HP 85044A/B Transmisslon/ReflectlonTest Sets The HP 85044A/B test sets provide the capability to measure the reflection and transmissioncharacteristics of 500. and 75 n devices, respectively. Specifications Summary Impedance: Frequency range: Directivi~: HP 85044A HP 85044 B 50n 75n 300 kHz to 3 GHz 300 kHz to 2 GHz 35 dB to 1.3GHz 35 dB to 1.3GHz 30 dB [0 3.0 GHz 30 dB to 2.0 GHz Typical Tracking Transmission magnitude, phasel.2.3~ 0.3 MHz [0 2.0 MHz ::t1.5dB, ::t100 2.0 MHz to F max ::=1.5 dB, :!:100 Reflection magnitude, phase"u: 0.3 MHz to 2.0 ~(Hz = 1.5dB. =.25" 2.0 MHz (0 F max ::=; 1.5 dB, :!: 100 Effective source matchu: (Test Ports): 0.3 MHz to 2.0 MHz 14dB 2.0 MHz [0 1.3GHz 20dB 1.3GHz to F max 16dB RF Connectors Precision 7 mm Test ports: All others: Physical Characteristics 50 .0.Type N (female) := 1.5 dB, ::t 100 ~ 1.5 dB, :::t100 := 1.5 dB, :: 250 :: 1.5 dB, ::t 100 14dB 17 dB 16 dB 75 0. Type N (fe male) 50 0. Type N (female) Size: 615 mm H x 101mm W x 204 mrn D (2.44 in x 7.5 in x 8.0 in) Weight: Net. 1.7 kg (3.8 lb); shipping, 2.0 kg (4.4 Ib) 'Degrees.speCifiedas deviationtram linearphase. lF is the uoperfrequencylimit of the assoclatedtest set. JCanbe improvedthroughAccuracyEnhancement. With compliments Helmut Singer Elektronik www.helmut-singer.de [email protected] fon +49 241 155 315 fax +49 241 152 066 Feldchen 16-24 D-52070 Aachen Germany