Portaldosimetrie mit ´Portal Vision´ Var i an’ sQAsol ut i on VARIAN Medical Systems © Varian Medical Systems, 2005 Die Vision Goal Verify the dose distribution Each IMRT field, without patient in beam PortalVision LC250 and aS500 Increase efficiency for the Physicist Commission IMRT sooner Ongoing pre-treatment QA VARIAN Medical Systems Reduce time to evaluate a plan Less time/plan = Verify more plans March 2005 Slide 2 Die Zukunft... Focus Asymmetric Jaws MLC / Custom Field Shaping Quantitative dosimetric QA of Patient Treatment, including 3D reconstruction of multi-field IMRT Patient Central Dose Absolute dose= 100 cGy delivered? VARIAN Medical Systems Treatment Couch March 2005 Slide 3 Traditional Dosimetry Tools Measurement tools VARIAN Medical Systems Ion chamber Diodes Film Analysis software Challenges Non-linear sensitometry of film Image registration; accuracy variations User errors Labor intensive, time consuming (2-4 hr. / plan) March 2005 Slide 4 Traditional Film Dosimetry Expose film Expose film Process film Process film Scan film Scan film H&D curve calibration Calibration VARIAN Medical Systems Import dose image Align images Compare results Planned Field Analysis March 2005 Slide 5 Portal Dosimetry PV Eclipse Expose film Expose imager Import dose image Process film Process film Scan film Scan film Align images Calibrate Imager Compare results PV Vision Calibration VARIAN Medical Systems Planned Field Analysis March 2005 Slide 6 Portal Dosimetry Dosimetry is a Process ‘ Por t al Dosi met r y’i sasystem capability (TPS, PV, VarisVision) ‘ Dosi met r yWor kspace’access in PV (standalone) and Vision Review Portal Dosimetry is a set of capabilities which provides Accumulation of relative dose at the PV imager position Dose evaluation, acquired dose vs. TPS-predicted dose For the application of Pre-treatment QA VARIAN Medical Systems Beam through air only (thru phantom = future capability) No patient in beam March 2005 Slide 7 Focus Portal Dosimetry Dosimetry Asymmetric Jaws DMLC Application Pre-Treatment Verification without Phantom FDA 510(k) Qualitative Comparison Patient Predicted Portal Dose Acquired Portal Dose Treatment Couch Portal Imager VARIAN Medical Systems March 2005 Slide 8 Portal Dosimetry Applications The Future? Machine QA pre-Treatment QA in vivo QA Focus Asymmetric Jaws MLC / Custom Field Shaping Q A T e s t Consistency P l a n Plan vs Actual T x Patient EPID/Film Portal Dose VARIAN Medical Systems Treatment Couch Portal March 2005 Dose Detector Slide 9 Treatment Methods Supported Treatment Type Overall Dose Prediction Dose Acquisition IMRT sliding windows Yes Yes Yes Step-and-shoot Yes Yes Yes Static field for machine QA Yes Yes Yes Dynamic wedge No No Yes VARIAN Medical Systems March 2005 Slide 10 Varis/Vision TPS TPS Fluence map + dose matrix Dosimetric Image VARIAN Medical Systems USER PROCESS Vision Vision EPID Portal Dose Prediction March 2005 Slide 11 Eclipse (TPS) role in Dosimetry Patient Treatment Plan Verification Plan Dose prediction file for each field Gantry 0 degrees for all fields (Plan angles possible) User-defined imager position (SAD) Result –predicted dose matrix ( image ) absorbed in PV cassette in ABSOLUTE values VARIAN Medical Systems March 2005 Slide 12 Portal Vision ( EPID) role in Dosimetry Calibration (Treat or Standalone PV) Calibration of detector at each energy, dose rate, and 2 distances Dose normalization for LC250 Acquisition (Treat or Standalone PV) Integrated Image acquisition mode and sequence template Automated Density Image to Dose Image conversion Off-line integration of split fields (multiple image) VARIAN Medical Systems March 2005 Slide 13 Portal Vision (EPID) Acquisition FDA 510(k) - Images courtesy of KFJ Hospital, Vienna Austria VARIAN Medical Systems March 2005 Slide 14 Portal Vision (EPID) Acquisition Reference Image VARIAN Medical Systems Integrated Dose March 2005 Slide 15 VarisVision (R&V) role in Dosimetry Dosimetry workspace in Review (Vision or Standalone PV) Dose Image Display Dose Images Alignment Dose Measurement & Evaluation Point dose Line Profile Dose difference Gamma evaluation Off-line integration of split fields (multiple image) VARIAN Medical Systems March 2005 Slide 16 Dosimetry Review: Dose Evaluation Traditional Dose Evaluation Tools: Line Profile + Point Dose VARIAN Medical Systems March 2005 Slide 17 Dosimetry Review: Dose Evaluation Gamma Evaluation showing good correlation Gamma > 1 VARIAN Medical Systems March 2005 Slide 18 Dosimetry Review: Dose Evaluation Gamma Evaluation showing mismatch GammaGamma >1 >1 VARIAN Medical Systems March 2005 Slide 19 Dosimetry Summary IMRT adoption easier and faster Facilitates IMRT QA Confirms that Clinac delivers the planned dose Pre-treatment QA in a fraction of the time vs. traditional methods Advantages Improved staff efficiency Costs Facilitates good patient care VARIAN Medical Systems , billings March 2005 Slide 20 Conclusion Acceptance testing, commissioning and QA of IMRT involves new concepts specific to IMRT. Standards and guidelines are not yet established. Your clinical judgment is important to patient safety. In general, you should expect IMRT planning and delivery systems to perform as well as 3DCRT planning and delivery systems. VARIAN Medical Systems March 2005 Slide 21 Vielen Dank für Ihre Aufmerksamkeit VARIAN Medical Systems
© Copyright 2025 Paperzz