British Journal of Anaesthesia 82 (6): 843–6 (1999) Oxygenator exhaust capnography as an index of arterial carbon dioxide tension during cardiopulmonary bypass using a membrane oxygenator M. J. O’Leary*, S. P. MacDonnell and C. N. Ferguson Department of Anaesthesia, St Bartholomew’s Hospital, West Smithfield, London EC1A 7BE, UK *Corresponding author: Intensive Care Unit, The St George Hospital, Gray Street, Kogarah, NSW 2207, Australia We have studied the relationship between the partial pressure of carbon dioxide in oxygenator exhaust gas (PECO2) and arterial carbon dioxide tension (PaCO2) during hypothermic cardiopulmonary bypass with non-pulsatile flow and a membrane oxygenator. A total of 172 paired measurements were made in 32 patients, 5 min after starting cardiopulmonary bypass and then at 15-min intervals. Additional measurements were made at 34°C during rewarming. The degree of agreement between paired measurements (PaCO2 and PECO2) at each time was calculated. Mean difference (d) was 0.9 kPa (SD 0.99 kPa). Results were analysed further during stable hypothermia (n530, d51.88, SD50.69), rewarming at 34°C (n522, d50, SD50.84), rewarming at normothermia (n548, d50.15, SD50.69) and with (n578, d50.62, SD50.99) or without (n591, d51.07, SD50.9) carbon dioxide being added to the oxygenator gas. The difference between the two measurements varied in relation to nasopharyngeal temperature if PaCO2 was not corrected for temperature (r250.343, P5,0.001). However, if PaCO2 was corrected for temperature, the difference between PaCO2 and PECO2 was not related to temperature, and there was no relationship with either pump blood flow or oxygenator gas flow. We found that measurement of carbon dioxide partial pressure in exhaust gases from a membrane oxygenator during cardiopulmonary bypass was not a useful method for estimating PaCO2. Br J Anaesth 1999; 82: 843–6 Keywords: partial pressure, carbon dioxide; partial pressure, arterial; monitoring, carbon dioxide; heart, cardiopulmonary bypass; equipment, membrane oxygenator Accepted for publication: January 26, 1999 Regular measurement of arterial carbon dioxide partial pressure (PaCO2) during cardiopulmonary bypass (CPB) is routine, usually by intermittent blood-gas analysis using a bench blood-gas analyser. Capnography is used routinely during anaesthesia before and after CPB to monitor endtidal carbon dioxide partial pressure, and its use in monitoring carbon dioxide partial pressure in exhaust gases from the oxygenator (PECO2) during CPB has been reported previously both in vitro and in vivo.1–3 These studies suggested a reasonably close relationship between PECO2 and PaCO2 during the cooling and stable hypothermic phases of CPB, but a far more variable relationship during rewarming. Thus oxygenator exhaust capnography is not used in clinical practice as a measurement technique. These studies involved CPB, using bubble oxygenators rather than modern membrane oxygenators. We investigated the relationship between PECO2 and PaCO2 during CPB using a membrane oxygenator. Close control of PaCO2 is important as poor control has been linked with poor psychomotor outcome after CPB.4 Although in-line blood-gas monitors which allow continuous monitoring of PaCO2 are available, they are expensive and may be difficult to use. If oxygenator exhaust capnography were an accurate monitor of PaCO2, it would be a simple, inexpensive and useful addition to patient management during CPB. Patients and methods We studied 32 adult patients undergoing elective cardiac surgery using CPB. All gave informed consent to participate in the study, which was approved by the East London and City Health Authority Research Ethics Committee. A standard anaesthetic technique was used. After premedication with papaveretum and hyoscine, anaesthesia was induced with thiopental and fentanyl, and neuromuscular block was produced with pancuronium. Ventilation was adjusted to give an end-tidal carbon dioxide partial pressure of 3.6–4.4 kPa. Anaesthesia was maintained with © British Journal of Anaesthesia O’Leary et al. 1% isoflurane and 70% nitrous oxide in oxygen, with incremental doses of fentanyl. After heparinization, CPB was commenced using a roller pump (Sarns 7000MDX, Sarns 3M Healthcare Group, Ann Arbor, MI, USA) and a membrane oxygenator (CML Duo, Cobe Cardiovascular Inc., Arvanda, CO, USA). Non-pulsatile pump flow at a rate of 2.4 litre min–1 m2 was used throughout bypass. Moderate hypothermia (32°C) and cold crystalloid cardioplegia were used for cardioprotection. Anaesthesia was maintained during CPB using 1% isoflurane delivered from a vaporizer in-line with the gas inlet to the oxygenator, and with incremental doses of fentanyl. Exhaust gas from the oxygenator was channelled from the two scavenge ports to a 100-cm length of elephant tubing incorporating a Heidbrink valve, to prevent membrane damage in the event of accidental occlusion of the gas outlet, and a sidestream capnograph connection. Care was taken to eliminate any leaks of exhaust gas before the sampling point and to prevent entrainment of air. The same capnograph was used throughout the study (Datex Normocap CO2 Monitor, Datex Instrumentarium Oy, Helsinki, Finland). The machine was calibrated according to the manufacturer’s instructions using British Oxygen Company (BOC) alpha standard gas cylinders, and was allowed to warm up for 1 h before use. Five minutes after institution of CPB and at 15-min intervals thereafter, an arterial blood sample was obtained and PaCO2 measured immediately using a bench blood-gas analyser in the corridor next to the operating theatre (IL1312 Blood Gas Manager, Instrumentation Laboratories SpA, Milan, Italy). The blood-gas analyser was calibrated to one point every 20 min and two points every 3 h, and checked against the manufacturer’s quality controls each day of use. Both uncorrected and temperature corrected results were recorded; temperature correction was performed automatically by the analyser using the following formula5: PCO2 (corrected)5PCO2 (measured)3100.019(T–37) where T5patient’s measured body temperature. At the same times, the capnograph reading, pump flow, gas flows and nasopharyngeal temperature were recorded. An additional sample was obtained and recordings made at 34°C during rewarming as a standard point between patients. A total of 172 paired measurements were made. The degree of agreement between paired measurements at each time was calculated using the method of Bland and Altman.6 Results When all measurements were considered together, the mean difference between the two measurements was large, whether uncorrected (0.9 (SD 0.99) kPa) or corrected for temperature (0.22 (0.89) kPa) (Figs 1, 2). When results obtained during the different phases of CPB (stable hypothermia, rewarming at 34°C and rewarming at normother- Fig 1 Difference in the partial pressure of carbon dioxide in oxygenator exhaust gas (PECO2) and arterial carbon dioxide tension (PaCO2) measured by a capnograph and bench blood-gas analyser at all times, plotted against the mean of the two measurements (PaCO2 was not corrected for nasopharyngeal temperature). Fig 2 Difference in the partial pressure of carbon dioxide in oxygenator exhaust gas (PECO2) and arterial carbon dioxide tension (PaCO2) measured by a capnograph and bench blood-gas analyser at all times, plotted against the mean of the two measurements (PaCO2 was corrected for nasopharyngeal temperature). Table 1 Relationship between PCO2 measured by a capnograph and bench blood-gas analyser during different phases of CPB (PCO2 not corrected and corrected for nasopharyngeal temperature). n5Number of observations, Mean5 mean difference between the two measurements, SD5SD of mean difference, 95% LA595% limits of agreement CPB phase Temp. uncorrected All measurements Stable hypothermia Rewarming, 34°C Rewarming, 36–38°C Temp. corrected All measurements Stable hypothermia Rewarming, 34°C Rewarming, 36–38°C n Mean SD 95% LA 172 30 32 48 0.9 1.88 0 0.15 0.99 0.69 0.84 0.69 –1.09 to 2.88 0.5 to 3.26 –1.28 to 1.28 –1.23 to 1.53 172 30 32 48 0.19 0.43 -0.49 0.22 0.77 0.72 0.65 0.71 –1.35 to 1.73 –1.07 to 1.87 –1.79 to 0.81 –1.2 to 1.64 mia) were analysed separately, there was still a large difference between the measurements whether uncorrected or corrected for temperature (Table 1). Results were analysed 844 Oxygenator exhaust capnography Table 2 Relationship between PCO2 measured by a capnograph and bench blood-gas analyser during CPB with and without carbon dioxide (CO2) added to the oxygenator. n5Number of observations, Mean5mean difference between the two measurements, SD5SD of mean difference, 95% AL595% limits of agreement No CO2 added CO2 added n Mean SD 95% AL 78 91 0.62 1.07 0.99 0.9 –1.36 to 2.6 –0.91 to 2.88 Fig 3 Relationship between the difference between the two measurements (partial pressure of carbon dioxide in oxygenator exhaust gas (PECO2) and arterial carbon dioxide tension (PaCO2)) and nasopharyngeal temperature (°C) (PaCO2 was not corrected for nasopharyngeal temperature). further for patients with and without carbon dioxide added to the oxygenator (Table 2); there was no improvement in agreement between the measurements. There appeared to be no pattern for the relationship between the two measurements, which may suggest a systematic error, and unlike in a previous study,3 there was no trend towards an improved relationship during any phase of CPB. At some times, PECO2 was higher than PaCO2 (in patients without carbon dioxide added to the oxygenator). In an attempt to define which factors may influence the difference between PECO2 and PaCO2 during CPB, and consequently explain our results, we examined the relationship between the difference and nasopharyngeal temperature, pump blood flow and oxygenator gas flow using simple regression analysis. The magnitude of the difference between the two measurements was found to correlate closely with nasopharyngeal temperature (r250.343, P,0.001), increasing as temperature decreased (Fig. 3), but there was no relationship if temperature correction was made (r250.002, P50.6). Changes in the difference between the measurements was not related to any of the other variables: for PaCO2 uncorrected for temperature vs pump blood flow and pump gas flow, r250.1, P50.19 and r252.301E–4, P50.84, respectively, whereas for PaCO2 corrected for temperature, the relationships were r25 2.664E–4, P50.84 (blood flow) and r250.015, P50.12 (gas flow). Discussion Our results showed a poor relationship between carbon dioxide tension in exhaust gas from a membrane oxygenator during CPB and PaCO2. Indeed, the relationship between the two measurements was extremely variable even during stable phases of CPB. This finding is in contrast with that from a previous study using a bubble oxygenator.3 Although the physical processes which occur during gas exchange are the same in an oxygenator as in the lung, namely convection, diffusion and chemical reaction, the diffusional distances in an oxygenator are much greater than those in the lung (100–500 µm compared with 10 µm) and the surface area for gas exchange considerably less (,10 m2 compared with approximately 70 m2). Gas exchange is thus much less efficient in the oxygenator, and the ventilating gases do not usually reach equilibrium with the bloodstream.7 None the less, one might expect a consistent relationship between gas tensions in the exhaust and arterial blood, and a previous study suggested that this may be the case, at least during some phases of CPB.3 There are important differences between bubble and membrane oxygenators with respect to transfer of carbon dioxide. In bubble oxygenators, oxygen uptake and carbon dioxide elimination are intrinsically linked and dependent on bubble size. The smaller the bubbles the more efficient oxygen uptake and the more inefficient carbon dioxide removal become. Most bubble oxygenators are designed to compromise between these factors, designing for bubble size to produce a ratio of carbon dioxide elimination to oxygen uptake of 0.8.8 Therefore, there is an obligatory link between oxygenation and carbon dioxide removal with bubble oxygenators such that it is not possible to vary oxygenation and ventilation independently. In contrast, membrane oxygenators allow variation of ventilation (that is PaCO2) by varying gas flow rate, and variation of oxygenation by varying the fraction of oxygen in gas supplied to the oxygenator. It is probable that the relationship between exhaust gas PCO2 and arterial PCO2 is more stable with bubble oxygenators. A further complication is the assumption that by using PECO2 as a measurement technique for PaCO2, a similar gas is being measured as when measuring end-tidal carbon dioxide tension during anaesthesia. However, there are fundamental differences between the relationship between carbon dioxide tension in the exhaust gas from a membrane oxygenator and PaCO2 and the relationship between endtidal carbon dioxide tension and PaCO2. In the latter, the end-tidal gas is assumed to be representative of alveolar gas. Differences between arterial and alveolar gas tensions are generally caused by ventilation–perfusion mismatching, whereas in the oxygenator, there are significant differences between gas and blood partial pressures. There is a significant barrier to diffusion between the gas and blood phase in an oxygenator compared with the lung. The partial pressure of carbon dioxide in the gas phase in both bubble and 845 O’Leary et al. membrane oxygenators varies significantly under normal operating conditions.6 The tension of carbon dioxide in the exhaust is influenced significantly by gas flow along and blood flow across the membrane. In contrast, in a bubble oxygenator the principal determinant of carbon dioxide elimination is bubble size.8 We attempted to control for the relationship between PECO2 and gas and blood flow through the oxygenator by analysing the results according to different flows, but this failed to improve the agreement between the measurements. We were also unable to demonstrate any consistent relationship between pump gas or blood flow and the difference between PECO2 and PaCO2. Not surprisingly, the difference between the measurements was increased during hypothermia if PaCO2 was not corrected for temperature; this finding reflects the increased solubility of carbon dioxide at low temperatures, the effect disappearing when temperature correction of blood gases was performed. Because of the large variability between measurements and lack of a systematic error, we considered a technical problem as an explanation for our results. Possible technical errors are incorrect calibration of the capnograph or bloodgas analyser or entrainment of air or other gases in the sampling tube. Calibration of the capnograph and blood-gas analyser were performed according to the manufacturer’s instructions. Fletcher pointed out that when results from a capnograph and blood-gas analyser are being compared, they should be cross-calibrated with the same gas.9 We did not perform a cross-calibration, but a discrepancy would have caused a systematic error and would not explain the wide variability of our results. Although dilution of sampled gases cannot be excluded completely as an explanation for our observations, we consider this unlikely as we ensured that all exhaust ports of the oxygenator were either sealed tightly or connected to our sampling chamber. The sampling tube was connected to the proximal end of 100 cm of elephant tubing which should be more than adequate to prevent air entrainment. In several patients without carbon dioxide added to the oxygenator, PECO2 was greater than PaCO2. This finding was surprising; if blood flow and gas flow are in opposite directions (the counter-current principle) it is theoretically possible that the maximum PECO2 may be equivalent to venous PCO2, but the Cobe oxygenator relies on a crosscurrent system with blood flow perpendicular to gas flow. Interestingly, this is somewhat similar to the bird lung, where air flows from the rear to the front through parabronchi and the gas exchanger is arranged along the parabronchi, perfused with blood according to a cross-current system. This system allows mixed blood leaving the lung to have a higher PO2 than expired air,10 but in our case it was the mixed exhaust gas which had a higher PCO2 than the mixed blood leaving the oxygenator. In summary, we have demonstrated that measurement of carbon dioxide partial pressure in the exhaust gas from the oxygenator during CPB was not a useful method for estimating PaCO2. Acknowledgements We thank Mr S. J. Edmondson for allowing us to study patients under his care, and Mr Stephen Clark, Senior Anaesthetic Technician, and the perfusion team for practical assistance. Dr M. J. O’Leary was supported by the Joint Research Board of St Bartholomew’s Hospital and BMI Columbia Healthcare Ltd. This study was presented as a poster at the Autumn Meeting of the Anaesthetic Research Society, 1996. References 1 Alston RP, McNichol J. Oxygenator exhaust capnography: An in vitro evaluation. J Cardiothorac Anesth 1988; 6: 798–802 2 Riley JB. Prediction of arterial blood PCO2 by measuring the ventilating gas exhaust PCO2 from a bubble oxygenator. J Extracorporeal Technol 1982; 14: 312–15 3 Zia M, Davies FW, Alston RP. Oxygenator exhaust capnography: A method of estimating arterial carbon dioxide tension during cardiopulmonary bypass. J Cardiothorac Vasc Anesth 1992; 6: 42–5 4 Nevin M, Adams S. Evidence for involvement of hypocapnia and hypoperfusion in aetiology of neurological deficit after cardiopulmonary bypass. Lancet 1987; ii: 1493–7 5 Operators Manual, Model IL1312 Blood Gas Manager. Viale Monza, Milan: Instrumentation Laboratory SpA 1991; Section 10: 12 6 Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; i: 307–10 7 Voorhees ME, Elgas R. Membrane and bubble oxygenators. In: Kay P, ed. Techniques in Extracorporeal Circulation, 3rd Edn. Oxford: Butterworth–Heinemann, 1992; 44–55 8 High K, Snider M, Bashein G. Principles of oxygenator function: Gas exchange, heat transfer, and blood–artificial surface interactions. In: Gravlee GP, Davis RF, eds. Cardiopulmonary Bypass Principles and Practice. Baltimore: Williams and Wilkins, 1993; 28–54 9 Fletcher R. In deadspace studies, capnograph and bloodgas analyser should be cross-calibrated with the same gas. Br J Anaesth 1995; 74: 485P 10 Scheid P, Piiper J. Cross-current gas exchange in avian lungs: effects of reversed parabronchial air flow in ducks. Respir Physiol 1972; 16: 304–12 846
© Copyright 2026 Paperzz