1116_1.pdf

Beam Optics for a Scanned Proton Beam at Loma Linda
University Medical Center
George Coutrakon, Jeff Hubbard, Peter Koss, Ed Sanders, Mona Panchal
Loma Linda University Medical Center
11234 Anderson Street
Loma Linda, CA 92354
Abstract
Beam scanning in proton therapy is a medical technique to lower the dose to healthy tissue while irradiating a tumor volume.
Scanned proton beams for proton radiation therapy require small beam sizes at the tumor location. In beam scanning, a small
beam usually less than 1 cm diameter is swept across the tumor volume with two magnets located several meters upstream of the
patient. In general, all proton beams in a therapy facility must be transported from the accelerator to the treatment rooms where
the scanning systems are located. This paper addresses the problem of transporting the beam without losses to the patient and
achieving a small beam at the tumor location in the patient. The strengths of the beam line quadrupoles were allowed to vary to
produce the desired beam sizes along the beam lines. Quadrupole strengths were obtained using the beam simulation program
TRANSPORT 1 originally from Stanford Linear Accelerator Center in Palo Alto, CA. An enhanced version of the original
program by Accel Soft Inc. in San Diego, CA has been used for these studies. Beam size measurements were used for
comparison with TRANSPORT to verify the predictions of TRANSPORT calculations.
maximum field gradient of 1.3 kG/cm. Three of the
patient treatment rooms contain rotating beam lines
called gantries which can direct the beam from any
azimuthal angle around the patient as shown in Fig.2.
This paper examines the quadrupole solutions for the
three gantries to provide a small beam at isocenter.
Introduction
Four rooms at Loma Linda are currently being used
to treat more than one hundred patients per day.2
The facility layout with the treatment rooms is shown
in Fig. 1. All quadrupoles in the switchyard and
gantries have an effective length of 29 cm and a
Fig. 1 The layout of the Loma Linda beam lines in the
“switchyard”. The long “spine” from the accelerator to the
Research room is referred to as the North-South line.
CP680, Application of Accelerators in Research and Industry: 17th Int'l. Conference, edited by J. L. Duggan and I. L. Morgan
© 2003 American Institute of Physics 0-7354-0149-7/03/$20.00
1116
Fig. 3 A graph of the minimum beam size which can be
expected for a hypothetical nozzle at the end of proton
range in the patient. Beam size contributions due to 1)
multiple Coulomb scattering from a “low mass” nozzle
upstream of the patient, 2) multiple coulomb scattering
from patient tissue upstream of the stopping depth of
protons in the patient, 3) the beam optics contribution due
to beam emittance from the accelerator and 4) the sum of
1), 2) and 3) added in quadrature. The beam optics
contribution, 2.6 mm, represents a goal for this study
which is less than the sum of the other two contributions
which can not be significantly reduced further.
Fig 2. The Loma Linda gantry3 with the proposed scanning
magnets in the “nozzle”. The entire magnet structure can
rotate 360 degrees about the patient to bring beam to
isocenter from any azimuth angle. The nozzle is the
segment between the last gantry bending magnet and the
patient where beam spreading occurs.
Beam Size requirements for scanned
proton beams
Since the beam entering into any treatment room is
usually less than two centimeters diameter and has an
approximately Gaussian shaped profile, some means
of increasing the area of the beam to a large uniform
size must be developed. Presently, Pb and lexan foils
with contoured shapes are used to scatter the beam
into a large diameter. When raster scanning is
implemented in the future, two scanning magnets
will deflect the small pencil beam across the tumor.
These devices which are used to spread the beam
into a large area are referred to as beam delivery
systems or “nozzles”.
Methods
The G1(Gantry 1) beam line as shown in Fig.1 uses
ten switchyard quads (SQ1 –SQ10) and eight gantry
quads (GQ1-GQ8). This beam line runs from the
extraction septum to the isocenter of G1.
TRANSPORT was used to calculate the quadrupole
strengths which best satisfy our constraints, at least
approximately, and predict the horizontal (X) and
vertical (Y) beam sizes in the beam lines and at
isocenter. The initial values of the emittances in X
and Y for each energy were taken from earlier
studies5 and used in TRANSPORT as the starting
values at the extraction septum. At 250 MeV, the
emittances (1 rms, unnormalized) that we used were
εx = 0.6 π mm-mrad and εy = 2.5 π mm-mrad.
Fig. 3 shows the contributions to beam size at the
end of the proton’s range as a function of the incident
proton energy. Some assumptions about nozzle
design and minimal mass in the beam path upstream
of the patient have been used to estimate the
scattering contribution from the nozzle. In general,
this contribution decreases with increasing proton
energy in accordance with the Highland4 formula for
multiple Coulomb scattering. The second beam size
contribution increases with energy because the
protons travel further in tissue before stopping. In
addition to these contributions, the minimum spot
size which can be achieved in vacuum at isocenter
must be included. This contribution should be
smaller than the quadratic sum of the other two to
maintain a small beam size. For the example shown
in Fig. 3, we have taken the desired beam size in
vacuum at isocenter to be approximately 2.6 mm =
σx = σy over the full range of energies (70 to 250
MeV) and gantry angles (00 to 3600).
To simplify the task, the gantry optics were studied
alone to find values of αx, αy, βx and βy at the input
of the gantry which could give small beam sizes (σx
and σy < 2.6 mm) at isocenter for all gantry angles.
Small dispersions (R16 and R36) at the nozzle are also
important so that beam position does not change
throughout the beam spill due to changing beam
momentum or magnet power supply fluctuations.
The eight quads on the gantry were allowed to vary
independently to achieve these constraints for each
trial of α and β. We tested input values of α from –
1.0 to +1.0 and β from 0.5 m to 5.0 m. For
simplicity, we kept αx = αy and βx = βy. The best
results obtained with these simulations are shown in
1117
Fig. 4 for α = -0.5 and β = 3 m which satisfies the
small beam constraints at all gantry angles. These
values became the match conditions from the
extraction septum to the entrance of G1. The
dispersion values for the gantry were R16 = 0.05 m
and R36 = 0.29 m which are quite small. The rms
momentum spread of the beam, ∆p/p, due to the RF
accelerating cavity is only 0.0075% at 250 MeV.
The beam sizes predicted from TRANSPORT were
then compared with measured data. Nine multi wire
ionization chambers (MWIC’s) were distributed
throughout the G1 beam line to acquire beam size
measurements. The X and Y emittances along with
their α and β values at the extraction septum were
then varied to obtain improved agreement between
TRANSPORT predicted beam sizes and actual
measurements. We then verified that the predicted
rms beam sizes at isocenter remained small.
Results
The results for 250 MeV in G1 at 2700 gantry angle
are shown in Fig. 5. The agreement is quite good in
both planes. The predicted beam size at isocenter
(which presently can not measured due to Pb foils in
the beam path at 3 meters from the patient) still falls
within our desired limit of 2.6 mm. For fixed quad
strengths, the predicted beam size at isocenter
changes slightly with gantry angle but is still less
than 2.6 mm. The emittance values used at the
septum in Fig. 5 are listed below in Table 1. Further
refinements, using these emittances will be done to
get a better match at the input to the gantries and
therefore a better match to the gantry model in Fig. 4.
Fig. 4 The predicted beam sizes through the gantry to
isocenter for 250 MeV at 2700. This result is based on a
search for the α’s and β’s at the gantry input which satisfy
the requirement for small beam size at isocenter and low
dispersion for the gantry dipole magnets. Predicted beam
sizes at other angles are similar to 2700.
Next, the ten switchyard quadrupoles were allowed
to vary to match the desired α’s and β’s at the gantry
input while achieving several other constraints. In
particular, we found that beam position stability
could be improved by reducing the dispersion R16
from the 900 switchyard bend magnet to the nozzle.
SQ7 through SQ 10 were varied to achieve this
constraint. Quads SQ1 through SQ6 were then
allowed to vary to achieve best match in α and β at
the gantry input. Also, SQ1 through SQ6 were varied
to achieve small values of R11 and R21 from the
septum to the nozzle. Smaller values of R11 and R21
have improved position stability at the nozzle when
intensity fluctuations in the accelerator exceeds 20 or
30% from spill to spill. For beam scanning, the
required intensity changes from spill to spill may be
20:1 which will make the stability issue even more
important. These constraints were approximately
achieved as follows:
R11 = 0.5 (septum to nozzle)
R21 = -0.2 mm/mrad (septum to nozzle)
R16 = 0.05 m (switchyard 900 bend to nozzle)
αx = -0.3 αy = -0.2 βx = 3.3 m βy = 3.1 m (at
input of G1)
Fig. 5 TRANSPORT predictions and measured beam sizes
for 250 MeV in G1 at 2700 gantry angle.
Table 1. Emittance parameters at the extraction septum for
the TRANSPORT calculations shown in Figs. 5 through 7.
150 MeV values are also included. The emittance values
have units of π mm-mrad (1 rms, unnormalized)
X axis
Y axis
Energy MeV
70
150
250
70
150
250
Emittance
0.3
0.3
0.3
2.5
2.0
1.5
Alpha
2.5
2.1
1.8
.8
.5
.2
Beta (m)
35
35
35
2.0
2.0
2.0
.014
0.0105
.0075
∆p/p in %
1118
Extending this optics solution to G2 and G3 was
achieved using the FODO cells in the North-South
line. Each FODO cell consists of a focussing
quadruple, a drift, a defocussing quadruple and
another drift. The α’s and β’s at the entrance to the
G1 900 bend magnet were noted and replicated at the
G2 900 bend and G3 900 bend by requiring a 1800
phase advance of the X and Y beam ellipses at these
positions in the beam line. The eight quadrupole
values on G1 and the SQ7 through SQ10 values were
then copied over to the G2 and G3 beam lines. One
result using this technique is shown in Fig. 6. The
quadrupole strengths were scaled down by
momentum to 70 MeV in the G2 beam line and beam
size measurements were taken as before.
TRANSPORT was initially run with the same α’s
and β’s at the septum as the 250 MeV beam. The
emittances were increased in proportion to1/p and
the momentum spread was increased to 0.014% (1
rms) as determined by the RF voltage in the
accelerating cavity. The emittance parameters were
then allowed to vary to achieve a good visual fit to
the beam size measurements as shown in Fig. 7. The
final 70 MeV emittance values used in Fig. 7 can be
seen in Table 1. The optics solution for 70 MeV was
also confirmed in G1 and G3 using the FODO cells
in the North-South beam line. Although the
emittance parameters are slightly different than for
250 MeV, the same beam constraints have been
achieved at 70 MeV. Subsequent tests at 150 MeV in
all three gantries showed similar results with good
agreement between TRANSPORT and measured
beam data.
Fig. 7 TRANSPORT predictions and measured beam sizes
in G2 for 70 MeV at 2700.
Conclusions
A method of finding an optics solution which
allows a small beam at the isocenter of a rotatable
gantry has been presented. These solutions have
already demonstrated enhanced position stability in
the passive nozzles in the Loma Linda gantries.
Using the FODO cells in the switchyard, nearly
identical optics can be achieved in all three gantries.
The solution can be applied to other energies by
scaling the quadrupole strengths according to the
beam’s momentum. Because this paper represents a
“works in progress”, more improvements may still be
necessary. First, there is an elliptical shape to the
beam at isocenter which will vary with gantry angle.
In theory, this will require additional studies, but in
practice, the multiple Coulomb scattering due to
nozzle material and patient tissue will remove most
of the ellipticity of the beam shape. Secondly, the
gantry optics is not truly identical between gantries
1, 2 and 3. Some variation in beam size can be seen
at isocenter from Fig. 5 and Fig. 6. Better matching
of the beam parameters at the switchyard dipoles and
gantry input should improve this. Finally, the
vertical beam size at isocenter in Fig 7 is slightly
larger (2.9 mm) than the design goal of 2.6 mm, but
it does not significantly alter the beam size prediction
at 70 MeV in Fig. 3. In summary, the work
presented here has demonstrated a good foundation
for scanned beam delivery systems in the future.
Fig. 6 TRANSPORT predictions and measured beam sizes
in G2 for 250 MeV at 2700.
1119
Acknowledgements
The authors wish to thank David Lesyna at Optivus
Technology, Inc. in San Bernardino, CA for doing
the multiple Coulomb scattering calculations leading
to the beam size analysis as shown in Fig.3. We
thank Stuart Wakefield for the figure drawings and
putting them into the proper style and format, also
Lynn Segobia and Dawn Huggins for typing the text
and preparing the paper in its final form. Additional
TRANSPORT calculations by Anthony Teran and
Jojit Torcedo are gratefully acknowledged. Finally,
we thank Dr. James Slater, Vice Chairman of the
Dept. of Radiation Medicine at LLUMC for
encouragement and support for all the teams working
on the beam scanning project at Loma Linda.
References
1.
2.
3.
4.
5.
The Reports Office, SLAC, P.O. Box 4349, Stanford, CA
94305, Ref. SLAC-R-95-462.
http://www.llu.edu/proton/history
A.M. Kohler, “Preliminary Design Study for a Corkscrew
Gantry.” Harvard Cyclotron Laboratory, Cambridge, MA,
internal report (1987).
V.L. Highland, Nuclear Instruments and Methods, Vol.
129, p.497 (1975).
M. Schulze, “Commissioning Results of the LLUMC
Beam Switchyard and Gantry”, Proceedings of the 1990
IEEE Particle Accelerator Conference, p.610.
1120