Use of Polystyrene-Block-Poly(MethylacrylateRandom-Octadecylate)-Block-Polystyrene as an On Demand Adhesive Schaun Ginesi Department of Chemical Engineering Honors Research Project Submitted to The HonorsCollege Approved: Accepted: ______________________ Date ________ Honors Project Sponsor (signed) __________________ Date _________ Department Head (signed) ______________________________ Honors Project Sponsor (printed) _______________________ Department Head (printed) ______________________ Date _______ Reader (signed) __________________ Date ________ Honors Faculty Advisor (signed) _________________________ Reader (printed) _____________________________ Honors Faculty Advisor (printed) ______________________ Date _______ Reader (signed) __________________ Date _________ Dean, HonorsCollege _________________________________ Reader (printed) Table of Contents Executive Summary ................................................................................................................................. 3 Introduction & Purpose ........................................................................................................................... 4 Background ............................................................................................................................................. 5 Experimental Procedure .......................................................................................................................... 8 Results & Discussion .............................................................................................................................. 13 Conclusion & Recommendations ........................................................................................................... 23 References ............................................................................................................................................ 26 2 Executive Summary Initial testing of a triblock copolymer with polystyrene end blocks and a poly(methylacrylaterandom-octadecylacrylate) mid-block(PS-b-PMA-r-PODA-b-PS) was performed using the JohnsonKendall-Roberts (JKR) method for measuring work of adhesion and adhesion strength to determine its feasibility for use as an on demand adhesive. Further descriptions of the polymer tested and the JKR test method are given in the Background section of this report. Testing of the PS-b-PMA-r-PODA-b-PS was performed at three temperatures, refrigerated, room temperature and elevated, to provide results on both sides of two thermal transition temperatures. The testing and calculations outlined in the Experimental Procedure section allowed for calculations of the system’s modulus from loading and unloading, the work of adhesion and adhesion strength, respectively. For PS-b-PMA-r-PODA-b-PS the average system‘s modulus were determined from loading to be 2.25x106± 2.15x105N/m2at the refrigerated temperature, 2.43x106 ± 1.81x105N/m2at room temperature, and 1.15x106± 1.08x106N/m2at the elevated temperature. The measured average work of adhesions of the polymer and a soft silicone lens at the refrigerated, room, and elevated temperatures were found to be 24.23 ± 5.05mJ/m2, 27.43 ± 4.31mJ/m2, and 154.84 ± 87.93 mJ/m2respectively. These values represent the thermodynamic minimum interaction energy of the two surfaces at the given temperature. For the same conditions respectively the average adhesion strengths were found to be 185.37 ± 34.83mJ/m2, 120.29 ± 14.82mJ/m2, and 698.72 ± 115.74mJ/m2. The adhesion strength is the actual interaction energy that must be overcome when separating the two surfaces in contact. In ideal situations this value should be equal to the work of adhesion. The adhesion strength is the value that is of greatest concern for this project based on the possible applications. When looking at these values it can be seen that there is a significant increase in the adhesion strength, over four and a half times larger, when increasing the temperature from room temperature to the elevated one. Analysis of variance (ANOVA) testing verified that this increase was not the result of random sample variation with greater than 99.9% confidence. There is also a change in the adhesion strength when moving from room temperature to the refrigerated temperature. This change was less significant than when going to the elevated temperature, but ANOVA testing verified with greater than 95% confidence that this was not due to random sample variation. It is believed that these changes in adhesion strength are the result of the thermal transitions undergone by the polymer when changing temperature and that the transition going from room temperature to the elevated temperature is more significant in effecting adhesive propertied. Because of the large increase in adhesive strength that is generated in reaction to an outside stimulus, PS-b-PMA-r-PODA-b-PS has shown adhesive on demand properties in this testing. Based on this project’s results it is believed that PS-b-PMA-r-PODA-b-PS is a feasible material for use in an on demand adhesive application. Further testing and research is needed to confirm its usability in specific applications. 3 Introduction &Purpose The goal of this project is to use a polymer material with specific cross linking properties to apply for use as an adhesive. Properties of the adhesive will be variable depending on the cross linking of the polymer chain. Ultimately it is desired to create an adhesive that only has adhesive properties within a set temperature range, and beyond that is not adhesive. Doing this will allow the polymer to have the adhesive properties when it is desired, or on demand. One current possible application for a material with these properties is for the use as a tape that can be applied while the polymer has adhesive properties, or is sticky, and then removed when the properties are changed. This would allow for the removal of the tape with little to know damage to the substance to which it was applied. A tape like this can be used in the medical industry for application to sensitive skin such as that of new born babies or burn victims. The tape could then be applied to the skin to hold on instruments or other important medical equipment without concern for damage of the skin upon removal. A tape with on demand adhesion properties has potential for use in almost any area in which tapes are applied for temporary use. This project has many potential applications and benefits to the greater community. 4 Background This project was centered on a polymer material where the crosslink density can be varied in a stepwise manner across a thermal transition. The polymer that was tested was a triblock co-polymer with long alkyl side chains on the middle block of the copolymer. Both end groups of this polymer were polystyrene, and the midblock consisted of poly(methylacrylate-random-octadecylacrylate)(PS-b-PMA-rPODA-b-PS). Figure 1 shows the structure of the polymer used. The nature of this block copolymer allowed for the polystyrene end blocks to physically crosslink the midblock which bridges the domain of the glassy end blocks. The random octadecylacrylate side chains on the midblock were able to crystallize at room temperature which created a secondary physical linking within the network. This polymer has already been published on as a shape memory polymer in Synthesis and Characterization of aPoly(styrene-block-methylacrylaterandom-octadecylacrylate-block-styrene) ShapeMemory ABA Triblock Copolymer by Fei and Cavicchi. In both applications, shape memory and adhesive on demand, it was the random octadecylacrylateside chains that made the desired behavior possible. The ability of the side chains to crystallize, with transition temperatures Tc=19°C and Tm=32°C (Fei & Cavicchi, 2010), allowed for a melting or crystallization transition to occur. This thermal transition created a change in the properties of the material. Figure 1 (Fei & Cavicchi, 2010) Adhesion testing for this project was done via the contact Johnson-Kendall-Roberts (JKR) method of measuring contact mechanics. The JKR method measures the work of adhesion of a semispherical elastic lens to a thin film substrate. Using the JKR method the lens was pressed onto the 5 thin film substrate using a cantilever while a microscope measured the contact area of the two and a balance measured the force applied to the substrate. The setup for this configuration is shown in Figure 2. In Figure 2 the microscope, substrate and balance are labeled. The orange semispherical section is the lens, the lens is attached to the gray bar that is the cantilever, and the contact area measured by the microscope can be seen as the flat part of the lens touching the substrate. Figure 2 (Chaudhury) Work of adhesion was calculated from the measurements of mass, which was used to estimate the force applied, from the balance and contact area from the lens. Equation 1 was used to generate the correlation between force, contact area and work of adhesion. Further explanation of how work of adhesion was calculated from Equation 1 is given in the Experimental Procedure section of this report. Equation 1 In Equation 1 a= radius of contact area R= radius of curvature of the lens E= overall Young’s modulus of system P= applied load W= work of adhesion As will be further explained in the Experimental Procedure section of this report, multiple measurements were taken with different materials to confirm measurement procedure was valid. One of the materials used as a substrate was Poly(N-isopropyl acrylamide) (pNIPAM) which is a thermoresponsive polymer used commonly in bioengineering applications (Cooperstein & Canavan, 2009). 6 There is a transition of pNIPAM at around 32°C (Cooperstein & Canavan, 2009), which causes the properties of the polymer to change over this temperature. Below this temperature pNIPAM is hydrophilic, while above this temperature the polymer is hydrophobic (Cooperstein & Canavan, 2009). This change affects the adhesive properties of the polymer among other properties. To determine if the differences between the different conditions were significant or caused by random deviation in the testing, analysis of variance (ANOVA) tests were performed on the data. The ANOVA test compares the variation within a group of data to the variation between two groups of data by calculating an F statistic. The F statistic is calculated based on the average value of the data, the variance of the data, and the number of data points. For a given confidence level a critical F statistic (Fcrit) exists such that if F is not greater than Fcrit the null hypothesis (that there is no statically significant difference between the two sets of data) fails to be rejected. In other words, if F is greater than F crit it can be said to the set confidence level that there is a significant difference between the two sets of data. The critical F statistic is calculated from the location of a value calculated using the confidence level and number of data points on a set distribution which is constant. Summaries of the results of ANOVA tests run are given in the Results & Discussion section. This allowed for the significance of the difference between the different conditions to be determined to a known confidence level. 7 Experimental Procedure Work of adhesion was measured using the JKR method of measuring surface contact mechanics. As outlined in the Background section, this method involves the use of a semispherical lens pressed onto a substrate with a thin film of the test material. For all tests run for this project the lens used was made of polydimethyl siloxane (PDMS). A microscope was set up over the lens so that the contact area could be clearly seen. The microscope output was sent to a computer screen for viewing and a program allowed the diameter of the contact area to be measured in pixels. The measurement in pixels was converted to meters by using a conversion factor found by measuring the number of pixels corresponding to 1mm on a ruler. Once this was completed, all measurements of contact diameter were taken using the computer program and recorded as the number of pixels to the nearest whole pixel. The applied load was measured using an electric balance under the thin film substrate and cantilever setup and measured the mass applied to the nearest .0001 g. Prior to the lens contacting the substrate the balance was zeroed. Once contact was made, the reading on the balance increased above zero corresponding to the applied force. Measurements were recorded as the mass given on the balance and converted to force by multiplying by the acceleration of gravity. Values for diameter and mass on the balance were recorded and converted to the desired values of radius and applied force in the correct units of meters and Newton’s respectively as an ordered pair. Measurements were taken at ten different applied forces and the corresponding values were recorded. This allowed for the creation of a ten point plot of the ordered pairs. Using Equation 1, this plot allowed for the calculation of Young’s modulus and work of adhesion. Equation 1 Equation 2 In Equation 2 x= the independent variable y= the dependent variable m= the slope of the line b= the intercept on the y-axis 8 Comparing Equations 1 & 2, it can be seen that Equation 1 is of the same form as Equation 2 if certain parameters and variables are grouped in the correct way. It is also observed that Equation 2 is the general equation of a line. Taking both of these observations to be true, then Equation 1 can be plotted using the ordered pairs for radius of contact area and applied load as a straight line. This is done by setting the following groupings equal to the variables and parameters in Equation 2. Equation 3 Equation 4 Equation 5 Equation 6 Plotting x vs. y derived in Equations 3 & 4, a line was generated. A linear regression was used to fit a straight line to this plot with a corresponding equation of the same for of Equation 2. The equation for this line gave the values of m and b for the given set of ordered pairs. Next Equations 5 & 6 were rearranged so that Young’s modulus (E) of the system and work of adhesion (W) could be calculated for the material using the m and b from the corresponding equation. Equation 7 Equation 8 Equations 7 & 8 were used to find the Young’s modulus in Newton’s per meter squared or Pascal (N/m2 or Pa) and work of adhesion in millijoules per square meter (N/m2) for each material at each condition. Each value received corresponds to one set of ten ordered pairs of radius of contact area and applied force. For each material at each condition several runs were performed giving values of E and W for each run. This allowed for an average value and standard deviation to be calculated for each material and condition. Another parameter value needed in Equation 3 so that E and W can be calculated is the radius of curvature of the lens (R). This was found using Equation 9 and the corresponding values found from a picture similar to that of Figure 3. In Equation 9, c is the length of a chord drawn across the lens and m 9 is the distance from the midpoint of that chord perpendicularly to the edge of the lens. A known magnification factor was used to find the actual values of c, m, and therefore R from the measurements taken from a print out. For each lens, multiple chords were drawn so that multiple R values could be calculated and an average value could be used for calculations. Equation 9 Figure 3 The Young’s modulus and work of adhesion were calculated for different materials and conditions. Control materials were run first so that the test procedure could be practiced. As stated before, all tests that were run used a PDMS lens. All of the substrate materials that were tested were either solvent cast or spin coated onto a silicon wafer. The first substrate used for initial testing was a PDMS sheet. Next practice runs were performed using the pNIPAM thermo-responsive polymer described in the Background section earlier. Testing with this polymer substrate was run at room temperature and at an elevated temperature above its transition temperature. The final material that was tested was the PS-b-PMA-r-PODA-b-PS polymer that was the focus of this project. This polymer was tested at room temperature, an elevated temperature, and at a refrigerated temperature. The 10 comparisons between these different conditions and why they were selected will be discussed further in the Results &Discussion section. The PDMS substrate (i.e. a thin film of PDMS) was used as a control material for the first rounds of testing to establish the procedure. Prior to finding the measurement technique that was used another technique was used. For all of these preliminary tests, PDMS was used as the material. PDMS was used for these setup tests because it was a readily available material in the lab that gave consistent results. It was also selected because there was a rough idea of the modulus and work of adhesion expected for this condition. This allowed the method to be tested on a material that was inexpensive and easily replaced if the method did not work as desired. Several runs on the PDMS substrate were performed using an initial test method. This method did not yield consistent results nor results that aligned with the expected values for the given conditions. All of the data collected during these runs was kept and compared to the data received using the improved test method. Once the final test method was confirmed and found to be repeatable the PDMS was used as the substrate to collect usable data for comparison. After the PDMS substrate was used to confirm the experimental procedure was repeatable, the next substrate was tested was the pNIPAM thermo-responsive polymer. Using this material allowed for experience to be gained in testing at two different temperatures and comparing results on either side of a transition temperature. Testing of the pNIPAM polymer was done at both room temperature and at an elevated temperature. The exact temperatures at which the tests were run were not recorded exactly; the onlyconcern was assuring that the temperature of the polymer was maintained above the transition temperature during testing. To ensure that the temperature of the polymer was maintained at a high enough temperature, the polymer substrate was attached to a small piece of metal which was heated on a hot plate prior to each run. The metal under the substrate held the temperature much longer than the silicon wafer alone would have been able to, providing more time to run the tests. 11 Multiple runs at both room temperature and the elevated temperature were performed. The results of each run at one temperature were averaged with the other runs at that temperature and the averages of the two temperatures were compared against each other. An interpretation of these results gave a chance to determine if there is a significant difference between the results at two different temperatures. The last set of tests run for this project was run on the PS-b-PMA-r-PODA-b-PS polymer that this project was centered around. Measurements were taken at three temperatures for this material, room temperature, an elevated temperature, and a refrigerated temperature. The temperature of the polymer was maintained in the same manner as was done for the pNIPAM polymer. The same piece of metal that was used for the pNIPAM polymer was used to maintain the temperature at both the high and low temperatures. A hot plate was used to elevate the temperature and the metal piece was placed in a freezer. Several runs were performed at each temperature and the results at each condition were averaged. Comparing the averages determined whether a significant difference existed between the works of adhesion at the three temperatures. The masses and diameters of contact were the values actually recorded while the tests were being run. After the runs were performed these values were entered into a spreadsheet which calculated the slope and intercept using Equations 5 & 6 and then the Young’s modulus and work of adhesion using Equations 7 & 8. The data was entered after each run so that the results could be verified. Table 1in the Results & Discussion section shows an example of the values calculated from a run. Once the results were entered and the values calculated the run was complete. The condition was changed, either material or temperature, only after enough runs were completed to give an average and standard deviation for that condition. After all of the results were collected, the results were analyzed to determine if the PS-b-PMA-r-PODA-b-PS polymer could be used in an adhesive on demand application. 12 Results & Discussion The results of all testing are separated by the slopes of the lines generated for loading and unloading. The work performed during loading (i.e. increasing the applied force)is known as work of adhesion and labeled W. Work of adhesion is the thermodynamic interaction energy between to two surfaces and minimum energy of interaction. Equation 1 is used to calculate the work of adhesion. When unloading(i.e. decreasing applied force),the force and contact radius are recorded to obtainthe interaction energy called adhesion strength and labeled G. To calculate the adhesion strength Equations 1 & 8 become Equations 10 & 11 respectively. In an ideal situation adhesion strength would be equal to the work of adhesion, but because no system is able to act completely ideally, G is generally larger than W for a given system. While W and G represent different values, both are calculated through the same method. These values differ from one another to varying degrees depending on the properties of the material. Equation 10 Equation 11 Different forces can act upon the lens depending on whether lens is being applied to the material (loading), or removed from the material (unloading). This project will be more concerned with the adhesion strength corresponding to unloading. As described in the Introduction & Purpose section, one possible future application of this material is a medical tape that will not damage the skin upon removal of the tape. In this application it is more important that a significant reduction in adhesion strength occur when removing the tape. When taking the measurements, a microscope was used to see the contact area of the lens on the substrate. As described earlier, the microscope output when to a computer to measure the diameter. Figures 4 & 5 are examples of what was seen while taking measurements with marks showing where the measurements for the diameter of the contact area were taken. Figure 4 is pNIPAM at room 13 temperature and the corresponding measurements are D=90 pixels and m=0.0580g. Figure 5 is pNIPAM at the elevated temperature with measurements taken of D=88 pixels and m=0.0384 g. Figure 4 Figure 5 Some of the important features outlined in Figures 4 & 5 include first the contact area in both figures are a light circle with a dark outline near the center of the figure. The light surrounding the contact area is the reflection of the light source off of the silicon that the polymer was spin coated onto. In both figures arcs can be seen where the contrast changes near two of the corners. This is the outside of the lens where it is attached to the cantilever. It is also noticeable that while these came from two 14 different conditions, there is no major difference between the two figures. In general the image seen through the microscope was consistent regardless of the condition being measured. As described in the Experimental Procedure section, Table 1 is an example of the spreadsheet that was used to enter in the data from testing and generate the ordered pairs used to plot the lines given by Equations 1 & 10. The data used to generate Table 1 came from the first run on the room temperature PS-b-PMA-r-PODA-b-PS polymer. Run 1 RT PS-b-PMA-r-PODA-b-PS R (m) 0.00148 Pixels/cm 4950 unload Load m (g) D (pixels) a (m) P (N) a^3 a^1.5 a^1.5/R P/a^1.5 0.0120 69 6.97E-05 1.18E-04 3.39E-13 5.82E-07 3.94E-04 202.11 0.0194 0.0330 71 74 7.17E-05 7.47E-05 1.90E-04 3.23E-04 3.69E-13 4.18E-13 6.07E-07 6.46E-07 4.11E-04 4.38E-04 313.04 500.43 0.0402 0.0461 76 79 7.68E-05 7.98E-05 3.94E-04 4.52E-04 4.52E-13 5.08E-13 6.73E-07 7.13E-07 4.56E-04 4.83E-04 585.71 633.78 0.0531 0.0614 82 84 8.28E-05 8.48E-05 5.20E-04 6.02E-04 5.68E-13 6.11E-13 7.54E-07 7.82E-07 5.11E-04 5.29E-04 690.32 769.89 0.0676 0.0792 86 88 8.69E-05 8.89E-05 6.62E-04 7.76E-04 6.56E-13 7.02E-13 8.10E-07 8.38E-07 5.48E-04 5.68E-04 818.23 926.15 0.0877 0.0781 90 90 9.09E-05 9.09E-05 8.59E-04 7.65E-04 7.51E-13 7.51E-13 8.67E-07 8.67E-07 5.87E-04 5.87E-04 991.55 883.01 0.0693 0.0600 89 88 8.99E-05 8.89E-05 6.79E-04 5.88E-04 7.27E-13 7.02E-13 8.52E-07 8.38E-07 5.77E-04 5.68E-04 796.76 701.63 0.0488 88 8.89E-05 4.78E-04 7.02E-13 8.38E-07 5.68E-04 570.66 0.0395 0.0235 87 86 8.79E-05 8.69E-05 3.87E-04 2.30E-04 6.79E-13 6.56E-13 8.24E-07 8.10E-07 5.58E-04 5.48E-04 469.89 284.45 0.0085 0.0003 84 82 8.48E-05 8.28E-05 8.33E-05 2.94E-06 6.11E-13 5.68E-13 7.82E-07 7.54E-07 5.29E-04 5.11E-04 106.58 3.90 -0.0097 79 7.98E-05 -9.51E-05 5.08E-13 7.13E-07 4.83E-04 -133.36 -0.0162 76 7.68E-05 -1.59E-04 4.52E-13 6.73E-07 4.56E-04 -236.03 Table 1 In Table 1 columns 2 & 3 labeled, m and D, contain the actual data collected while running the test. Columns 8 & 9, labeled a^1.5/R and P/a^1.5, are the values corresponding to the y and x values respectively of the ordered pairs used to plot Equation 1. All of the columns in between were used to 15 convert the data into the ordered pairs. The value labeled R at the top of Table 1 is the radius of curvature of the lens described earlier. Next to this value is the conversion factor used to convert the number of pixels measured into the a value needed in meters. Looking over the data collected for this run it is seen that some of the values of mass are negative. While physically mass cannot be negative, these values are still valid measurements in this case. These masses are relative to the zeroing of the balance before any load was applied and can therefore be negative if the substrate is being lifted off of the balance slightly. Like is occurring in this case, the lens and the surface are held under tension as opposed to compression when the force is positive. The adhesion of the polymer to the lens is strong enough to lift the substrate slightly off of the balance. When the ordered pairs that were generated are plotted, the resulting graphis shown in Figure 6. Run 1 RT PS-b-PMA-r-PODA-b-PS 7.0E-04 y = 1.042E-07x + 5.026E-04 R² = 9.192E-01 6.0E-04 a^1.5/R 5.0E-04 y = 2.593E-07x + 3.258E-04 R² = 9.673E-01 4.0E-04 Load 3.0E-04 unload 2.0E-04 1.0E-04 -400 0.0E+00 -200 0 200 400 600 800 1000 1200 P/a^1.5 Figure 6 Figure 6 is the plot of the ordered pairs that were generated in Table 1. As predicted by Equation 1, the resulting plots can be fit, with R2 values over .9, to a straight line. The line of best fit that is created for each set of data is of the same for as Equations 1 & 2 and can therefore be used with 16 Equations 7, 8& 11 to calculate the Young’s modulus and either work of adhesion or adhesion strength. This was done using a spreadsheet setup like that shown by Table 2. m Run 1 RT PS-b-PMA-r-PODA-b-PS Load Unload b m b 2.593E-07 3.258E-04 m=9/(16E) 2.17E+06 Pa (N/m2) E 1.042E-07 5.026E-04 m=9/(16E) E b=3/4(6*pi*W/E)^.5 W 21.72 mJ/m 5.40E+06 Pa (N/m2) b=3/4(6*pi*W/E)^.5 2 G 128.65 mJ/m2 Table 2 All of the runs that were performed generated a table similar to Table 2. Each condition’s results were grouped together and averages and standard deviations were taken for all of the conditions. Table 3 shows the average and standard deviation values for adhesion strength and work of adhesion for each condition that was tested. Table 4 has the average and standard deviation values of Young’s modulus for each condition. Average (mJ/m2) Stdev 38.16 5.08 38.08 6.22 Loading Condition PDMS RT pNIPAM Unloading W G Hot pNIPAM RT PS-b-PMA-r-PODA-b-PS Hot PS-b-PMA-r-PODA-b-PS Cold PS-b-PMA-r-PODA-b-PS PDMS RT pNIPAM Hot pNIPAM RT PS-b-PMA-r-PODA-b-PS Hot PS-b-PMA-r-PODA-b-PS Cold PS-b-PMA-r-PODA-b-PS 35.07 27.43 154.84 24.23 115.61 80.39 58.50 120.29 698.72 185.37 8.79 4.31 87.93 5.05 12.61 11.66 7.09 14.82 115.74 34.83 Table 3 17 Loading Unloading Condition PDMS RT pNIPAM Hot pNIPAM Average (Pa) 7.02E+05 2.69E+06 2.31E+06 RT PS-b-PMA-r-PODA-b-PS Stdev 1.78E+05 5.17E+05 2.66E+05 2.43E+06 1.81E+05 Hot PS-b-PMA-r-PODA-b-PS E Cold PS-b-PMA-r-PODA-b-PS PDMS RT pNIPAM Hot pNIPAM RT PS-b-PMA-r-PODA-b-PS Hot PS-b-PMA-r-PODA-b-PS 1.15E+06 2.25E+06 1.37E+06 3.88E+06 2.92E+06 5.25E+06 2.51E+06 1.08E+06 2.15E+05 5.16E+05 2.36E+05 2.60E+05 6.17E+05 3.40E+06 E Cold PS-b-PMA-r-PODA-b-PS 7.48E+06 1.08E+06 Table 4 Looking over the results in Table 3 there are several important observations that can be made. As discussed earlier, it can be seen that the adhesion strength of all of the conditions is greater than the work of adhesion for the same condition. While comparing the W and G values from a given condition it can also be seen that the extent to which G is larger than W varies. Looking at the work of adhesion for PDMS, RT pNIPAM, and hot pNIPAM, the values are within 10 percent of each other. When comparing the adhesion strength of these same conditions it is seen that the values vary by around 100 percent of the lowest value. This shows that there are different degrees of non-ideality in the systems at different conditions. This is what causes the G for PDMS to be around 3 times larger than the W for PDMS while the G for the hot pNIPAM was around 1.5 times the W. Another observation that can be made about the results in Table 3 is that the deviations for the conditions at temperatures other than room temperature are larger than the deviations of the room temperature conditions. During testing it was noticed that generating consistent results at the elevated temperature was more difficult than at room temperature. This is reflected in the results. It is believed that that cause of the larger deviation comes from the setup of the test system. The block of metal used to maintain the temperature was elevated to the level of the cantilever by being placed on plastic petri 18 dishes. It is believed that the elevated temperature of metal block caused some warping of the plastic petri dishes which caused the reading of the scale to fluctuate more. At the refrigerated temperature it is believed small amounts of condensation accumulated mass on the metal plate while measurements were being taken causing the mass to fluctuate more during testing. This is believed to be the reason why it was more difficult to receive consistent results at the temperatures other than room temperature. Because of the nature of why the PDMS substrate was tested, the exact results are not the most important part of it being tested. The average modulus value from testing was compared to an expected value for the material found from previous testing. It is important to notice the standard deviations of the tests run on the PDMS substrate are low relative to the averages. This was the goal of testing on the PDMS, to reduce the variation in the measurements prior to testing the PS-b-PMA-rPODA-b-PS polymer. Based on the results for the PDMS and all other substrates it is believed that the use of PDMS as a substrate was successful. The next substrate used was the pNIPAM polymer which was tested at both room temperature and an elevated temperature. Comparing the works of adhesion at the two temperatures, the values are within 10 percent of one another. This shows that thermodynamically the interaction energies at the two temperatures does not change on either side of the transition temperature. There is a more noticeable difference between the adhesion strengths at the two temperatures. An ANOVA test was performed at confidence levels of 95% and 99% and the results are summarized in Table 5. pNIPAM Confidence F Fcrit Conclusion 95% 12.24 5.59 A statistically significant difference exists between the two groups 99% 12.24 12.25 No statistically significant difference exists between the two groups Table 5 In Table 5 the two groups being compared to calculate the F statistic are the adhesion strengths at room temperature and the results at the elevated temperature. It can be seen that the F statistic is 19 unchanged by changing the confidence level only the Fcrit changes. This is as expected when recalling the values used to calculate both values. The most important part of Table 5 is that the difference between the two groups is significant to 95% confidence however at a 99% confidence level the F statistic is less than Fcrit and therefore the difference is no longer statistically significant. These results of the testing on the pNIPAM polymer were considered very successful. The results show that it is possible to get a significant change in the adhesive properties of a thermally responsive polymer when measuring on either side of the transition temperature. These results also show that there is a certain confidence level above which it can no longer be said that the difference between the two groups is statistically significant. This helped to interpret the results testing the PS-b-PMA-r-PODA-b-PS polymer. The final substrate used for testing was the PS-b-PMA-r-PODA-b-PS polymer that was the desired test material. This was tested at three different temperatures as opposed to the two that the pNIPAM was tested because broader results were desired. It was originally believed that the elevated temperature would cause the desired change in adhesive properties. Because the Tg of PMA is 10°C, the refrigerated temperature was measured to determine the effect of the transition on the adhesive properties. The results for PS-b-PMA-r-PODA-b-PS showed a definite change in the adhesive properties when at an elevated temperature. The change in properties was less definitive when tested at the refrigerated temperature. The work of adhesion when going from room temperature to the elevated temperature changed by over 460%, while it only changed by around 13% when going to the refrigerated temperature from room temperature. This shows that the transition that occurred between room temperature and the elevated temperature was thermodynamically more significant than the transition at 15°C. The change in adhesion strength when the temperatures were changed was even more pronounced. The change in work of adhesion from room temperature to the elevated temperature was over 480% or over 575 mJ/m2. ANOVA tests that compared the adhesion strengths 20 between room temperature and the elevated temperature and room temperature and the refrigerated temperature show the significance of the differences. A summary of these tests are given in Table 6 below. PS-b-PMA-r-PODA-b-PS Temperatures Confidence F Fcrit 95% 104.47 6.61 99% 104.47 16.26 RT & Hot RT % Cold 99.9% 104.47 47.18 95% 11.83 5.99 99% 11.83 13.75 Conclusion A statistically significant difference exists between the two groups A statistically significant difference exists between the two groups A statistically significant difference exists between the two groups A statistically significant difference exists between the two groups No statistically significant difference exists between the two groups Table 6 In Table 6 it can be seen that the jump in adhesion strength from room temperature is statistically significant with over 99.9% confidence. This means that one can be over 99.9% sure that this change is because of the thermal transition of the polymer and not random sampling variation. There is less confidence that the change in adhesion strength going from room temperature to the refrigerated temperature. There is less than 99% confidence that this change is not caused by random sample variation. One strange trend that can be seen in the results in Table 1 is that the adhesion strength of PSb-PMA-r-PODA-b-PS decreases going from the refrigerated temperature to room temperature and then increases greatly from room temperature to the elevated temperature. It was expected that as the temperature increased from one condition to the next, the trend would not change direction. The cause of this is likely do to the nature of the transition that occurred between each of three temperatures. The results that were obtained for PS-b-PMA-r-PODA-b-PS work of adhesion do overall show what was expected and desired. The significant change when changing temperature means that this material has adhesive properties that can be changed via outside stimulus. Being able to control this 21 stimulus would allow the adhesive properties to be changed on demand. The results of this testing have proven the feasibility of PS-b-PMA-r-PODA-b-PS as an adhesive on demand. 22 Conclusion& Recommendations Use of the JKR method of measuring work of adhesion and adhesion strength has shown that PS-b-PMA-r-PODA-b-PS may have a feasible application as an adhesive on demand. The test results show that there is a sharp increase in the adhesion strength of the polymer going from room temperature to an elevated temperature. ANOVA testing confirms that this change is not caused by random sample variation. Testing on another material with known thermal responsiveness showed that there can be a significant change in adhesive properties cause by a thermal transition of a polymer. It is believed that the reversible thermal transition of PS-b-PMA-r-PODA-b-PS when increasing temperature caused the increase in adhesion strength. This would allow the polymer to be used in an on demand adhesive application and be repeatedly applied and removed. While testing performed for this project shows initial feasibility of a PS-b-PMA-r-PODA-b-PS polymer to be used in this application additional testing still needs to be performed to confirm this material can be used in desired applications. Further testing or research must be conducted to find the thermal transition temperature more exactly. If the polymer is to be used as a medical tape the transition temperature must be below the surface temperature of the human body. This must be the case to make the adhesion strength high enough when attached to the skin. Depending on the exact application, the desired transition temperature may wish to be adjusted. This may be possible through different changes to the polymer structure. The current form of the polymer has 18 carbon long side chains on the midblock. Adjusting the length of these chains could be one way to change the crystalline properties of the polymer. This would likely shift the temperature at which the crystallization occurs and the adhesion strength on both sides of the transition. Increasing or decreasing the frequency of the side chains could also change the amount of crystallization that can occur. Increasing the frequency of the side chains would likely increase the number of crosslinks up to a certain point. This change would likely also affect the transition temperature and adhesive properties. 23 A third way that these properties could be shifted would be to change the degree of polymerization of the three blocks. Changing the lengths of the blocks in the polymer would change the amount of primary and secondary physical linking. By changing one or a combination of these polymer properties it could be possible to move the transition temperature into the desired range. This might also prove to be a way to increase or decrease the adhesion strength either above or below the transition temperature. It is recommended that there be more research into these possible changes to increase the usability of PS-b-PMA-r-PODA-b-PS in a wider variety of applications. In the case being tested, temperature change was the outside stimulus driving the change in adhesive properties. While changes to the structure of the polymer would likely change the temperature range over which the thermal transition occurs; they may also be able to change the stimulus required to trigger the property change. There are other possible stimuli that may be more desirable to cause the adhesive property change. One stimuli that is currently being researched further is the application of ultraviolet (UV) light. If UV light can be used to trigger a transition in the polymer it can also be used as the stimulus that changes adhesive properties. This would be beneficial because the application of UV light would be less likely to do damage to sensitive skin or other surfaces to which the adhesive is applied. Any other stimuli that could be found to generate a transition in the polymer could be applied to causing the on demand adhesive properties of this or a similar polymer. Additional testing that needs to be performed is testing using lenses of materials other than PDMS. If this is to be used in medical applications a lens covered with skin or modified with proteins found on skin. This would create test conditions more similar to the desired application and give a better indication of whether the material is feasible to use on skin. It is recommended that other lens materials be tested to see the affect this has on the adhesion strength. Initial testing that has been completed shows that PS-b-PMA-r-PODA-b-PS has adhesive on demand properties. More testing and research needs to be performed to show that this polymer can be 24 used as a tape in any application especially in a medical application. All of the results received while completing this project show that this material is feasible to use in the adhesive on demand application. 25 References 1. Chaudhury, M. K. (n.d.). General Adhesion: Methods of Contact Mechanics. Retrieved from http://polymers.nist.gov/combi/NCMC-2_Presentations/02-mChaudhury.PDF 2. Cooperstein, M. A., & Canavan, H. E. (2009). Biological Cell Detachment from Poly(N-isopropyl acrylamide) and Its Applications. Langmuir Invited Featured Article, 7695-7707. 3. Fei, P., & Cavicchi, K. A. (2010). Synthesis and Characterization of a Poly(styrene-blockmethylacrylate-random-octadecylacrylate-block-styrene) Shape Memory ABA Triblock Copolymer. Applied Materials & Interfaces, 2797–2803. 4. Silberzan, P., Perutz, S., & Kramer, E. J. (1994). Study of the Self-Adhesion Hysteresis of a Siloxane Elastomer Using the JKR Method. 2466-2470. 26
© Copyright 2026 Paperzz