HP 85044A/B Transmisslon/Reflectlon Test Sets 50 n 75 n

HP 85044A/B
Transmisslon/ReflectlonTest Sets
The HP 85044A/B test sets provide the capability to measure the
reflection and transmissioncharacteristics of 500. and 75 n devices,
respectively.
Specifications Summary
Impedance:
Frequency range:
Directivi~:
HP 85044A
HP 85044 B
50n
75n
300 kHz to 3 GHz 300 kHz to 2 GHz
35 dB to 1.3GHz 35 dB to 1.3GHz
30 dB [0 3.0 GHz 30 dB to 2.0 GHz
Typical Tracking
Transmission magnitude, phasel.2.3~
0.3 MHz [0 2.0 MHz
::t1.5dB, ::t100
2.0 MHz to F max
::=1.5 dB, :!:100
Reflection magnitude, phase"u:
0.3 MHz to 2.0 ~(Hz
= 1.5dB. =.25"
2.0 MHz (0 F max
::=;
1.5 dB,
:!:
100
Effective source matchu: (Test Ports):
0.3 MHz to 2.0 MHz
14dB
2.0 MHz [0 1.3GHz
20dB
1.3GHz to F max
16dB
RF Connectors
Precision 7 mm
Test ports:
All others:
Physical Characteristics
50 .0.Type N
(female)
:= 1.5 dB, ::t 100
~ 1.5 dB, :::t100
:= 1.5 dB, :: 250
:: 1.5 dB, ::t 100
14dB
17 dB
16 dB
75 0. Type N
(fe male)
50 0. Type N
(female)
Size: 615 mm H x 101mm W x 204 mrn D (2.44 in x 7.5 in x 8.0 in)
Weight: Net. 1.7 kg (3.8 lb); shipping, 2.0 kg (4.4 Ib)
'Degrees.speCifiedas deviationtram linearphase.
lF is the uoperfrequencylimit of the assoclatedtest set.
JCanbe improvedthroughAccuracyEnhancement.
With compliments
Helmut Singer Elektronik
www.helmut-singer.de [email protected]
fon +49 241 155 315
fax +49 241 152 066
Feldchen 16-24 D-52070 Aachen Germany