The XcelVap® Evaporation/ Concentration System Pressure Profiling Zoe Grosser and William Jones, Horizon Technology, Salem, NH USA Key Words XcelVap, Pressure Profiling, Evaporation, Concentration Programmable pressure profiling in an evaporation system allows the gas pressure to be gradually increased as the evaporation cycle progresses. The low flow rate early in the process promotes gradual evaporation under the vortex gas flow (Figure 1). As the solution evaporates and more space is available in the vessel, the gas flow is increased to further speed evaporation. Evaporation proceeds optimally, avoiding splashing and eliminating the possibility of cross contamination and poor recoveries. The pressure profile is a linear ramp, so the initial gas pressure and the final gas pressure (24 psi max) are spread over the time set for the process to take place. Once this ramp has been developed the sample can be run unattended and the audible alarm will signal when the ramp is complete and the evaporation cycle is done. Figure 2 shows the software for setting the pressure profile ramp. Simply set the initial and final pressure and the software will automatically calculate a ramp to gradually increase the gas flow over the time set for the evaporation. Figure 1. Gas vortex flow The initial pressure should be determined by testing a sample in the glassware chosen for the run by gradually increasing the pressure until a gentle vortex circular motion is seen on the surface, depicted in Figure 3. This pressure will be used as the initial pressure and the final pressure should also be determined for the final extract volume with the maximum allowable being 24 psi. Figure 4 compares the time needed for evaporation of ethyl acetate, using a constant pressure vs. a ramped pressure profile for 7-mL extract evaporated to dryness. The sample evaporated over 30% faster using the ramped pressure profile. The entire process was handled without operator intervention. Figure 5 shows an example of an evaporation process for a 200-mL dichloromethane sample evaporated in a bath at 40oC and with three different gas ramp profiles. The isobaric pressure profile is shown for comparison, but is generally only used on older equipment not offering a ramping option. The ramp set using the criteria described for the small volume sample was used for this larger-volume sample and the time for the Figure 2. Pressure Profile Software Screen Figure 3. Gas vortex looking into the sample from the top Page 2 desired evaporation is significantly shorter. Another alternative to further customize the pressure profile and speed evaporation is to use a two-step process. To start, the ramp is used until the amount of solvent being evaporated is at a safe level where an increase in the gas flow will not cause splattering. Then, the gas flow is increased to the maximum pressure and maintained for the remained of the run. This involves an additional interaction by the operator to reset the profile to continued maximum pressure when the alarm sounds, indicating the ramp part of the program is finished. This program should be developed experimentally to determine if a ramp or step gives the best performance for the solvent, sample size and glassware. This may not be practical on a routine basis, but when time is of the essence for a series of samples, this is another alternative to consider, especially for a larger volume of solvent. Although operator intervention is necessary to change the profile when the audible alarm sounds after the first 10 minutes of the evaporation process, the step profile does reduce the evaporation time. The XcelVap system offers several options for automatic gas increase during the evaporation process. The laboratory can choose the profile appropriate for their needs, based on time, solvent, and demand. Figure 4. Comparison of evaporation using ramped pressure and isobaric pressure profiles for a 7-mL sample evaporated to dryness Figure 5. Comparison of isobaric pressure run with pressure profile ramp and manual stepped ramp for a 200-mL sample TN0021405-01 16 Northwestern Drive, Salem, NH 03079 USA ▪ Tel: (603) 893-3663 ▪ Email: [email protected]
© Copyright 2024 Paperzz