................................. 18 3.1.4 Pressure and Temperature screen .................................................................................... 20 3.1.5 Help system ...................................................................................................................20 3.1.6 Calibration .....................................................................................................................21 3.1.7 BACK button .................................................................................................................. 21 Startup .......................................................................................................................... 23 4.1 4.2 Preparation for use .......................................................................................................................23 Startup .........................................................................................................................................23 Quick Start Guide i Contents ii Rosemount CT5100 Plan 1 Plan Topics covered in this chapter: • • • 1.1 Unpacking the Emerson CT5100 Continuous Gas Analyzer (CGA) Mounting the CT5100 Continuous Gas Analyzer Detailed system specifications Unpacking the Emerson CT5100 Continuous Gas Analyzer (CGA) This procedure requires a minimum of two people to safely remove the equipment from the shipping container. WARNING! LIFTING HAZARD The equipment weighs 50.35 kg (111lbs) and requires a two man lift. Local health and safety procedures must be followed. Failure to follow these procedures may result in serious personal injury. WARNING! TRANSPORTATION HAZARD Use safety approved lifting equipment. The user must ensure the equipment is tested, meets the lifting ratings for the weight of the equipment, and is in good operational condition. Failure to verify equipment meets the lifting ratings and is in good operational condition may cause injury to personnel or damage the CT5100 CGA. Quick Start Guide 1. Visually inspect the exterior of the CT5100 CGA for signs of damage, corrosion, gas leaks, or signs of previously overheating. 2. Report anything found to the maintenance organization. 3. Attach suitably rated and test lifting slings to the safety engineered lifting eye bolts mounted on top of the CT5100 CGA. 4. One person should carefully guide the equipment from the horizontal to vertical position while the other person lifts the equipment. 5. Use safety approved and tested lifting equipment to remove the CT5100 from the shipping container. 1 Plan Figure 1-1: CT5100 - rear view showing lifting eyes and 4 off Ø10.5 mounting holes 6. 2 Place the CT5100 CGA on a solid, level surface and prepare to wall mount the analyzer. Rosemount CT5100 Plan 1.2 Mounting the CT5100 Continuous Gas Analyzer This procedure requires two people to safely move and mount the CT5100 CGA. Quick Start Guide 1. Use safety approved equipment to lift the CT5100 CGA from the stable platform. 2. One person guides the unit into position as the other person carefully lifts the CT5100 CGA. 3. Use four M8 (3/8 inch) fasteners to attach the CT5100 CGA to the wall mount brackets (Items A and B). 4. Make sure the bolts are secure. Do not over tighten the fasteners. 5. Remove the double swivel lifting eyes (Figure 1-1). 6. The lifting eye connection threads must be adequately protected with plastic grommets and or a suitable grease. 3 Plan Figure 1-2: CT5100 CGA mounting holes A. CT5100 CGA right side mounting holes B. CT5100 CGA left side mounting holes 4 7. Remove the lifting eye bolts from the CT5100 CGA after lifting is completed and store them in an appropriate case or container per their manufacturer's recommendation. 8. Apply lubricant or anti-seize to the threads of the plug bolts and install the plug bolts in the tapped holes in the CT5100 CGA to keep the holes free of debris and to prevent corrosion. Rosemount CT5100 Plan Figure 1-3: CT5100 CGA Top view plug bolt placement A. Left side eye bolt plug placement B. Right side eye bolt plug placement Quick Start Guide 5 Plan 1.2.1 Tools required The CT5100 CGA is shipped as a complete unit and requires minimal tools for installation. The minimum tools required are: 1.3 • 2 (14 mm) spanner wrenches (or 2 off adjustable spanners) for gas fittings • Small flat head screwdriver for wiring electrical contacts • Large flat head screwdriver to open main enclosure • Suitable spanner wrench for securing mounting bolts Detailed system specifications Table 1-1 gives the physical characteristics of the CT5100 CGA. Schematic diagrams of the sensor and mounting points are shown in Figure 1-1 and Figure 1-2. Table 1-2 gives the general characteristics of the instrument. Table 1-1: Physical characteristics CT5100 CGA Value Units Comment Gas analyzer dimensions 575 x 286 x 781 mm Length x Width x Height 22.64 x 11.25 x 30.8 in. Gas analyzer weight 50.35 kg 111 lb. Table 1-2: General characteristics CT5100 CGA Value Units Comment Instrumentation supply voltage 100 or 240 V AC 50/60 HZ +/- 10% Peak power consumption 500 W Max consumption per gas analyzer Inlet gas port connector 6 mm or 1/4 in. Swagelok type, factory configured, specify on order Outlet (Exhaust) gas port connector 6 mm or 1/4 in. Swagelok type, factory configured, specify on order Measurement result signals 4 to 20 10/100 mA 4 or 8 channel outputs, specify on order Mbps Ethernet Table 1-3 gives the environmental characteristics of the CT5100 CGA. 6 Rosemount CT5100 Plan Table 1-3: Environmental characteristics Environmental characteristic Value Units Comment Operating temperature range 5 to 45 °C Ambient temperature 41 to 113 °F 50 to 195 °C 122 to 383 °F Sample gas moisture content 20 % Sample gas particulate density 5 mg/m3 Sample gas particulate size 10 μm IP code 66 Sensor humidity range 10 to 95 % Operating altitude 0 to 2,000 m Sample gas temperature range Quick Start Guide Factory set, specify on order Maximum IP to IEC 60529 Relative humidity (non-condensing) at 45 °C (113 °F) 7 Plan 8 Rosemount CT5100 Install 2 Install Topics covered in this chapter: • • • 2.1 System overview Gas inputs and outputs Electrical/electronic inputs and outputs System overview A complete CT5100 CGA, as shown in Figure 2-1, consists of a gas handling system, the CT5100 CGA, and the associated interconnecting wiring and gas piping. Measurement data from the CT5100 CGA can be displayed in the process control center. The CT5100 CGA is supplied by Emerson. The gas handling system may be provided by either you or Emerson, depending upon the specific installation. The circuit breakers used to control the application of electrical power to the CT5100 CGA, the interconnecting wires, and gas piping are provided by you. In Figure 2-1, the items supplied by us are colored blue; the items supplied by you are colored purple. The green Gas Handling System may be provided by you or us. Table 2-1 lists the main items of the system. Figure 2-1: CT5100 CGA installation There is no sample conditioning provided within the CT5100 CGA; the sample gas must be brought within the parameters shown in Table 1-3 before entering the CT5100 CGA. Quick Start Guide 9 Install Table 2-1: Main items of the CT5100 CGA installation 2.2 Item Name or description Supplied by Part number Quantity Notes 1 CT5100 CGA Emerson CT5100 1 2 CT5100 CGA software package, version 4.0 Emerson N/A. 1 3 Gas handling system Customer (optionally by Emerson) Customer choice or Emerson 1 4 Heated gas sample line hose Customer Customer choice 1 5 Exhaust line hose (for sample gas) Customer Customer choice 1 6 Reference gas cylinders Customer (instrument gas) for calibration purposes Customer choice Dependent upon number of gases being measured 7 Pressure regulator Customer Customer choice 1 per gas cylinder Required for calibration 8 Pneumatic T-piece Customer Customer choice 1 Required for calibration 9 Excess flow line Customer Customer choice 1 Required for calibration 10 Power cables to CT5100 Customer CGA Customer choice 1 11 Cables from CT5100 CGA to control center Customer Customer choice 1 12 Main circuit breaker Customer Customer choice Version described in manual Software is embedded in system. Gas inputs and outputs The CT5100 CGA has one gas input and one gas output (Figure 2-2). 10 1. The gas sample that is to be measured for impurities enters the instrument through the sample gas input port (A) located on top of the CT5100 CGA. 2. Once the gas sample has been examined for impurities, it is expelled from the instrument through the sample gas return port (B). Rosemount CT5100 Install Figure 2-2: Gas inlet and outlet connectors A. B. C. D. 3. Sample gas input port Sample gas return port Top cover of CT5100 CGA (cell compartment) Laser/electrical compartment The sample supply line must be heated all the way to the sample gas input port on the CT5100 CGA to prevent condensation forming at any point in the sample supply line. DANGER! HAZARDOUS GASES The product stream that the CT5100 CGA is examining may be hazardous even at low concentrations. Special care should be taken to ensure that the sample gas return port either returns the sample gas to the product stream or discharges the sample gas to a location that will not cause a hazard. Hazardous gases may cause death. Quick Start Guide 11 Install 2.3 Electrical/electronic inputs and outputs Figure 2-3: Electrical / electronic connectors A. B. C. Analog/digital conduit Analog/digital conduit Power conduit Electrical/electronic connections to the CT5100 CGA are made through three electrical conduits located on the underside of the instrument, as shown in Figure 2-3. Use the wiring diagram to make the electrical user connections as shown below. 12 Rosemount CT5100 Install Terminal Note: Maximum number of user terminals shown. These may be reduced dependant on number and type of Analog/Digital outputs required. Function 1 SENSOR SYSTEM SUPPLY (L) 2 SENSOR SYSTEM SUPPLY (N) 3 EARTH 4 PURGE ALARM CONTACTS (Not used) 5 PURGE SYSTEM SUPPLY (L) (Not used) 6 PURGE SYSTEM SUPPLY (N) (Not used) 7 EARTH 8 PURGE SOLENOID (L) (Not used) 9 PURGE SOLENOID (N) (Not used) 10 EARTH 11 DIGITAL OUTPUT 1 12 DIGITAL OUTPUT 2 13 DIGITAL OUTPUT 3 14 DIGITAL OUTPUT 4 15 DIGITAL OUTPUT 5 16 DIGITAL OUTPUT 6 17 DIGITAL OUTPUT 7 18 DIGITAL OUTPUT 8 19 DIGITAL OUTPUT 9 20 DIGITAL OUTPUT 10 21 DIGITAL OUTPUT 11 22 DIGITAL OUTPUT 12 23 ANALOG OUTPUT 1 24 ANALOG OUTPUT 2 25 ANALOG OUTPUT 3 26 ANALOG OUTPUT 4 27 ANALOG OUTPUT 5 28 ANALOG OUTPUT 6 29 ANALOG OUTPUT 7 30 ANALOG OUTPUT 8 31 STATUS OUTPUT 1 (Check function) 32 STATUS OUTPUT 2 (Maintenance Required) 33 STATUS OUTPUT 3 (Out of Specification) 34 STATUS OUTPUT 4 (Failed) Electrical power is applied to the instrument through the power conduit (C). Quick Start Guide 13 Install The power supply is 110 to 240 Vac, 50/60 Hz ± 10%. AC to DC power converters inside the instrument automatically adjust in response to the input voltage level and ensure that the correct DC voltage is available inside the instrument. The instrument is electrically protected by an internal 5 A, 250 VA fast acting fuse on the instrumentation electrical supply line and an internal 2 A, 250 VA fuse on the purge electrical supply line. CAUTION! You must ensure that the mains supply cable used is of a suitable rating for the unit power requirements. Failure to do so may result in personal injury and/or damage to persons and/or property. The digital outputs conduit (B) provides an Ethernet output from the instrument that may be used for downloading data for failure diagnosis purposes or for downloading data to the process control center. The results of the gas analysis are output from the instrument through the 4-20 mA analog or Modbus outputs and sent to your process control center. The optional digital outputs provide fault indications to your process control center. Each digital output is connected to a normally closed relay, located inside the CT5100 CGA, which will open in response to the detection of a specific fault. The possible causes of a fault indication are: 1. The sample gas concentration is outside of specification, i.e., the sample gas concentration has exceeded the measurement range of the instrument. 2. The CT5100 CGA is out of specification or has developed a fault. NOTICE The installation of the gas analyzer shall be in accordance with National Electrical Code (NEC). 14 Rosemount CT5100 Navigating the Displays and Indicators 3 Navigating the Displays and Indicators 3.1 Controls and indicators The controls and indicators on the CT5100 CGA are described in Section 3.1.1 of this manual. 3.1.1 Front panel controls and indicators The CT5100 CGA is configured from the two control displays located on the front panel (Figure 3-1). Quick Start Guide 15 Navigating the Displays and Indicators Figure 3-1: Front panel A. Display controller NOTICE On/Off circuit breakers There are no On/Off switches on the CT5100 CGA. The application of electrical power to the instrument is controlled through an external circuit breaker. The circuit breaker is a simple 2-pole on/off circuit breaker that must be set to On to permit the safe operation of the CT5100 CGA. Operation of the CT5100 CGA is controlled primarily through the display controller (Figure 3-2). 16 Rosemount CT5100 Navigating the Displays and Indicators Figure 3-2: Display controller 3.1.2 Display controller Figure 3-3: Display controller buttons A. B. C. D. E. F. G. LCD display Configurable button Scroll button Configurable button Configurable button Scroll button Configurable button Operation of the CT5100 CGA is controlled through the six buttons on the display controller: Figure 3-3. The LCD display (A) can be used to display: Quick Start Guide 1. The gas concentration measurements obtained 2. The operating temperature and pressure 17 Navigating the Displays and Indicators 3. Help screens 4. Step-by-step calibration and diagnostics The two scroll buttons (C and F) are used to scroll through the information on the LCD display. The right-hand scroll button (C) is used to scroll up, and the left-hand scroll button (F) is used to scroll down. The other four buttons (B, D, E, and G) are configured to perform different functions according to which the software screen is shown on the LCD display. 3.1.3 Gas sensor main screen When the CT5100 CGA is switched on, at the end of the startup procedure, the gas sensor main screen (Figure 3-4) appears. The gas sensor main screen is the screen that is normally displayed. NOTICE The gas concentrations shown in the following screenshots may be different from those shown in your particular CT5100 CGA. They indicate the functionality of the software, which it the same regardless of the gases or gas concentrations being measured. Figure 3-4: Gas sensor main screen The gas sensor main screen displays the gas concentration measurements obtained by the CT5100 CGA. In the example shown in Figure 3-4, the gases nitric oxide (NO), nitrogen dioxide (NO2), oxygen (O2), carbon monoxide (CO), and carbon dioxide (CO2) are being measured, and for each gas the concentration detected is in parts per million (ppm) or percentage as applicable. At the end of the startup procedure, the gas measurements initially appear as 0.00 ppm until the first readings are taken. After a few seconds, the initial gas concentrations are displayed. 18 Rosemount CT5100 Navigating the Displays and Indicators The gas sensor main screen also shows the status of the CT5100 CGA. In the example shown in Figure 3-5, the instrument is Running and OK (i.e., no faults have been identified). This area of the display shows any errors detected by the software. On the software screens, highlighted items are links to other screens in the software. To access a screen, press the button next to the highlighted item. Figure 3-5: Gas sensor main screen buttons A. B. C. D. E. F. G. MENU button MENU text PAGE text PAGE button HELP button HELP text STATUS button PAGE (C) is a link between the gas sensor main screen and the Pressure and Temperature screen (described in Section 5.9). Press the PAGE button (D) to toggle between these two screens. HELP (F)is a link to the Help system. Press the HELP button to go to the HELP screen (described in Section 3.1.5). MENU (B) is a link to the main menu of the software. Press the MENU button (A) to go to the MAIN MENU screen (described in Section 3.1.6). On the gas sensor main screen, the STATUS button (G) has no function when the CT5100 CGA is operating correctly. If, however, the software detects a fault, an error message is displayed. Press the STATUS button to get further information on the error. Quick Start Guide 19 Navigating the Displays and Indicators 3.1.4 Pressure and Temperature screen The Pressure and Temperature screen (Figure 3-6) shows pressure and temperature measurements taken inside the CT5100 CGA. Figure 3-6: Pressure and Temperature screen The Cell reading is the temperature, in °C, of the analysis cell. The Gas reading is the temperature, in °C, of the gas within the analysis cell. The Pres reading is the pressure, measured in Torr, inside the analysis cell. NOTICE A Torr is a non-SI unit of pressure defined as 1/760 of standard atmospheric pressure and is equal to the fluid pressure of 1 mm of mercury. CAUTION! UNSERVICEABLE EQUIPMENT If the Pressure and Temperature screen does not display measurements similar to those shown in Figure 3-6, contact your local service representative for guidance. 3.1.5 Help system The CT5100 CGA software includes a context-sensitive help system. Press the HELP button, which is available on most of the software screens, to open the help system. The help system contains a number of different help screens, each conveying a different message. As the help system is context-sensitive, when the HELP button is pressed, the help screen that appears is the one that is most appropriate to the software function engaged when the HELP button was pressed. Figure 3-7 shows an example of a help screen. 20 Rosemount CT5100 Navigating the Displays and Indicators Figure 3-7: Example of a help screen 3.1.6 Calibration To access the Main menu (Figure 3-8), press the MENU button on either side of the gas sensor main screen (Figure 3-4) or the Pressure and Temperature screen (Figure 5-7). The Main menu is used for calibration, diagnostics, fault finding, downloading data, and shutting down the CT5100 CGA. Figure 3-8: Main menu 3.1.7 BACK button On most of the software screens, the top left-hand button (Figure 3-9) is configured as a BACK button. Press the BACK button to return to the previous screen. Quick Start Guide 21 Navigating the Displays and Indicators Figure 3-9: BACK button A. 22 BACK button Rosemount CT5100 Startup 4 Startup Topics covered in this chapter: • • 4.1 Preparation for use Startup Preparation for use The CT5100 CGA must be installed and fully commissioned prior to startup. CAUTION! EQUIPMENT DAMAGE. Do not power up or try to operate the CT5100 CGA unless it is physically secure and all electrical and pneumatic connections to the CT5100 CGA are in place. Before commencing the startup process, ensure that electrical power, sample gas handling facilities, and any calibration gases that are required are available to the CT5100 CGA. Failure to perform pre-system startup checks may cause damage to equipment. WARNING! BURN HAZARD Some parts of the CT5100 CGA may be heated to 165 °C (329 °F). To prevent burns, do not touch any of the hot parts. All components of a CT5100 CGA are hot unless it has been switched off and allowed to cool down. Before fitting, removing, or performing any maintenance on the CT5100 CGA, make sure that it has been switched off and allowed to cool for at least two hours. Before performing any maintenance on, or in the vicinity of, the analysis cell, allow the CT5100 CGA to cool for at least twelve hours, as the analysis cell is insulated against heat loss. When handling the CT5100 CGA, always use suitable protective gloves. Personal injury and/or damage to property may result if these safety precautions are not observed. These precautions are especially important when working at heights. If a burn is received, seek medical treatment immediately. 4.2 Startup NOTICE The gases shown in the screenshots and the measurements thereof may be different from those shown in your particular CT5100 CGA. They indicate the functionality of the software, which is the same regardless of the gases being measured. Quick Start Guide 23 Startup NOTICE To stop the startup procedure at any time, set the main circuit breaker to Off. To start up the CT5100 CGA, perform the following steps: 24 1. Visually inspect the exterior of the CT5100 CGA for signs of damage, corrosion, gas leaks, or overheating. Report anything found to the maintenance organization. 2. Make sure that the top cover is fitted to the CT5100 CGA. If it is not, report it to the maintenance organization and do not proceed further until the top cover has been fitted. 3. Make sure that the door to the electrical compartment is closed and locked. If the door cannot be closed and locked, report it to the maintenance organization and do not proceed further until the door has been repaired. 4. Make sure that the gas handling system is turned off. 5. Make sure that the external main circuit breaker is set to Off. 6. Visually examine the gas ports (if necessary, refer to Figure 4-1) to make sure that the sample supply line and the sample return line are correctly attached. Rosemount CT5100 Startup Figure 4-1: CT5100 CGA Gas Supply Lines A. B. C. D. E. F. G. Quick Start Guide Sample gas input port Sample gas return port Top cover of CT5100 CGA (cell compartment) Laser/electrical compartment Power conduit Analog/digital conduit Analog/digital conduit 7. Visually check that the electrical connections have been made to the three conduit hubs at the base of the instrument. See Figure 4-1. 8. Set the main circuit breaker to On. 25 Startup The CT5100 CGA commences its automatic startup. The control PC that forms part of the instrument is configured to automatically load the necessary gas sensor software and configuration files. The startup sequence commences automatically under software control. After a few seconds, the Gas Sensor main screen (Figure 4-2) appears on the display controller. If it does not, report the fault to the maintenance organization. Figure 4-2: Gas sensor main screen 9. Start up the system for returning the sample gas. 10. Start up the gas handling system that conditions the sample gas before it is fed into the CT5100 CGA. At the end of the startup procedure, the gas measurements initially appear as 0.00 ppm until the first readings are taken. After a few seconds, the initial gas concentrations are displayed. The startup procedure is now complete. 26 Rosemount CT5100 Startup Quick Start Guide 27 D-7010-0052 Rev A 2016 Cascade Technologies Emerson Process Management Glendevon House Castle Business Park Stirling, FK9 4TZ T +44 1786 447 721 F +44 1786 475 822 [email protected] ©2016 Rosemount. All rights reserved. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners.
© Copyright 2025 Paperzz