joint-stock company Natural gas underground storage SPECIFICATION of gas chromatographs for the investment project – Exchange of chromatographs MS Number of pages: Malacky, July 2013 3 1 Gas chromatograph – process The procurement subject is a supply of 2 gas process chromatographs with control units. Required parameters: an analysis of the composition of natural gas according to the standard ISO 6974:2000 measured elements – in mol. % depicted to 3 decimal places: 1. methane (C1) 2. ethane (C2) 3. propane (C3) 4. n-butane (n-C4) 5. i-butane, (i-C4) 6. n-pentane (n-C5) 7. neo-pentane (neo-C5) 8. i-pentane (i-C5) 9. hexane and higher hydrocarbons in total (C6+) 10. nitrogen (N2) 11. carbon dioxide (CO2) 12. oxygen (O2) the calculation of physical quantities according to the standard ISO 6976: a) combustion heat, Hs kWh/Sm3 depiction 3 decimal places b) calorific value, Hi kWh/Sm3 depiction 3 decimal places c) relative density, d depiction 4 decimal places d) absolute density, kg/Sm3 depiction 4 decimal places e) Wobbe index, Ws kWh/Sm3 depiction 3 decimal places an analysis of measured samples sequentially in a closed cycle (sequence selectable by the operator) interval of an analysis not more than 15 minutes automatic and manual calibration. Automatic calibration – pre-optional by operator (differences for repeated calibrations with a known calibration mixture and the calculated GCV must be within a range of max 0.2% relative) 2 inputs for measured gas, 1 input for calibration gas analog and digital outputs archiving of measured data – minimum and maximum values and the calculation of average values 1÷12. all measured elements of the gas, except for item 7. (neo-pentane) - combustion heat, Hs kWh/Sm3 - calorific value, Hi kWh/Sm3 - absolute density, ρ kg/Sm3 - relative density, d - Wobbe index, Ws kWh/Sm3 reproducibility and repeatability according to the standard ISO 6974: 2000, part 5 EEx preparation for the zone 1 the chromatographic system is to comply with the requirements of the standard ISO 6974: 2000, part 1 and part 2 the analytic method of determination is to comply with the requirements of the standard ISO 6974:2000, part 5 for determination of the compressibility factor to apply standards SGERG–88, AGA NX 19 and AGA 8–92 communication by the protocol Modbus in the regime Master-Slave and the regime Slavemaster – both concurrently output for a printing machine and display (print and the visualization of immediate values) Exchange of chromatographs MS page 2 Combustion heat – for the calculation shall apply the values of combustion heat in kWh/Sm3 components of gas based on chart 5 column 25/15°C and the compressibility factor calculated under Chart 2 column 15°C/101.325 kPa standard ISO 6976, currently applicable publication in the SR. The results of calculations shall be rounded up to 3 decimal places. It must be possible to switch reference conditions into values (25/20°C and 25/0°C). The reference conditions for MS Láb IV: Pn = 101.325 kPa and Tn 15°C (switchable to 20°C and 0°C). The type of the chromatographs must have the confirmation of the type applicable in the Slovak Republic. 2 Gas chromatograph – sulphur The procurement subject is a supply of 1 process gas chromatograph with a control unit. Required parameters: determination of sulphur mixtures in natural gas under ISO 19739:2004 measured elements – scope of measurement: a) hydrogen sulphide (H2S) 0 ÷ 7,5 mg/Sm3 b) carbonoxylsulfide (COS) 0 ÷ 15 mg/Sm3 c) metylmerkaptane (CH3SH) 0 ÷ 50 mg/Sm3 Calculated elements: d) total sulphur 0-150 mg/Sm3 values for a calibration gas mixture: e) hydrogen sulphide (H2S) 5 mg/Sm3 f) carbonoxylsulfide (COS) 15 mg/Sm3 g) metylmerkaptane (CH3SH) 50 mg/Sm3 analysis of measured samples sequentially in a closed cycle (sequence selectable by the operator) interval of the analysis max. 10 minutes automatic and manual calibration. Automatic calibration – pre-optional by operator Accuracy of calibration (allowance from a calibration gas mixture): - H2S ±5% - COS ±2% - CH3SH ±5% one input for measured gas, two inputs for calibration gas analog and digital outputs archiving of measured data, the calculation of max., min. and average values and the archiving thereof EEx preparation for the zone 1 Communication by the protocol Modbus in the regime Master-Slave output for a printing machine and display (print and the visualization of immediate values) Exchange of chromatographs MS page 3
© Copyright 2026 Paperzz