22nd International Symposium on Plasma Chemistry July 5-10, 2015; Antwerp, Belgium Measurement of N(4S) and N(2P) absolute densities by optical emission spectroscopy A. Salmon1,2, G.D. Stancu1,2, M. Cirisan3 and C.O. Laux1,2 1 CNRS, UPR 288, E.M2.C, Grande Voie des Vignes, 92295 Châtenay-Malabry, France 2 Ecole Centrale Paris, Grande Voie des Vignes, 92295 Châtenay-Malabry, France 3 PlasmaBiotics, 215 rue Jean-Jacques Rousseau, 92136 Issy les Moulineaux, France Abstract: Bactericidal reactive species N(4S) and N(2P) were measured by optical emission spectroscopy in the afterglow of a nitrogen pulsed discharge at atmospheric pressure. Two methods were used and validated: absolute and kinetic methods. The newly developed kinetic method gives results close to those obtained with the absolute calibration technique. Keywords: emission spectroscopy, absolute calibration, nitrogen afterglow, nitrogen atom density, nitrogen metastable 1. Introduction Decontamination of heat-sensitive medical devices (e.g. endoscopes) is still challenging nowadays. Conventional methods have the disadvantage to damage the material, so the development of new disinfection techniques is necessary. Use of the afterglow of nitrogen pulsed plasma at atmospheric pressure is promising for decontamination at ambient temperature [1]. This configuration was studied by several research groups and revealed that the main bactericidal action was due to the chemical species produced by the plasma [2]. To investigate further the role of chemical species, measurement of absolute densities is necessary. Optical emission spectroscopy (OES) is a common technique used to measure the relative densities of excited states. However, to measure the absolute ground state density of reactive species by OES, a calibration technique and a good knowledge of the kinetics are necessary. In nitrogen plasma, the mechanisms are sufficiently well-known to use such an approach. In Ref. [3], kinetic mechanisms associated with the afterglow emission were used to measure N, N 2 (A) and O. Several calibration techniques were used in the literature. In Ref. [3], calibration was done using an NO titration technique, but this method is intrusive and experimentally difficult to use and leads to large uncertainties at atmospheric pressure [4]. In Ref. [5] a calibration technique based on the emission signal resulting from excitation transfer of argon metastable atoms to nitrogen molecules was used to measure N(2P) density. Here, we compare the use of two methods to measure the axial distribution of the nitrogen ground state N(4S) and metastable state N(2P) in the afterglow of a pulsed nitrogen discharge at atmospheric pressure. relative emission intensity. The absolute method [7] is based on measurements of the absolute emission intensity of N(2P) O-9-3 state and N 2 (B,11) state following nitrogen atoms recombination (reaction (R2) in Table 1). So, this method requires absolute calibration of the optical device. The kinetic method is based on the study of the temporal decay of N 2 (B,11) emission intensity and therefore only relied on relative intensities First, we will present both methods in details, then the experimental set-up. Finally we will discuss the experimental results and the use of the two methods in other configurations. 2. Description of the methods 2.1. Kinetic method The kinetic method is a two-step technique which requires to work in a pure nitrogen afterglow. From the N 2 (B,v=11) state emission in arbitrary units and using a simplified kinetic scheme, one can measure the N density profile in this configuration, as explained in section 2.1.1. The main reactions and the associated constants are listed in Tables 1 and 2. In section 2.1.2., the method is extended to other wavelengths using relative calibration. At this stage, the emission can be studied with different gas mixtures. 2.1.1. N measurement N 2 (B,11) state is mainly populated by the recombination reaction (R2) and depleted by quenching with N 2 (R3). Thus, assuming steady state for the population of N 2 (B,11), and using (R2-R3) one can easily find : [N2 (B, 11)] = k2 [N(4 S)] 2 k3 (1) Considering the emission intensity of N 2 (B,11) at 580 nm, equation (1) becomes: I(580) = K[N(4 S)] 2 (2) 1 with, k2 hc K = π΄580 C(580) k 3 580x 10β9 (3) (4) where V is the flow velocity, z is the position along the post-discharge tube and 1/[π(4 π)]0 is a constant. Using equations (2) and (4), we obtain that I(580)β1/2 is linear along the post-discharge tube: I(580) β1/2 =K β1/2 1 2k1 [N2 ] οΏ½ 4 + zοΏ½ [N( S)]0 V (6) where G(Ξ») corresponds to the spectral sensitivity of the optical device while b includes geometrical parameters such as the emission column length and the solid angle. This assumption was previously used in [8]. Using equation (3), the constant b can be calculated while the factor G(Ξ») is obtained by relative calibration. We now apply this approach to measure the N(2P) density from the forbidden line emission at 346.6 nm following the transition N(2 P βΆ4 S), whose probability A346 equals 0.005 s-1. The emission intensity can be written as follows: π΄580 π2 346 πΊ(580) πΌ(346.6) π΄346 π3 580 πΊ(346) πΎ (8) Table 1. Reaction mechanisms. N° Reactions R2 N(4 S)+N(4 S)+N2 β N2 (X, A, B) + N2 R3 N2 (B, 11) + N2 βN2 + N2 R1 N(4 S)+N(4 S)+N2 β N2 (B, 11) + N2 Table 2. Reaction rate coefficients. N° R1 R2 2.1.2. Extension of the method to other wavelengths We now assume that the calibration factor can be written as follows: 2 [N(2 P)] = (5) From the slope of I(580)β1/2 , it is possible to determine the constant K. The density of nitrogen atoms can finally be obtained using equation (2) and keeping arbitrary units for I(580). The obtained values are only dependent on the validity of the kinetic mechanism used and on the uniformity of the flow velocity across the tube section. However they are independent of the choice of constants k 2 and k 3 because at this stage, only equations (2) and (5) are used. C(Ξ») = b G(Ξ») (7) Using equations (3) and (6), equation (7) gives: where A 580 (=77600 s-1) is the Einstein coefficient of spontaneous emission associated to the transition N2 (B, 11 βΆ A, 7), h is the Planck constant, c is the speed of light, and C(580) is the calibration factor of the optical device including acquisition parameters and spectral sensitivity. In pure nitrogen afterglow, nitrogen atoms are mainly lost by three-body recombination (reaction R1 in Table 1). Assuming steady state and considering reaction (R1), we obtain: 1 1 2k1 [N2 ] = + z 4 4 [N( S)] [N( S)]0 V hc C(346.6)[N(2 P)] 346x 10β9 I(346.6) = A346 R3 Reaction coefficient 4.4 x 10-33 cm6.s-1 1.7 x 10-32 cm6.s-1 1.5 x 10-32 cm6.s-1 4.4 x 10-34 cm6.s-1 Ref. [9] [10] [11] [3] 2.8 x 10-11 cm3.s-1 [12] 2.2. Absolute method The absolute method is based on the measurement of the intensity of the volumetric plasma source in absolute units which is defined as follows: πΌ(π) = 1 πΏβππ΄21 π2 π(π) 4π (9) where L is the emission column length, A 21 is the Einstein coefficient of spontaneous emission, Ο(Ξ») is the normalized line profile, π2 is the density of the upper state and βπ is the photon energy. Here, for the comparison with the kinetic method, we only considered the emissivity integrated on the column emission length, thus giving results averaged on the tube section. The absolute intensity is difficult to obtain because it depends on the solid angle, the length of collection volume and the detector response. For this reason, a calibration lamp of known emissivity is used in place of the plasma source while the acquisition parameters are kept constant. From this measurement, conversion from arbitrary units to absolute units can be done. Then, using an absolute spectroscopic model (Specair) to fit the measured intensity, the densities of the excited states can be calculated. Here, the N 2 (B,11) state density is calculated from its measured absolute emission intensity at 580 nm and using the Specair software. N(4S) density is then calculated O-9-3 using equation (1). For N(2P), the density is directly calculated by Specair from the forbidden line emission at 346.6 nm measured in absolute units. 3. Experimental set-up The experimental set-up is shown in Figure 1. The plasma is generated by an industrial reactor working in a point-to-point geometry (Plasmabiotics, Inpulse One V3). Both electrodes are encapsulated in a ceramic box. The gap between the electrodes (few mm) is fed with pure nitrogen gas (Air Liquide, azote U) at a flow rate of 30 slpm, so that the flow velocity in the post-discharge tube is about 40 m.s-1. The afterglow emission is collected using two off-axis parabolic (OAP) mirrors and focused on the entrance slit of the spectrometer (Acton 2500i). The spectra are recorded using a ICCD camera (PI-MAX 2). Fig. 2. Calibration curve for the kinetic method 4. Results and discussion The longitudinal profiles of atomic nitrogen density measured using kinetic and absolute method are shown in Figure 3. For the kinetic method, three recombination coefficients were tested. The coefficients measured in [10] and [11] gave results in good agreement with those obtained by absolute method. On the contrary, the constant measured in [9], which was used for kinetic method in [6], leads to a large difference in the obtained values. Indeed, according to the kinetic method, the N(4S) density increases linearly with the constant k 1 whereas the N(2P) density increases quadratically. So, the measurement is strongly sensitive to the value chosen for this constant. Comparison with the absolute method encourages us to choose the value given in Refs. [10] and [11]. Fig. 1. Schematic of the experimental set-up. For absolute calibration, the light emitted by a calibrated tungsten ribbon lamp of known radiance traceable to NIST standards was recorded keeping the same conditions as for the acquisitions from the discharge afterglow. The signal was corrected from a significant contribution of the stray light. The spectra were then fitted using the Specair software [13]. For the kinetic method, emission of N 2 (B,v=11) at 580 nm was recorded along the post-discharge tube. In Figure 2 the calibration curve shows a linear profile as expected in equation (5). The non-linear region is due to a change in the losses mechanisms of nitrogen atoms. Three-body recombination is not dominant anymore, so equation (4) is not valid in this region. The slope of the calibration curve was measured using a linear fit and was used to derive N(4S) density according to equation (2). Then, relative calibration data of a tungsten ribbon lamp were used to measure N(2P) density according to relation (8). Fig. 3. N(4S) density distribution along the post-discharge tube. In Figure 4, the longitudinal profile of N(2P) is shown. The observed differences between kinetic and absolute methods may be due to the imperfect knowledge of the constant k 1 . Additional research is in progress to explain O-9-3 3 this behavior. Globally, the results give the same order of magnitude as those obtained in [5]. [7] G.D. Stancu, F. Kaddouri, D.A. Lacoste and C.O. Laux, J. Phys. D: Appl. Phys., 43, 124002 (2010) [8] C. Yubero, M. C. García, M. D. Calzada, Optica Applicata, Vol. XXXVIII, No. 2, 2008 [9] I. A. Kossyi, A. Yu. Kostinsky, A. A. Matveyev and V. P. Silakov, Plasma Sources Sci. Technol. 1, 207-220 (1992) [10] P. Harteck, R.R. Reeves and G. Manella, J. Chem. Phys., 29 , 608 (1958) [11] J.T. Herron, J.L. Franklin, P. Bradt and V.H. Dibeler, J. Chem. Phys., 30, 4, 879 (1959) [12] A. Ricard, J. Tétreault and J. Hubert., Phys. B: At. Mol. Opt. Phys., 24 , 1115 (1991) [13] C. O. Laux, T. G. Spence, C. H. Kruger and R. N. Zare, Plasma Sources Sci. Technol., 12, 125 (2003) Fig. 4. N(2P) density distribution along the post-discharge tube 5. Conclusions Absolute densities of N(2P) and N(4S) along the postdischarge tube of a pulsed nitrogen discharge were measured by OES using absolute and kinetic methods. Both techniques give close results if the volume recombination constant of nitrogen atoms k 1 measured in [10] and [11] is used in the kinetic method. Absolute calibration requires an absolutely calibrated lamp, and an absolute intensity spectroscopic model. On the contrary, the kinetic method is much easier to use because it only requires relative intensities. However it cannot provide radial resolution and can only be used in an axisymmetric tube. This calibration-free technique is still promising to measure the densities of species inside small-diameter tubes such as plasma jets. 6. Acknowledgments This research was supported by DGA grant DESDEMONA (RAPID contract EJ nr. 2101 241 872) 7. References [1] J. Ehlbeck, U. Schnabel, M. Polak, J. Winter,Th Von Woedtke, R. Brandenburg, T. von dem Hagen and K-D Weltmann, J. Phys. D: Appl. Phys., 44, 013002 (2011) [2] M. Mols, H. Mastwijk, M. Nierop Groot and T. Abee, Journal of Applied Microbiology, 115, 689-702 (2013) [3] A. Ricard, S. Oh and V. Guerra, Plasma Sources Sci. Technol., 22 035009 (2013) [4] P. Vasina, V. Kudrle, A. Tálsky, P. Botos, M. Mrázková and M. Mesko, Plasma Sources Sci. Technol. 13 668β674 (2004) [5] E. Eslami and N. Sadeghi, Eur. Phys. J. Appl. Phys., 43, 93β102 (2008) [6] A-M. Pointu, A. Ricard, E. Odic and M. Ganciu, Plasma Process. Polym., 5, 559 (2008) 4 O-9-3
© Copyright 2026 Paperzz