What is Spectroscopy? • Spectroscopy: a general term for the science that deals with the interactions of various types of radiation (electromagnetic radiation and other forms of energy) with matter • Spectrometry and spectrometric methods: -a large group of analytical methods that are based on atomic and molecular spectroscopy -the measurement of the intensity of radiation with a photoelectric transducer or other type of electronic device. • Colorimetry: based on absorption of visible light. Ch. 18 – Ch. 20 Spectrophotometry Properties of Electromagnetic Radiation Classical sinusoidal wave model • Plane-polarized electromagnetic radiation of wavelength λ, propagating along the x axis. • The electric field of the plane-polarized light is confined to a single plane. • Ordinary, unpolarized light has electric field components in all planes. υ ×λ = c c = 2.998 ×108 m/s •υ: frequency (Hz) •λ: wavelength •c: the velocity of radiation in vacuum Properties of Electromagnetic Radiation Particle model • When EM radiation is emitted or absorbed, a permanent transfer of energy from the emitting object or to the absorbing medium occurs. • It’s necessary to treat EM radiation not as a collection of waves but rather as a stream of discrete particles called photons or quanta. • The discovery of the photoelectric effect in the late nineteenth century. E = hυ E = hcυ~ h = 6.626 × 10 −34 J ⋅ s υ~ = 1 / λ (wave number) Electromagnetic Spectrum Example: How much is the energy of CO2 increased per mole when it absorbs IR at 2300 cm-1? ΔE = h υ = h c λ = hcυ~ = (6.626 × 10 J ⋅ s)(2.998 × 10 m/s)(2300 cm )(100 cm/m) = 4.6 × 10 J -34 8 -1 - 20 ( 4.6 × 10 J)(6.022 × 10 ) = 28 kJ/mol - 20 23 E = hcυ~ υ~ = 1 / λ (wave number) 1 A Spectrum of White Light Line Spectrum A spectrum produced by a luminous gas or vapor and appearing as distinct lines characteristic of the various elements constituting the gas. Emission Spectrum Hydrogen Line Emission Spectrum The spectrum of bright lines, bands, or continuous radiation characteristic of and determined by a specific emitting substance subjected to a specific kind of excitation. Ground State The state of least possible energy in a physical system, as of elementary particles. Also called ground level. Excited State Being at an energy level higher than the ground state. 2 Methods of Interaction Absorption Spectrum • Light shinning on a sample causes electrons to be excited from the ground state to an excited state • The wavelengths of that energy are removed from transmitted spectra M + hν → M * M * → M + heat (excited state) (relaxation) •Absorption of light increases the energy of a substance. •Emission of light decreases its energy. Types of Molecular Transitions Interactions of Radiation and Matter Electronic transition – electron between two orbitals (UV-vis). hν = Energy Difference of Orbitals Vibrational transitions – vibrational states associated with bonds holding molecule. Rotational transitions – changes in rotational states, about its center of gravity. Overall change in Energy with an orbital of a molecule is: E = Eelectronic + Evibrational + Erotational Interactions of Radiation and Matter Single-beam Absorbance Experiment grating, prism or filter Irradiance out Lamp or laser Irradiance in Vibrational energy transferred Pathlength Irradiance (P): the energy per unit area in the light beam (W/m2). Relaxation by emitting heat Transmittance (T): the fraction of original light not absorbed by the sample. T = P/P0 Relaxation by emitting photons Monochromatic light: consists of “one color” (one wavelength). Chromophore: the part of a molecule responsible for light absorption 3 Derivation of Beer’s Law Derivation of Beer’s Law P ∫ P0 dP = − ∫ βcdx P b 0 After integrate both sides: dP = − β Pcdx ln dP: the change in irradiance β: a constant of proportionality c: concentration dP ∫ = − ∫ βcdx P dP = − βcdx P P b P0 0 P P = 2.303 × log = − βcb P P 0 log Absorbance and Beer’s Law β P = cb = εcb P 2.303 0 − log T = A = εcb Relation between Transmittance and Absorbance • Absorbance (A): the amount light absorbed by the sample is related to transmittance (T): ⎛P ⎞ A = log⎜ 0 ⎟ = − log T ⎝P⎠ • Beer’s law relates the absorbance of a chemical to its concentration: ⎛P ⎞ A = log⎜ 0 ⎟ = − log T ⎝P⎠ P/Po 1 0.1 0.01 Aλ = ε λ bc b : the pathlength, typically in cm, and c is the concentration of the chemical species in M ε : the molar absorptivity, the unit that tells how much light is absorbed for a given wavelength. ε has units of M-1 cm-1 Absorption Methods - Beer’s Law A = abc = εbc where a => absorptivity b => path length c => concentration ε => molar absorptivity 0 %T 100 10 1 A 0 1 2 Absorption Methods - Beer’s Law A = abc = εbc A c • The light being shined on the sample must be monochromatic (one color or wavelength) • The analyte must not be participate in a concentration dependent equilibrium 4 1.0 1 - 10 mg l 0.8 r = 0.8791 -1 • Real Limitations – high concentrated solutions, concentrated electrolyte solutions (proximity alters molecular absorption). NO 3 -N 0.6 0.4 Spectroscopic Procedure • Have a spectrometer and cuvette(s) – Single-beam instrument has one sample holder (swap blank and sample) – Double-beam instrument splits light output between two holders (measure blank and sample) – A baseline spectrum is a spectrum of a reference solution (solvent or reagent blank) 0.2 0.0 0 2 4 6 [NO 3-N] (mg l -1 8 10 ) HNO + H O ←⎯→ H O + NO − 2 2 3 2 • Chemical Limitations – absorbing species participate in association or dissociation reactions, e.g. weak acids in concentrated solutions, complexation. Spectroscopic Procedure • Keep the absorbance reading of the sample below 1. Spectroscopic Procedure • Try to do an analysis at the λmax – Sensitivity is greatest at maximum absorbance – Curve is relatively flat in case the monochromator drifts and is off by a little in wavelength – % transmittance is related logarithmically with concentration (from 1-99% transmittance you can detect ~ 2 orders of magnitude in analyte concentration) – Any orders of magnitude greater than that will be detected in the range of 0-1% T. λmax.= 620 nm 0.5 550 nm 0.4 • Dilute the solution so that the transmittance reading is not maxed out in that region (for accuracy) 0.3 510 nm 0.2 Absorbance Abs Limitations to Beer’s Law 0.1 0.0 -0.1 -0.2 350 400 450 500 550 600 650 700 Wavelength (nm) Example: Find the absorbance and transmittance of a 0.0220 M solution with a molar absorptivity of 15.5 M-1 cm-1 in a 2.00 cm pathlength cell Example: Consider a solution of benzene in hexane (28.8 mg in 250 mL solution). If the absorbance of your 1 cm cell is 0.266 at λ=200nm, what is the ε value for benzene (FW: 78.11)? Aλ = ε λ bc = (15.5 M -1cm -1 )(2.00 cm)(0.0220 M) = 0.68 log T = − A T = 10 = 0.21 − 0.68 21% of the irradiated light emerges from the analyte sample c= 28.8mg C6H6 250 ml sol’n 1g 1000ml 1000mg 1L c=[C6H6]=1.321x10-3M 1mol C6H6 78.11 g C6H6 ε=A /cb =A/c ε=0.266/1.321x10-3 5 Example: A 22.5 mg sample of anthracene (C14H10, MW=178.23) was dissolved in 100 mL ethanol and diluted by 50% four times. The absorbance reading of a 1.0 cm cell of this solution at 266 nm was still 0.93. What is ε? [C14 H10 ] = (0.0225 g)/(178.22 g/mol) × ⎛⎜ 1 ⎞⎟ 0.1 L ε= Wavelengths and Color 4 ⎝2⎠ = 7.89 × 10 −5 M A 0.93 = = 1.2 × 10 4 M -1cm −1 bc (1.0)(7.89 × 10 −5 ) Analysis of Mixtures MixtureAbs = ε xb[ x] + εY b[Y ] + ...... A1 = ε 1 (λ1 )c1b + ε 2 (λ1 )c 2 b A2 = ε 1 (λ2 )c1b + ε 2 (λ2 )c 2 b Isosbestic points • Good evidence of presence of only 2 species which interchange between themselves (e.g. indicator dye with 2 states) •What are the knowns and unknowns in the above Eq.? •Measure Abs at more wavelengths than there are components in the mixture. •More wavelengths increase the accuracy. Emission Processes Absorption vs. Emission Spectra • Luminescence – emission of light from any excited state of a molecule. more sensitive than absorption measurements – Fluorescence – emission of photon during transition between states with same spin quantum #’s (e.g. S1 →S0 ) – Phosphorescence - emission of photon during transition between states with different spin quantum #’s (e.g. T1 →S0 ) 6 Emission Spectroscopy Experiment Absorption vs. Emission Spectra Emission Intensity at Low Concentration ⎛P ⎞ A = log⎜ ⎟ = ε bc ⎝P⎠ Irradiance striking central region : P = P 10 ε Components of Optical Instruments 0 ex − ' 0 Additional distance b : P = P 10 ε ' − ' 2 Single Beam Instruments ex b1 c 0 ex b 2 c 0 Emission intensity is proportional to the irradiance absorbed in the central region of the cell : I = k ( P − P ) ' ' ' ' 0 Double Beam Instruments Emission intensity emerging from the cell : I = I 10 ε ' − ' = k P 10 ' 0 em b 3 c −ε ex b1 c = k ( P − P )10 ε ' ' − ' = k ( P 10 ε (1 − 10 )10 − ' em b 3 c 0 −ε ex b 2 c 0 ex b1 c − P 10 ε 10 ε − − ex b1 c 0 ex b 2 c )10 ε − em b 3 c −ε em b 3 c (ε b c ln 10) (ε b c ln 10) − + ... 2! 3! I = kP0 c ∴ At low concentration : k P ε b c ln 10 ⇒ Q10 −ε ex b 2 c 2 = 1 − ε b c ln10 + ex 2 ex 3 ex 2 2 ' 0 ex 2 Instrumentation - Basic Components Intensity of a Tungsten Filament and a Deuterium Arc Lamp (1) Light source -Tungsten filament lamp (Visible: 320 - 2500 nm); Deuterium arc lamp (UV: 200-400 nm); Nernst glower, globar (IR region). 7 Instrumentation - Basic Components (1) Light source -Tungsten filament lamp (Visible: 320 - 2500 nm); Deuterium arc lamp (UV: 200-400 nm); Nernst glower, globar (IR region). (2) Monochromators/Filters: select a narrow band of wavelengths to pass on to the sample or detector. -Gratings & Prisms: disperses light into its component wavelengths -Filters: filter (remove) wide bands of radiation from a signal; Colored glass Instrumentation - Basic Components Grating vs. Prism •Grating: a reflective or transmissive optical component with a series of closely ruled lines; can bend light (diffraction) •Prism: can bend light (refraction) Cells for Spectrophotometry (1) Light source -Tungsten filament lamp (Visible: 320 - 2500 nm); Deuterium arc lamp (UV: 200-400 nm); Nernst glower, globar (IR region). (2) Monochromators/Filters: select a narrow band of wavelengths to pass on to the sample or detector. -Gratings & Prisms: disperses light into its component wavelengths -Filters: filter (remove) wide bands of radiation from a signal; Colored glass (3) Sample containers (cuvettes or cells) -Glass, Plastic & Quartz (usually 1 cm); KBr/NaCl (IR) Instrumentation - Basic Components Detector Response (1) Light source -Tungsten filament lamp (Visible: 320 - 2500 nm); Deuterium arc lamp (UV: 200-400 nm); Nernst glower, globar (IR region). (2) Monochromators/Filters: select a narrow band of wavelengths to pass on to the sample or detector. -Gratings & Prisms: disperses light into its component wavelengths -Filters: filter (remove) wide bands of radiation from a signal; Colored glass (3) Sample containers (cuvettes or cells) -Glass, Plastic & Quartz (usually 1 cm); KBr/NaCl (IR) (4) Detector: produces an electric signal when it is struck by photons -Phototubes, Photomultiplier tubes, & Photodiode array •A function of wavelength of incident light •The greater the sensitivity, the greater the current of voltage produced by the detector for a given incident irradiance of photons. 8
© Copyright 2026 Paperzz