The Antek Advantages: The Antek Advantages:

Nitrogen & Sulfur GC
7090 SERIES
DETECTORS
www.paclp.com
The Antek Advantages:
Experienced
we are the experts at nitrogen and sulfur analysis
technology with over 25 years experience
Recognized
methods of choice by regulatory agencies and
method writing institutions
Responsive
our experienced staff of highly trained chemists will
help you solve your analytical problems or fit the
7090 Series to your application
Cost-effective
Antek’s GC detectors offer a low-cost alternative for
selective multi-element detection
Nitrogen & Sulfur GC
7090 SERIES
DETECTORS
The leadership continues…
Building upon years of innovation, analytical leadership, and customer satisfaction, Antek sets the
standard for simultaneous nitrogen and sulfur GC detection.
Utilizing accuracy and precision of our Pyro-chemiluminescent™ technology, the 7090NS Nitrogen/
Sulfur GC Detector delivers equimolar response for all nitrogen- and sulfur-containing compounds.
Chemiluminescence is the clean, fast, interference-free method for determining all bound nitrogen and
sulfur in a variety of matrices.
R-S + R-N + O2 —> SO2 + NO + other oxides
R-S + R-N + H2 —> H2S + NO + other reduced species
H2S + O3 —> SO2* —> SO2 + hυ (200–400 nm)
NO + O3 —> NO2* —> NO2 + hυ’ (600–900 nm)
Nitrogen, Nitrosamines, Sulfur
Antek Instruments, Inc. offers the most advanced
detectors for GC, SFC, and SimDis. With the Model
7090’s versatility, users can buy one detector now—
nitrogen or sulfur—and add the other later.
The Model 7090N Chemiluminescent Nitrogen Detector is specific for nitrogen compounds and easily
converts to nitrosamine-selective mode with out any
hardware change. The Model 7090S Sulfur Chemiluminescent Detector is specific for sulfur compounds.
Our Model 7090NS offers unprecedented simultaneous detection of nitrogen and sulfur compounds.
Conventional element enhanced detectors are less
selective and therefore still have problems with
interfering compounds. The 7090 Series eliminates
those problems by detectingnitrogen and/or sulfur
compounds only—without halogen,phosphorous, or
hyrocarbon interferences that have complicated this
work in the past. This unique detector also gives a
linear response, and has sensitivity that is stable over
time, providing more reliable results.
Antek’s 7090 Series retrofits with most GCs on the market.
Contact PAC today
with your analytical needs — let us provide your
“Real World Solution”
Product Features
Patented Pyro-chemiluminescent™
nitrogen and sulfur technology
add to FID for one injection—three results
highly selective for nitrogen and sulfur
equimolar response for all nitrogen and sulfur
bearing compounds
Maximum uptime
% Devition from Mean
one injection—two results
5
0
-5
-10
1
long-term operation on one calibration
14” dual zone furnace for optimal oxidation
and reduction
complete sample combustion enhances short- and
long-term stability, equimolarity, and selectivity
long ceramic tube life: 6–12 months
oil-less vacuum pump operates maintenance-free
600
for several years
500
15
injections
30
700
CS2
H2S
MeSH
DMS
EtSH
COS
Thiophene
iBUSH
nBUSH
Thiophenol
400
Versatile
nitrogen, sulfur or simultaneous nitrogen/sulfur—
buy one detector, add a second later
compatible with most commercially available GCs
electronic mass flow control for all gases
modular designed detectors and electronics
300
200
100
0
Satisfaction guaranteed
application solutions by experienced, knowledgable
technical staff
one year limited warranty on parts and labor
in-depth training courses offered at corporate
CS2
MeSH
EtSH
DMS
COS
nBUSH
iBUSH
Thiophene
H2S
Thiophenol
headquarters
D
FID
Sulfur
Nitrogen
Detector Response
Detector Response
Regular unleaded gasoline
Sulfur
Nitrogen
32 minutes
40 minutes
Nitrogen & Sulfur GC
Heavy diesel
7090 SERIES
FI
Right: The Model 7090 GCNSD is ideal for a variety
of applications including
agriculture, biotechnology,
biochemical, food, solid
phase synthesis, organic
synthesis, petrochemical,
pharmaceutical, and
polymer/plastics.
45
DETECTORS
10
Stability
45 injections at 9.24 ppm
S
solidpartners
provensolutions
solidpartners
provensolutions
U.S.A.
PAC, LP | 8824 Fallbrook Drive | Houston, Texas 77064
T: +1 800.444.TEST | O: +1 281.940.1803 | F: +1 281.580.0719
[email protected] | [email protected]
Product Specifications
Detection Method
Pyro-chemiluminescence™
Performance
Sensitivity: <3 pgN/sec or <1 pgS/sec (individually optimized);
<30 pgN/sec and <10 pgS/sec (simultaneous detection)
FRANCE
BP 70285 | Verson | 14653 CARPIQUET Cedex
T: +33 0 231 264 300 | F: +33 0 321 266 293
[email protected]| [email protected]
Selectivity: >107 to hydrocarbon and nitrogen/sulfur by weight
Linearity: >10 4
Equimolarity: <±10%
GERMANY
Badstrasse 3-5 | P.O.Box 1241 | D-97912 Lauda-Königshofen,
T: +49 9343 6400 | F: +49 9343 640 101
[email protected]| [email protected]
Gases
“Chromatographic” or “Zero Grade”
Dry oxygen: 2–20 cc/min 99.75%, 1.5 bar (20 psig)
Dry hydrogen: 50–400 cc/min 99.995%, 1.5 bar (20 psig)
SINGAPORE
61 Science Park Road | #03-09/10 The Galen
Singapore Science Park III | Singapore 117525
T: +65 6412 0890 | F: +65 6412 0899
[email protected] | [email protected]
Outputs
1 mV, 1 V, 10 V
Combustion Zone Temperature
950ºC–1100ºC in 1ºC increments
Power
115 vac 50/60 Hz 12 Amp or
230 vac 50/60 Hz 6 Amp
The NETHERLANDS
P.O.Box 10.054 | 3004 AB Rotterdam
Innsbruckweg 35 | 3047 AG Rotterdam
T: +31 10 462 4811 | F: +31 10 462 6330
[email protected] | [email protected]
Physical
Dimensions: 36 cm wide (14”) x 59 cm deep (23”) x
61 cm high (24”)
Weight: 28 kg (62 lb)
RUSSIA
Shabolovka Street | 34, Bldg. 2 | 115419 Moscow
T: +7 495 617 10 86 | F: +7 495 913 97 65
[email protected] | [email protected]
CHINA
Room 1003, Sunjoy Mansion | No. 6 RiTan Rd.
Chao Yang District | Beijing 100020
T: +86 10 650 72236 | F: +86 10 650 72454
[email protected] | [email protected]
Patents and Approvals
U.S. Patents: # 6458328, #7018845, #5916523, #6143245,
#6075609, #6636314
Ordering Information
Model 7090N Nitrogen Detector: designed for sensitive, selective, equimolar quantitation of
nitrogen compounds. Accessories include ceramic tubes, operation/service manual, fittings and
gas line tubing required for system operation. Requires vacuum pump and mounting kit.
INDIA
1036 Regus | Trade Center, Level 1
Bandra (E) - 400 051 | Mumbai
T: +91 22 40 700 447 / 700 | F: +91 22 40 700 800
[email protected] | [email protected]
Model 7090S Sulfur Detector: designed for sensitive, selective, equimolar quantitation of sulfur
compounds. Features large 14” furnace, facilitating complete combustion. Accessories include
ceramic tubes, operation/service manual, fittings and gas line tubing required for system
operation. Complies with GPA 2199-99, ASTM D 5504 and D 5623. Requires vacuum pump and
mounting kit.
MIDDLE EAST
Villa #8/1, Street #9 | POBox 1009174 | Abu Dhabi
T: +971 2 446 9671 | F: +971 2 446 9672
[email protected] | [email protected]
SOUTH KOREA
#621 World Vision Building | 24-2, Youido-dong
Seoul 150-010
T: +82 2785 3900 | F: +82 2785 3977
[email protected] | [email protected]
Model 7090NS Nitrogen/Sulfur Detector: designed for simultaneous quantitation of nitrogen and
sulfur compounds. Features large 14” furnace, which facilitates complete combustion and
enhances sensitivity and equimolarity. Accessories include ceramic tubes, operation/service
manual, fittings and gas line tubing required for system operation. Complies with ASTM D 5504
and D 5623. Requires vacuum pump and mounting kit.
THAILAND
26th Floor, M. Thai Tower | All Seasons Place
87 Wireless Road | Lumpini, Phatumwan | Bangkok 10330
T: +66 2627 9410 | F: +662627 9401
[email protected] | [email protected]
Nitrogen Detector Kit: for addition to Model 7090S Sulfur Detector. P/N 92154
PAC Authorized Representatives are also located in most countries
worldwide. For more information visit www.paclp.com
Sulfur Detector Kit: for addition to Model 7090N Nitrogen Detector. P/N 92153
ANTEK by PAC
Vacuum Pump: low maintenance, high capacity vacuum pump. Oil-less for safe oxygen
operation. Includes vacuum kit. P/N 71304
Leading Elemental Analysis and Chromatography
Since 1967, Antek has been a global leader in lab and
on-line elemental analysis instrumentation and custom
chromatography instruments that detect sulfur and
nitrogen/sulfur compounds. The Antek line also includes
process analyzers for monitoring total nitrogen and sulfur
in liquids, gases and LPGs and determination of nitrogen,
sulfur, total hydrocarbon, and aromatics for the quality
assurance for CO2.
Mounting Kit: includes all parts necessary for connection to specific GC. Please provide make and
model of GC.
Software: is purchased separately.
®
by
USA
.
by
FRANCE
®
by
by
.
GERMANY
.
NE THERLANDS
.
UAE
.
RUSSIA
by
.
CHINA
.
SINGAPORE
by
.
SOUTH KOREA
.
THAILAND
.
by
INDI A
Copyright 2011 PAC L.P. All rights reserved - I-7090-99 2011/2