Application Note 01142 Wavelength Accuracy – Measurement and Effect on Performance in UV-Visible Spectrophotometry Introduction Key Words • Wavelength Accuracy • Mercury Light Source • Deuterium Light Source • Emission Lines • Holmium Filter • Didymium Filter • Rare Earth Oxides Any quality control procedure will require that the performance of the instrument be checked on specified occasions. Verification of the wavelength accuracy will form part of any such performance check. This application note details how the wavelength accuracy of a Thermo Electron Corporation UV-Visible spectrophotometer may be checked and recalibrated if necessary. Several methods are listed below, in order of preference. Methods 1 and 2 are primary methods employing fundamental constants, and should be used when available. 1. Emission lines from a Mercury source. 2. Emission lines from a Deuterium source. 3. Calibrated traceable Holmium & Didymium glasses. 4. Rare Earth Oxides in acid solution. • Calibration • NTRM Factors Affecting Wavelength Accuracy • NIST Physical Shock Accuracy may be affected if the spectrophotometer is jolted violently, dropped, or subjected to severe or prolonged vibration. • CRM • NPL • ISO/IEC Standard 17025 • Neutral Density Filters • Wavelength Calibration • Holmium Oxide in Perchloric Acid • Calibrating Solutions Temperature Problems may be found if the operating temperature is significantly different from the temperature at which the spectrophotometer was last calibrated. All instruments are calibrated at room temperature and wavelength differences of between 0.1 and 0.3 nm per 10°C are not uncommon. It is therefore essential that instruments are warmed up to normal operating temperature before use and before performing any instrument verification test. Replacement of Key Optical Components The instrument will need recalibrating if any of the key optical components are replaced. Mechanical Wear Thermo UV-Visible spectrophotometers have been designed to minimize the effects of wear. Therefore this should not be a problem. Requirements for Wavelength Standards A standard method should be: • Readily available. • Easily assembled or prepared. • Easy and safe to handle. • Suited to the optical properties of the instrument e.g. spectral bandwidth, etc. • Unaffected by environmental conditions, e.g. good chemical stability, small temperature coefficients, etc.1 Instrumentation Application 1 Use of Emission Lines from a Mercury Source Mercury sources are available as optional accessories for the following Thermo spectrophotometers: INSTRUMENT Nicolet® Evolution™ 300 Nicolet Evolution 500 UV Series 300 and 500 The mercury lamp fits into the third lamp position within the optical system of the instrument, and is powered from the instrument power supply. Once installed it can be left in place and used when required. Method: The emission spectrum of a low-pressure mercury lamp has a number of intense lines that cover a large part of the UV and visible range. The frequencies of these lines are fundamental physical properties of the source element, and therefore the wavelengths are invariant. The mercury emission spectrum is a primary wavelength standard. nm 253.65 296.73 404.66 435.84 546.07 760.95 871.68 To obtain an emission spectrum, in addition to selecting the required wavelength range, it is important to select appropriate parameters for the scan method. Suitable values are as follows: LOCAL CONTROL SOFTWARE Scan Type Mode Bandwidth Speed Data Interval Lamp Change Standard Scan I (Intensity Mode) 0.5 nm (if available) 30 nm/min 0.1 nm Hg VISION SOFTWARE Type Data Mode Bandwidth Scan Speed Data Interval Lamp Change Standard Intensity 0.5 nm (if available) 30 nm/min 0.1 nm Mercury When using an instrument with a photomultiplier (PMT) detector, the FLAT EHT profile must be used for this procedure. Sample beam intensity (l units) 1.0 253.65 Sample beam intensity (l) 0.9 0.8 0.7 0.6 435.84 0.5 365.02 0.4 656.1 200000 486.0 150000 100000 50000 0 470 404.66 0.3 250000 490 510 530 546.07 0.2 550 570 590 610 630 650 670 Wavelength (nm) 0.1 576.96 Figure 2: Deuterium lamp emission spectrum 0.0 200 250 300 350 400 450 500 550 600 Wavelength (nm) Figure 1: A mercury emission spectrum (rescaled) Most of the energy in the mercury spectrum is emitted in the 253.65 nm line, the intensity of which is much greater than the other lines. When the spectrum is first scanned only the 253.65 nm line will be apparent, because the vertical scale will have been expanded to accommodate it. When the scan is completed the spectrum can be rescaled as follows, in order to see all the peaks. Select MANIPULATE then rescale. Set GRAPH HIGH to 1 and GRAPH LOW to 0. The top of the line at 253.65 nm is off the scale but the lower energy lines are now clearly visible. Values for the wavelengths can be obtained from the Peak Table. The permitted tolerances for wavelength accuracy are quoted in the specifications of Thermo UV-Visible spectrophotometers. The mercury lamp accessory enables the user to demonstrate that the wavelength accuracy of the instrument is within specification quickly and conveniently, whenever necessary. If the required tolerances are not met, a wavelength calibration is necessary. Under normal circumstances this would only be expected following replacement of a major optical component. The mercury lamp calibration remeasures all seven relevant emission lines, then recalculates and replaces the software compensation equation. The permitted tolerances for wavelength accuracy are quoted in the specifications of Thermo UV-Visible spectrophotometers. If the required tolerances are not met, a wavelength calibration is necessary. Under normal circumstances this would only be expected following replacement of a major optical component. The Deuterium lamp recalibration uses the Deuterium emission line at 656.1 nm to adjust the software compensation equation and is intended to remove the effects of small movements of optical components and shifts due to temperature, shock or vibration. If the errors found are linear and no more than 0.5 nm then a Deuterium calibration can be used. The process will take about five minutes and is fully described in the appropriate software manual. This option only applies where an existing calibration is present. Application 3 Use of Rare Earth Glasses A calibrated/traceable set of holmium & didymium glasses is available as follows: 9423 185 03111 Set of 2 calibrated/traceable wavelength filters • Certified Reference Materials (CRM’s). • Holmium filter traceable to National Physical Laboratory (NPL) and National Institute of Standards and Technology (NIST) via ISO/IEC Standard 17025 calibration processes. Application 2 • Didymium filter traceable to NPL only via ISO/IEC Standard 17025 calibration process. Use of Emission Lines from a Deuterium Source • Supplied with calibration certificate. Wavelength verification and recalibration can be performed using the deuterium lamp in UV-Visible instruments having a deuterium source (e.g. Nicolet Evolution 500, UV Series, Spectronic® UV Helios range). • UKAS accreditation number 0521. Method: The emission lines produced from the deuterium lamp are again characteristic of the source element; and as a fundamental physical standard, these wavelengths are invariant. However, unlike the mercury source, only two characteristic lines are usable. These are at 486.0 and 656.1 nm. The emission spectrum (over the desired range) can be easily obtained by selecting scan method parameters as above but with lamp change set to Deuterium (D2). When using an instrument with a photomultiplier (PMT) detector, the FLAT EHT profile must be used for this procedure. • Suitable for use with any appropriate Visible or UV-Visible spectrophotometer. • Includes wooden storage box. Further details are also available in our Standards for UV-Visible Spectrometry publication.2 NIST Traceable Reference Materials 9423 CRM 9300E NIST Traceable Reference Material (NTRM) filter set 1 • Comprises 3 neutral density filters, a clear filter holder and a wavelength check filter. • Suitable for checking the photometric calibration of UV-Visible spectrophotometers in absorbance or transmittance. • Supplied with calibration certificate. • Includes filters of nominal values 0.5, 0.7 and 1A at 546.1 nm. • Calibration values are given at 5 wavelengths. • Holmium glass wavelength filter with certified wavelengths. • Includes transport case. 9423 CRM 1930E NTRM filter set 2 • Comprises 3 neutral density filters, a clear filter holder and a wavelength check filter. Wavelength accuracy checking using a traceable holmium oxide glass filter can be conveniently carried out using one of the following performance qualification accessories: PART NUMBER DESCRIPTION 9423 UV6 1400E Calibration Validation Carousel (NPL) for Spectronic Helios Alpha/Beta Intelligent Calibration Validation Carousel (NPL) for Nicolet Evolution 300 Calibration Validation Carousel (NIST) for Spectronic Helios Alpha/Beta Intelligent Calibration Validation Carousel (NIST) for Nicolet Evolution 300 Calibration Validation Unit (NPL) for Nicolet Evolution 500/UV Series Calibration Validation Unit (NIST) for Nicolet Evolution 500/UV Series 10 01 0601 9423 UV6 1460E • Suitable for checking the photometric calibration of UV-Visible spectrophotometers in Absorbance or transmittance. 10 01 0701 • Supplied with calibration certificate. 9423 UV6 1250E • Includes filters of nominal values 0.3, 1.5 and 2A at 546.1 nm. 9423 UV6 1260E • Calibration values are given at 5 wavelengths. • Holmium glass wavelength filter with certified wavelengths. Application 4 • Includes transport case. Use of Rare Earth Oxide Solutions 9423 CRM 9400E NTRM filter sets 1 and 2 Method: This technique uses solutions of rare earth oxide, prepared by dissolution in acid media. The most frequently used is holmium oxide in perchloric acid. The reference wavelengths have been well-defined3, and a suitable solution is available from various sources. Once the solution has been obtained, the spectrum (over the desired range) can be quickly obtained by selecting the normal scan mode. This procedure does not allow for recalibration of the instrument, but its speed of implementation compared to Applications 1 & 2 makes it ideal for routine checking of the wavelength calibration. NB: The peaks produced when the holmium compound is in solution are much sharper than those produced by the glasses. They are therefore much more sensitive to spectral bandwidth. On a variable bandwidth instrument, 0.5 nm should be used. Compensation tables are available for use at other known bandwidths.3 • Suitable for checking the photometric calibration of UV-Visible spectrophotometers in absorbance or transmittance. • Supplied with calibration certificate. • Includes filters of nominal value 0.3, 0.5, 0.7, 1.0, 1.5, and 2.0A at 546.1 nm. • Calibration values are given at 5 wavelengths. • Holmium glass wavelength filter with certified wavelengths. • Includes transport case. Method: This technique uses glasses manufactured by fusing the appropriate rare earth oxide in silica. The two most frequently used are holmium and didymium (a mixture of neodymium and praseodymium). Manufacturing can cause batch-to-batch variation in these standards. This is overcome by calibrating each standard glass against certified reference glasses from the NPL and NIST (Holmium) or NPL only (Didymium) via an ISO/IEC standard 17025 procedure. With the appropriate filter available the spectrum (over the desired range) can be quickly obtained by selecting the normal scan mode. This procedure does not allow for recalibration of the instrument, but its speed of implementation compared to Applications 1 & 2 makes it ideal for routine checking of the wavelength calibration. NB: On a variable bandwidth instrument, 0.5 nm should be used as the spectral bandwidth. Results: 536.4 1.4 1.2 640.3 1.0 Absorbance • Comprises 6 neutral density filters, clear filter holders and a wavelength check filter. 0.8 278.1 361.2 485.1 0.6 0.4 249.7 385.3 0.2 0 245 295 345 395 445 495 Wavelength (nm) Figure 3: Holmium oxide solution in perchloric acid 545 595 645 Far UV Wavelength Standard This standard consists of a solution of a rare earth oxide dissolved in dilute sulfuric acid, supplied in a permanently heat-sealed Far UV quartz cell. The absorbance spectrum of this rare earth contains a series of characteristic peaks extending into the Far UV region. Specification Material Solution of rare earth oxide in dilute sulfuric acid. Consists of one sealed cell with certified peaks at spectral bandwidth values of 0.1, 0.2, 0.5, 1.0, 1.5, 2.0 and 3.0 nm. Usable Range 200 – 300 nm, instruments with spectral bandpass less than 5 nm 9423 UV9 5540E UV Wavelength filter Thermo Electron Corporation has direct subsidiary offices in North America, Europe and Japan. To complement these direct subsidiaries, we maintain a network of representative organizations throughout the world. Use this reference list or visit our Web site to locate the representative nearest you. References Australia Tel: +61 (0)2 9898 1244 1. C. Burgess & A. Knowles, UV Spectrometry Group, Standards in Absorption Spectrometry, Chapman & Hall, ISBN 0-412-22470-4. Austria Tel: +43 (0)1 333 50 34 0 2. Data Sheet 18, “Standards for UV-Visible Spectrometry” (Thermo Electron Corporation). Belgium Tel: +32 3890 4770 3. Holmium Oxide Solution Wavelength Standard From 240 to 640 nm, SRM 2034 NBS Special Publication, 260-102. Canada Tel: +1 905 890 1034 China Thermo ARL: Tel: +86 10 6833 6715 Thermo Elemental: Tel: +86 10 6592 0232 Thermo Nicolet: Tel: +86 10 6597 3388 ext 2912 France Tel: +33 (0)1 39 30 53 00 Germany Tel: +49 (0)6102 3671 0 Italy Tel: +39 02 6601 6351 Japan Thermo Elemental: Tel: +81 774 201245 Thermo Nicolet: Tel: +81 45 450 1112 Netherlands Tel: +31 76 5724840 Nordic Thermo Nicolet: Tel: +358 9 3291 00 Thermo ARL: Tel: +46 (0)8 556 468 31 South Africa Tel: +27 (0) 11 570 1840 Spain Tel: +34 (91) 657 49 30 www.thermo.com/spectronic USA: 5225 Verona Road Madison, WI 53711-4495 Tel: +1 800 201 8132, +1 608 276 6100 Fax: +1 608 273 5046 Email: [email protected] UK: 8 Mercers Row Cambridge, CB5 8HY, UK Tel: +44 (0)1223 345410 Fax: +44 (0)1223 345411 Email: [email protected] Switzerland: En Vallaire Ouest C case postale, CH-1024, Ecublens Tel: +41 (0)21 694 71 11 Fax: +41 (0)21 694 71 12 Email: [email protected] ©2002 Thermo Electron Corp. All rights reserved worldwide. We make no warranties, expressed or implied, in this product summary, and information is subject to change without notice. All product and company names are property of their respective owners. AN_S0368
© Copyright 2026 Paperzz