INSTRUCTIONS FOR USE Mg Magnesium VITROS Chemistry Products Mg Slides 825 5093 192 1204 Rx ONLY Intended Use For in vitro diagnostic use only. VITROS Chemistry Products Mg Slides quantitatively measure magnesium (Mg) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Summary and Explanation of the Test Magnesium is predominantly an intracellular cation and is essential in enzyme reactions. Magnesium deficiency may cause weakness, tremors, tetany, and convulsions. Hypomagnesemia is associated with hypocalcemia, alcoholism, some types of malnutrition, malabsorption, chronic hemodialysis, and pregnancy. Increased serum magnesium concentrations occur in patients with renal failure, dehydration, and Addison’s disease. 1 Principles of the Procedure The VITROS Mg Slide method is performed using the VITROS Mg Slides and the VITROS Chemistry Products Calibrator Kit 1 on VITROS 250/350/95/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. The VITROS Mg Slide is a multilayered, analytical element coated on a polyester support. A drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Magnesium (both free and protein-bound) from the sample then reacts with the formazan dye derivative in the reagent layer; the high magnesium affinity of the dye dissociates magnesium from binding proteins. The resulting magnesium-dye complex causes a shift in the dye absorption maximum. The amount of dye complex formed is proportional to the magnesium concentration present in the sample and is measured by reflection density. Test Type and Conditions Test Type Colormetric VITROS System 5600, 4600, 5,1 FS, 950, 250/350 Approximate Incubation Time 5 minutes Temperature Wavelength Reaction Sample Volume 37 °C (98.6 °F) 630 nm 5 µL NOTE: The reaction sample volume depends on the format of the slide and is determined automatically by the analyzer. Slides with coating numbers <3201 require a 10 µL reaction sample volume. Not all products and systems are available in all countries. Reaction Scheme chelator Mg2 + Ca+2 Mg+2 + formazan dye derivative Mg+2 + Ca+2-chelator complex pH 9.75 Mg+2-dye complex Warnings and Precautions For in vitro diagnostic use only. WARNING: Version 10.0 Take care when handling materials and samples of human origin. Since no test method can offer complete assurance that infectious agents are absent, consider all clinical specimens, controls, and calibrators potentially infectious. Handle specimens, solid and liquid waste, and test components in accordance with local Pub. No. MP2-47_EN 1 of 14 Mg INSTRUCTIONS FOR USE Magnesium Reagents regulations and CLSI Guideline M29 2 or other published biohazard safety guidelines. For specific warnings and precautions for calibrators, quality control materials, and other components, refer to the Instructions for Use for the appropriate VITROS product, or to other manufacturer’s product literature. Reagents Slide Diagram Slide Ingredients Reactive Ingredients per cm2 1. Upper slide mount 2. Spreading layer (TiO2) 3. Reagent layer • calcium chelator • buffer, pH 9.75 • formazan dye 1,2-bis(o-aminophenoxy)ethane-N,N,N',N'-tetraacetic acid (calcium chelator) 242 µg and 1,5-bis(2-hydroxy-3,5dichlorophenyl)-3-cyanoformazan (dye) 38 µg. 4. Support layer 5. Lower slide mount Other Ingredients Pigment, binders, buffer, dye solubilizer, surfactants, crosslinking agent and stabilizer. Reagent Handling Caution: • • Do not use slide cartridges with damaged or incompletely sealed packaging. Inspect the packaging for signs of damage. Be careful when opening the outer packaging with a sharp instrument so as to avoid damage to the individual product packaging. Reagent Preparation IMPORTANT: The slide cartridge must reach room temperature, 18–28 °C (64–82 °F), before it is unwrapped and loaded into the slide supply. 1. Remove the slide cartridges from storage. 2. Warm the wrapped cartridge at room temperature for 60 minutes. 3. Unwrap and load the cartridge into the slide supply. Note: Load the cartridges within 24 hours after they reach room temperature, 18–28 °C (64–82 °F). Reagent Storage and Stability VITROS Mg Slides are stable until the expiration date on the carton when they are stored and handled as specified. Do not use beyond the expiration date. Reagent Slide Cartridges Opened Frozen On-analyzer On-analyzer Storage Condition ≤-18 °C (≤0 °F) System turned on System turned off Stability Until expiration date ≤ 2 weeks ≤ 2 hours Verify performance with quality control materials: • If the system is turned off for more than 2 hours. • After reloading cartridges that have been removed from the slide supply and stored for later use. Specimen Collection, Preparation and Storage Caution: 2 of 14 Protective gloves manufactured with magnesium stearate (talc) powders may cause elevated test results because of the contamination of sample handling supplies (for example, pipette tips, transfer pipettes, sample cups and caps). Supplies that have come in contact with powdered gloves may subsequently contaminate the test specimen during sample metering. Pub. No. MP2-47_EN Version 10.0 INSTRUCTIONS FOR USE Mg Specimen Collection, Preparation and Storage Note: Magnesium Gloves labeled as “powder-free” may contain some contaminating powder agents on the inside of the glove. Specimens Recommended • • Serum Plasma: 3 Heparin Urine • IMPORTANT: Certain collection devices have been reported to affect other analytes and tests. 4 Owing to the variety of specimen collection devices available, Ortho-Clinical Diagnostics is unable to provide a definitive statement on the performance of its products with these devices. Confirm that your collection devices are compatible with this test. Specimens Not Recommended • Plasma: 5 – EDTA – Fluoride oxalate – Citrate • Urine: 6 Boric acid with sodium formate as a preservative Serum and Plasma Specimen Collection and Preparation Collect specimens using standard laboratory procedures. 7, Note: 8 For details on minimum fill volume requirements, refer to the operating instructions for your system. Patient Preparation No special patient preparation is necessary. Special Precautions • Hemolyzed specimens can cause falsely elevated results due to intracellular magnesium levels. 9 • Centrifuge specimens and remove the serum or plasma from the cellular material as soon as possible after collection. 1 Specimen Handling and Storage • • Handle and store specimens in stoppered containers to avoid contamination and evaporation. Mix samples by gentle inversion and bring to room temperature, 18–28 °C (64–82 °F), prior to analysis. Specimen Storage and Stability: Serum and Plasma 5 Storage Room temperature Refrigerated Frozen Temperature 18–28 °C (64–82 °F) 2–8 °C (36–46 °F) ≤-18 °C (≤0 °F) Stability ≤ 1 week ≤ 1 week ≤ 1 month Urine Specimen Collection and Preparation • • • Collect specimens using standard laboratory procedures. 6 Use metal-free containers for collection and dilution. Keep refrigerated until analysis. Note: For details on minimum fill volume requirements, refer to the operating instructions for your system. Patient Preparation No special patient preparation is necessary. Special Precautions Urine specimens must be pretreated prior to processing. Refer to “Specimen Pretreatment” for instructions. Specimen Handling and Storage • • Version 10.0 Handle and store specimens in stoppered containers to avoid contamination and evaporation. Mix samples by gentle inversion and bring to room temperature, 18–28 °C (64–82 °F), prior to analysis. Pub. No. MP2-47_EN 3 of 14 Mg INSTRUCTIONS FOR USE Magnesium Specimen Pretreatment Specimen Storage and Stability: Urine 5 Storage Room temperature Refrigerated Frozen Temperature 18–28 °C (64–82 °F) 2–8 °C (36–46 °F) ≤-18 °C (≤0 °F) Stability ≤ 1 week ≤ 1 week ≤ 1 month Specimen Pretreatment Urine Acidification Procedure IMPORTANT: The acidification procedure is intended for patient specimens only. Do not adjust the pH of quality-control or calibration materials. IMPORTANT: If running multiple assays from a single acidified sample, ensure that the pH of the sample is appropriate for all assays programmed. 1. Thoroughly mix the urine specimen. 2. Pipette a 5 mL aliquot of urine into a separate test tube. 3. Dropwise, add 6 N HCl, mixing thoroughly after the addition of each drop. Check the pH frequently to obtain a pH between 3.0 and 4.0. 4. If pH is <3.0, discard and begin with another 5 mL aliquot. Predilution Procedure IMPORTANT: If using a VITROS 250/350 Chemistry System, VITROS 5,1 FS/4600 Chemistry System or VITROS Integrated System in On-Analyzer Dilution Mode, do not manually dilute samples for analysis and do not multiply by a dilution factor after analysis. Refer to the operating instructions for your system for more information on the On-Analyzer Dilution Procedure. 1. Dilute the pH-adjusted urine using 1 part of pH-adjusted urine and 5 parts of reagent-grade water to give a sixfold dilution. 2. Analyze the diluted sample and multiply the reported result by 6 to obtain the urine magnesium concentration in the original sample. Testing Procedure Materials Provided VITROS Chemistry Products Mg Slides Materials Required but Not Provided • • • • • • VITROS Chemistry Products Calibrator Kit 1 Quality control materials, such as VITROS Chemistry Products Performance Verifier I and II for serum and plasma tests Isotonic saline or reagent-grade water 6 N HCI VITROS Chemistry Products FS Diluent Pack 2 (BSA/Saline) or VITROS Chemistry Products FS Diluent Pack 3 (Specialty Diluent/Water) (for on-analyzer dilution of serum and plasma samples) VITROS Chemistry Products FS Diluent Pack 3 (Specialty Diluent/Water) (for on-analyzer dilution of urine samples) Operating Instructions • • Check reagent inventories at least daily to ensure that quantities are sufficient for the planned workload. For additional information, refer to the operating instructions for your system. IMPORTANT: Bring all fluids and samples to room temperature, 18–28 °C (64–82 °F), prior to analysis. Sample Dilution Serum and Plasma If samples show magnesium concentrations that exceed the system’s measuring (reportable or dynamic) range: 4 of 14 Pub. No. MP2-47_EN Version 10.0 INSTRUCTIONS FOR USE Mg Calibration Magnesium Manual Sample Dilution 1. Dilute sample with an equal volume of isotonic saline or reagent-grade water. 2. Reanalyze. 3. Multiply the results by 2 to obtain an estimate of the original sample’s magnesium concentration. On-Analyzer Sample Dilution (VITROS Integrated, VITROS 5,1 FS/4600 and 250/350 Systems only) Refer to the operating instructions for your system for more information on the On-Analyzer Dilution Procedure. For VITROS Integrated and 5,1 FS/4600 Chemistry Systems, use VITROS Chemistry Products FS Diluent Pack 2 or VITROS Chemistry Products FS Diluent Pack 3 for the dilution. Urine If samples show magnesium concentrations that are above the upper end or below the lower end of the system’s measuring (reportable or dynamic) range: VITROS 250/350 and 950 Systems 1. Dilute 1 part sample with reagent-grade water as indicated in the table. 2. Sample Concentration * Dilution Factor 9 parts reagent-grade water 10 < lower end** 2 parts reagent-grade water 3 > upper pH-adjusted urine Diluent/Volume end* Above the upper end of the reportable range ** Below the lower end of the reportable range 2. Reanalyze. 3. Multiply the results by the dilution factor to obtain an estimate of the original sample’s magnesium concentration. On-Analyzer Sample Dilution (VITROS Integrated and 5,1 FS/4600 Systems only) Refer to the operating instructions for your system for more information on the On-Analyzer Dilution Procedure. Use VITROS Chemistry Products FS Diluent Pack 3 for the dilution. Calibration Required Calibrators VITROS Chemistry Products Calibrator Kit 1 Note: The same VITROS Calibrator Kit is used to calibrate both serum and urine magnesium. However, specific supplementary assigned values (SAVs) are applied for each body fluid. Calibrator Preparation, Handling, and Storage Refer to the Instructions for Use for VITROS Calibrator Kit 1. IMPORTANT: Do not adjust the pH of calibration materials for use with the test for urine magnesium. Calibration Procedure Refer to the operating instructions for your system. When to Calibrate Calibrate: • When the slide lot number changes. • When critical system parts are replaced due to service or maintenance. • When government regulations require. For example, in the USA, CLIA regulations require calibration or calibration verification at least once every six months. The VITROS Mg test may also need to be calibrated: • If quality control results are consistently outside acceptable range. • After certain service procedures have been performed. For additional information, refer to the operating instructions for your system. Calculations Reflectance from the slide is measured at 630 nm after the fixed incubation time. Once a calibration has been performed for each slide lot, magnesium concentration in unknown samples can be determined using the software-resident endpoint colorimetric math model and the response obtained from each unknown test slide. Version 10.0 Pub. No. MP2-47_EN 5 of 14 Mg INSTRUCTIONS FOR USE Magnesium Quality Control Validity of a Calibration Calibration parameters are automatically assessed by the system against a set of quality parameters detailed in the Coefficients and Limits screen (for VITROS Integrated and 5,1 FS/4600 Systems, see the Review Assay Data screen). Failure to meet any of the pre-defined quality parameters results in a failed calibration. The calibration report should be used in conjunction with quality control results to determine the validity of a calibration. Measuring (Reportable or Dynamic) Range Conventional Units (mg/dL) 0.20–10.0 1.20–60.0 Serum Urine* *After SI Units (mmol/L) 0.08–4.11 0.49–24.68 Alternate Units (mEq/L) 0.16–8.23 0.99–49.37 multiplying by a 6x dilution factor. For out-of-range samples, refer to “Sample Dilution.” Traceability of Calibration Values assigned to the VITROS Chemistry Products Calibrator Kit 1 for magnesium are traceable to the Certified NIST (National Institute of Standards and Technology) Reference Material, SRM® (Standard Reference Material) 929. The OrthoClinical Diagnostics calibration laboratory uses SRM® 929 to calibrate the flame atomic absorption spectroscopy method 10 to support magnesium value assignment for VITROS Calibrator Kit 1. Quality Control Quality Control Material Selection IMPORTANT: • • VITROS Performance Verifiers are recommended for use with the VITROS Chemistry and Integrated Systems. Evaluate the performance of other commercial control fluids for compatibility with this test before using for quality control. Control materials other than VITROS Performance Verifiers may show a difference when compared with other magnesium methods if they: – Depart from a true human matrix. – Contain high concentrations of preservatives, stabilizers, or other nonphysiological additives. Do not use control materials stabilized with ethylene glycol. Urine For urine specimens, use commercially available urine control materials. Quality Control Procedure Recommendations • • • • • • Choose control levels that check the clinically relevant range. Analyze quality control materials in the same manner as patient samples, before or during patient sample processing. To verify system performance, analyze control materials: – After calibration. – According to local regulations or at least once each day that the test is being performed. – After specified service procedures are performed. Refer to the operating instructions for your system. If control results fall outside your acceptable range, investigate the cause before deciding whether to report patient results. For general quality control recommendations, refer to Statistical Quality Control for Quantitative Measurements: Principles and Definitions; Approved Guideline-Third Edition 11 or other published guidelines. For additional information, refer to the operating instructions for your system. Urine • • • Do not adjust the pH of quality-control materials. Dilute sixfold as for a patient sample. Select urine as the fluid when programming the system to run urine controls. IMPORTANT: 6 of 14 If using a VITROS 250/350 Chemistry System, VITROS 5,1 FS/4600 Chemistry System or VITROS Integrated System in On-Analyzer Dilution Mode, do not manually dilute samples for analysis and do not multiply by a dilution factor after analysis. Refer to the operating instructions for more information on the OnAnalyzer Dilution Procedure. Pub. No. MP2-47_EN Version 10.0 INSTRUCTIONS FOR USE Mg Results Magnesium Quality Control Material Preparation, Handling, and Storage Refer to the Instructions for Use for VITROS Chemistry Products Performance Verifier I and II or to other manufacturer's product literature. Results Reporting Units and Unit Conversion The VITROS Chemistry and Integrated Systems may be programmed to report Mg results in conventional, SI, or alternate units. Conventional Units mg/dL SI Units mmol/L (mg/dL x 0.4114) Alternate Units mEq/L (mg/dL x 0.8229) Limitations of the Procedure Known Interferences • • • The VITROS Mg Slide method was screened for interfering substances following NCCLS Protocol EP7 12 The substances listed in the table, when tested at the concentrations indicated, caused the bias shown. Erythrocytes in urine can produce falsely elevated results due to intracellular magnesium levels. For substances that were tested and did not interfere, refer to “Specificity.” Interferent* Serum and Plasma Calcium Urine Inorganic phosphorus Calcium Sodium fluoride Interferent Concentration Magnesium Concentration Conv. SI (mg/dL) (mmol/L) Average Bias Conv. SI (mg/dL) (mmol/L) 20 mg/dL (5.0 mmol/L) 2.3 0.9 +0.3 +0.12 300 mg/dL (96.8 mmol/L) 7.2 3.0 -2.4 -1.00 300 mg/dL 40 mg/dL 10 mg/mL (74.9 mmol/L) (10.0 mmol/L) (238 mmol/L) 20.6 21.6 7.0 8.5 8.9 2.9 +7.0 +0.5 -3.0 +2.9 +0.2 -1.2 *It is possible that other interfering substances may be encountered. These results are representative; however, your results may differ somewhat due to test-to-test variation. The degree of interference at concentrations other than those listed might not be predictable. Other Limitations Certain drugs and clinical conditions are known to alter magnesium concentrations in vivo. For additional information, refer to one of the published summaries. 9, 14 Expected Values Reference Interval The serum reference interval is the central 95% of results from an internal study of 288 apparently healthy individuals from a working population (58 females and 230 males). The urine reference interval is based on an external study. 1 Serum Urine (24-Hour) *Magnesium Conventional Units (mg/dL) 1.6–2.3 73.0–122.0 mg/day* SI Units (mmol/L) 0.7–1.0 3.0–5.0 mmol/day** Alternate Units (mEq/L) 1.3–1.9 6.0–10.0 mEq/day*** concentration (mg/dL) x 24-hour volume (dL) = mg/day. To convert mg/day to g/day divide by 1000. **Magnesium concentration (mmol/L) x 24-hour volume (L) = mmol/day ***Magnesium concentration (mEq/L) x 24-hour volume (L) = mEq/day Each laboratory should confirm the validity of these intervals for the population it serves. Version 10.0 Pub. No. MP2-47_EN 7 of 14 INSTRUCTIONS FOR USE Mg Magnesium Performance Characteristics Performance Characteristics Method Comparison The plots and tables show the results of a comparison of serum and urine samples analyzed on the VITROS 750 System with those analyzed using the atomic absorption comparative method. 10 Testing followed NCCLS Protocol EP9. 15 The tables, for serum and urine respectively, summarize the results of regression analyses of data obtained from measurements with both the VITROS 250 and 950 Systems compared to measurements with the VITROS 750 System. Regression analyses of data for serum and urine samples measured on the VITROS 5,1 FS System compared to measurements with the VITROS 950 System are also provided. In addition, the tables for serum and urine summarize the regression analysis of comparisons of results for serum, plasma and urine samples measured on the VITROS 5600 Integrated System and the VITROS 5,1 FS Chemistry System. Testing followed NCCLS Protocol EP9. 16 Serum SI Units VITROS 750 System (mg/dL) VITROS 750 System (mmol/L) Conventional Units Comparative Method: Atomic Absorption (mg/dL) Comparative Method: Atomic Absorption (mmol/L) Conventional Units (mg/dL) SI Units (mmol/L) n Slope Correlation Coefficient 750 vs. comparative method 160 1.01 0.999 0.3–9.6 -0.02 0.10 0.1–3.9 -0.01 0.04 250 vs. 750 88 0.99 1.000 0.6–9.7 0.00 0.08 0.2–4.0 0.00 0.03 950 vs. 750 126 1.00 1.000 0.4–9.5 -0.01 0.03 0.2–3.9 0.00 0.01 5,1 FS vs. 950 124 0.99 1.000 0.2–9.2 +0.03 0.05 0.1–3.8 +0.01 0.02 5600 vs. 5,1 FS 110 1.00 0.999 0.5–9.7 +0.01 0.13 0.2–4.0 0.00 0.05 Range of Sample Conc Intercept Sy.x Range of Sample Conc. Intercept Sy.x Analytical processing hardware and software algorithms on the VITROS 4600 Chemistry System are designed to the same specifications as those applied to the VITROS 5,1 FS Chemistry System. Assay performance on the VITROS 4600 System has been demonstrated to be comparable to that on the VITROS 5,1 FS System. All performance characteristics for VITROS 5,1 FS System are therefore applicable to the VITROS 4600 System. 8 of 14 Pub. No. MP2-47_EN Version 10.0 Mg INSTRUCTIONS FOR USE Performance Characteristics Magnesium Urine SI Units VITROS 750 System (mg/dL) VITROS 750 System (mmol/L) Conventional Units Comparative Method: Atomic Absorption (mg/dL) Comparative Method: Atomic Absorption (mmol/L) Conventional Units (mg/dL) SI Units (mmol/L) n Slope Correlation Coefficient 750 vs. comparative method 156 1.00 0.998 1.4–58.8 +0.54 0.91 0.6–24.2 +0.22 0.37 250 vs. 750 89 0.99 0.999 2.1–53.6 -0.05 0.41 0.9–22.1 -0.02 0.17 950 vs. 750 159 1.01 1.000 2.0–59.5 -0.06 0.14 0.8–24.5 -0.02 0.06 5,1 FS vs. 950 108 1.01 1.000 1.3–55.1 -0.14 0.40 0.5–22.7 -0.06 0.16 5600 vs. 5,1 FS 108 1.02 1.000 1.4–57.8 -0.12 0.34 0.6–23.8 -0.05 0.14 Range of Sample Conc. Intercept Sy.x Range of Sample Conc. Intercept Sy.x Analytical processing hardware and software algorithms on the VITROS 4600 Chemistry System are designed to the same specifications as those applied to the VITROS 5,1 FS Chemistry System. Assay performance on the VITROS 4600 System has been demonstrated to be comparable to that on the VITROS 5,1 FS System. All performance characteristics for VITROS 5,1 FS System are therefore applicable to the VITROS 4600 System. Precision Precision was evaluated with quality control materials on VITROS 250, 750, 950, and 5,1 FS Systems following NCCLS Protocol EP5. 17 Precision was also evaluated with quality control materials on VITROS 5600 Integrated System following NCCLS protocol EP5. 18 The data presented are a representation of test performance and are provided as a guideline. Variables such as sample handling and storage, reagent handling and storage, laboratory environment, and system maintenance can affect reproducibility of test results. Version 10.0 Pub. No. MP2-47_EN 9 of 14 Mg INSTRUCTIONS FOR USE Magnesium Performance Characteristics Serum Conventional Units (mg/dL) 250 750 950 5,1 FS 5600 * SI Units (mmol/L) Within Lab Within Lab SD** CV%** Mean Conc. Within Day SD* Within Lab SD** Mean Conc. Within Day SD* 2.1 0.03 0.05 0.9 0.01 0.02 2.4 79 20 4.5 0.04 0.08 1.9 0.02 0.03 1.7 78 20 2.2 0.03 0.04 0.9 0.01 0.02 1.6 90 23 4.7 0.05 0.06 1.9 0.02 0.02 1.3 91 23 2.3 0.03 0.03 1.0 0.01 0.01 1.3 90 23 4.8 0.05 0.06 2.0 0.02 0.02 1.2 92 23 2.1 0.02 0.04 0.9 0.01 0.01 1.7 88 22 4.3 0.03 0.04 1.8 0.01 0.02 0.9 89 22 1.8 0.01 0.03 0.7 0.00 0.01 1.7 92 23 4.2 0.03 0.07 1.7 0.01 0.03 1.7 92 23 No. Observ. No. Days Within Day precision was determined using two runs/day with at least two replications. ** Within Lab precision was determined using a single lot of slides and calibrating weekly. Analytical processing hardware and software algorithms on the VITROS 4600 Chemistry System are designed to the same specifications as those applied to the VITROS 5,1 FS Chemistry System. Assay performance on the VITROS 4600 System has been demonstrated to be comparable to that on the VITROS 5,1 FS System. All performance characteristics for VITROS 5,1 FS System are therefore applicable to the VITROS 4600 System. Precision for Mg: Urine Conventional Units (mg/dL) System 250 750 950 5,1 FS 5600 * SI Units (mmol/L) Within Lab Within Lab CV%** SD** Mean Conc. Within Day SD* Within Lab SD** Mean Conc. Within Day SD* 6.2 0.06 0.17 2.5 0.03 0.07 9.2 0.09 0.16 3.8 0.04 11.2 0.08 0.18 4.6 15.4 0.09 0.20 7.0 0.07 32.9 No. Observ. No. Days 2.7 88 22 0.07 1.7 88 22 0.03 0.08 1.6 88 22 6.3 0.04 0.08 1.3 88 22 0.18 2.9 0.03 0.07 2.6 90 23 0.25 0.83 13.5 0.10 0.34 2.5 86 22 7.2 0.12 0.18 3.0 0.05 0.07 2.9 90 23 33.0 0.18 0.72 13.6 0.07 0.30 2.3 86 23 5.6 0.09 0.16 2.3 0.04 0.07 2.9 91 23 13.3 0.11 0.20 5.5 0.04 0.08 1.5 93 23 4.1 0.13 0.20 1.7 0.05 0.08 4.9 92 23 11.4 0.17 0.23 4.7 0.07 0.09 2.0 92 23 Within Day precision was determined using two runs/day with two to three replications. ** Within Lab precision was determined using a single lot of slides and calibrating weekly. Analytical processing hardware and software algorithms on the VITROS 4600 Chemistry System are designed to the same specifications as those applied to the VITROS 5,1 FS Chemistry System. Assay performance on the VITROS 4600 System has been demonstrated to be comparable to that on the VITROS 5,1 FS System. All performance characteristics for VITROS 5,1 FS System are therefore applicable to the VITROS 4600 System. Specificity Substances That Do Not Interfere The substances listed in the table were tested with VITROS Mg Slides, following NCCLS Protocol EP7 12 and found not to interfere, bias <0.2 mg/dL (<0.08 mmol/L), at the concentrations shown. 10 of 14 Pub. No. MP2-47_EN Version 10.0 INSTRUCTIONS FOR USE Mg References Compound Acetaminophen Magnesium Concentration Compound 5 mg/dL 331 µmol/L Aluminum 300 µg/dL 111.3 µmol/L Bilirubin 27 mg/dL 462 µmol/L Calcium 16 mg/dL 4 mmol/L Copper Concentration Lithium 1 mEq/L 1 mmol/L Nickel 15 µg/dL 2.6 µmol/L Potassium 8 mEq/L 8 mmol/L Salicylate 30 mg/dL 2.2 mmol/L 156 mEq/L 156 mmol/L 500 µg/dL 78.5 µmol/L Sodium Cyclosporin 15 µg/dL 125 nmol/L Total protein 3 g/dL 30 g/L Hemoglobin 1000 mg/dL 10 g/L Total protein 10 g/dL 100 g/L Iron 500 µg/dL 89.5 µmol/L Triglycerides 1200 mg/dL 13.5 mmol/L Intralipid 800 mg/dL 8 g/L Urea nitrogen 100 mg/dL 35.7 mmol/L 5 mg/dL 561 µmol/L Zinc 300 µg/dL 45.9 µmol/L L-Alanine References 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. Version 10.0 Tietz NW (ed). Fundamentals of Clinical Chemistry. ed. 3. Philadelphia: WB Saunders; 521-524; 1987. CLSI. Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline – Third Edition. CLSI document M29-A3 (ISBN 1-56238-567-4). CLSI, 940 West Valley Road, Suite 1400, Wayne, PA 19087-1898 USA; 2005. Doumas BT, et al. Differences Between Values for Plasma and Serum in Tests Performed in the Ektachem 700 XR Analyzer, and Evaluation of “Plasma Separator Tubes (PST).” Clin. Chem. 35:151–153; 1989. Calam RR. Specimen Processing Separator Gels: An Update. J Clin Immunoassay. 11:86-90; 1988. Clinical Laboratory Handbook for Patient Preparation and Specimen Handling. Fascicle VI: Chemistry/Clinical Microscopy. Northfield, IL: College of American Pathologists; 1992. NCCLS. Urinalysis and Collection, Transportation, and Preservation of Urine Specimens; Approved Guideline. NCCLS Document GP16. CLSI, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA; 1995. CLSI. Procedures for the Collection of Diagnostic Blood Specimens by Venipuncture; Approved Standard-Sixth Edition. CLSI document H3-A6 (ISBN 1-56238-650-6). CLSI, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA; 2007. NCCLS. Procedures and Devices for the Collection of Diagnostic Capillary Blood Specimens; Approved Standard— Fifth Edition. NCCLS document H4-A5 [ISBN 1-56238-538-0]. CLSI, 940 West Valley Road, Suite 1400, Wayne, PA 19087-1898 USA, 2004. Young DS. Effects of Preanalytical Variables on Clinical Laboratory Tests. ed. 2. Washington D.C.: AACC Press; 3-370; 1997. Kaplan L, Pesce A. Clinical Chemistry: Theory, Analysis, and Correlation. CV Mosby; 1069; 1984. NCCLS. Statistical Quality Control for Quantitative Measurements: Principles and Definitions; Approved Guideline – Third Edition. CLSI document C24-A3 (ISBN 1-56238-613-1). CLSI, 940 West Valley Road, Suite 1400, Wayne, PA 19087-1898 USA; 2006. NCCLS. Interference Testing in Clinical Chemistry. NCCLS Document EP7. CLSI, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA; 1986. Young DS. Effects of Drugs on Clinical Laboratory Tests. ed. 4. Washington D.C.: AACC Press; 1995. Friedman RB. Young DS. Effects of Disease on Clinical Laboratory Tests. Washington, D.C.: AACC Press; 1990. NCCLS. Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline. NCCLS Document EP9. CLSI, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA: NCCLS; 1995. NCCLS. Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline. NCCLS document EP9-A2 [ISBN 1-56238-472-4]. CLSI, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA; 2002. NCCLS. User Evaluation of Precision Performance with Clinical Chemistry Devices. NCCLS Document EP5. CLSI, 940 West Valley Road, Suite 1400, Wayne, Pennsylvania 19087-1898 USA; 1992. NCCLS. Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Second Edition. NCCLS document EP5-A2 [ISBN 1-56238-542-9]. CLSI, 940 West Valley Road, Suite 1400, Wayne, PA 19087-1898 USA; 2004. Pub. No. MP2-47_EN 11 of 14 Mg INSTRUCTIONS FOR USE Magnesium Glossary of Symbols Glossary of Symbols Revision History 12 of 14 Date of Revision 2015-10-12 Version 10.0 2014-09-05 2012-02-28 2010-11-01 9.0 8.0 7.0 2008-10-22 6.0 2005-12-15 2004-12-23 5.0 4.0 Description of Technical Changes* • Prescription Use statement added. • Updated EC Representative address. • Added USA to legal manufacture address Glossary of Symbols: added Date of Manufacture Glossary of Symbols: updated • Added information for the VITROS 4600 Chemistry System • Precision – Urine: 5,1 FS: Corrected No. Observ. • Added information for the VITROS 5600 Integrated System • Test Type and Conditions – Added statement • Method Comparison – Added information on sample types • References – Updated • Glossary of Symbols – Updated • Minor wording and formatting changes Acidification procedure — added “Important” Note that the Date of Revision for this version (4.0) was corrected to the actual date of 2004-12-23 after release. • Materials Required But Not Provided – updated data • Sample Dilution, Urine – updated data Pub. No. MP2-47_EN Version 10.0 INSTRUCTIONS FOR USE Mg Revision History Date of Revision 2004-09-13 Version 3.0 2002-12-16 2.0 2002APR19 1.0 – English only Magnesium Description of Technical Changes* • Added VITROS 5,1 FS Chemistry System • Specimen Requirements, Special Precautions – wording update • Reference Interval – updated data • Precision, 250 and 950 Systems – updated values • Specificity – added Intralipid; updated Bilirubin, Hemoglobin • References – corrected data • Glossary of Symbols – updated data • New organization and sections consistent with IVD Directive • Test Type and Conditions - added sample drop volume for slides with coating numbers 3201 and above • Specimens Recommended - plasma: updated to heparin. • Specimens Not Recommended - urine: updated • Reportable (Dynamic) Range - urine: corrected alternate units • Reference Interval - 24 hour urine: corrected unit • Known Interfering Substances - urine: removed statement that preservatives cause negative biases • Known Interfering Substances table - serum and plasma: removed inorganic phosphorous; urine: added sodium fluoride; updated inorganic phosphorous and calcium • Method Comparison - serum: updated all comparisons and plots; urine: updated 750 System versus comparative method, 250 versus 750 Systems, and plots • Precision: serum: updated data for 250 and 750 System; urine: updated data for 750 System • Specificity - updated data for total protein • References – added 2, 3, 4, 12, and 15 New format, technically equivalent to 11/96. * The change bars indicate the position of a technical amendment to the text with respect to the previous version of the document. When this Instructions For Use is replaced, sign and date below and retain as specified by local regulations or laboratory policies, as appropriate. Signature Version 10.0 Obsolete Date Pub. No. MP2-47_EN 13 of 14 Mg INSTRUCTIONS FOR USE Magnesium Revision History Ortho-Clinical Diagnostics Felindre Meadows Pencoed Bridgend CF35 5PZ United Kingdom Ortho-Clinical Diagnostics, Inc. 100 Indigo Creek Drive Rochester, NY 14626 USA VITROS is a trademark of Ortho-Clinical Diagnostics, Inc. © Ortho-Clinical Diagnostics, Inc., 2002-2015 14 of 14 Pub. No. MP2-47_EN Version 10.0
© Copyright 2026 Paperzz