PDS 725

Quantum™ QC, 8 Peak Beads
B E A D S A B OV E T H E R E ST
Product Data Sheet 725
DESCRIPTION
Quantum™ QC is a 8 peak multi-intensity, multi-fluorescent standard that is intended for use as an in-depth tool for daily cytometer QC, and is appropriate
for use with all lasers and detectors. It may be used to determine detection thresholds, understand resolution, and assess and track linearity of detectors. It
can aid in providing confidence that the system is suitable for use, or alert operators to potential problems before samples are run.
100
Count
~5µm
Bottle 1 (mixed, 7 fluorescent intensity populations),1 x 107 beads / mL
Bottle B (Blank), 2 x 106 beads / mL
De-ionized water + 0.01% Tween®20 + 2mM NaN3
0
Mean Diameter:
Particle Concentration:
Storage Buffer:
50
CHARACTERISTICS
150
Quantum™ QC may also be used for instrument set-up to achieve standardized PMT settings and define the window of analysis for relevant detectors.
PROCEDURE
102
103
104
105
FITC
Daily QC:
Quantum™ QC may be used on a routine basis to assess the resolution, linearity and threshold for most detectors on the cytometer. Our QuickCal Linearity
tool provides templates for 7 detectors: VIO-450, VIO-525, FITC, PE, PE-Cy™5, APC, and APC-Cy™7.
1. Go to bangslabs.com/quickcal to download the QuickCal®- Linearity template.
2. Select the file for download.
3. Select the appropriate version based on the resolution of your instrument:
• Semi-Log (e.g. 1024 channels)
• Log / Log (e.g. FACSDiva)
4. The template will download to your computer.
5. Adjust PMT voltages for each detector to standard daily settings.
6. Place one drop of the blank (bottle B) and one drop of the mixed fluorescent population (bottle 1) in a tube.
Dilute with ~800µL DI water. (The Blank may also be run separately from the fluorescent populations to
improve resolution.)
7. Run on LOW to achieve a count rate of 100-150 events/sec; low run rates aid in achieving lowest
fluorescence CV’s.
8. For each detector, gate on individual peaks, and record channel values (Median or Geo. Mean). Enter these
values in the appropriate tables of the QuickCal® Linearity template. A Detection Threshold and Regression
Figure 1
Coefficient (r2) will be calculated for each detector.
Bangs Laboratories, Inc.
QuickCal® Linearity v 2.3-L
Quantum™ QC
Instrument ID:
Date Acquired:
Date Entered:
725 / 11446
Vio-450
Enter
Filter:
Enter
PMT voltage:
Bead
Intensity %
Blank
0
Bead #1
20.10
Bead #2
25.58
Bead #3
38.25
Bead #4
52.37
Bead #5
75.16
Bead #6
90.32
Bead #7
100.00
Regression Coefficient (r2)
Detection Threshold (%)
Channel
Data incomplete
Data incomplete
FITC
Enter
Filter:
Enter
PMT voltage:
Bead
Intensity %
Blank
0
Bead #1
12.44
Bead #2
22.78
Bead #3
32.46
Bead #4
49.50
Bead #5
70.08
Bead #6
84.80
Bead #7
100.00
2
Regression Coefficient (r )
Detection Threshold (%)
Rqd7 beads
Mod Factor
Bead
x^2
y y^2
x*y Blank
0 0 0 0 0
count0
m= n
b= n
corr=
Channel
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Data incomplete
Data incomplete
Rqd
7 beads
Mod Factor
x^2y y^2x*y
0 0 0 0 0
count0
m= n
b= n
corr=
PE
Channel
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Data incomplete
Data incomplete
PE-Cy™5
Enter
Filter:
Enter
PMT voltage:
Bead
Intensity %
Blank
0
Bead #1
Bead #2
Bead #3
26.82
Bead #4
39.67
Bead #5
58.48
Bead #6
75.20
Bead #7
100.00
2
Regression Coefficient (r )
Detection Threshold (%)
Vio-525
Enter
Enter
Intensity %
0
Bead #1
9.45
Bead #2
15.57
Bead #3
29.76
Bead #4
44.82
Bead #5
70.04
Bead #6
87.13
Bead #7
100.00
Regression Coefficient (r2)
Detection Threshold (%)
Filter:
PMT voltage:
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Filter:
Enter
PMT voltage:
Rqd7 beads
Mod Factor
Bead
Intensity %
x^2
y y^2
x*y Blank
0
Bead #1
12.46
Bead #2
21.91
Bead #3
32.48
Bead #4
48.73
Bead #5
69.55
Bead #6
84.16
Bead #7
100.00
2
Regression Coefficient (r )
0 0 0 0 0
count0
Detection Threshold (%)
m= n
b= n
corr=
Channel
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Data incomplete
Data incomplete
Rqd
7 beads
Mod Factor
x^2y y^2x*y
0 0 0 0 0
count0
m= n
b= n
corr=
APC
Channel
Enter
Enter
Enter
Enter
Enter
Enter
Data incomplete
Data incomplete
Enter
Filter:
Enter
PMT voltage:
Rqd5 beads
Mod Factor
Bead
Intensity %
x^2
y y^2
x*y Blank
0
Bead #1
6.95
Bead #2
11.50
Bead #3
19.63
Bead #4
29.66
Bead #5
46.27
Bead #6
65.69
Bead #7
100.00
2
Regression Coefficient (r )
0 0 0 0 0
count0
Detection Threshold (%)
m= n
b= n
corr=
Channel
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Enter
Data incomplete
Data incomplete
Rqd
7 beads
Mod Factor
x^2y y^2x*y
0 0 0 0 0
count0
m= n
b= n
corr=
APC-Cy™7
Enter
Filter:
Enter
PMT voltage:
Bead
Intensity %
Blank
0
Bead #1
Bead #2
9.65
Bead #3
18.48
Bead #4
28.13
Bead #5
44.13
Bead #6
63.57
Bead #7
100.00
Regression Coefficient (r2)
Detection Threshold (%)
Channel
Enter
Rqd
6 beads
Mod Factor
x^2y y^2x*y
Enter
Enter
Enter
Enter
Enter
Enter
Data incomplete
Data incomplete
0 0 0 0 0
count0
m= n
b= n
corr=
Instrument Set up:
Quantum™ QC may be used to set up the window of analysis for detectors by positioning a specific peak at a target channel value.
Percentage assignments were made as an LSR II (FACSDiVa) with the following optics.
Dichroic
mirror
Bandpass filter
VIO 450
-
450/50
VIO 525
555 LP
525/50
FITC
505 LP
530/30
PE
555 LP
575/26
PE - Cy™ 5
635 LP
670/14
-
660/20
735 LP
780/60
Laser
405nm
488nm
633nm
Detector
APC
APC - Cy™7
NOTE: Due to the nature of fluorescence detectors (cytometer) and
fluorescence carryover (fluorophores), we anticipate varying resolution
capabilities for different detectors. This is reflected in the example
histograms provided in “Expected Results,” page 2, as well as the
QuickCal Linearity template (Fig. 1), which features channels in which the
blank and a minimum of 5 fluorescence peaks were resolved.
9025 Technology Dr. Fishers, IN 46038 • www.bangslabs.com • [email protected] • 317.570.7020 • fax 371.570.7034 PDS 725
Rev. #005, Active: 11/Nov/2016
Page 1 of 2
PDS 725
Figure 2 : Expected Results (LSRII, FACSDiVa) Note: For additional detectors ( e.g. U V and green lasers), see PDS 725A www.bangslabs.com.
25
50
Count
75
100
200
150
100
0
50
SSC-H QC (1.000)
250
Quantum QC
100
FSC - H
150
200
102
250
(x 1.000)
103
104
105
104
105
104
105
104
105
PE-Cy5
40
30
10
20
Count
50
60
70
50
103
104
105
0
102
102
VIO 450
103
150
100
Count
50
102
103
104
0
0
50
25
Count
75
100
PE-Cy7
105
102
VIO 525
103
100
Count
50
100
50
0
0
Count
150
150
APC
102
103
104
105
102
FITC
NOTES
Manually shake bottle before use. The prepared tube may be briefly (pulse)
vortexed if needed to increase % singlets.
STORAGE AND STABILITY
Store at 2-8˚C. Freezing, drying, or centrifuging particles may result in
irreversible aggregation. This product is warranted to be stable for 12
months from the date of purchase, provided that it is handled in accordance
with the manufacturer’s recommendations. The product should be stored in
its opaque bottle to safeguard against photobleaching.
SAFETY
This particle suspension contains sodium azide. Sodium azide may react
with lead and copper plumbing to form explosive metal azides. Upon
disposal of material, flush with a large volume of water to prevent azide
accumulation. Please consult the Safety Data Sheet for more information.
103
PE
TRADEMARKS
1.
2.
3.
4.
Quantum™ QC is a trademark of Bangs Laboratories, Inc.
QuickCal® is a registered trademark of Bangs Laboratories, Inc.
Tween® is a registered trademark of ICI Americas, Inc.
Cy™ is a trademark of GE Healthcare Limited.
This product is for research use only and is not intended for use in
humans or for in vitro diagnostic use.
ORDERING INFORMATION
Cat. Code
725
Description
Quantum™ QC, 8 Peak Beads
Size
5mL
Order online anytime at www.bangslabs.com.
9025 Technology Dr. Fishers, IN 46038 • www.bangslabs.com • [email protected] • 317.570.7020 • fax 371.570.7034 PDS 725
Rev. #005, Active: 14/Nov/2016
Page 2 of 2